Effective dose range for dental cone beam computed tomography scanners

Author keywords

Cone beam computed tomography; Dentomaxillofacial; Effective radiation dose; Thermoluminescent dosimetry

Indexed keywords

ARTICLE; COMPUTED TOMOGRAPHY SCANNER; CONE BEAM COMPUTED TOMOGRAPHY; DOSE RESPONSE; PRIORITY JOURNAL; SALIVARY GLAND; THYROID GLAND; TOOTH RADIOGRAPHY;

EID: 84855604583     PISSN: 0720048X     EISSN: 18727727     Source Type: Journal    
DOI: 10.1016/j.ejrad.2010.11.028     Document Type: Article
Times cited : (451)

References (22)
  • 1
    • 33644639722 scopus 로고    scopus 로고
    • Clinical applications of cone-beam computed tomography in dental practice
    • W.C. Scarfe, A.G. Farman, and P. Sukovic Clinical applications of cone-beam computed tomography in dental practice J Can Dent Assoc 72 1 2006 75 85
    • (2006) J Can Dent Assoc , vol.72 , Issue.1 , pp. 75-85
    • Scarfe, W.C.1    Farman, A.G.2    Sukovic, P.3
  • 2
    • 0032915266 scopus 로고    scopus 로고
    • Balancing patient dose and image quality
    • DOI 10.1016/S0969-8043(98)00021-9, PII S0969804398000219
    • C.J. Martin, D.G. Sutton, and P.F. Sharp Balancing patient dose and image quality Appl Radiat Isot 50 1 1999 1 19 (Pubitemid 29048988)
    • (1999) Applied Radiation and Isotopes , vol.50 , Issue.1 , pp. 1-19
    • Martin, C.J.1    Sutton, D.G.2    Sharp, P.F.3
  • 4
    • 0003733263 scopus 로고    scopus 로고
    • International Commission on Radiological Protection ICRP Publication 103. Ann ICRP 37. Oxford, UK: Pergamon Press
    • International Commission on Radiological Protection. Recommendations of the International Commission on Radiological Protection. ICRP Publication 103. Ann ICRP 37. Oxford, UK: Pergamon Press, 2007.
    • (2007) Recommendations of the International Commission on Radiological Protection
  • 5
    • 34948887091 scopus 로고    scopus 로고
    • Effective dose: How should it be applied to medical exposures?
    • DOI 10.1259/bjr/25922439
    • C.J. Martin Effective dose: how should it be applied to medical exposures? Br J Radiol 80 956 2007 639 647 (Pubitemid 47517809)
    • (2007) British Journal of Radiology , vol.80 , Issue.956 , pp. 639-647
    • Martin, C.J.1
  • 6
    • 46849113662 scopus 로고    scopus 로고
    • Effective dose: A flawed concept that could and should be replaced
    • DOI 10.1259/bjr/22942198
    • D.J. Brenner Effective dose: a flawed concept that could and should be replaced Br J Radiol 81 967 2008 521 523 (Pubitemid 351957735)
    • (2008) British Journal of Radiology , vol.81 , Issue.967 , pp. 521-523
    • Brenner, D.J.1
  • 7
    • 77953352476 scopus 로고    scopus 로고
    • Methods of determining the effective dose in dental radiology
    • A. Thilander-Klang, and E. Helmrot Methods of determining the effective dose in dental radiology Radiat Prot Dosimetry 139 1-3 2010 306 309
    • (2010) Radiat Prot Dosimetry , vol.139 , Issue.13 , pp. 306-309
    • Thilander-Klang, A.1    Helmrot, E.2
  • 8
    • 0021691472 scopus 로고
    • Estimation of mean organ doses in diagnostic radiology from Rando phantom measurements
    • W. Huda, and G.A. Sandison Estimation of mean organ doses in diagnostic radiology from Rando phantom measurements Health Phys 47 3 1984 463 467 (Pubitemid 15216923)
    • (1984) Health Physics , vol.47 , Issue.3 , pp. 463-467
    • Huda, W.1    Sandison, G.A.2
  • 9
    • 69949169325 scopus 로고    scopus 로고
    • Comparison between effective radiation dose of CBCT and MSCT scanners for dentomaxillofacial applications
    • M. Loubele, R. Bogaerts, and E. Van Dijck Comparison between effective radiation dose of CBCT and MSCT scanners for dentomaxillofacial applications Eur J Radiol 71 3 2009 461 468
    • (2009) Eur J Radiol , vol.71 , Issue.3 , pp. 461-468
    • Loubele, M.1    Bogaerts, R.2    Van Dijck, E.3
  • 10
    • 62549111002 scopus 로고    scopus 로고
    • Absorbed and effective doses from cone beam volumetric imaging for implant planning
    • T. Okano, Y. Harata, and Y. Sugihara Absorbed and effective doses from cone beam volumetric imaging for implant planning Dentomaxillofac Radiol 38 2 2009 79 85
    • (2009) Dentomaxillofac Radiol , vol.38 , Issue.2 , pp. 79-85
    • Okano, T.1    Harata, Y.2    Sugihara, Y.3
  • 11
    • 69249102664 scopus 로고    scopus 로고
    • Dosimetry and image quality of four dental cone beam computed tomography scanners compared with multislice computed tomography scanners
