Evaluation of a novel continuous glucose measurement device in patients with diabetes mellitus across the glycemic range

Author keywords

Blood glucose; Continuous glucose monitoring; Glucose clamp method; Hyperglycemia; Hypoglycemia

Indexed keywords

ADULT; ARTICLE; AUTOMATION; BLOOD; BLOOD GLUCOSE MONITORING; CALIBRATION; CLINICAL TRIAL; DEVICES; DIABETES MELLITUS; FEMALE; GENETIC PROCEDURES; GLUCOSE BLOOD LEVEL; GLUCOSE CLAMP TECHNIQUE; HUMAN; INSTRUMENTATION; MALE; METABOLISM; METHODOLOGY; MIDDLE AGED; REPRODUCIBILITY; RETROSPECTIVE STUDY; STANDARD;

EID: 84855410538     PISSN: None     EISSN: 19322968     Source Type: Journal    
DOI: 10.1177/193229681100500406     Document Type: Article
Times cited : (11)

References (22)
  • 1
    • 0037278485 scopus 로고    scopus 로고
    • Unrecognized hypo- and hyperglycemia in well-controlled patients with type 2 diabetes mellitus: The results of continuous glucose monitoring
    • Hay LC, Wilmshurst EG, Fulcher G. Unrecognized hypo- and hyperglycemia in well-controlled patients with type 2 diabetes mellitus: the results of continuous glucose monitoring. Diabetes Technol Ther. 2003;5(1):19-26.
    • (2003) Diabetes Technol Ther , vol.5 , Issue.1 , pp. 19-26
    • Hay, L.C.1    Wilmshurst, E.G.2    Fulcher, G.3
  • 2
    • 1442324399 scopus 로고    scopus 로고
    • Improved glucose excursions using an implantable real-time continuous glucose sensor in adults with type 1 diabetes
    • Garg SK, Schwartz S, Edelman SV. Improved glucose excursions using an implantable real-time continuous glucose sensor in adults with type 1 diabetes. Diabetes Care. 2004;27(3):734-8.
    • (2004) Diabetes Care , vol.27 , Issue.3 , pp. 734-738
    • Garg, S.K.1    Schwartz, S.2    Edelman, S.V.3
  • 3
    • 18144421498 scopus 로고    scopus 로고
    • Continuous glucose monitoring: Roadmap for 21st century diabetes therapy
    • Klonoff DC. Continuous glucose monitoring: roadmap for 21st century diabetes therapy. Diabetes Care. 2005;28(5):1231-9.
    • (2005) Diabetes Care , vol.28 , Issue.5 , pp. 1231-1239
    • Klonoff, D.C.1
  • 4
    • 0032511583 scopus 로고    scopus 로고
    • Intensive blood-glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes (UKPDS 33)
    • UK Prospective Diabetes Study (UKPDS) Group
    • UK Prospective Diabetes Study (UKPDS) Group. Intensive blood-glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes (UKPDS 33). Lancet. 1998;352(9131):837-53.
    • (1998) Lancet , vol.352 , Issue.9131 , pp. 837-853
  • 5
    • 0027370108 scopus 로고
    • The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus
    • The Diabetes Control and Complications Trial Research Group
    • The Diabetes Control and Complications Trial Research Group. The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med. 1993;329(14):977-86.
    • (1993) N Engl J Med , vol.329 , Issue.14 , pp. 977-986
  • 7
    • 78349281594 scopus 로고    scopus 로고
    • Continuous glucose monitoring system in a rural intensive care unit: A pilot study evaluating accuracy and acceptance
    • Jacobs B, Phan K, Bertheau L, Dogbey G, Schwartz F, Shubrook J. Continuous glucose monitoring system in a rural intensive care unit: a pilot study evaluating accuracy and acceptance. J Diabetes Sci Technol. 2010;4(3):636-44.
    • (2010) J Diabetes Sci Technol , vol.4 , Issue.3 , pp. 636-644
    • Jacobs, B.1    Phan, K.2    Bertheau, L.3    Dogbey, G.4    Schwartz, F.5    Shubrook, J.6
  • 8
    • 77951121063 scopus 로고    scopus 로고
    • Comparison of the clinical information provided by the FreeStyle Navigator continuous interstitial glucose monitor versus traditional blood glucose readings
    • McGarraugh GV, Clarke WL, Kovatchev BP. Comparison of the clinical information provided by the FreeStyle Navigator continuous interstitial glucose monitor versus traditional blood glucose readings. Diabetes Technol Ther. 2010;12(5):365-71.
    • (2010) Diabetes Technol Ther , vol.12 , Issue.5 , pp. 365-371
    • McGarraugh, G.V.1    Clarke, W.L.2    Kovatchev, B.P.3
  • 9
    • 58149389636 scopus 로고    scopus 로고
    • Evaluating the accuracy, reliability, and clinical applicability of continuous glucose monitoring (CGM): Is CGM ready for real time?
    • Mazze RS, Strock E, Borgman S, Wesley D, Stout P, Racchini J. Evaluating the accuracy, reliability, and clinical applicability of continuous glucose monitoring (CGM): Is CGM ready for real time? Diabetes Technol Ther. 2009;11(1):11-8.
    • (2009) Diabetes Technol Ther , vol.11 , Issue.1 , pp. 11-18
    • Mazze, R.S.1    Strock, E.2    Borgman, S.3    Wesley, D.4    Stout, P.5    Racchini, J.6
