Volumn 4, Issue 3, 2010, Pages 562-570

Impact of blood glucose self-monitoring errors on glucose variability, risk for hypoglycemia, and average glucose control in type 1 diabetes: An in silico study

Author keywords

Blood glucose; Diabetes; Glycemic control; Hypoglycemia; Meter accuracy; Simulations

Indexed keywords

ARTICLE; BLOOD; BLOOD GLUCOSE MONITORING; COMPUTER SIMULATION; DRUG SELF ADMINISTRATION; GLUCOSE BLOOD LEVEL; HUMAN; HYPOGLYCEMIA; INSULIN DEPENDENT DIABETES MELLITUS; RISK FACTOR;

EID: 78149382135     PISSN: None     EISSN: 19322968     Source Type: Journal    
DOI: 10.1177/193229681000400309     Document Type: Article
Times cited : (94)

References (40)
  • 1
    • 0027370108 scopus 로고
    • The effect of intensive treatment of diabetes on the development and progression of long-term complications of insulin-dependent diabetes mellitus
    • The Diabetes Control and Complications Trial Research Group
    • The Diabetes Control and Complications Trial Research Group. The effect of intensive treatment of diabetes on the development and progression of long-term complications of insulin-dependent diabetes mellitus. N Engl J Med. 1993;329(14):977-86.
    • (1993) N Engl J Med , vol.329 , Issue.14 , pp. 977-986
  • 2
    • 0027952951 scopus 로고
    • Mortality and treatment side-effects during long-term intensified conventional insulin treatment in the Stockholm Diabetes Intervention Study
    • Reichard P, Phil M. Mortality and treatment side effects during long-term intensified conventional insulin treatment in the Stockholm Diabetes Intervention study. Diabetes. 1994;43(2):313-7. (Pubitemid 24033687)
    • (1994) Diabetes , vol.43 , Issue.2 , pp. 313-317
    • Reichard, P.1    Pihl, M.2
  • 3
    • 0032511583 scopus 로고    scopus 로고
    • Intensive bloodglucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes
    • UK Prospective Diabetes Study Group (UKPDS)
    • UK Prospective Diabetes Study Group (UKPDS). Intensive bloodglucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes. Lancet. 1998;352(9131):837-53.
    • (1998) Lancet , vol.352 , Issue.9131 , pp. 837-853
  • 4
    • 0020674603 scopus 로고
    • Identification of type I diabetic patients at increased risk for hypoglycemia during intensive therapy
    • White NH, Skor DA, Cryer PE, Levandoski L, Santiago JV. Identification of type I diabetic patients at increased risk for hypoglycemia during intensive therapy. N Engl J Med. 1983;308(9):485-91. (Pubitemid 13195475)
    • (1983) New England Journal of Medicine , vol.308 , Issue.9 , pp. 485-491
    • White, N.H.1    Skor, D.A.2    Cryer, P.E.3
  • 5
    • 0021853211 scopus 로고
    • Glucose counterregulation, hypoglycemia, and intensive insulin therapy in diabetes mellitus
    • Cryer PE, Gerich JE. Glucose counterregulation, hypoglycemia, and intensive therapy of diabetes mellitus. N Engl J Med. 1985;313(4):232-41. (Pubitemid 15068639)
    • (1985) New England Journal of Medicine , vol.313 , Issue.4 , pp. 232-241
    • Cryer, P.E.1    Gerich, J.E.2
  • 6
    • 0024269080 scopus 로고
    • Effect of intensive insulin therapy on glycemic thresholds for counterregulatory hormone release
    • Amiel SA, Sherwin RS, Simonson DC, Tamborlane WV. Effect of intensive insulin therapy on glycemic thresholds for counterregulatory hormone release. Diabetes. 1988;37(7):901-7. (Pubitemid 19043718)
    • (1988) Diabetes , vol.37 , Issue.7 , pp. 901-907
    • Amiel, S.A.1    Sherwin, R.S.2    Simonson, D.C.3    Tamborlane, W.V.4
  • 7
    • 0023216411 scopus 로고
    • Defective glucose counterregulation after strict glycemic control of insulin-dependent diabetes mellitus
    • Amiel SA, Tamborlane WV, Simonson DC, Sherwin RS. Defective glucose counterregulation after strict glycemic control of insulin-dependent diabetes mellitus. N Engl J Med. 1987;316(22):1376-83. (Pubitemid 17079681)
