Cardiac resynchronisation therapy optimisation strategies: Systematic classification, detailed analysis, minimum standards and a roadmap for development and testing

(16)  Sohaib, S M Afzal a   Whinnett, Zachary I a   Ellenbogen, Kenneth A b   Stellbrink, Christoph c   Quinn, T Alexander e,l   Bogaard, Margot D j   Bordachar, Pierre g   Van Gelder, Berry M f   Van Geldorp, Irene E d   Linde, Cecilia h   Meine, Mathias j   Prinzen, Frits W d   Turcott, Robert G i   Spotnitz, Henry M e   Wichterle, Dan k   Francis, Darrel P a  


Author keywords

AV optimisation; Cardiac resynchronization therapy; Heart failure

Indexed keywords

ACCURACY; CARDIAC RESYNCHRONIZATION THERAPY; CLASSIFICATION; CLINICAL PROTOCOL; COST BENEFIT ANALYSIS; DISEASE EXACERBATION; DOPPLER ECHOCARDIOGRAPHY; HEALTH CARE DELIVERY; HUMAN; NON INVASIVE PROCEDURE; PATIENT INFORMATION; PATIENT SATISFACTION; PHOTOELECTRIC PLETHYSMOGRAPHY; PREDICTION; PRIORITY JOURNAL; RELIABILITY; REPRODUCIBILITY; REVIEW; TREATMENT PLANNING;

EID: 84889089069     PISSN: 01675273     EISSN: 18741754     Source Type: Journal    
DOI: 10.1016/j.ijcard.2013.10.069     Document Type: Review
Times cited : (33)

References (60)
  • 1
    • 84879782216 scopus 로고    scopus 로고
    • Effect of atrioventricular and ventriculoventricular delay optimization on clinical and echocardiographic outcomes of patients treated with cardiac resynchronization therapy: A meta-analysis
    • D. Auger, U. Hoke, J.J. Bax, E. Boersma, and V. Delgado Effect of atrioventricular and ventriculoventricular delay optimization on clinical and echocardiographic outcomes of patients treated with cardiac resynchronization therapy: a meta-analysis Am Heart J 166 1 2013 20 29
    • (2013) Am Heart J , vol.166 , Issue.1 , pp. 20-29
    • Auger, D.1    Hoke, U.2    Bax, J.J.3    Boersma, E.4    Delgado, V.5
  • 3
    • 33750302936 scopus 로고    scopus 로고
    • Haemodynamic effects of changes in atrioventricular and interventricular delay in cardiac resynchronisation therapy show a consistent pattern: Analysis of shape, magnitude and relative importance of atrioventricular and interventricular delay
    • Z.I. Whinnett, J.E. Davies, and K. Willson Haemodynamic effects of changes in atrioventricular and interventricular delay in cardiac resynchronisation therapy show a consistent pattern: analysis of shape, magnitude and relative importance of atrioventricular and interventricular delay Heart 92 2006 1628 1634
    • (2006) Heart , vol.92 , pp. 1628-1634
    • Whinnett, Z.I.1    Davies, J.E.2    Willson, K.3
  • 4
    • 84881257640 scopus 로고    scopus 로고
    • ESC guidelines on cardiac pacing and cardiac resynchronization therapy: The Task Force on Cardiac Pacing and Resynchronization Therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA)
    • M. Brignole, A. Auricchio, and G. Baron-Esquivias ESC guidelines on cardiac pacing and cardiac resynchronization therapy: the Task Force on Cardiac Pacing and Resynchronization Therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA) Eur Heart J 34 29 2013 2281 2329
    • (2013) Eur Heart J , vol.34 , Issue.29 , pp. 2281-2329
    • Brignole, M.1    Auricchio, A.2    Baron-Esquivias, G.3
  • 5
    • 84865632707 scopus 로고    scopus 로고
    • EHRA/HRS expert consensus statement on cardiac resynchronization therapy in heart failure: Implant and follow-up recommendations and management
    • European Heart Rhythm Association, European Society of Cardiology, and Heart Rhythm Society
    • European Heart Rhythm Association, European Society of Cardiology, and Heart Rhythm Society EHRA/HRS expert consensus statement on cardiac resynchronization therapy in heart failure: implant and follow-up recommendations and management Heart Rhythm 9 2012 1524 1576
    • (2012) Heart Rhythm , vol.9 , pp. 1524-1576
  • 6
    • 80051707529 scopus 로고    scopus 로고
    • Atrioventricular and interventricular delay optimization in cardiac resynchronization therapy: Physiological principles and overview of available methods
