Volumn 38, Issue 6, 2012, Pages 971-977

Intrasubject repeatability of corneal morphology measurements obtained with a new Scheimpflug photography-based system

Author keywords

[No Author keywords available]

Indexed keywords

ADOLESCENT; ADULT; AGED; ANTERIOR EYE CHAMBER DEPTH; ANTERIOR EYE SEGMENT; ARTICLE; CHILD; CORNEA; CORNEA THICKNESS; CORRELATION ANALYSIS; EYE EXAMINATION; EYE PHOTOGRAPHY; FEMALE; HUMAN; HUMAN EXPERIMENT; MALE; NORMAL HUMAN; PACHYMETRY; PLACIDO DISK TECHNOLOGY; POSTERIOR EYE CHAMBER; PRIORITY JOURNAL; SCHEIMPFLUG PHOTOGRAPHY; SCHOOL CHILD; SPAIN;

EID: 84861566302     PISSN: 08863350     EISSN: 18734502     Source Type: Journal    
DOI: 10.1016/j.jcrs.2011.12.029     Document Type: Article
Times cited : (56)

References (26)
  • 1
    • 29544435798 scopus 로고    scopus 로고
    • The shape of the anterior and posterior surface of the aging human cornea
    • M. Dubbelman, V.A.D.P. Sicam, and G.L. van der Heijde The shape of the anterior and posterior surface of the aging human cornea Vision Res 46 2006 993 1001
    • (2006) Vision Res , vol.46 , pp. 993-1001
    • Dubbelman, M.1    Sicam, V.A.D.P.2    Van Der Heijde, G.L.3
  • 2
    • 33750315860 scopus 로고    scopus 로고
    • Repeatability and reproducibility of posterior corneal curvature measurements by combined scanning-slit and Placido-disc topography after LASIK
    • M.J. Maldonado, J.C. Nieto, M. Díez-Cuenca, and D.P. Piñero Repeatability and reproducibility of posterior corneal curvature measurements by combined scanning-slit and Placido-disc topography after LASIK Ophthalmology 113 2006 1918 1926
    • (2006) Ophthalmology , vol.113 , pp. 1918-1926
    • Maldonado, M.J.1    Nieto, J.C.2    Díez-Cuenca, M.3    Piñero, D.P.4
  • 4
    • 0034899732 scopus 로고    scopus 로고
    • Real-time optical coherence tomography of the anterior segment at 1310 nm
    • Available at: Accessed January 28, 2012
    • S. Radhakrishnan, A.M. Rollins, J.E. Roth, S. Yazdanfar, V. Westphal, D.S. Bardenstein, and J.A. Izatt Real-time optical coherence tomography of the anterior segment at 1310 nm Arch Ophthalmol 119 2001 1179 1185 Available at: http://archopht.ama-assn.org/cgi/reprint/119/8/1179 Accessed January 28, 2012
    • (2001) Arch Ophthalmol , vol.119 , pp. 1179-1185
    • Radhakrishnan, S.1    Rollins, A.M.2    Roth, J.E.3    Yazdanfar, S.4    Westphal, V.5    Bardenstein, D.S.6    Izatt, J.A.7
  • 5
    • 37449017944 scopus 로고    scopus 로고
    • Anterior segment biometry with 2 imaging technologies: Very-high-frequency ultrasound scanning versus optical coherence tomography
    • D.P. Piñero, A.B. Plaza, and J.L. Alió Anterior segment biometry with 2 imaging technologies: very-high-frequency ultrasound scanning versus optical coherence tomography J Cataract Refract Surg 34 2008 95 102
    • (2008) J Cataract Refract Surg , vol.34 , pp. 95-102
    • Piñero, D.P.1    Plaza, A.B.2    Alió, J.L.3
  • 6
    • 78751489525 scopus 로고    scopus 로고
    • Repeatability of automatic measurements performed by a dual Scheimpflug analyzer in unoperated and post-refractive surgery eyes
    • G. Savini, M. Carbonelli, P. Barboni, and K.J. Hoffer Repeatability of automatic measurements performed by a dual Scheimpflug analyzer in unoperated and post-refractive surgery eyes J Cataract Refract Surg 37 2011 302 309
    • (2011) J Cataract Refract Surg , vol.37 , pp. 302-309
