Noninvasive continuous versus intermittent arterial pressure monitoring: Evaluation of the vascular unloading technique (CNAP device) in the emergency department

Author keywords

Diastolic blood pressure; Emergency medicine; Mean arterial pressure; Photoplethysmography; Systolic blood pressure; Volume clamp method

Indexed keywords

AGED; ARTERIAL PRESSURE; ARTICLE; BLOOD PRESSURE MEASUREMENT; COMPARATIVE STUDY; EMERGENCY HEALTH SERVICE; EQUIPMENT; EQUIPMENT DESIGN; FEMALE; FOLLOW UP; HUMAN; HYPOTENSION; MALE; MIDDLE AGED; PATHOPHYSIOLOGY; PHOTOELECTRIC PLETHYSMOGRAPHY; PHYSIOLOGIC MONITORING; PHYSIOLOGY; PROSPECTIVE STUDY; REPRODUCIBILITY;

EID: 84893009023     PISSN: None     EISSN: 17577241     Source Type: Journal    
DOI: 10.1186/1757-7241-22-8     Document Type: Article
Times cited : (32)

References (23)
  • 2
    • 0031951612 scopus 로고    scopus 로고
    • Indications and complications of arterial catheter use in surgical or medical intensive care units: analysis of 4932 patients
    • Frezza EE, Mezghebe H. Indications and complications of arterial catheter use in surgical or medical intensive care units: analysis of 4932 patients. Am Surg 1998, 64:127-131.
    • (1998) Am Surg , vol.64 , pp. 127-131
    • Frezza, E.E.1    Mezghebe, H.2
  • 3
    • 33746367766 scopus 로고    scopus 로고
    • Continuous non-invasive blood pressure monitoring using concentrically interlocking control loops
    • 10.1016/j.compbiomed.2005.04.003, 16483562
    • Fortin J, Marte W, Grullenberger R, Hacker A, Habenbacher W, Heller A, Wagner C, Wach P, Skrabal F. Continuous non-invasive blood pressure monitoring using concentrically interlocking control loops. Comput Biol Med 2006, 36:941-957. 10.1016/j.compbiomed.2005.04.003, 16483562.
    • (2006) Comput Biol Med , vol.36 , pp. 941-957
    • Fortin, J.1    Marte, W.2    Grullenberger, R.3    Hacker, A.4    Habenbacher, W.5    Heller, A.6    Wagner, C.7    Wach, P.8    Skrabal, F.9
  • 4
    • 77955573200 scopus 로고    scopus 로고
    • Continuous non-invasive arterial pressure measurement: evaluation of CNAP device during vascular surgery
    • Biais M, Vidil L, Roullet S, Masson F, Quinart A, Revel P, Sztark F. Continuous non-invasive arterial pressure measurement: evaluation of CNAP device during vascular surgery. Ann Fr Anesth Reanim 2010, 29:530-535.
    • (2010) Ann Fr Anesth Reanim , vol.29 , pp. 530-535
    • Biais, M.1    Vidil, L.2    Roullet, S.3    Masson, F.4    Quinart, A.5    Revel, P.6    Sztark, F.7
  • 5
    • 77956019814 scopus 로고    scopus 로고
    • Precision and accuracy of a new device (CNAPTM) for continuous non-invasive arterial pressure monitoring: assessment during general anaesthesia
    • 10.1093/bja/aeq143, 20627878
    • Jeleazcov C, Krajinovic L, Munster T, Birkholz T, Fried R, Schuttler J, Fechner J. Precision and accuracy of a new device (CNAPTM) for continuous non-invasive arterial pressure monitoring: assessment during general anaesthesia. Br J Anaesth 2010, 105:264-272. 10.1093/bja/aeq143, 20627878.
    • (2010) Br J Anaesth , vol.105 , pp. 264-272
    • Jeleazcov, C.1    Krajinovic, L.2    Munster, T.3    Birkholz, T.4    Fried, R.5    Schuttler, J.6    Fechner, J.7
  • 6
    • 84858387796 scopus 로고    scopus 로고