    • P. Houthuizen, F.A. Bracke, and B.M. van Gelder Atrioventricular and interventricular delay optimization in cardiac resynchronization therapy: physiological principles and overview of available methods Heart Fail Rev 16 2011 263 276
    • (2011) Heart Fail Rev , vol.16 , pp. 263-276
    • Houthuizen, P.1    Bracke, F.A.2    Van Gelder, B.M.3
  • 7
    • 0033108072 scopus 로고    scopus 로고
    • Determination of the optimal atrioventricular delay in DDD pacing. Comparison between echo and peak endocardial acceleration measurements
    • P. Ritter, L. Padeletti, L. Gillio-Meina, and G. Gaggini Determination of the optimal atrioventricular delay in DDD pacing. Comparison between echo and peak endocardial acceleration measurements Europace 1 1999 126 130
    • (1999) Europace , vol.1 , pp. 126-130
    • Ritter, P.1    Padeletti, L.2    Gillio-Meina, L.3    Gaggini, G.4
  • 8
    • 34548434671 scopus 로고    scopus 로고
    • Optimizing the programation of cardiac resynchronization therapy devices in patients with heart failure and left bundle branch block
    • B. Vidal, M. Sitges, and A. Marigliano Optimizing the programation of cardiac resynchronization therapy devices in patients with heart failure and left bundle branch block Am J Cardiol 100 2007 1002 1006
    • (2007) Am J Cardiol , vol.100 , pp. 1002-1006
    • Vidal, B.1    Sitges, M.2    Marigliano, A.3
  • 9
    • 0034893807 scopus 로고    scopus 로고
    • CARE-HF Study Steering Committee and Investigators. The CARE-HF Study (CArdiac REsynchronisation in Heart Failure study): Rationale, design and end-points
    • J.G. Cleland, J.C. Daubert, and E. Erdmann CARE-HF Study Steering Committee and Investigators. The CARE-HF Study (CArdiac REsynchronisation in Heart Failure study): rationale, design and end-points Eur J Heart Fail 3 2001 481 489
    • (2001) Eur J Heart Fail , vol.3 , pp. 481-489
    • Cleland, J.G.1    Daubert, J.C.2    Erdmann, E.3
  • 10
    • 1642561531 scopus 로고    scopus 로고
    • A fast and simple echocardiographic method of determination of the optimal atrioventricular delay in patients after biventricular stimulation
    • J. Meluzín, M. Novák, and J. Müllerová A fast and simple echocardiographic method of determination of the optimal atrioventricular delay in patients after biventricular stimulation Pacing Clin Electrophysiol 27 2004 58 64
    • (2004) Pacing Clin Electrophysiol , vol.27 , pp. 58-64
    • Meluzín, J.1    Novák, M.2    Müllerová, J.3
  • 11
    • 84889094469 scopus 로고    scopus 로고
    • Applicability of the iterative technique for cardiac resynchronization therapy optimization: Full-disclosure, 50-sequential-patient dataset of transmitral Doppler traces, with implications for future research design and guidelines
    • 10.1093/europace/eut257
    • S. Jones, M.J. Shun-Shin, G.D. Cole, A. Sau, K. March, S. Williams, A. Kyriacou, A.D. Hughes, J. Mayet, M. Frenneaux, C.H. Manisty, Z.I. Whinnett, and D.P. Francis Applicability of the iterative technique for cardiac resynchronization therapy optimization: full-disclosure, 50-sequential-patient dataset of transmitral Doppler traces, with implications for future research design and guidelines Europace 2013 Sep 25 10.1093/europace/eut257
    • (2013) Europace
    • Jones, S.1    Shun-Shin, M.J.2    Cole, G.D.3    Sau, A.4    March, K.5    Williams, S.6    Kyriacou, A.7    Hughes, A.D.8    Mayet, J.9    Frenneaux, M.10    Manisty, C.H.11    Whinnett, Z.I.12    Francis, D.P.13
  • 12
    • 26444455795 scopus 로고    scopus 로고
    • Cardiac resynchronization with sequential biventricular pacing for the treatment of moderate-to-severe heart failure
    • A.R. León, W.T. Abraham, and S. Brozena Cardiac resynchronization with sequential biventricular pacing for the treatment of moderate-to-severe heart failure J Am Coll Cardiol 46 2005 2298 2304
    • (2005) J Am Coll Cardiol , vol.46 , pp. 2298-2304
    • León, A.R.1    Abraham, W.T.2    Brozena, S.3
  • 13
    • 78650700438 scopus 로고    scopus 로고