    • Savini, G.1    Carbonelli, M.2    Barboni, P.3    Hoffer, K.J.4
  • 7
    • 77349104354 scopus 로고    scopus 로고
    • Repeatability of corneal power and wavefront aberration measurements with a dual-Scheimpflug Placido corneal topographer
    • L. Wang, M. Shirayama, and D.D. Koch Repeatability of corneal power and wavefront aberration measurements with a dual-Scheimpflug Placido corneal topographer J Cataract Refract Surg 36 2010 425 430
    • (2010) J Cataract Refract Surg , vol.36 , pp. 425-430
    • Wang, L.1    Shirayama, M.2    Koch, D.D.3
  • 8
    • 57749189820 scopus 로고    scopus 로고
    • Intraobserver and interobserver repeatability of curvature and aberrometric measurements of the posterior corneal surface in normal eyes using Scheimpflug photography
    • D.P. Piñero, C. Saenz González, and J.L. Alió Intraobserver and interobserver repeatability of curvature and aberrometric measurements of the posterior corneal surface in normal eyes using Scheimpflug photography J Cataract Refract Surg 35 2009 113 120
    • (2009) J Cataract Refract Surg , vol.35 , pp. 113-120
    • Piñero, D.P.1    Saenz González, C.2    Alió, J.L.3
  • 9
    • 37349056419 scopus 로고    scopus 로고
    • Anterior segment biometry with the Pentacam: Comprehensive assessment of repeatability of automated measurements
    • H. Shankar, D. Taranath, C.T. Santhirathelagan, and K. Pesudovs Anterior segment biometry with the Pentacam: comprehensive assessment of repeatability of automated measurements J Cataract Refract Surg 34 2008 103 113
    • (2008) J Cataract Refract Surg , vol.34 , pp. 103-113
    • Shankar, H.1    Taranath, D.2    Santhirathelagan, C.T.3    Pesudovs, K.4
  • 10
    • 61449231646 scopus 로고    scopus 로고
    • Reliability and repeatability of the Pentacam on corneal curvatures
    • Available at: Accessed January 28, 2012
    • D. Chen, and A.K.C. Lam Reliability and repeatability of the Pentacam on corneal curvatures Clin Exp Optom 92 2009 110 118 Available at: http://onlinelibrary.wiley.com/doi/10.1111/j.1444-0938.2008.00336.x/pdf Accessed January 28, 2012
    • (2009) Clin Exp Optom , vol.92 , pp. 110-118
    • Chen, D.1    Lam, A.K.C.2
  • 11
    • 57749191496 scopus 로고    scopus 로고
    • Repeatability, reproducibility, and agreement characteristics of rotating Scheimpflug photography and scanning-slit corneal topography for corneal power measurement
    • T. Kawamorita, H. Uozato, K. Kamiya, L. Bax, K. Tsutsui, D. Aizawa, and K. Shimizu Repeatability, reproducibility, and agreement characteristics of rotating Scheimpflug photography and scanning-slit corneal topography for corneal power measurement J Cataract Refract Surg 35 2009 127 133
    • (2009) J Cataract Refract Surg , vol.35 , pp. 127-133
    • Kawamorita, T.1    Uozato, H.2    Kamiya, K.3    Bax, L.4    Tsutsui, K.5    Aizawa, D.6    Shimizu, K.7
  • 12
    • 52249090464 scopus 로고    scopus 로고
    • Comparison and reproducibility of corneal thickness and curvature readings obtained by the Galilei and the Orbscan II analysis systems
    • N. Menassa, C. Kaufmann, M. Goggin, O.M. Job, L.M. Bachmann, and M.A. Thiel Comparison and reproducibility of corneal thickness and curvature readings obtained by the Galilei and the Orbscan II analysis systems J Cataract Refract Surg 34 2008 1742 1747
    • (2008) J Cataract Refract Surg , vol.34 , pp. 1742-1747
