Design of a scanning tunneling microscope for in situ observation of the interactions of molecular beams with surfaces

Author keywords

[No Author keywords available]

Indexed keywords

COLLISION-INDUCED MOBILITY; HELIUM-SEEDED ARGON BEAM; MOLECULAR-BEAM SOURCE; ULTRAHIGH VACUUM SCANNING TUNNELING MICROSCOPE (UHV-STM);

EID: 26444486990     PISSN: 00346748     EISSN: None     Source Type: Journal    
DOI: 10.1063/1.2005467     Document Type: Article
Times cited : (10)

References (28)
  • 22
    • 26444440236 scopus 로고    scopus 로고
    • note
    • Because thermal fluctuations of the STM must be avoided, some care is necessary when filling the liquid nitrogen trap. We have found that a minimum time of 6 h is required to allow the STM to equilibrate back to room temperature after filling the liquid nitrogen trap, and before the next experiment is performed.
  • 26
    • 26444523267 scopus 로고    scopus 로고
    • note
    • Fourier processing was used to remove the effects of 60 Hz interference in these images caused by the nearby solenoid gate valve in Chamber III. Switching to a dc-driven solenoid has since removed this source of instrumental noise.