    • Primary results from the SmartDelay determined AV optimization: A comparison to other AV delay methods used in cardiac resynchronization therapy (SMART-AV) trial: A randomized trial comparing empirical, echocardiography- guided, and algorithmic atrioventricular delay programming in cardiac resynchronization therapy
    • K.A. Ellenbogen, M.R. Gold, and T.E. Meyer Primary results from the SmartDelay determined AV optimization: a comparison to other AV delay methods used in cardiac resynchronization therapy (SMART-AV) trial: a randomized trial comparing empirical, echocardiography-guided, and algorithmic atrioventricular delay programming in cardiac resynchronization therapy Circulation 122 2010 2660 2668
    • (2010) Circulation , vol.122 , pp. 2660-2668
    • Ellenbogen, K.A.1    Gold, M.R.2    Meyer, T.E.3
  • 14
  • 15
    • 81055140537 scopus 로고    scopus 로고
    • Cardiac resynchronization therapy optimization using noninvasive cardiac output measurement
    • F.Z. Khan, M.S. Virdee, and J. Hutchinson Cardiac resynchronization therapy optimization using noninvasive cardiac output measurement Pacing Clin Electrophysiol 34 2011 1527 1536
    • (2011) Pacing Clin Electrophysiol , vol.34 , pp. 1527-1536
    • Khan, F.Z.1    Virdee, M.S.2    Hutchinson, J.3
  • 16
    • 31944445724 scopus 로고    scopus 로고
    • Correlation of echo-Doppler optimization of atrioventricular delay in cardiac resynchronization therapy with invasive hemodynamics in patients with heart failure secondary to ischemic or idiopathic dilated cardiomyopathy
    • A.H. Jansen, F.A. Bracke, and J.M. van Dantzig Correlation of echo-Doppler optimization of atrioventricular delay in cardiac resynchronization therapy with invasive hemodynamics in patients with heart failure secondary to ischemic or idiopathic dilated cardiomyopathy Am J Cardiol 97 2006 552 557
    • (2006) Am J Cardiol , vol.97 , pp. 552-557
    • Jansen, A.H.1    Bracke, F.A.2    Van Dantzig, J.M.3
  • 17
    • 23944520865 scopus 로고    scopus 로고
    • Optimized biventricular pacing in atrioventricular block after cardiac surgery
    • G. Berberian, T.A. Quinn, and J.P. Kanter Optimized biventricular pacing in atrioventricular block after cardiac surgery Ann Thorac Surg 80 2005 870 875
    • (2005) Ann Thorac Surg , vol.80 , pp. 870-875
    • Berberian, G.1    Quinn, T.A.2    Kanter, J.P.3
  • 18
    • 79952993787 scopus 로고    scopus 로고
    • Optimized temporary biventricular pacing acutely improves intraoperative cardiac output after weaning from cardiopulmonary bypass: A sub-study of a randomized clinical trial
    • D.Y. Wang, M.E. Richmond, and T.A. Quinn Optimized temporary biventricular pacing acutely improves intraoperative cardiac output after weaning from cardiopulmonary bypass: a sub-study of a randomized clinical trial J Thorac Cardiovasc Surg 141 2011 1002 1008
    • (2011) J Thorac Cardiovasc Surg , vol.141 , pp. 1002-1008
    • Wang, D.Y.1    Richmond, M.E.2    Quinn, T.A.3
  • 19
    • 77956520362 scopus 로고    scopus 로고
    • Real-time stroke volume measurements for the optimization of biventricular pacing parameters
    • J.M. Dizon, T.A. Quinn, and S.E. Cabreriza Real-time stroke volume measurements for the optimization of biventricular pacing parameters Europace 12 2010 1270 1274
    • (2010) Europace , vol.12 , pp. 1270-1274
    • Dizon, J.M.1    Quinn, T.A.2    Cabreriza, S.E.3
  • 20
    • 53349161639 scopus 로고    scopus 로고
    • Doppler echocardiographic methods for optimization of the atrioventricular delay during cardiac resynchronization therapy
    • A.D. Waggoner, L. de las Fuentes, and V.G. Davila-Roman Doppler echocardiographic methods for optimization of the atrioventricular delay during cardiac resynchronization therapy Echocardiography 25 2008 1047 1055
    • (2008) Echocardiography , vol.25 , pp. 1047-1055
    • Waggoner, A.D.1    De Las Fuentes, L.2    Davila-Roman, V.G.3
  • 21
    • 80051707128 scopus 로고    scopus 로고
    • When is an optimization not an optimization? Evaluation of clinical implications of information content (signal-to-noise ratio) in optimization of cardiac resynchronization therapy, and how to measure and maximize it