    • Menassa, N.1    Kaufmann, C.2    Goggin, M.3    Job, O.M.4    Bachmann, L.M.5    Thiel, M.A.6
  • 13
    • 33847163129 scopus 로고    scopus 로고
    • Intrasession and intersession repeatability of the Pentacam system on posterior corneal assessment in the normal human eye
    • D. Chen, and A.K.C. Lam Intrasession and intersession repeatability of the Pentacam system on posterior corneal assessment in the normal human eye J Cataract Refract Surg 33 2007 448 454
    • (2007) J Cataract Refract Surg , vol.33 , pp. 448-454
    • Chen, D.1    Lam, A.K.C.2
  • 14
    • 27644542761 scopus 로고    scopus 로고
    • Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound
    • Available at: Accessed January 28, 2012
    • B. Lackner, G. Schmidinger, S. Pieh, M.A. Funovics, and C. Skorpik Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound Optom Vis Sci 82 2005 892 899 Available at: http://www.oculus.de/chi/downloads/dyn/sonstige/sonstige/lackner-pachymetry.pdf Accessed January 28, 2012
    • (2005) Optom Vis Sci , vol.82 , pp. 892-899
    • Lackner, B.1    Schmidinger, G.2    Pieh, S.3    Funovics, M.A.4    Skorpik, C.5
  • 15
    • 78751497805 scopus 로고    scopus 로고
    • Pachymetric measurements with a new Scheimpflug photography-based system; Intraobserver repeatability and agreement with optical coherence tomography pachymetry
    • M. Milla, D.P. Piñero, F. Amparo, and J.L. Alió Pachymetric measurements with a new Scheimpflug photography-based system; intraobserver repeatability and agreement with optical coherence tomography pachymetry J Cataract Refract Surg 37 2011 310 316
    • (2011) J Cataract Refract Surg , vol.37 , pp. 310-316
    • Milla, M.1    Piñero, D.P.2    Amparo, F.3    Alió, J.L.4
  • 16
    • 0035140259 scopus 로고    scopus 로고
    • Power vector analysis of the optical outcomes of refractive surgery
    • L.N. Thibos, and D. Horner Power vector analysis of the optical outcomes of refractive surgery J Cataract Refract Surg 27 2001 80 85
    • (2001) J Cataract Refract Surg , vol.27 , pp. 80-85
    • Thibos, L.N.1    Horner, D.2
  • 17
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • Available at: Accessed January 28, 2012
    • J.M. Bland, and D.G. Altman Statistical methods for assessing agreement between two methods of clinical measurement Lancet 1 1986 307 310 Available at: http://www-users.york.ac.uk/∼mb55/meas/ba.pdf Accessed January 28, 2012
    • (1986) Lancet , vol.1 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 18
    • 0029975750 scopus 로고    scopus 로고
    • Statistical notes. Measurement error and correlation coefficients
    • correction, 744. Available at: Accessed January 28, 2012. Correction available at: http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2352111/pdf/bmj00560- 0056b.pdf Accessed January 28, 2012
    • J.M. Bland, and D.G. Altman Statistical notes. Measurement error and correlation coefficients BMJ 313 1996 41 42 correction, 744. Available at: http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2351452/pdf/bmj00549-0045.pdf Accessed January 28, 2012. Correction available at: http://www.ncbi.nlm.nih.gov/ pmc/articles/PMC2352111/pdf/bmj00560-0056b.pdf Accessed January 28, 2012
    • (1996) BMJ , vol.313 , pp. 41-42
    • Bland, J.M.1    Altman, D.G.2
  • 20
    • 80051861463 scopus 로고    scopus 로고