    • (1987) New England Journal of Medicine , vol.316 , Issue.22 , pp. 1376-1383
    • Amiel, S.A.1    Tamborlane, W.V.2    Simonson, D.C.3    Sherwin, R.S.4
  • 8
    • 0026504154 scopus 로고
    • Iatrogenic hypoglycemia as a cause of hypoglycemia-associated autonomic failure in IDDM. A vicious cycle
    • Cryer PE. Iatrogenic hypoglycemia as a cause of hypoglycemia-associated autonomic failure in IDDM. A vicious cycle. Diabetes. 1992;41(3):255-60.
    • (1992) Diabetes , vol.41 , Issue.3 , pp. 255-260
    • Cryer, P.E.1
  • 9
    • 0028607457 scopus 로고
    • Banting Lecture. Hypoglycemia: The limiting factor in the management of IDDM
    • Cryer PE. Banting Lecture. Hypoglycemia: the limiting factor in the management of IDDM. Diabetes. 1994;43(11):1378-89.
    • (1994) Diabetes , vol.43 , Issue.11 , pp. 1378-1389
    • Cryer, P.E.1
  • 10
    • 0036020278 scopus 로고    scopus 로고
    • Hypoglycaemia: The limiting factor in the glycaemic management of Type I and Type II diabetes
    • DOI 10.1007/s00125-002-0822-9
    • Cryer PE. Hypoglycaemia: the limiting factor in the glycaemic management of type I and type II diabetes. Diabetologia. 2002;45(7):937-48. (Pubitemid 34762188)
    • (2002) Diabetologia , vol.45 , Issue.7 , pp. 937-948
    • Cryer, P.E.1
  • 11
    • 0027519381 scopus 로고
    • Lessons from the diabetes control and complications trial
    • Santiago JV. Lessons from the Diabetes Control and Complications Trial. Diabetes. 1993;42(11):1549-54. (Pubitemid 23318108)
    • (1993) Diabetes , vol.42 , Issue.11 , pp. 1549-1554
    • Santiago, J.V.1
  • 12
    • 77951733801 scopus 로고    scopus 로고
    • A review of standards and statistics used to describe blood glucose monitor performance
    • Krouwer JS, Cembrowski GS. A review of standards and statistics used to describe blood glucose monitor performance. J Diabetes Sci Technol. 2010;4(1):75-83.
    • (2010) J Diabetes Sci Technol , vol.4 , Issue.1 , pp. 75-83
    • Krouwer, J.S.1    Cembrowski, G.S.2
  • 16
    • 40249097584 scopus 로고    scopus 로고
    • Evaluating the Accuracy of Modern Glucose Meters
    • DOI 10.1016/S1557-0843(08)80004-8, PII S1557084308800048
    • Bergenstal RM. Evaluating the accuracy of modern glucose meters. Insulin. 2008;3:5-14. (Pubitemid 351331167)
    • (2008) Insulin , vol.3 , Issue.1 , pp. 5-14
    • Bergenstal, R.M.1
  • 18
    • 57349104419 scopus 로고    scopus 로고
    • Self-monitoring of blood glucose (SMBG) in insulin- and non-insulin-using adults with diabetes: Consensus recommendations for improving SMBG accuracy, utilization, and research
    • Hirsch IB, Bode BW, Childs BP, Close KL, Fisher WA, Gavin JR, Ginsberg BH, Raine CH, Verderese CA. Self-monitoring of blood glucose (SMBG) in insulin- and non-insulin-using adults with diabetes: consensus recommendations for improving SMBG accuracy, utilization, and research. Diabetes Technol Ther. 2008;10(6):419-39.
    • (2008) Diabetes Technol Ther , vol.10 , Issue.6 , pp. 419-439
    • Hirsch, I.B.1    Bode, B.W.2    Childs, B.P.3    Close, K.L.4    Fisher, W.A.5    Gavin, J.R.6    Ginsberg, B.H.7    Raine, C.H.8    Verderese, C.A.9
  • 19
    • 77954685027 scopus 로고    scopus 로고
    • Analytical and clinical performance of blood glucose monitors
    • Boren SA, Clarke WL. Analytical and clinical performance of blood glucose monitors. J Diabetes Sci Technol. 2010;4(1):84-97.
    • (2010) J Diabetes Sci Technol , vol.4 , Issue.1 , pp. 84-97
    • Boren, S.A.1    Clarke, W.L.2
  • 20
    • 67749100556 scopus 로고    scopus 로고
    • Predicted blood glucose from insulin administration based on values from miscoded glucose meters
    • Raine CH, Pardo S, Parkes JL. Predicted blood glucose from insulin administration based on values from miscoded glucose meters. J Diabetes Sci Technol. 2008;2(4):557-62.