    • P.A. Pabari, K. Willson, and B. Stegemann When is an optimization not an optimization? Evaluation of clinical implications of information content (signal-to-noise ratio) in optimization of cardiac resynchronization therapy, and how to measure and maximize it Heart Fail Rev 16 2011 277 290
    • (2011) Heart Fail Rev , vol.16 , pp. 277-290
    • Pabari, P.A.1    Willson, K.2    Stegemann, B.3
  • 22
    • 84864677885 scopus 로고    scopus 로고
    • Fully automatable, reproducible, noninvasive simple plethysmographic optimization: Proof of concept and potential for implantability
    • A. Kyriacou, P.A. Pabari, and Z.I. Whinnett Fully automatable, reproducible, noninvasive simple plethysmographic optimization: proof of concept and potential for implantability Pacing Clin Electrophysiol 35 2012 948 960
    • (2012) Pacing Clin Electrophysiol , vol.35 , pp. 948-960
    • Kyriacou, A.1    Pabari, P.A.2    Whinnett, Z.I.3
  • 23
    • 84874541261 scopus 로고    scopus 로고
    • How to reliably deliver narrow individual-patient error bars for optimization of pacemaker AV or VV delay using a "pick-the-highest" strategy with haemodynamic measurements
    • D.P. Francis How to reliably deliver narrow individual-patient error bars for optimization of pacemaker AV or VV delay using a "pick-the- highest" strategy with haemodynamic measurements Int J Cardiol 163 3 2013 221 225
    • (2013) Int J Cardiol , vol.163 , Issue.3 , pp. 221-225
    • Francis, D.P.1
  • 24
    • 72549101553 scopus 로고    scopus 로고
    • Simultaneous variation of ventricular pacing site and timing with biventricular pacing during acute ventricular failure improves function by interventricular assist
    • T.A. Quinn, S.E. Cabreriza, M.E. Richmond, A.D. Weinberg, J.W. Holmes, and H.M. Spotnitz Simultaneous variation of ventricular pacing site and timing with biventricular pacing during acute ventricular failure improves function by interventricular assist Am J Physiol Heart Circ Physiol 297 2009 H2220 H2226
    • (2009) Am J Physiol Heart Circ Physiol , vol.297
    • Quinn, T.A.1    Cabreriza, S.E.2    Richmond, M.E.3    Weinberg, A.D.4    Holmes, J.W.5    Spotnitz, H.M.6
  • 25
    • 78650435075 scopus 로고    scopus 로고
    • Comparison of a non-invasive arterial pulse contour technique and echo Doppler aorta velocity-time integral on stroke volume changes in optimization of cardiac resynchronization therapy
    • I.E. van Geldorp, T. Delhaas, and B. Hermans Comparison of a non-invasive arterial pulse contour technique and echo Doppler aorta velocity-time integral on stroke volume changes in optimization of cardiac resynchronization therapy Europace 13 2011 87 95
    • (2011) Europace , vol.13 , pp. 87-95
    • Van Geldorp, I.E.1    Delhaas, T.2    Hermans, B.3
  • 27
    • 84856599968 scopus 로고    scopus 로고
    • Acute beneficial hemodynamic effects of a novel 3D-echocardiographic optimization protocol in cardiac resynchronization therapy
    • C. Sonne, L. Bott-Flügel, and S. Hauck Acute beneficial hemodynamic effects of a novel 3D-echocardiographic optimization protocol in cardiac resynchronization therapy PLoS One 7 2012 e30964
    • (2012) PLoS One , vol.7 , pp. 30964
    • Sonne, C.1    Bott-Flügel, L.2    Hauck, S.3
  • 28
    • 20244368664 scopus 로고    scopus 로고
    • Echocardio-graphic examination of atrioventricular and interventricular delay optimization in cardiac resynchronization therapy
    • M.C. Porciani, C. Dondina, and R. Macioce Echocardio-graphic examination of atrioventricular and interventricular delay optimization in cardiac resynchronization therapy Am J Cardiol 95 2005 1108 1110
    • (2005) Am J Cardiol , vol.95 , pp. 1108-1110
    • Porciani, M.C.1    Dondina, C.2    Macioce, R.3
  • 29
    • 55249123455 scopus 로고    scopus 로고
    • Interventricular delay interval optimization in cardiac resynchronization therapy guided by echocardiography versus guided by electrocardiographic QRS interval width
