메뉴 건너뛰기




Volumn 31 IFMBE, Issue , 2010, Pages 1435-1438

Formula for elastic radial stiffness of the tubular vascular stent

Author keywords

formula; internal force; radial stiffness; vascular stent

Indexed keywords

AXIAL FORCES; BENDING BEAM; CIRCUMFERENTIAL DISPLACEMENT; EQUILIBRIUM EQUATION; FORMULA; INTERNAL FORCE; MECHANICAL PERFORMANCE; RADIAL DISPLACEMENTS; RADIAL STIFFNESS; SYMMETRY CONDITIONS; THEORETICAL MODELS; VASCULAR STENTS;

EID: 77957982142     PISSN: 16800737     EISSN: None     Source Type: Conference Proceeding    
DOI: 10.1007/978-3-642-14515-5_366     Document Type: Conference Paper
Times cited : (2)

References (6)
  • 1
    • 0034930155 scopus 로고    scopus 로고
    • Mechanical property of coronary stents determined by using finite element analysis
    • Etave, F., Finet, G., Boivin, M. et al. (2001) Mechanical property of coronary stents determined by using finite element analysis. Journal of Biomechanics 34: 1065-1075
    • (2001) Journal of Biomechanics , vol.34 , pp. 1065-1075
    • Etave, F.1    Finet, G.2    Boivin, M.3
  • 2
    • 20444475661 scopus 로고    scopus 로고
    • Cardiovascular stent design and vessel stresses: A finite element analysis
    • Lally, C., Dolan, F., Prendergast, P. J. (2005) Cardiovascular stent design and vessel stresses: a finite element analysis. Journal of Biomechanics 38: 1574-1581
    • (2005) Journal of Biomechanics , vol.38 , pp. 1574-1581
    • Lally, C.1    Dolan, F.2    Prendergast, P.J.3
  • 3
    • 10444258936 scopus 로고    scopus 로고
    • A predictive study of the mechanical behaviour of coronary stents by computer modeling
    • Migliavacca, F., Petrini, L., Montanari, V. et al. (2005) A predictive study of the mechanical behaviour of coronary stents by computer modeling. Medical Engineering & Physics 27: 13-18
    • (2005) Medical Engineering & Physics , vol.27 , pp. 13-18
    • Migliavacca, F.1    Petrini, L.2    Montanari, V.3
  • 4
    • 38349062270 scopus 로고    scopus 로고
    • Realistic finite element-based stent design: The impact of balloon folding
    • De Beule, M., Mortier, P., Carlier, S. G. et al. (2008) Realistic finite element-based stent design: The impact of balloon folding. Journal of Biomechanics 41: 383-389
    • (2008) Journal of Biomechanics , vol.41 , pp. 383-389
    • De Beule, M.1    Mortier, P.2    Carlier, S.G.3
  • 5
    • 0027567438 scopus 로고
    • A study of the geometrical and mechanical properties of a self-expanding met allic stent-theory and experiment
    • Jedwab, M. R., Clerc, C. O. (1993) A study of the geometrical and mechanical properties of a self-expanding met allic stent-theory and experiment. Journal of Applied Biomaterials 4: 77-85
    • (1993) Journal of Applied Biomaterials , vol.4 , pp. 77-85
    • Jedwab, M.R.1    Clerc, C.O.2
  • 6
    • 71249143884 scopus 로고    scopus 로고
    • Mechanical formula for the plastic limit pressure of stent during expansion
    • Yang Jie, Huang Nan, (2009) Mechanical formula for the plastic limit pressure of stent during expansion. Acta Mechanica Sinica, 25: 795-801.
    • (2009) Acta Mechanica Sinica , vol.25 , pp. 795-801
    • Jie, Y.1    Nan, H.2


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.