    • (2008) J Diabetes Sci Technol , vol.2 , Issue.4 , pp. 557-562
    • Raine, C.H.1    Pardo, S.2    Parkes, J.L.3
  • 21
    • 0035130886 scopus 로고    scopus 로고
    • Quality specifications for glucose meters: Assessment by simulation modeling of errors in insulin dose
    • Boyd JC, Bruns DE. Quality specifications for glucose meters: assessment by simulation modeling of errors in insulin dose. Clin Chem. 2001;47(2):209-14. (Pubitemid 32142918)
    • (2001) Clinical Chemistry , vol.47 , Issue.2 , pp. 209-214
    • Boyd, J.C.1    Bruns, D.E.2
  • 22
    • 71549156881 scopus 로고    scopus 로고
    • Monte Carlo simulation in establishing analytical quality requirements for clinical laboratory tests: Meeting clinical needs
    • Boyd JC, Bruns DE. Monte Carlo simulation in establishing analytical quality requirements for clinical laboratory tests: meeting clinical needs. Methods Enzymol. 2009;467:411-33.
    • (2009) Methods Enzymol , vol.467 , pp. 411-433
    • Boyd, J.C.1    Bruns, D.E.2
  • 25
    • 69949146908 scopus 로고    scopus 로고
    • In silico preclinical trials: A proof of concept in closed-loop control of type 1 diabetes
    • Kovatchev BP, Breton M, Dalla Man CD, Cobelli C. In silico preclinical trials: a proof of concept in closed-loop control of type 1 diabetes. J Diabetes Sci Technol. 2009;3(1):44-55.
    • (2009) J Diabetes Sci Technol , vol.3 , Issue.1 , pp. 44-55
    • Kovatchev, B.P.1    Breton, M.2    Dalla Man, C.D.3    Cobelli, C.4
  • 29
    • 48149095259 scopus 로고    scopus 로고
    • Translating the hemoglobin A1c assay into estimated average glucose values
    • A1c-Derived Average Glucose Study Group
    • Nathan DM, Kuenen J, Borg R, Zheng H, Schoenfeld D, Heine RJ; A1c-Derived Average Glucose Study Group. Translating the hemoglobin A1c assay into estimated average glucose values. Diabetes Care. 2008;31(8):1473-8.
    • (2008) Diabetes Care , vol.31 , Issue.8 , pp. 1473-1478
    • Nathan, D.M.1    Kuenen, J.2    Borg, R.3    Zheng, H.4    Schoenfeld, D.5    Heine, R.J.6
  • 32
    • 70349240850 scopus 로고    scopus 로고
    • Statistical tools to analyze continuous glucose monitor data
    • Clarke W, Kovatchev B. Statistical tools to analyze continuous glucose monitor data. Diabetes Technol Ther. 2009;11 Suppl 1:S45-54.
    • (2009) Diabetes Technol Ther , vol.11 , Issue.SUPPL. 1
    • Clarke, W.1    Kovatchev, B.2
  • 35
    • 0242300709 scopus 로고    scopus 로고
    • Archimedes: A trial-validate model of diabetes
    • Eddy DM, Schlessinger L. Archimedes: a trial-validate model of diabetes. Diabetes Care. 2003;26(11):3093-101.
    • (2003) Diabetes Care , vol.26 , Issue.11 , pp. 3093-3101
    • Eddy, D.M.1    Schlessinger, L.2
  • 36
    • 0242300707 scopus 로고    scopus 로고
    • Validation of the Archimedes Diabetes Model
    • DOI 10.2337/diacare.26.11.3102
    • Eddy DM, Schlessinger L. Validation of the archimedes diabetes model. Diabetes Care. 2003;26(11):3102-10. (Pubitemid 37339505)
    • (2003) Diabetes Care , vol.26 , Issue.11 , pp. 3102-3110
    • Eddy, D.M.1    Schlessinger, L.2
  • 37
    • 77249175519 scopus 로고    scopus 로고
    • Overnight closed-loop insulin delivery with model predictive control: Assessment of hypoglycemia and hyperglycemia risk using simulation studies
    • Wilinska ME, Budiman ES, Taub MB, Elleri D, Allen JM, Acerini CL, Dunger DB, Hovorka R. Overnight closed-loop insulin delivery with model predictive control: assessment of hypoglycemia and hyperglycemia risk using simulation studies. J Diabetes Sci Technol. 2009;3(5):1109-20.
    • (2009) J Diabetes Sci Technol , vol.3 , Issue.5 , pp. 1109-1120
    • Wilinska, M.E.1    Budiman, E.S.2    Taub, M.B.3    Elleri, D.4    Allen, J.M.5    Acerini, C.L.6    Dunger, D.B.7    Hovorka, R.8


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.