    • M. Bertini, M. Ziacchi, and M. Biffi Interventricular delay interval optimization in cardiac resynchronization therapy guided by echocardiography versus guided by electrocardiographic QRS interval width Am J Cardiol 102 2008 1373 1377
    • (2008) Am J Cardiol , vol.102 , pp. 1373-1377
    • Bertini, M.1    Ziacchi, M.2    Biffi, M.3
  • 30
    • 59349116485 scopus 로고    scopus 로고
    • Optimization of cardiac resynchronization therapy: Echocardiographic vs semiautomatic device algorithms
    • M. Anselmino, M. Antolini, C. Amellone, E. Piovano, R. Massa, and G. Trevi Optimization of cardiac resynchronization therapy: echocardiographic vs semiautomatic device algorithms Congest Heart Fail 15 2009 14 18
    • (2009) Congest Heart Fail , vol.15 , pp. 14-18
    • Anselmino, M.1    Antolini, M.2    Amellone, C.3    Piovano, E.4    Massa, R.5    Trevi, G.6
  • 31
    • 77952983318 scopus 로고    scopus 로고
    • Rationale and design of a randomized clinical trial to assess the safety and efficacy of frequent optimization of cardiac resynchronization therapy: The Frequent Optimization Study Using the QuickOpt Method (FREEDOM) trial
    • W.T. Abraham, D. Gras, C.M. Yu, L. Guzzo, M.S. Gupta, and FREEDOM Steering Committee Rationale and design of a randomized clinical trial to assess the safety and efficacy of frequent optimization of cardiac resynchronization therapy: the Frequent Optimization Study Using the QuickOpt Method (FREEDOM) trial Am Heart J 159 6 2010 944 948.e1
    • (2010) Am Heart J , vol.159 , Issue.6
    • Abraham, W.T.1    Gras, D.2    Yu, C.M.3    Guzzo, L.4    Gupta, M.S.5
  • 32
    • 80052489114 scopus 로고    scopus 로고
    • Results from the FREEDOM Trial - Assess the safety and efficacy of frequent optimization of cardiac resynchronization therapy
    • W.T. Abraham, D. Gras, and C.M. Yu Results from the FREEDOM Trial - assess the safety and efficacy of frequent optimization of cardiac resynchronization therapy Heart Rhythm 7 2010 S1 S524
    • (2010) Heart Rhythm , vol.7
    • Abraham, W.T.1    Gras, D.2    Yu, C.M.3
  • 33
    • 33846342565 scopus 로고    scopus 로고
    • Acute evaluation of programmer-guided AV/PV and VV delay optimization comparing an IEGM method and echocardiogram for cardiac resynchronization therapy in heart failure patients and dual-chamber ICD implants
    • J.H. Baker II, J. McKenzie III, and S. Beau Acute evaluation of programmer-guided AV/PV and VV delay optimization comparing an IEGM method and echocardiogram for cardiac resynchronization therapy in heart failure patients and dual-chamber ICD implants J Cardiovasc Electrophysiol 18 2007 185 191
    • (2007) J Cardiovasc Electrophysiol , vol.18 , pp. 185-191
    • Baker, I.I.J.H.1    McKenzie III, J.2    Beau, S.3
  • 34
    • 34247512160 scopus 로고    scopus 로고
    • A prospective comparison of AV delay programming methods for hemodynamic optimization during cardiac resynchronization therapy
    • M.R. Gold, I. Niazi, and M. Giudici A prospective comparison of AV delay programming methods for hemodynamic optimization during cardiac resynchronization therapy J Cardiovasc Electrophysiol 18 2007 490 496
    • (2007) J Cardiovasc Electrophysiol , vol.18 , pp. 490-496
    • Gold, M.R.1    Niazi, I.2    Giudici, M.3
  • 35
    • 84889089666 scopus 로고    scopus 로고
    • Boston Scientific Corporation Boston Scientific Corporation Minnesota, USA
    • Boston Scientific Corporation In-service: SmartDelay®. CRM3-5765-1110 2010 Boston Scientific Corporation Minnesota, USA
    • (2010) In-service: SmartDelay®. CRM3-5765-1110
  • 36
    • 2442479695 scopus 로고    scopus 로고
    • Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure
    • M.R. Bristow, L.A. Saxon, and J. Boehmer Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure N Engl J Med 350 2004 2140 2150
    • (2004) N Engl J Med , vol.350 , pp. 2140-2150
    • Bristow, M.R.1    Saxon, L.A.2    Boehmer, J.3
  • 37
    • 84861322305 scopus 로고    scopus 로고
    • A novel algorithm for individualized cardiac resynchronization therapy: Rationale and design of the adaptive cardiac resynchronization therapy trial
