Volumn 12, Issue 11, 2010, Pages 847-857

Quality of glucose measurement with blood glucose meters at the point-of-care: Relevance of interfering factors

Author keywords

[No Author keywords available]

Indexed keywords

GLUCOSE; MALTOSE; PARACETAMOL; PERITONEAL DIALYSIS FLUID;

EID: 78649386363     PISSN: 15209156     EISSN: None     Source Type: Journal    
DOI: 10.1089/dia.2010.0076     Document Type: Review
Times cited : (56)

References (46)
  • 1
    • 77953064609 scopus 로고    scopus 로고
    • Finger-stick glucose monitoring: Issues of accuracy and specificity
    • Olansky L, Kennedy L: Finger-stick glucose monitoring: issues of accuracy and specificity. Diabetes Care 2010;33:948-949.
    • (2010) Diabetes Care , vol.33 , pp. 948-949
    • Olansky, L.1    Kennedy, L.2
  • 3
    • 77953072186 scopus 로고    scopus 로고
    • Review of adverse events associated with false glucose readings measured by GDH-PQQ-based glucose test strips in the presence of interfering sugars
    • Frias JP, Lim CG, Ellison JM, Montandon CM: Review of adverse events associated with false glucose readings measured by GDH-PQQ-based glucose test strips in the presence of interfering sugars. Diabetes Care 2010;33:728-729.
    • (2010) Diabetes Care , vol.33 , pp. 728-729
    • Frias, J.P.1    Lim, C.G.2    Ellison, J.M.3    Montandon, C.M.4
  • 5
    • 78649367357 scopus 로고    scopus 로고
    • Baxter Healthcare Corp accessed June 30, 2008
    • Baxter Healthcare Corp.: Extraneal (Icodextrin) Peritoneal Dialysis Solution. 2006. www.baxter.com/products/renal/downloads/extraneal-pi.pdf (accessed June 30, 2008).
    • (2006) Extraneal (Icodextrin) Peritoneal Dialysis Solution
  • 6
    • 78649389845 scopus 로고    scopus 로고
    • accessed June 30, 2008
    • Octapharma Pharmazeutika: Octagam® 5%, 2005. www. fda.gov/cber/label/igivoct020306LB.pdf (accessed June 30, 2008).
    • (2005) Octapharma Pharmazeutika: Octagam® 5%
  • 7
    • 78649386296 scopus 로고    scopus 로고
    • Cangene Corp WinRho® SDF. (accessed June 30, 2008)
    • Cangene Corp.: Rh0 (D) Immune Globulin, Intravenous (Human), WinRho® SDF. 2008. www.winrho.com/pdfs/WinRhoPI.pdf (accessed June 30, 2008).
    • (2008) Rh0 (D) Immune Globulin, Intravenous (Human)
  • 9
    • 2542465496 scopus 로고    scopus 로고
    • fda.gov/scripts/cder/drugsat fda/index. cfm, accessdata accessed June 30, 2008
    • Food and Drug Administration: Drugs@FDA; Drug Details. 2008. accessdata.fda.gov/scripts/cder/drugsat fda/index. cfm (accessed June 30, 2008).
    • (2008) Food and Drug Administration: Drugs@FDA; Drug Details
  • 10
    • 78649385401 scopus 로고    scopus 로고
    • Baxter Healthcare Corp accessed June 30, 2008
    • Baxter Healthcare Corp.: Adept®, Adhesion Reduction Solution (4% Icodextrin). 2008. www.baxter.com/products/biopharmaceuticals/downloads/Adept- IFU.pdf (accessed June 30, 2008).
    • (2008) Adept®, Adhesion Reduction Solution (4% Icodextrin).
  • 11
    • 78649384607 scopus 로고    scopus 로고
    • Food and Drug Administration accessed June 30, 2008
    • Food and Drug Administration: FDA New Device Approval. 2006. www.fda.gov/cdrh/mda/docs/p050011.html (accessed June 30, 2008).
    • (2006) FDA New Device Approval.
  • 14
    • 34548306928 scopus 로고    scopus 로고
    • Interference of maltose, icodextrin, galactose, or xylose with some blood glucose monitoring systems
    • Schleis TG: Interference of maltose, icodextrin, galactose, or xylose with some blood glucose monitoring systems. Pharmacotherapy 2007;27:1313-1321.
    • (2007) Pharmacotherapy , vol.27 , pp. 1313-1321
    • Schleis, T.G.1
  • 15
    • 34548680752 scopus 로고    scopus 로고
    • Plasma-equivalent glucose at the point-of-care: Evaluation of Roche Accu-Chek Inform and Abbott Precision PCx glucose meters
    • Ghys T, Goedhuys W, Spincemaille K, Gorus F, Gerlo E: Plasma-equivalent glucose at the point-of-care: evaluation of Roche Accu-Chek Inform and Abbott Precision PCx glucose meters. Clin Chim Acta 2007;386:63-68.
    • (2007) Clin Chim Acta , vol.386 , pp. 63-68
