A comprehensive evaluation of the performance of the test strip technology for OneTouch Verio glucose meter systems

Author keywords

[No Author keywords available]

Indexed keywords

GLUCOSE;

EID: 84864663289     PISSN: 15209156     EISSN: 15578593     Source Type: Journal    
DOI: 10.1089/dia.2011.0260     Document Type: Article
Times cited : (14)

References (18)
  • 1
    • 78651338445 scopus 로고    scopus 로고
    • Standards of medical care in diabetes-2011
    • American Diabetes Association
    • American Diabetes Association: Standards of medical care in diabetes-2011. Diabetes Care 2011;34(Suppl 1):S11-S61.
    • (2011) Diabetes Care , vol.34 , Issue.1
  • 2
    • 70350209515 scopus 로고    scopus 로고
    • Factors affecting blood glucose monitoring: Sources of errors in measurement
    • Ginsberg BH: Factors affecting blood glucose monitoring: sources of errors in measurement. J Diabetes Sci Technol 2009;3:903-913.
    • (2009) J. Diabetes. Sci. Technol. , vol.3 , pp. 903-913
    • Ginsberg, B.H.1
  • 3
    • 77955480844 scopus 로고    scopus 로고
    • Regulatory controversies surround blood glucose monitoring devices
    • Klonoff DC: Regulatory controversies surround blood glucose monitoring devices. J Diabetes Sci Technol 2010;4:231-235.
    • (2010) J. Diabetes. Sci. Technol. , vol.4 , pp. 231-235
    • Klonoff, D.C.1
  • 4
    • 84864702968 scopus 로고    scopus 로고
    • Improved glucose measurements using a thin layer cell and tri-pulse electrochemical detection
    • Hodges AM, Chatelier RC: Improved glucose measurements using a thin layer cell and tri-pulse electrochemical detection. Clin Chem 2008;54:1934.
    • (2008) Clin. Chem. , vol.54 , pp. 1934
    • Hodges, A.M.1    Chatelier, R.C.2
  • 5
    • 57349104419 scopus 로고    scopus 로고
    • Self-monitoring of blood glucose SMBG in insulin-and non-insulin-using adults with diabetes: Consensus recommendations for improving SMBG accuracy utilization and research
    • Hirsch IB, Bode BW, Childs BP, Close KL, Fisher WA, Gavin JR, Ginsberg BH, Raine CH, Verderese CA: Self-monitoring of blood glucose (SMBG) in insulin-and non-insulin-using adults with diabetes: consensus recommendations for improving SMBG accuracy, utilization, and research. Diabetes Technol Ther 2008;10:419-439.
    • (2008) Diabetes Technol. Ther. , vol.10 , pp. 419-439
    • Hirsch, I.B.1    Bode, B.W.2    Childs, B.P.3    Close, K.L.4    Fisher, W.A.5    Gavin, J.R.6    Ginsberg, B.H.7    Raine, C.H.8    Verderese, C.A.9
  • 6
    • 84864678385 scopus 로고    scopus 로고
    • Clinical patterns of low blood glucose identified by a pattern algorithm may predict increased risk of severe hypoglycemia in the following 24-hour period abstract
    • Lee-Davey J, Alexander, Raja P: Clinical patterns of low blood glucose identified by a pattern algorithm may predict increased risk of severe hypoglycemia in the following 24-hour period [abstract]. Diabetes Technol Ther 2011;13:264-265.
    • (2011) Diabetes Technol. Ther. , vol.13 , pp. 264-265
    • Lee-Davey, J.1    Alexander Raja, P.2
  • 8
    • 84864695252 scopus 로고    scopus 로고
    • Hematocrit compensation in electrochemical blood glucose monitoring systems
    • Teodorczyk M, Cardosi M, Setford S: Hematocrit compensation in electrochemical blood glucose monitoring systems. J Diabetes Sci Technol 2012;6:648-655.
    • (2012) J. Diabetes. Sci. Technol. , vol.6 , pp. 648-655
    • Teodorczyk, M.1    Cardosi, M.2    Setford, S.3
  • 9
    • 0004217479 scopus 로고    scopus 로고
    • E In Vitro Diagnostic Test Systems-Requirements for Blood-Glucose Monitoring Systems for Self-Testing in Managing Diabetes Mellitus Geneva: International Organization for Standardization 2003
    • International Organization for Standardization: ISO 15197:2003(E): In Vitro Diagnostic Test Systems-Requirements for Blood-Glucose Monitoring Systems for Self-Testing in Managing Diabetes Mellitus. Geneva: International Organization for Standardization, 2003.