    • H. Krum, B. Lemke, and D. Birnie A novel algorithm for individualized cardiac resynchronization therapy: rationale and design of the adaptive cardiac resynchronization therapy trial Am Heart J 163 2012 747 752
    • (2012) Am Heart J , vol.163 , pp. 747-752
    • Krum, H.1    Lemke, B.2    Birnie, D.3
  • 38
    • 84883253370 scopus 로고    scopus 로고
    • Clinical outcomes with synchronized left ventricular pacing: Analysis of the adaptive CRT trial
    • D. Birnie, B. Lemke, and K. Aonuma Clinical outcomes with synchronized left ventricular pacing: analysis of the adaptive CRT trial Heart Rhythm 10 9 2013 1368 1374
    • (2013) Heart Rhythm , vol.10 , Issue.9 , pp. 1368-1374
    • Birnie, D.1    Lemke, B.2    Aonuma, K.3
  • 39
    • 80053389479 scopus 로고    scopus 로고
    • Adjusting the timing of left-ventricular pacing using electrocardiogram and device electrograms
    • Y. Khaykin, D. Exner, D. Birnie, J. Sapp, S. Aggarwal, and A. Sambelashvili Adjusting the timing of left-ventricular pacing using electrocardiogram and device electrograms Europace 13 2011 1464 1470
    • (2011) Europace , vol.13 , pp. 1464-1470
    • Khaykin, Y.1    Exner, D.2    Birnie, D.3    Sapp, J.4    Aggarwal, S.5    Sambelashvili, A.6
  • 40
    • 78249259383 scopus 로고    scopus 로고
    • Optimal atrioventricular delay in CRT patients can be approximated using surface electrocardiography and device electrograms
    • R.C. Jones, T. Svinarich, and A. Rubin Optimal atrioventricular delay in CRT patients can be approximated using surface electrocardiography and device electrograms J Cardiovasc Electrophysiol 21 2010 1226 1232
    • (2010) J Cardiovasc Electrophysiol , vol.21 , pp. 1226-1232
    • Jones, R.C.1    Svinarich, T.2    Rubin, A.3
  • 41
    • 0033562889 scopus 로고    scopus 로고
    • How high can a correlation coefficient be? Effects of limited reproducibility of common cardiological measures
    • D.P. Francis, A.J. Coats, and D.G. Gibson How high can a correlation coefficient be? Effects of limited reproducibility of common cardiological measures Int J Cardiol 69 1999 May 15 185 189
    • (1999) Int J Cardiol , vol.69 , pp. 185-189
    • Francis, D.P.1    Coats, A.J.2    Gibson, D.G.3
  • 42
    • 50449103135 scopus 로고    scopus 로고
    • The optimized V-V interval determined by interventricular conduction times versus invasive measurement by LVdP/dtMAX
    • B.M. van Gelder, A. Meijer, and F.A. Bracke The optimized V-V interval determined by interventricular conduction times versus invasive measurement by LVdP/dtMAX J Cardiovasc Electrophysiol 19 2008 939 944
    • (2008) J Cardiovasc Electrophysiol , vol.19 , pp. 939-944
    • Van Gelder, B.M.1    Meijer, A.2    Bracke, F.A.3
  • 43
    • 84867859165 scopus 로고    scopus 로고
    • Investigation of a novel algorithm for synchronized left-ventricular pacing and ambulatory optimization of cardiac resynchronization therapy: Results of the adaptive CRT trial
    • D.O. Martin, B. Lemke, and D. Birnie Investigation of a novel algorithm for synchronized left-ventricular pacing and ambulatory optimization of cardiac resynchronization therapy: results of the adaptive CRT trial Heart Rhythm 9 2012 1807 1814
    • (2012) Heart Rhythm , vol.9 , pp. 1807-1814
    • Martin, D.O.1    Lemke, B.2    Birnie, D.3
  • 45
    • 33745343276 scopus 로고    scopus 로고
    • Determination of optimal atrioventricular delay for cardiac resynchronization therapy using acute non-invasive blood pressure
    • Z.I. Whinnett, J.E. Davies, and K. Willson Determination of optimal atrioventricular delay for cardiac resynchronization therapy using acute non-invasive blood pressure Europace 8 2006 358 366
    • (2006) Europace , vol.8 , pp. 358-366
    • Whinnett, Z.I.1    Davies, J.E.2    Willson, K.3
  • 46
    • 0037012355 scopus 로고    scopus 로고
    • Cardiac resynchronization therapy restores optimal atrioventricular mechanical timing in heart failure patients with ventricular conduction delay
