Volumn 97, Issue 3, 2010, Pages 390-394

The accuracy of frameless stereotactic intracranial radiosurgery

Author keywords

Frameless stereotactic radiosurgery; Hidden target test; Intrafraction motion

Indexed keywords

ACCURACY; ARTICLE; CLINICAL ARTICLE; FRAMELESS STEREOTACTIC INTRACRANIAL RADIOSURGERY; HUMAN; IMAGING SYSTEM; PHANTOM; PRIORITY JOURNAL; RADIATION DOSE FRACTIONATION; RANDOM ERROR; ROBOTICS; STEREOTACTIC RADIOSURGERY; SYSTEMATIC ERROR;

EID: 78651500696     PISSN: 01678140     EISSN: None     Source Type: Journal    
DOI: 10.1016/j.radonc.2010.06.012     Document Type: Article
Times cited : (59)

References (20)
  • 1
    • 78651494618 scopus 로고
    • The application accuracy of stereotactic frames
    • R.J. Maciunas, R.L. Galloway, and J.W. Latimer The application accuracy of stereotactic frames Neurosurgery 179 1994 571 575
    • (1994) Neurosurgery , vol.179 , pp. 571-575
    • MacIunas, R.J.1    Galloway, R.L.2    Latimer, J.W.3
  • 3
    • 15944381594 scopus 로고    scopus 로고
    • Treatment accuracy of fractionated stereotactic radiotherapy
    • S. Kumar, K. Burke, and C. Nalder Treatment accuracy of fractionated stereotactic radiotherapy Radiother Oncol 74 2005 53 59
    • (2005) Radiother Oncol , vol.74 , pp. 53-59
    • Kumar, S.1    Burke, K.2    Nalder, C.3
  • 4
    • 0028786274 scopus 로고
    • A precision cranial immobilization system for conformal stereotactic fractionated radiation therapy
    • S.J. Rosenthal, K.P. Gall, and M. Jackson A precision cranial immobilization system for conformal stereotactic fractionated radiation therapy Int J Radiat Oncol Biol Phys 33 1995 1239 1245
    • (1995) Int J Radiat Oncol Biol Phys , vol.33 , pp. 1239-1245
    • Rosenthal, S.J.1    Gall, K.P.2    Jackson, M.3
  • 5
    • 0036463773 scopus 로고    scopus 로고
    • Accuracy of a relocatable stereotactic radiotherapy head frame evaluated by use of a depth helmet
    • K.E. Burton, S.J. Thomas, and D. Whitney Accuracy of a relocatable stereotactic radiotherapy head frame evaluated by use of a depth helmet Clin Oncol 14 2002 31 39
    • (2002) Clin Oncol , vol.14 , pp. 31-39
    • Burton, K.E.1    Thomas, S.J.2    Whitney, D.3
  • 6
    • 0033497405 scopus 로고    scopus 로고
    • Considerations of fractionated stereotactic radiotherapy in patients with dentures
    • H. Ashamalla, P. Ross, and N.C. Ikoro Considerations of fractionated stereotactic radiotherapy in patients with dentures J Radiosurg 2 1999 89 93
    • (1999) J Radiosurg , vol.2 , pp. 89-93
    • Ashamalla, H.1    Ross, P.2    Ikoro, N.C.3
  • 7
    • 0038030020 scopus 로고    scopus 로고
    • Quality assurance of a system for improved target localization and patient set-up that combines real-time infrared tracking and stereoscopic X-ray imaging
    • D. Verellen, G. Soete, and N. Linthout Quality assurance of a system for improved target localization and patient set-up that combines real-time infrared tracking and stereoscopic X-ray imaging Radiother Oncol 67 2003 129 141
    • (2003) Radiother Oncol , vol.67 , pp. 129-141
    • Verellen, D.1    Soete, G.2    Linthout, N.3
  • 8
    • 0347625753 scopus 로고    scopus 로고
    • A phantom study on the positioning accuracy of the Novalis body system
    • H. Yan, F. Yin, and J.H. Kim A phantom study on the positioning accuracy of the Novalis body system Med Phys 30 2003 3052 3060
    • (2003) Med Phys , vol.30 , pp. 3052-3060
    • Yan, H.1    Yin, F.2    Kim, J.H.3
  • 9
    • 58149202436 scopus 로고    scopus 로고
    • Novalis frameless image-guided noninvasive radiosurgery: Initial experience
    • R.E. Wurm, S. Erbel, and I. Schwenkert Novalis frameless image-guided noninvasive radiosurgery: initial experience Neurosurgery 62 2008 A11 A18
    • (2008) Neurosurgery , vol.62
    • Wurm, R.E.1    Erbel, S.2    Schwenkert, I.3
  • 10
    • 77950338949 scopus 로고    scopus 로고
    • A clinical comparison of patient setup and intra-fraction motion using frame-based radiosurgery versus a frameless image-guided radiosurgery system for intracranial lesions
    • N. Ramakrishna, F. Rosca, and S. Friesen A clinical comparison of patient setup and intra-fraction motion using frame-based radiosurgery versus a frameless image-guided radiosurgery system for intracranial lesions Radiother Oncol 95 2010 109 115
    • (2010) Radiother Oncol , vol.95 , pp. 109-115
    • Ramakrishna, N.1    Rosca, F.2    Friesen, S.3
  • 11
    • 70049096052 scopus 로고    scopus 로고