    • (2003) International Organization for Standardization: ISO 15197
  • 10
    • 84864649154 scopus 로고    scopus 로고
    • Clinical and laboratory standards institute
    • 2nd ed. Publication number CLSI EP07-A2. Wayne, PA: Clinical and Laboratory Standards Institute
    • Clinical and Laboratory Standards Institute: Interference Testing in Clinical Chemistry, Approved Guideline, 2nd ed. Publication number CLSI EP07-A2. Wayne, PA: Clinical and Laboratory Standards Institute, 2005.
    • (2005) Interference Testing in Clinical Chemistry Approved Guideline
  • 11
    • 0001844948 scopus 로고
    • Kinetics of icodextrin
    • Davies D: Kinetics of icodextrin. Perit Dial Int 1994;14(Suppl 2):S45-S50.
    • (1994) Perit. Dial. Int. , vol.14 , Issue.2
    • Davies, D.1
  • 12
    • 84872616897 scopus 로고    scopus 로고
    • accessed January 19
    • Mayo Clinic: Tolazamide (Oral Route). www.mayoclinic .com/health/drug- information/DR602751/DSECTION =proper-use (accessed January 19, 2012).
    • (2012) Mayo Clinic Tolazamide Oral Route
  • 13
    • 0031926062 scopus 로고    scopus 로고
    • Fully automated assay for total homocysteine cysteine cysteinylglycine glutathione cysteamine and 2-mercaptopropionylglycine in plasma and urine
    • Pastore A, Massoud R, Motti C, Lo Russo A, Fucci G, Cortese C, Federici G: Fully automated assay for total homocysteine, cysteine, cysteinylglycine, glutathione, cysteamine, and 2-mercaptopropionylglycine in plasma and urine. Clin Chem 1998;44:825-832.
    • (1998) Clin. Chem. , vol.44 , pp. 825-832
    • Pastore, A.1    Massoud, R.2    Motti, C.3    Lo Russo, A.4    Fucci, G.5    Cortese, C.6    Federici, G.7
  • 14
    • 78349287412 scopus 로고    scopus 로고
    • The food and drug administration is now preparing to establish tighter performance requirements for blood glucose monitors
    • Klonoff DC: The Food and Drug Administration is now preparing to establish tighter performance requirements for blood glucose monitors. J Diabetes Sci Technol 2010;4: 499-504.
    • (2010) J. Diabetes. Sci. Technol. , vol.4 , pp. 499-504
    • Klonoff, D.C.1
  • 15
    • 78149382135 scopus 로고    scopus 로고
    • Impact of blood glucose selfmonitoring errors on glucose variability, risk for hypoglycemia and average glucose control in type 1 diabetes: An in silico study
    • Breton MD, Kovatchev BP: Impact of blood glucose selfmonitoring errors on glucose variability, risk for hypoglycemia, and average glucose control in type 1 diabetes: an in silico study. J Diabetes Sci Technol 2010;4:562-570.
    • (2010) J. Diabetes. Sci. Technol. , vol.4 , pp. 562-570
    • Breton, M.D.1    Kovatchev, B.P.2
  • 16
    • 0003945589 scopus 로고
    • Burtis CA, Ashwood ER, eds. 2nd ed. Philadelphia: W.B. Saunders Co.
    • Burtis CA, Ashwood ER, eds.: Tietz Textbook of Clinical Chemistry, 2nd ed. Philadelphia: W.B. Saunders Co., 1994.
    • (1994) Tietz Textbook of Clinical Chemistry
  • 18
    • 79951717523 scopus 로고    scopus 로고
    • Structured self-monitoring of blood glucose significantly reduces A1C levels in poorly controlled noninsulin-treated type 2 diabetes: Results from the structured testing program study
    • Polonsky WH, Fisher L, Schikman CH, Hinnen DA, Parkin CG, Jelsovsky Z, Petersen B, Schweitzer M, Wagner RS: Structured self-monitoring of blood glucose significantly reduces A1C levels in poorly controlled, noninsulin-treated type 2 diabetes: results from the Structured Testing Program study. Diabetes Care 2011;34:262-267.
    • (2011) Diabetes Care , vol.34 , pp. 262-267
    • Polonsky, W.H.1    Fisher, L.2    Schikman, C.H.3    Hinnen, D.A.4    Parkin, C.G.5    Jelsovsky, Z.6    Petersen, B.7    Schweitzer, M.8    Wagner, R.S.9


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.