  • 10
    • 0033759474 scopus 로고    scopus 로고
    • Can interstitial glucose assessment replace blood glucose measurements?
    • Rebrin K, Steil GM. Can interstitial glucose assessment replace blood glucose measurements? Diabetes Technol Ther. 2000;2(3):461-72.
    • (2000) Diabetes Technol Ther , vol.2 , Issue.3 , pp. 461-472
    • Rebrin, K.1    Steil, G.M.2
  • 13
    • 58449116859 scopus 로고    scopus 로고
    • Comparison of accuracy and safety of the SEVEN and the Navigator continuous glucose monitoring systems
    • Garg SK, Smith J, Beatson C, Lopez-Baca B, Voelmle M, Gottlieb PA. Comparison of accuracy and safety of the SEVEN and the Navigator continuous glucose monitoring systems. Diabetes Technol Ther. 2009;11(2):65-72.
    • (2009) Diabetes Technol Ther , vol.11 , Issue.2 , pp. 65-72
    • Garg, S.K.1    Smith, J.2    Beatson, C.3    Lopez-Baca, B.4    Voelmle, M.5    Gottlieb, P.A.6
  • 15
    • 0033855803 scopus 로고    scopus 로고
    • A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose
    • Parkes JL, Slatin SL, Pardo S, Ginsberg BH. A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose. Diabetes Care. 2000;23(8):1143-8.
    • (2000) Diabetes Care , vol.23 , Issue.8 , pp. 1143-1148
    • Parkes, J.L.1    Slatin, S.L.2    Pardo, S.3    Ginsberg, B.H.4
  • 16
    • 27644587693 scopus 로고    scopus 로고
    • The original Clarke Error Grid Analysis (EGA)
    • Clarke WL. The original Clarke Error Grid Analysis (EGA). Diabetes Technol Ther. 2005;7(5):776-9.
    • (2005) Diabetes Technol Ther , vol.7 , Issue.5 , pp. 776-779
    • Clarke, W.L.1
  • 17
    • 0002331186 scopus 로고
    • Dermal toxicity
    • Staff of the Division of Pharmacology, Food and Drug Administration, Department of Health, Education and Welfare, eds. Austin: Association of Food and Drug Officials of the United States Business Office, Bureau of Food and Drugs, Texas State Department of Health
    • Draize JH. Dermal toxicity. In: Staff of the Division of Pharmacology, Food and Drug Administration, Department of Health, Education and Welfare, eds. Appraisal of the safety of chemicals in foods, drugs and cosmetics. Austin: Association of Food and Drug Officials of the United States Business Office, Bureau of Food and Drugs, Texas State Department of Health; 1959, 46-59.
    • (1959) Appraisal of the Safety of Chemicals in Foods, Drugs and Cosmetics , pp. 46-59
    • Draize, J.H.1
  • 19
    • 67650248749 scopus 로고    scopus 로고
    • Clinical review: Realistic expectations and practical use of continuous glucose monitoring for the endocrinologist
    • Hirsch IB. Clinical review: realistic expectations and practical use of continuous glucose monitoring for the endocrinologist. J Clin Endocrinol Metab. 2009;94(7):2232-8.
    • (2009) J Clin Endocrinol Metab , vol.94 , Issue.7 , pp. 2232-2238
    • Hirsch, I.B.1
  • 20
    • 77953066924 scopus 로고    scopus 로고
    • Delays in minimally invasive continuous glucose monitoring devices: A review of current technology
    • Keenan DB, Mastrototaro JJ, Voskanyan G, Steil GM. Delays in minimally invasive continuous glucose monitoring devices: a review of current technology. J Diabetes Sci Technol. 2009;3(5):1207-14.
    • (2009) J Diabetes Sci Technol , vol.3 , Issue.5 , pp. 1207-1214
    • Keenan, D.B.1    Mastrototaro, J.J.2    Voskanyan, G.3    Steil, G.M.4
  • 21
    • 25644435718 scopus 로고    scopus 로고
    • Evaluating the clinical accuracy of two continuous glucose sensors using continuous glucose-error grid analysis
    • Clarke WL, Anderson S, Farhy L, Breton M, Gonder-Frederick L, Cox D, Kovatchev B. Evaluating the clinical accuracy of two continuous glucose sensors using continuous glucose-error grid analysis. Diabetes Care. 2005;28(10):2412-7.
    • (2005) Diabetes Care , vol.28 , Issue.10 , pp. 2412-2417
    • Clarke, W.L.1    Anderson, S.2    Farhy, L.3    Breton, M.4    Gonder-Frederick, L.5    Cox, D.6    Kovatchev, B.7
  • 22
    • 77953054702 scopus 로고    scopus 로고
    • Accuracy of the SEVEN continuous glucose monitoring system: Comparison with frequently sampled venous glucose measurements
    • Zisser HC, Bailey TS, Schwartz S, Ratner RE, Wise J. Accuracy of the SEVEN continuous glucose monitoring system: comparison with frequently sampled venous glucose measurements. J Diabetes Sci Technol. 2009;3(5):1146-54.
    • (2009) J Diabetes Sci Technol , vol.3 , Issue.5 , pp. 1146-1154
    • Zisser, H.C.1    Bailey, T.S.2    Schwartz, S.3    Ratner, R.E.4    Wise, J.5


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.