Longitudinal evaluation of compressor/nebulizer performance

Author keywords

Aerosol; Breathing simulation; Compressor; Cystic fibrosis; Flow; Nebulizer; Particle size; Pressure

Indexed keywords

SALBUTAMOL; AEROSOL; BRONCHODILATING AGENT;

EID: 84928106762     PISSN: 00201324     EISSN: 19433654     Source Type: Journal    
DOI: 10.4187/respcare.02776     Document Type: Article
Times cited : (13)

References (25)
  • 1
    • 65849522497 scopus 로고    scopus 로고
    • Cystic fibrosis
    • O'Sullivan BP, Freedman SD. Cystic fibrosis. Lancet 2009; 373(9678): 1891-1904.
    • (2009) Lancet , vol.373 , Issue.9678 , pp. 1891-1904
    • O'Sullivan, B.P.1    Freedman, S.D.2
  • 2
    • 80053375975 scopus 로고    scopus 로고
    • Bethesda, MD: Cystic Fibrosis Foundation
    • Cystic Fibrosis Foundation Patient Registry. Annual data report to the Center Directors. Bethesda, MD: Cystic Fibrosis Foundation; 2010.
    • (2010) Annual data report to the Center Directors
  • 3
    • 66849089934 scopus 로고    scopus 로고
    • Cystic fibrosis: Pathogenesis and future treatment strategies
    • Ratjen FA. Cystic fibrosis: pathogenesis and future treatment strategies. Respir Care 2009; 54(5): 595-605.
    • (2009) Respir Care , vol.54 , Issue.5 , pp. 595-605
    • Ratjen, F.A.1
  • 6
  • 7
    • 69549135253 scopus 로고    scopus 로고
    • Inhaled medication and inhalation devices for lung disease in patients with cystic fibrosis: A European consensus
    • Heijerman H, Westerman E, Conway S, Touw D, Döring G; consensus working group. Inhaled medication and inhalation devices for lung disease in patients with cystic fibrosis: a European consensus. J Cyst Fibros 2009; 8(5): 295-315.
    • (2009) J Cyst Fibros , vol.8 , Issue.5 , pp. 295-315
    • Heijerman, H.1    Westerman, E.2    Conway, S.3    Touw, D.4    Döring, G.5
  • 8
    • 0028129568 scopus 로고
    • Effect of aerosolized recombinant human DNase on exacerbations of respiratory symptoms and on pulmonary function in patients with cystic fibrosis. The Pulmozyme Study Group
    • Fuchs HJ, Borowitz DS, Christiansen DH, Morris EM. Nash ML, Ramsey BW, et al. Effect of aerosolized recombinant human DNase on exacerbations of respiratory symptoms and on pulmonary function in patients with cystic fibrosis. The Pulmozyme Study Group. N Engl J Med 1994; 331 (10): 637-642.
    • (1994) N Engl J Med , vol.331 , Issue.10 , pp. 637-642
    • Fuchs, H.J.1    Borowitz, D.S.2    Christiansen, D.H.3    Morris, E.M.4    Nash, M.L.5    Ramsey, B.W.6
  • 9
    • 78650036519 scopus 로고    scopus 로고
    • A technical feasibility study of dornase alfa delivery with eFlow vibrating membrane nebulizers: Aerosol characteristics and physicochemical stability
    • Scherer T, Geller DE, Owyang L, Tservesta M, Keller M, Boden N, et al. A technical feasibility study of dornase alfa delivery with eFlow vibrating membrane nebulizers: aerosol characteristics and physicochemical stability. J Pharm Sci 2011; 100(1): 98-109.
    • (2011) J Pharm Sci , vol.100 , Issue.1 , pp. 98-109
    • Scherer, T.1    Geller, D.E.2    Owyang, L.3    Tservesta, M.4    Keller, M.5    Boden, N.6
  • 10
    • 84928168053 scopus 로고    scopus 로고
    • [package insert]. Genentech, South San Francisco, CA
    • Pulmozyme [package insert]. Genentech, South San Francisco, CA; 2010.
    • (2010) Pulmozyme
  • 12
    • 79952763932 scopus 로고    scopus 로고
    • Aerosol drug delivery: Developments in device design and clinical use
    • Dolovich MB, Dhand R. Aerosol drug delivery: developments in device design and clinical use. Lancet 2011; 377(9770): 1032-1045.
    • (2011) Lancet , vol.377 , Issue.9770 , pp. 1032-1045
    • Dolovich, M.B.1    Dhand, R.2
  • 14
    • 70349923912 scopus 로고    scopus 로고
    • Output and aerosol properties of 5 nebulizer/compressor systems with arformoterol inhalation solution
    • Bauer A, McGlynn P, Bovet LL, Mims PL, Curry LA, Hanrahan JP. Output and aerosol properties of 5 nebulizer/compressor systems with arformoterol inhalation solution. Respir Care 2009: 54(10): 1342-1347.
    • (2009) Respir Care , vol.54 , Issue.10 , pp. 1342-1347
    • Bauer, A.1    McGlynn, P.2    Bovet, L.L.3    Mims, P.L.4    Curry, L.A.5    Hanrahan, J.P.6
  • 15
    • 0029763316 scopus 로고    scopus 로고
    • Medication nebulizer performance: Effects of diluent volume, nebulizer flow, and nebulizer brand
    • Hess D, Fisher D, Williams P, Pooler S, Kacmarek RM. Medication nebulizer performance: effects of diluent volume, nebulizer flow, and nebulizer brand. Chest 1996; 110(2): 498-505.