    • Clinical validation of a continuous non-invasive haemodynamic monitor (CNAP™ 500) during general anaesthesia
    • 10.1093/bja/aer499, 22307242
    • Hahn R, Rinosl H, Neuner M, Kettner SC. Clinical validation of a continuous non-invasive haemodynamic monitor (CNAP™ 500) during general anaesthesia. Br J Anaesth 2012, 108:581-585. 10.1093/bja/aer499, 22307242.
    • (2012) Br J Anaesth , vol.108 , pp. 581-585
    • Hahn, R.1    Rinosl, H.2    Neuner, M.3    Kettner, S.C.4
  • 7
    • 84856303645 scopus 로고    scopus 로고
    • Investigation of the agreement of a continuous non-invasive arterial pressure device in comparison with invasive radial artery measurement
    • 10.1093/bja/aer394, 22171358
    • Ilies C, Bauer M, Berg P, Rosenberg J, Hedderich J, Bein B, Hinz J, Hanss R. Investigation of the agreement of a continuous non-invasive arterial pressure device in comparison with invasive radial artery measurement. Br J Anaesth 2012, 108:202-210. 10.1093/bja/aer394, 22171358.
    • (2012) Br J Anaesth , vol.108 , pp. 202-210
    • Ilies, C.1    Bauer, M.2    Berg, P.3    Rosenberg, J.4    Hedderich, J.5    Bein, B.6    Hinz, J.7    Hanss, R.8
  • 8
    • 84864270411 scopus 로고    scopus 로고
    • A comparison of a continuous noninvasive arterial pressure (CNAP) monitor with an invasive arterial blood pressure monitor in the cardiac surgical ICU
    • 10.4103/0971-9784.97973, 22772511
    • Jagadeesh AM, Singh NG, Mahankali S. A comparison of a continuous noninvasive arterial pressure (CNAP) monitor with an invasive arterial blood pressure monitor in the cardiac surgical ICU. Ann Card Anaesth 2012, 15:180-184. 10.4103/0971-9784.97973, 22772511.
    • (2012) Ann Card Anaesth , vol.15 , pp. 180-184
    • Jagadeesh, A.M.1    Singh, N.G.2    Mahankali, S.3
  • 11
    • 84866354749 scopus 로고    scopus 로고
    • Level of agreement between Nexfin non-invasive arterial pressure with invasive arterial pressure measurements in children
    • 10.1093/bja/aes295, 22907342
    • Garnier RP, van der Spoel AG, Sibarani-Ponsen R, Markhorst DG, Boer C. Level of agreement between Nexfin non-invasive arterial pressure with invasive arterial pressure measurements in children. Br J Anaesth 2012, 109:609-615. 10.1093/bja/aes295, 22907342.
    • (2012) Br J Anaesth , vol.109 , pp. 609-615
    • Garnier, R.P.1    van der Spoel, A.G.2    Sibarani-Ponsen, R.3    Markhorst, D.G.4    Boer, C.5
  • 12
    • 31444450219 scopus 로고    scopus 로고
    • An accuracy evaluation of the T-Line Tensymeter (continuous noninvasive blood pressure management device) versus conventional invasive radial artery monitoring in surgical patients
    • 10.1213/01.ane.0000194873.52453.bd, 16428547
    • Janelle GM, Gravenstein N. An accuracy evaluation of the T-Line Tensymeter (continuous noninvasive blood pressure management device) versus conventional invasive radial artery monitoring in surgical patients. Anesth Analg 2006, 102:484-490. 10.1213/01.ane.0000194873.52453.bd, 16428547.
    • (2006) Anesth Analg , vol.102 , pp. 484-490
    • Janelle, G.M.1    Gravenstein, N.2
  • 13
    • 39149084640 scopus 로고    scopus 로고
    • An evaluation of the T-Line Tensymeter continuous noninvasive blood pressure device during induced hypotension