    • Simulation of intrafraction motion and overall geometric accuracy of a frameless intracranial radiosurgery process
    • V. Feygelman, L. Walker, and P. Chinnaiyan Simulation of intrafraction motion and overall geometric accuracy of a frameless intracranial radiosurgery process J Appl Clin Med Phys 9 2008 2828
    • (2008) J Appl Clin Med Phys , vol.9 , pp. 2828
    • Feygelman, V.1    Walker, L.2    Chinnaiyan, P.3
  • 12
    • 31844439099 scopus 로고    scopus 로고
    • Six dimensional analysis with daily stereoscopic X-ray imaging of intrafraction patient motion in head and neck treatments using five points fixation masks
    • N. Linthout, D. Verellen, and K. Tournel Six dimensional analysis with daily stereoscopic X-ray imaging of intrafraction patient motion in head and neck treatments using five points fixation masks Med Phys 33 2006 504 513
    • (2006) Med Phys , vol.33 , pp. 504-513
    • Linthout, N.1    Verellen, D.2    Tournel, K.3
  • 13
    • 38149009727 scopus 로고    scopus 로고
    • Time dependence of intrafraction patient motion assessed by repeat stereoscopic imaging
    • M.S. Hoogeman, J.J. Nuyttens, and P.C. Levendag Time dependence of intrafraction patient motion assessed by repeat stereoscopic imaging Int J Radiat Oncol Biol Phys 70 2008 609 618
    • (2008) Int J Radiat Oncol Biol Phys , vol.70 , pp. 609-618
    • Hoogeman, M.S.1    Nuyttens, J.J.2    Levendag, P.C.3
  • 14
    • 0035156119 scopus 로고    scopus 로고
    • Infrared patient positioning for stereotactic radiosurgery of extracranial tumors
    • L.T. Wang, T.D. Solberg, and P.M. Medlin Infrared patient positioning for stereotactic radiosurgery of extracranial tumors Comput Biol Med 31 2001 101 111
    • (2001) Comput Biol Med , vol.31 , pp. 101-111
    • Wang, L.T.1    Solberg, T.D.2    Medlin, P.M.3
  • 15
    • 0030767978 scopus 로고    scopus 로고
    • Quality assurance for dynamic multileaf collimator modulated fields using a fast beam imaging system
    • L. Ma, P.B. Geis, and A.L. Boyer Quality assurance for dynamic multileaf collimator modulated fields using a fast beam imaging system Med Phys 24 1997 1213 1220
    • (1997) Med Phys , vol.24 , pp. 1213-1220
    • Ma, L.1    Geis, P.B.2    Boyer, A.L.3
  • 16
    • 0023875610 scopus 로고
    • A system for stereotactic radiosurgery with a linear accelerator
    • W. Lutz, K.R. Winston, and N. Maleki A system for stereotactic radiosurgery with a linear accelerator Int J Radiat Oncol Biol Phys 14 1988 373 381
    • (1988) Int J Radiat Oncol Biol Phys , vol.14 , pp. 373-381
    • Lutz, W.1    Winston, K.R.2    Maleki, N.3
  • 17
    • 33845325809 scopus 로고    scopus 로고
    • 2D/3D image fusion for accurate target localization and evaluation of a mask based stereotactic system in fractionated stereotactic radiotherapy of cranial lesions
    • J.Y. Jin, K. Faber, and T. Mikkelsen 2D/3D image fusion for accurate target localization and evaluation of a mask based stereotactic system in fractionated stereotactic radiotherapy of cranial lesions Med Phys 33 2006 4557 4566
    • (2006) Med Phys , vol.33 , pp. 4557-4566
    • Jin, J.Y.1    Faber, K.2    Mikkelsen, T.3
  • 18
    • 71849099108 scopus 로고    scopus 로고
    • The geometric accuracy of frameless stereotactic radiosurgery using 6D robotic couch system
    • T. Takakura, T. Mizowaki, and M. Nakata The geometric accuracy of frameless stereotactic radiosurgery using 6D robotic couch system Phys Med Biol 55 2010 1 10
    • (2010) Phys Med Biol , vol.55 , pp. 1-10
    • Takakura, T.1    Mizowaki, T.2    Nakata, M.3
  • 19
    • 67049107491 scopus 로고    scopus 로고
    • Evaluation of image-guided positioning for frameless intracranial radiosurgery
    • M. Lamba, J.C. Breneman, and R.E. Warnick Evaluation of image-guided positioning for frameless intracranial radiosurgery In J Radiat Oncol Biol Phys 74 2009 913 919
    • (2009) J Radiat Oncol Biol Phys , vol.74 , pp. 913-919
    • Lamba, M.1    Breneman, J.C.2    Warnick, R.E.3
  • 20
    • 69549122373 scopus 로고    scopus 로고
    • Assessment of spatial uncertainties in the radiotherapy process with the Novalis system
    • N. Hayashi, Y. Obata, and Y. Uchiyama Assessment of spatial uncertainties in the radiotherapy process with the Novalis system Int J Radiat Oncol Biol Phys 75 2009 549 557
    • (2009) Int J Radiat Oncol Biol Phys , vol.75 , pp. 549-557
    • Hayashi, N.1    Obata, Y.2    Uchiyama, Y.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.