    • Repeatability of automatic measurements by a new Scheimpflug camera combined with Placido topography
    • G. Savini, P. Barboni, M. Carbonelli, and K.J. Hoffer Repeatability of automatic measurements by a new Scheimpflug camera combined with Placido topography J Cataract Refract Surg 37 2011 1809 1816
    • (2011) J Cataract Refract Surg , vol.37 , pp. 1809-1816
    • Savini, G.1    Barboni, P.2    Carbonelli, M.3    Hoffer, K.J.4
  • 22
    • 33645992147 scopus 로고    scopus 로고
    • The topography of the central and peripheral cornea
    • Available at: Accessed January 28, 2012
    • S.A. Read, M.J. Collins, L.G. Carney, and R.J. Franklin The topography of the central and peripheral cornea Invest Ophthalmol Vis Sci 47 2006 1404 1415 Available at: http://www.iovs.org/cgi/reprint/47/4/1404 Accessed January 28, 2012
    • (2006) Invest Ophthalmol Vis Sci , vol.47 , pp. 1404-1415
    • Read, S.A.1    Collins, M.J.2    Carney, L.G.3    Franklin, R.J.4
  • 23
    • 35348960981 scopus 로고    scopus 로고
    • Reproducibility and repeatability of central corneal thickness measurement in keratoconus using the rotating Scheimpflug camera and ultrasound pachymetry
    • U. de Sanctis, A. Missolungi, B. Mutani, L. Richiardi, and F.M. Grignolo Reproducibility and repeatability of central corneal thickness measurement in keratoconus using the rotating Scheimpflug camera and ultrasound pachymetry Am J Ophthalmol 144 2007 712 718
    • (2007) Am J Ophthalmol , vol.144 , pp. 712-718
    • De Sanctis, U.1    Missolungi, A.2    Mutani, B.3    Richiardi, L.4    Grignolo, F.M.5
  • 24
    • 37349081956 scopus 로고    scopus 로고
    • Reliability of peripheral corneal pachymetry with the Oculus Pentacam
    • [letter] reply by R Khoramnia, TM Rabsilber, GU Auffarth, 8. Letter available at: Accessed Januaray 28, 2012
    • H. Shankar, and K. Pesudovs Reliability of peripheral corneal pachymetry with the Oculus Pentacam [letter] J Cataract Refract Surg 34 2008 7 reply by R Khoramnia, TM Rabsilber, GU Auffarth, 8. Letter available at: http://www.pesudovs.com/konrad/Docs/Shankar%20letter.pdf Accessed Januaray 28, 2012
    • (2008) J Cataract Refract Surg , vol.34 , pp. 7
    • Shankar, H.1    Pesudovs, K.2
  • 25
    • 37449000474 scopus 로고    scopus 로고
    • Corneal diameter measurements by corneal topography and angle-to-angle measurements by optical coherence tomography: Evaluation of equivalence
    • D.P. Piñero, A.B. Plaza Puche, and J.L. Alió Corneal diameter measurements by corneal topography and angle-to-angle measurements by optical coherence tomography: evaluation of equivalence J Cataract Refract Surg 34 2008 126 131
    • (2008) J Cataract Refract Surg , vol.34 , pp. 126-131
    • Piñero, D.P.1    Plaza Puche, A.B.2    Alió, J.L.3
  • 26
    • 33750479829 scopus 로고    scopus 로고
    • Internal anterior chamber diameter using optical coherence tomography compared with white-to-white distances using automated measurements
    • T. Kohnen, M.C. Thomala, M. Cichocki, and A. Strenger Internal anterior chamber diameter using optical coherence tomography compared with white-to-white distances using automated measurements J Cataract Refract Surg 32 2006 1809 1813
    • (2006) J Cataract Refract Surg , vol.32 , pp. 1809-1813
    • Kohnen, T.1    Thomala, M.C.2    Cichocki, M.3    Strenger, A.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.