    • 10.1111/j.1365-2044.2007.05369.x, 18289239
    • Szmuk P, Pivalizza E, Warters RD, Ezri T, Gebhard R. An evaluation of the T-Line Tensymeter continuous noninvasive blood pressure device during induced hypotension. Anaesthesia 2008, 63:307-312. 10.1111/j.1365-2044.2007.05369.x, 18289239.
    • (2008) Anaesthesia , vol.63 , pp. 307-312
    • Szmuk, P.1    Pivalizza, E.2    Warters, R.D.3    Ezri, T.4    Gebhard, R.5
  • 14
    • 84863085965 scopus 로고    scopus 로고
    • Noninvasive continuous beat-to-beat radial artery pressure via TL-200 applanation tonometry
    • 10.1007/s10877-012-9336-2, 22258303
    • Dueck R, Goedje O, Clopton P. Noninvasive continuous beat-to-beat radial artery pressure via TL-200 applanation tonometry. J Clin Monit Comput 2012, 26:75-83. 10.1007/s10877-012-9336-2, 22258303.
    • (2012) J Clin Monit Comput , vol.26 , pp. 75-83
    • Dueck, R.1    Goedje, O.2    Clopton, P.3
  • 15
    • 84866743712 scopus 로고    scopus 로고
    • The T-Line TL-200 system for continuous non-invasive blood pressure measurement in medical intensive care unit patients
    • 10.1007/s00134-012-2617-x, 22744376
    • Saugel B, Fassio F, Hapfelmeier A, Meidert AS, Schmid RM, Huber W. The T-Line TL-200 system for continuous non-invasive blood pressure measurement in medical intensive care unit patients. Intensive Care Med 2012, 38:1471-1477. 10.1007/s00134-012-2617-x, 22744376.
    • (2012) Intensive Care Med , vol.38 , pp. 1471-1477
    • Saugel, B.1    Fassio, F.2    Hapfelmeier, A.3    Meidert, A.S.4    Schmid, R.M.5    Huber, W.6
  • 16
    • 84880550339 scopus 로고    scopus 로고
    • Non-invasive continuous arterial pressure measurement based on radial artery tonometry in the intensive care unit: a method comparison study using the T-Line TL-200pro device
    • 10.1093/bja/aet025, 23491946
    • Saugel B, Meidert AS, Hapfelmeier A, Eyer F, Schmid RM, Huber W. Non-invasive continuous arterial pressure measurement based on radial artery tonometry in the intensive care unit: a method comparison study using the T-Line TL-200pro device. Br J Anaesth 2013, 111:185-190. 10.1093/bja/aet025, 23491946.
    • (2013) Br J Anaesth , vol.111 , pp. 185-190
    • Saugel, B.1    Meidert, A.S.2    Hapfelmeier, A.3    Eyer, F.4    Schmid, R.M.5    Huber, W.6
  • 17
    • 84887402327 scopus 로고    scopus 로고
    • Evaluation of the radial artery applanation tonometry technology for continuous noninvasive blood pressure monitoring compared with central aortic blood pressure measurements in patients with multiple organ dysfunction syndrome
    • 10.1016/j.jcrc.2013.06.012, 23910893
    • Meidert AS, Huber W, Hapfelmeier A, Schofthaler M, Muller JN, Langwieser N, Wagner JY, Schmid RM, Saugel B. Evaluation of the radial artery applanation tonometry technology for continuous noninvasive blood pressure monitoring compared with central aortic blood pressure measurements in patients with multiple organ dysfunction syndrome. J Crit Care 2013, 28:908-912. 10.1016/j.jcrc.2013.06.012, 23910893.
    • (2013) J Crit Care , vol.28 , pp. 908-912
    • Meidert, A.S.1    Huber, W.2    Hapfelmeier, A.3    Schofthaler, M.4    Muller, J.N.5    Langwieser, N.6    Wagner, J.Y.7    Schmid, R.M.8    Saugel, B.9