    • Ghys, T.1    Goedhuys, W.2    Spincemaille, K.3    Gorus, F.4    Gerlo, E.5
  • 18
    • 37849039167 scopus 로고    scopus 로고
    • Brussels: International Diabetes Federation
    • rd ed. Brussels: International Diabetes Federation, 2006.
    • (2006) rd Ed
  • 19
    • 33644847768 scopus 로고    scopus 로고
    • Effect of adverse storage conditions on performance of glucometer test strips
    • Bamberg R, Schulman K, MacKenzie M, Moore J, Olchesky S: Effect of adverse storage conditions on performance of glucometer test strips. Clin Lab Sci 2005;18:203-209.
    • (2005) Clin Lab Sci , vol.18 , pp. 203-209
    • Bamberg, R.1    Schulman, K.2    MacKenzie, M.3    Moore, J.4    Olchesky, S.5
  • 20
    • 33847699409 scopus 로고    scopus 로고
    • Performance of glucose dehydrogenase (GDH) based and glucose oxidase (GOX) based blood glucose meter systems at moderately high altitude
    • Bilen H, Kilicaslan A, Akcay G, Albayrak F: Performance of glucose dehydrogenase (GDH) based and glucose oxidase (GOX) based blood glucose meter systems at moderately high altitude. J Med Eng Technol 2007;31:152-156.
    • (2007) J Med Eng Technol , vol.31 , pp. 152-156
    • Bilen, H.1    Kilicaslan, A.2    Akcay, G.3    Albayrak, F.4
  • 21
    • 42149100490 scopus 로고    scopus 로고
    • Accuracy of bedside glucometry in critically ill patients: Influence of clinical characteristics and perfusion index
    • Desachy A, Vuagnat AC, Ghazali AD, Baudin OT, Longuet OH, Calvat SN, Gissot V: Accuracy of bedside glucometry in critically ill patients: influence of clinical characteristics and perfusion index. Mayo Clin Proc 2008;83:400-405.
    • (2008) Mayo Clin Proc , vol.83 , pp. 400-405
    • Desachy, A.1    Vuagnat, A.C.2    Ghazali, A.D.3    Baudin, O.T.4    Longuet, O.H.5    Calvat, S.N.6    Gissot, V.7
  • 23
    • 58449090580 scopus 로고    scopus 로고
    • Prevalence of interfering substances with point-of-care glucose testing in a community hospital
    • Eastham JH, Mason D, Barnes DL, Kollins J: Prevalence of interfering substances with point-of-care glucose testing in a community hospital.AmJHealth Syst Pharm 2009;66:167-170.
    • (2009) Am J Health Syst Pharm , vol.66 , pp. 167-170
    • Eastham, J.H.1    Mason, D.2    Barnes, D.L.3    Kollins, J.4
  • 25
    • 33846947575 scopus 로고    scopus 로고
    • Icodextrin: A major problem for glucose dehydrogenase-based glucose point of care testing systems
    • Floré K, Delanghe J: Icodextrin: a major problem for glucose dehydrogenase-based glucose point of care testing systems. Acta Clin Belg 2006;61:351-354.
    • (2006) Acta Clin Belg , vol.61 , pp. 351-354
    • Floré, K.1    Delanghe, J.2
  • 26
    • 25844510878 scopus 로고    scopus 로고
    • A commercial whole blood glucose biosensor with a low sensitivity to hematocrit based on an impregnated porous carbon electrode
    • Forrow NJ, Bayliff SW: A commercial whole blood glucose biosensor with a low sensitivity to hematocrit based on an impregnated porous carbon electrode. Biosens Bioelectron 2005;21:581-587.
    • (2005) Biosens Bioelectron , vol.21 , pp. 581-587
    • Forrow, N.J.1    Bayliff, S.W.2
  • 29
    • 18244365107 scopus 로고    scopus 로고
    • Interference between Extraneal peritoneal dialysis and the Accu-Chek blood glucose monitor
    • Hoftman N: Interference between Extraneal peritoneal dialysis and the Accu-Chek blood glucose monitor. Anesthe-siology 2005;102:871.
    • (2005) Anesthe-siology , vol.102 , pp. 871
    • Hoftman, N.1
  • 30
    • 0033969498 scopus 로고    scopus 로고
    • The inaccuracy of venous and capillary blood glucose measurement using reagent strips in the newborn period and the effect of haematocrit
    • Hussain K, Sharief N: The inaccuracy of venous and capillary blood glucose measurement using reagent strips in the newborn period and the effect of haematocrit. Early Hum Dev 2000;57:111-121.
    • (2000) Early Hum Dev , vol.57 , pp. 111-121
    • Hussain, K.1    Sharief, N.2
  • 32