    • A. Auricchio, J. Ding, and J.C. Spinelli Cardiac resynchronization therapy restores optimal atrioventricular mechanical timing in heart failure patients with ventricular conduction delay J Am Coll Cardiol 39 2002 1163 1169
    • (2002) J Am Coll Cardiol , vol.39 , pp. 1163-1169
    • Auricchio, A.1    Ding, J.2    Spinelli, J.C.3
  • 47
    • 0035910061 scopus 로고    scopus 로고
    • Effect of resynchronization therapy stimulation site on the systolic function of heart failure patients
    • C. Butter, A. Auricchio, and C. Stellbrink Effect of resynchronization therapy stimulation site on the systolic function of heart failure patients Circulation 104 2001 3026 3029
    • (2001) Circulation , vol.104 , pp. 3026-3029
    • Butter, C.1    Auricchio, A.2    Stellbrink, C.3
  • 48
    • 33751169368 scopus 로고    scopus 로고
    • Left ventricular pacing site-timing optimization during biventricular pacing using a multi-electrode patch
    • G. Berberian, S.E. Cabreriza, T.A. Quinn, C.A. Garofalo, and H.M. Spotnitz Left ventricular pacing site-timing optimization during biventricular pacing using a multi-electrode patch Ann Thorac Surg 82 2006 2292 2294
    • (2006) Ann Thorac Surg , vol.82 , pp. 2292-2294
    • Berberian, G.1    Cabreriza, S.E.2    Quinn, T.A.3    Garofalo, C.A.4    Spotnitz, H.M.5
  • 49
    • 77956509700 scopus 로고    scopus 로고
    • Can optimization of pacing settings compensate for a non-optimal left ventricular pacing site?
    • M.D. Bogaard, P.A. Doevendans, and G.E. Leenders Can optimization of pacing settings compensate for a non-optimal left ventricular pacing site? Europace 12 2010 1262 1269
    • (2010) Europace , vol.12 , pp. 1262-1269
    • Bogaard, M.D.1    Doevendans, P.A.2    Leenders, G.E.3
  • 50
    • 84864671000 scopus 로고    scopus 로고
    • Precision of a parabolic optimum calculated from noisy biological data, and implications for quantitative optimization of biventricular pacemakers (cardiac resynchronization therapy)
    • D.P. Francis Precision of a parabolic optimum calculated from noisy biological data, and implications for quantitative optimization of biventricular pacemakers (cardiac resynchronization therapy) Appl Math 2 2011 1497 1506
    • (2011) Appl Math , vol.2 , pp. 1497-1506
    • Francis, D.P.1
  • 51
    • 84865760157 scopus 로고    scopus 로고
    • A randomized pilot study of optimization of cardiac resynchronization therapy in sinus rhythm patients using a peak endocardial acceleration sensor vs standard methods
    • P. Ritter, P.P. Delnoy, and L. Padeletti A randomized pilot study of optimization of cardiac resynchronization therapy in sinus rhythm patients using a peak endocardial acceleration sensor vs. standard methods Europace 14 2012 1324 1333
    • (2012) Europace , vol.14 , pp. 1324-1333
    • Ritter, P.1    Delnoy, P.P.2    Padeletti, L.3
  • 52
    • 33645734785 scopus 로고    scopus 로고
    • Tailoring cardiac resynchronization therapy using interventricular asynchrony. Validation of a simple model
    • X.A. Verbeek, A. Auricchio, and Y. Yu Tailoring cardiac resynchronization therapy using interventricular asynchrony. Validation of a simple model Am J Physiol Heart Circ Physiol 290 2006 H968 H977
    • (2006) Am J Physiol Heart Circ Physiol , vol.290
    • Verbeek, X.A.1    Auricchio, A.2    Yu, Y.3
  • 53
    • 84863643031 scopus 로고    scopus 로고
    • Vectorcardiography as a tool for easy optimization of cardiac resynchronization therapy in canine left bundle branch block hearts
    • C.J. van Deursen, M. Strik, and L.M. Rademakers Vectorcardiography as a tool for easy optimization of cardiac resynchronization therapy in canine left bundle branch block hearts Circ Arrhythm Electrophysiol 5 2012 544 552
    • (2012) Circ Arrhythm Electrophysiol , vol.5 , pp. 544-552
    • Van Deursen, C.J.1    Strik, M.2    Rademakers, L.M.3
  • 54
    • 84867394360 scopus 로고    scopus 로고
    • Cardiac resynchronization therapy beyond nominal settings: Who needs individual programming of the atrioventricular and interventricular delay?