  • 18
    • 60049095571 scopus 로고    scopus 로고
    • Continuous non-invasive arterial pressure technique improves patient monitoring during interventional endoscopy
    • 2631161, 19173016
    • Siebig S, Rockmann F, Sabel K, Zuber-Jerger I, Dierkes C, Brunnler T, Wrede CE. Continuous non-invasive arterial pressure technique improves patient monitoring during interventional endoscopy. Int J Med Sci 2009, 6:37-42. 2631161, 19173016.
    • (2009) Int J Med Sci , vol.6 , pp. 37-42
    • Siebig, S.1    Rockmann, F.2    Sabel, K.3    Zuber-Jerger, I.4    Dierkes, C.5    Brunnler, T.6    Wrede, C.E.7
  • 19
    • 84865264710 scopus 로고    scopus 로고
    • Detection of hypotension during Caesarean section with continuous non-invasive arterial pressure device or intermittent oscillometric arterial pressure measurement
    • 10.1093/bja/aes224, 22798273
    • Ilies C, Kiskalt H, Siedenhans D, Meybohm P, Steinfath M, Bein B, Hanss R. Detection of hypotension during Caesarean section with continuous non-invasive arterial pressure device or intermittent oscillometric arterial pressure measurement. Br J Anaesth 2012, 109:413-419. 10.1093/bja/aes224, 22798273.
    • (2012) Br J Anaesth , vol.109 , pp. 413-419
    • Ilies, C.1    Kiskalt, H.2    Siedenhans, D.3    Meybohm, P.4    Steinfath, M.5    Bein, B.6    Hanss, R.7
  • 20
    • 0017032294 scopus 로고
    • Contribution to the continuous indirect blood pressure measurement
    • Penaz J, Voigt A, Teichmann W. Contribution to the continuous indirect blood pressure measurement. Z Gesamte Inn Med 1976, 31:1030-1033.
    • (1976) Z Gesamte Inn Med , vol.31 , pp. 1030-1033
    • Penaz, J.1    Voigt, A.2    Teichmann, W.3
  • 21
    • 34347375489 scopus 로고    scopus 로고
    • Agreement between methods of measurement with multiple observations per individual
    • 10.1080/10543400701329422, 17613642
    • Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat 2007, 17:571-582. 10.1080/10543400701329422, 17613642.
    • (2007) J Biopharm Stat , vol.17 , pp. 571-582
    • Bland, J.M.1    Altman, D.G.2
  • 22
    • 80052213181 scopus 로고    scopus 로고
    • Noninvasive continuous or intermittent blood pressure and heart rate patient monitoring in the ED
    • 10.1016/j.ajem.2011.05.014, 21802881
    • Nowak RM, Sen A, Garcia AJ, Wilkie H, Yang JJ, Nowak MR, Moyer ML. Noninvasive continuous or intermittent blood pressure and heart rate patient monitoring in the ED. Am J Emerg Med 2011, 29:782-789. 10.1016/j.ajem.2011.05.014, 21802881.
    • (2011) Am J Emerg Med , vol.29 , pp. 782-789
    • Nowak, R.M.1    Sen, A.2    Garcia, A.J.3    Wilkie, H.4    Yang, J.J.5    Nowak, M.R.6    Moyer, M.L.7
  • 23
    • 0003769166 scopus 로고    scopus 로고
    • ANSI/AAMI SP10:2002. American national standard for manual, electronic, or automated sphygmomanometers
    • ANSI/AAMI SP10:2002. American national standard for manual, electronic, or automated sphygmomanometers Association for the Advancement of Medical Instrumentation 2002, ANSI/AAMI SP10:2002. American national standard for manual, electronic, or automated sphygmomanometers.
    • (2002) Association for the Advancement of Medical Instrumentation


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.