    • 0033966742 scopus 로고    scopus 로고
    • Point-of-care glucose testing: Effects of critical care variables, influence of reference instruments, and a modular glucose meter design
    • Louie RF, Tang Z, Sutton, DV, Lee JH, Kost GJ: Point-of-care glucose testing: effects of critical care variables, influence of reference instruments, and a modular glucose meter design. Arch Pathol Lab Med 2000;124:257-266.
    • (2000) Arch Pathol Lab Med , vol.124 , pp. 257-266
    • Louie, R.F.1    Tang, Z.2    Sutton, D.V.3    Lee, J.H.4    Kost, G.J.5
  • 33
    • 0034851188 scopus 로고    scopus 로고
    • Important causes of hypoglycaemia in patients with diabetes on peritoneal dialysis
    • Mehmet S, Quan G, Thomas S, Goldsmith D: Important causes of hypoglycaemia in patients with diabetes on peritoneal dialysis. Diabet Med 2001;18:679-682.
    • (2001) Diabet Med , vol.18 , pp. 679-682
    • Mehmet, S.1    Quan, G.2    Thomas, S.3    Goldsmith, D.4
  • 34
    • 26944464560 scopus 로고    scopus 로고
    • Evaluation of the Glucometer Elite XL device for screening for neonatal hypoglycaemia
    • Michel A, Kuster H, Krebs A, Kadow I, Paul W, Nauck M, Fusch C: Evaluation of the Glucometer Elite XL device for screening for neonatal hypoglycaemia. Eur J Pediatr 2005;164:660-664.
    • (2005) Eur J Pediatr , vol.164 , pp. 660-664
    • Michel, A.1    Kuster, H.2    Krebs, A.3    Kadow, I.4    Paul, W.5    Nauck, M.6    Fusch, C.7
  • 38
    • 0035515892 scopus 로고    scopus 로고
    • Effect of ambient temperature on analytical and clinical performance of a blood glucose monitoring system: Omnitest Sensor glucose meter
    • Nawawi H, Sazali BS, Kamaruzaman BH, Yazid TN, Jemain AA, Ismail F, Khalid BA: Effect of ambient temperature on analytical and clinical performance of a blood glucose monitoring system: Omnitest Sensor glucose meter. Ann Clin Biochem 2001;38:676-683.
    • (2001) Ann Clin Biochem , vol.38 , pp. 676-683
    • Nawawi, H.1    Sazali, B.S.2    Kamaruzaman, B.H.3    Yazid, T.N.4    Jemain, A.A.5    Ismail, F.6    Khalid, B.A.7
  • 39
    • 0033970614 scopus 로고    scopus 로고
    • Effects of exposure at an altitude of 3,000m on performance of glucose meters
    • Pecchio O, Maule S, Migliardi M, Trento M, Veglio M: Effects of exposure at an altitude of 3,000m on performance of glucose meters. Diabetes Care 2000;23:129-131.
    • (2000) Diabetes Care , vol.23 , pp. 129-131
    • Pecchio, O.1    Maule, S.2    Migliardi, M.3    Trento, M.4    Veglio, M.5
  • 40
    • 0021921071 scopus 로고
    • In vitro drug interference with home blood-glucose-measurement systems
    • Rice GK, Galt KA: In vitro drug interference with home blood-glucose-measurement systems. Am J Hosp Pharm 1985;42:2202-2207.
    • (1985) Am J Hosp Pharm , vol.42 , pp. 2202-2207
    • Rice, G.K.1    Galt, K.A.2
  • 41
    • 0034005547 scopus 로고    scopus 로고
    • Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyzer
    • Tang Z, Du X, Louie RF, Kost GJ: Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyzer. Am J Clin Pathol 2000;113:75-86. (Pubitemid 30217342)
    • (2000) American Journal of Clinical Pathology , vol.113 , Issue.1 , pp. 75-86
    • Tang, Z.1    Du, X.2    Louie, R.F.3    Kost, G.J.4
  • 43
    • 0035095386 scopus 로고    scopus 로고
    • Blood glucose meter performance under hyperbaric oxygen conditions
    • Vote DA, Doar O, Moon RE, Toffaletti JG: Blood glucose meter performance under hyperbaric oxygen conditions. Clin Chim Acta 2001;305:81-87.
    • (2001) Clin Chim Acta , vol.305 , pp. 81-87
    • Vote, D.A.1    Doar, O.2    Moon, R.E.3    Toffaletti, J.G.4
  • 44
    • 0029286475 scopus 로고
    • The effect of haematocrit and sample temperature on the Glucotide/Glucometer 4 blood glucose assay system
    • Wiener K: The effect of haematocrit and sample temperature on the Glucotide/Glucometer 4 blood glucose assay system. Diabet Med 1995;12:362-363.
    • (1995) Diabet Med , vol.12 , pp. 362-363
    • Wiener, K.1


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.