    • M.D. Bogaard, M. Meine, A.E. Tuinenburg, B. Maskara, P. Loh, and P.A. Doevendans Cardiac resynchronization therapy beyond nominal settings: who needs individual programming of the atrioventricular and interventricular delay? Europace 14 2012 1746 1753
    • (2012) Europace , vol.14 , pp. 1746-1753
    • Bogaard, M.D.1    Meine, M.2    Tuinenburg, A.E.3    Maskara, B.4    Loh, P.5    Doevendans, P.A.6
  • 55
    • 84867367814 scopus 로고    scopus 로고
    • Atrioventricular and interventricular delay optimization and response quantification in biventricular pacing: Arrival of reliable clinical algorithms and research protocols, and how to distinguish them from unreliable counterparts
    • B. Stegemann, and D.P. Francis Atrioventricular and interventricular delay optimization and response quantification in biventricular pacing: arrival of reliable clinical algorithms and research protocols, and how to distinguish them from unreliable counterparts Europace 14 2012 1679 1683
    • (2012) Europace , vol.14 , pp. 1679-1683
    • Stegemann, B.1    Francis, D.P.2
  • 57
    • 84881474693 scopus 로고    scopus 로고
    • Choosing between velocity-time-integral ratio and peak velocity ratio for calculation of the dimensionless index (or aortic valve area) in serial follow-up of aortic stenosis
    • J.A. Finegold, C.H. Manisty, F. Cecaro, N. Sutaria, J. Mayet, and D.P. Francis Choosing between velocity-time-integral ratio and peak velocity ratio for calculation of the dimensionless index (or aortic valve area) in serial follow-up of aortic stenosis Int J Cardiol 167 2013 1524 1531
    • (2013) Int J Cardiol , vol.167 , pp. 1524-1531
    • Finegold, J.A.1    Manisty, C.H.2    Cecaro, F.3    Sutaria, N.4    Mayet, J.5    Francis, D.P.6
  • 58
    • 84880924293 scopus 로고    scopus 로고
    • A systematic approach to designing reliable VV optimization methodology: Assessment of internal validity of echocardiographic, electrocardiographic and haemodynamic optimization of cardiac resynchronization therapy
    • A. Kyriacou, M.E. Li Kam Wa, P.A. Pabari, B. Unsworth, R. Baruah, K. Willson, N.S. Peters, P. Kanagaratnam, A.D. Hughes, J. Mayet, Z.I. Whinnett, and D.P. Francis A systematic approach to designing reliable VV optimization methodology: assessment of internal validity of echocardiographic, electrocardiographic and haemodynamic optimization of cardiac resynchronization therapy Int J Cardiol 167 2013 954 964
    • (2013) Int J Cardiol , vol.167 , pp. 954-964
    • Kyriacou, A.1    Li Kam Wa, M.E.2    Pabari, P.A.3    Unsworth, B.4    Baruah, R.5    Willson, K.6    Peters, N.S.7    Kanagaratnam, P.8    Hughes, A.D.9    Mayet, J.10    Whinnett, Z.I.11    Francis, D.P.12
  • 59
    • 84858326999 scopus 로고    scopus 로고
    • The acute effects of changes to AV delay on BP and stroke volume: Potential implications for design of pacemaker optimization protocols
    • C.H. Manisty, A. Al-Hussaini, and B. Unsworth The acute effects of changes to AV delay on BP and stroke volume: potential implications for design of pacemaker optimization protocols Circ Arrhythm Electrophysiol 5 2012 122 130
    • (2012) Circ Arrhythm Electrophysiol , vol.5 , pp. 122-130
    • Manisty, C.H.1    Al-Hussaini, A.2    Unsworth, B.3
  • 60
    • 84869089072 scopus 로고    scopus 로고
    • Response of mean arterial pressure to temporary biventricular pacing after chest closure during cardiac surgery
    • B.J. Rubinstein, D.Y. Wang, and S.E. Cabreriza Response of mean arterial pressure to temporary biventricular pacing after chest closure during cardiac surgery J Thorac Cardiovasc Surg 144 2012 1445 1452
    • (2012) J Thorac Cardiovasc Surg , vol.144 , pp. 1445-1452
    • Rubinstein, B.J.1    Wang, D.Y.2    Cabreriza, S.E.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.