    • (1996) Chest , vol.110 , Issue.2 , pp. 498-505
    • Hess, D.1    Fisher, D.2    Williams, P.3    Pooler, S.4    Kacmarek, R.M.5
  • 16
    • 84928110670 scopus 로고
    • Comparison of twenty three nebulizer/compressor combinations for domiciliary use
    • Smith EC, Denyer J, Kendrick AH. Comparison of twenty three nebulizer/compressor combinations for domiciliary use. Eur Respir J 1995; 89(3): 157-159.
    • (1995) Eur Respir J , vol.89 , Issue.3 , pp. 157-159
    • Smith, E.C.1    Denyer, J.2    Kendrick, A.H.3
  • 17
    • 0035006433 scopus 로고    scopus 로고
    • The equivalence of compressor pressure-flow relationships with respect to jet nebulizer aerosolization characteristics
    • Standaert TA, Bohn SE, Aitken ML, Ramsey B. The equivalence of compressor pressure-flow relationships with respect to jet nebulizer aerosolization characteristics. J Aerosol Med 2001; 14(1): 31-42.
    • (2001) J Aerosol Med , vol.14 , Issue.1 , pp. 31-42
    • Standaert, T.A.1    Bohn, S.E.2    Aitken, M.L.3    Ramsey, B.4
  • 18
    • 0033918213 scopus 로고    scopus 로고
    • The choice of compressor effects the aerosol parameters and the delivery of tobramycin from a single model nebulizer
    • Standaert TA, Vandevanter D, Ramsey BW, Vasiljev M, Nardella P, Gmur D, et al. The choice of compressor effects the aerosol parameters and the delivery of tobramycin from a single model nebulizer. J Aerosol Med 2000; 13(2): 147-153.
    • (2000) J Aerosol Med , vol.13 , Issue.2 , pp. 147-153
    • Standaert, T.A.1    Vandevanter, D.2    Ramsey, B.W.3    Vasiljev, M.4    Nardella, P.5    Gmur, D.6
  • 20
    • 0344413659 scopus 로고    scopus 로고
    • The choice of a compressor for the aerosolisation of tobramycin (TOBI) with the PARI LC PLUS reusable nebuliser
    • de Boer AH, Hagedoorn P. Frijlink HW. The choice of a compressor for the aerosolisation of tobramycin (TOBI) with the PARI LC PLUS reusable nebuliser. Int J Pharm 268(1-2): 59-69, 2003.
    • (2003) Int J Pharm , vol.268 , Issue.1-2 , pp. 59-69
    • De Boer, A.H.1    Hagedoorn, P.2    Frijlink, H.W.3
  • 21
    • 0031813115 scopus 로고    scopus 로고
    • Effects of repetitive use and cleaning techniques of disposable jet nebulizers on aerosol generation
    • Standaert TA, Morlin GL. Williams-Warren J, Joy P, Pepc MS, Weber A, Ramsey BW. Effects of repetitive use and cleaning techniques of disposable jet nebulizers on aerosol generation. Chest 1998; 114(2): 577-586.
    • (1998) Chest , vol.114 , Issue.2 , pp. 577-586
    • Standaert, T.A.1    Morlin, G.L.2    Williams-Warren, J.3    Joy, P.4    Pepc, M.S.5    Weber, A.6    Ramsey, B.W.7
  • 22
    • 78649940446 scopus 로고    scopus 로고
    • Optimization of a procedure used to measure aerosol characteristics of nebulized solutions using a cooled next generation impactor
    • Berlinski A, Hayden JB. Optimization of a procedure used to measure aerosol characteristics of nebulized solutions using a cooled next generation impactor. J Aerosol Med Pulm Drug Deliv 2010; 23(6): 397-404.
    • (2010) J Aerosol Med Pulm Drug Deliv , vol.23 , Issue.6 , pp. 397-404
    • Berlinski, A.1    Hayden, J.B.2
  • 23
    • 70849117729 scopus 로고    scopus 로고
    • The influence of breathing pattern during nebulization on the delivery of arformoterol using a breath simulator
    • Bauer A, McGlynn P, Bovet LL, Mims PL, Curry LA, Hanrahan JP. The influence of breathing pattern during nebulization on the delivery of arformoterol using a breath simulator. Respir Care 2009; 54(11): 1488-1492.
    • (2009) Respir Care , vol.54 , Issue.11 , pp. 1488-1492
    • Bauer, A.1    McGlynn, P.2    Bovet, L.L.3    Mims, P.L.4    Curry, L.A.5    Hanrahan, J.P.6
  • 24
    • 77449113062 scopus 로고    scopus 로고
    • Effect of face mask static dead volume, respiratory rate and tidal volume on inhaled albuterol delivery
    • Chavez A, McCracken A, Berlinski A. Effect of face mask static dead volume, respiratory rate and tidal volume on inhaled albuterol delivery. Pediatr Pulmonol 2010; 45(3): 224-229.
    • (2010) Pediatr Pulmonol , vol.45 , Issue.3 , pp. 224-229
    • Chavez, A.1    McCracken, A.2    Berlinski, A.3
  • 25
    • 0031843329 scopus 로고    scopus 로고
    • Drug output from nebulizers is dependent on the method of measurement
    • Barry PW, O'Callaghan C. Drug output from nebulizers is dependent on the method of measurement. Eur Respir J 1998; 12(2): 463-466.
    • (1998) Eur Respir J , vol.12 , Issue.2 , pp. 463-466
    • Barry, P.W.1    O'Callaghan, C.2


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.