    • A. Suomalainen, T. Kiljunen, Y. Käser, J. Peltola, and M. Kortesniemi Dosimetry and image quality of four dental cone beam computed tomography scanners compared with multislice computed tomography scanners Dentomaxillofac Radiol 38 6 2009 367 378
    • (2009) Dentomaxillofac Radiol , vol.38 , Issue.6 , pp. 367-378
    • Suomalainen, A.1    Kiljunen, T.2    Käser, Y.3    Peltola, J.4    Kortesniemi, M.5
  • 12
    • 45949102819 scopus 로고    scopus 로고
    • Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology
    • J.B. Ludlow, and M. Ivanovic Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology Oral Surg Oral Med Oral Pathol Oral Radiol Endod 106 1 2008 106 114
    • (2008) Oral Surg Oral Med Oral Pathol Oral Radiol Endod , vol.106 , Issue.1 , pp. 106-114
    • Ludlow, J.B.1    Ivanovic, M.2
  • 13
    • 43049170124 scopus 로고    scopus 로고
    • Cone-beam computed tomography for routine orthodontic treatment planning: A radiation dose evaluation
    • M.A. Silva, U. Wolf, F. Heinicke, A. Bumann, H. Visser, and E. Hirsch Cone-beam computed tomography for routine orthodontic treatment planning: a radiation dose evaluation Am J Orthod Dentofacial Orthop 133 5 2008 640.e1-5
    • (2008) Am J Orthod Dentofacial Orthop , vol.133 , Issue.5
    • Silva, M.A.1    Wolf, U.2    Heinicke, F.3    Bumann, A.4    Visser, H.5    Hirsch, E.6
  • 14
    • 33746041477 scopus 로고    scopus 로고
    • Dosimetry of 3 CBCT devices for oral and maxillofacial radiology: CB Mercuray, NewTom 3G and i-CAT
    • DOI 10.1259/dmfr/14340323
    • J.B. Ludlow, L.E. Davies-Ludlow, S.L. Brooks, and W.B. Howerton Dosimetry of 3 CBCT devices for oral and maxillofacial radiology: CB Mercuray NewTom 3G and i-CAT Dentomaxillofac Radiol 35 4 2006 219 226 (Pubitemid 44076876)
    • (2006) Dentomaxillofacial Radiology , vol.35 , Issue.4 , pp. 219-226
    • Ludlow, J.B.1    Davies-Ludlow, L.E.2    Brooks, S.L.3    Howerton, W.B.4
  • 16
    • 58849163662 scopus 로고    scopus 로고
    • Effective dose from cone beam CT examinations in dentistry
    • Roberts JA, Drage NA, Davies J, Thomas DW. Effective dose from cone beam CT examinations in dentistry. Br J Radiol 2009;82(973):35-40.
    • (2009) Br J Radiol , vol.82 , Issue.973 , pp. 35-40
    • Roberts, J.A.1    Drage, N.A.2    Davies, J.3    Thomas, D.W.4
  • 17
    • 48649089199 scopus 로고    scopus 로고
    • Dosimetry of the cone beam computed tomography Veraviewepocs 3D compared with the 3D Accuitomo in different fields of view
    • Hirsch E, Wolf U, Heinicke F, Silva MA. Dosimetry of the cone beam computed tomography Veraviewepocs 3D compared with the 3D Accuitomo in different fields of view. Dentomaxillofac Radiol 2008;37(5):268-73.
    • (2008) Dentomaxillofac Radiol , vol.37 , Issue.5 , pp. 268-273
    • Hirsch, E.1    Wolf, U.2    Heinicke, F.3    Silva, M.A.4
  • 19
    • 33645235271 scopus 로고    scopus 로고
    • Radiation dose vs. image quality for low-dose CT protocols of the head for maxillofacial surgery and oral implant planning
    • Loubele M, Jacobs R, Maes F, et al. Radiation dose vs. image quality for low-dose CT protocols of the head for maxillofacial surgery and oral implant planning. Radiat Prot Dosimetry 2006;117(1-3):211-6.
    • (2006) Radiat Prot Dosimetry , vol.117 , Issue.1-3 , pp. 211-216
    • Loubele, M.1    Jacobs, R.2    Maes, F.3
  • 20
    • 1342311471 scopus 로고    scopus 로고
    • Comparison of radiation levels from computed tomography and conventional dental radiographs
    • D.C. Ngan, O.P. Kharbanda, J.P. Geenty, and M.A. Darendeliler Comparison of radiation levels from computed tomography and conventional dental radiographs Aust Orthod J 19 2 2003 67 75
    • (2003) Aust Orthod J , vol.19 , Issue.2 , pp. 67-75
    • Ngan, D.C.1    Kharbanda, O.P.2    Geenty, J.P.3    Darendeliler, M.A.4
  • 21
    • 52749087498 scopus 로고    scopus 로고
    • Patient risk related to common dental radiographic examinations: The impact of 2007 International Commission on Radiological Protection recommendations regarding dose calculation
    • J.B. Ludlow, L.E. Davies-Ludlow, and S.C. White Patient risk related to common dental radiographic examinations: the impact of 2007 International Commission on Radiological Protection recommendations regarding dose calculation J Am Dent Assoc 139 9 2008 1237 1243
    • (2008) J Am Dent Assoc , vol.139 , Issue.9 , pp. 1237-1243
    • Ludlow, J.B.1    Davies-Ludlow, L.E.2    White, S.C.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.