Which device is more accurate to determine the stability mobility of dental implants? A human cadaver study

Author keywords

Bone loss; Dental implants; Mobility; Peri implantitis; Stability

Indexed keywords

ALVEOLAR BONE LOSS; ARTICLE; CADAVER; COMPARATIVE STUDY; ELECTROMAGNETIC FIELD; EQUIPMENT DESIGN; HUMAN; INSTRUMENTATION; MAGNETISM; MANDIBLE; MATERIALS TESTING; METHODOLOGY; PATHOPHYSIOLOGY; TOOTH IMPLANTATION; TOOTH PROSTHESIS; TOOTH SOCKET; TRANSDUCER; VIBRATION;

EID: 77954402442     PISSN: 0305182X     EISSN: 13652842     Source Type: Journal    
DOI: 10.1111/j.1365-2842.2009.02038.x     Document Type: Article
Times cited : (11)

References (35)
  • 1
    • 42149147336 scopus 로고    scopus 로고
    • Implant stability measurements using resonance frequency analysis: biological and biomechanical aspects and clinical implications
    • Sennerby L, Meredith N. Implant stability measurements using resonance frequency analysis: biological and biomechanical aspects and clinical implications. Periodontol 2000 2008, 47:51-66.
    • (2008) Periodontol 2000 , vol.47 , pp. 51-66
    • Sennerby, L.1    Meredith, N.2
  • 4
    • 0344010698 scopus 로고    scopus 로고
    • Correlation of insertion torques with bone mineral density from dental quantitative CT in the mandible
    • Beer A, Gahleitner A, Holm A, Tschabitscher M, Homolka P. Correlation of insertion torques with bone mineral density from dental quantitative CT in the mandible. Clin Oral Implants Res 2003, 14:616-620.
    • (2003) Clin Oral Implants Res , vol.14 , pp. 616-620
    • Beer, A.1    Gahleitner, A.2    Holm, A.3    Tschabitscher, M.4    Homolka, P.5
  • 5
    • 0035489399 scopus 로고    scopus 로고
    • A retrospective analysis of factors associated with multiple implant failures in maxillae
    • Ekfeldt A, Christiansson U, Eriksson T. A retrospective analysis of factors associated with multiple implant failures in maxillae. Clin Oral Implants Res 2001, 12:462-467.
    • (2001) Clin Oral Implants Res , vol.12 , pp. 462-467
    • Ekfeldt, A.1    Christiansson, U.2    Eriksson, T.3
  • 6
    • 0032151752 scopus 로고    scopus 로고
    • Assessment of implant stability as a prognostic determinant
    • Meredith N. Assessment of implant stability as a prognostic determinant. Int J Prosthodont 1998, 11:491-501.
    • (1998) Int J Prosthodont , vol.11 , pp. 491-501
    • Meredith, N.1
  • 7
    • 34547700240 scopus 로고    scopus 로고
    • Clinical study on the primary stability of two dental implant systems with resonance frequency analysis
    • Rabel A, Kohler SG, Schmidt-Westhausen AM. Clinical study on the primary stability of two dental implant systems with resonance frequency analysis. Clin Oral Investig 2007, 11:257-265.
    • (2007) Clin Oral Investig , vol.11 , pp. 257-265
    • Rabel, A.1    Kohler, S.G.2    Schmidt-Westhausen, A.M.3
  • 8
    • 33750554083 scopus 로고    scopus 로고
    • Insertion torque and resonance frequency analysis of dental implant systems in an animal model with loaded implants
    • Al Nawas B, Wagner W, Grotz KA. Insertion torque and resonance frequency analysis of dental implant systems in an animal model with loaded implants. Int J Oral Maxillofac Implants 2006, 21:726-732.
    • (2006) Int J Oral Maxillofac Implants , vol.21 , pp. 726-732
    • Al Nawas, B.1    Wagner, W.2    Grotz, K.A.3
  • 9
    • 0030155796 scopus 로고    scopus 로고
    • Influence of mono- and bicortical anchorage on the integration of titanium implants. A study in the rabbit tibia
    • Ivanoff CJ, Sennerby L, Lekholm U. Influence of mono- and bicortical anchorage on the integration of titanium implants. A study in the rabbit tibia. Int J Oral Maxillofac Surg 1996, 25:229-235.
    • (1996) Int J Oral Maxillofac Surg , vol.25 , pp. 229-235
    • Ivanoff, C.J.1    Sennerby, L.2    Lekholm, U.3
  • 10
    • 0026167719 scopus 로고
    • Early failures in 4,641 consecutively placed Branemark dental implants: a study from stage 1 surgery to the connection of completed prostheses
    • Friberg B, Jemt T, Lekholm U. Early failures in 4,641 consecutively placed Branemark dental implants: a study from stage 1 surgery to the connection of completed prostheses. Int J Oral Maxillofac Implants 1991, 6:142-146.
    • (1991) Int J Oral Maxillofac Implants , vol.6 , pp. 142-146
    • Friberg, B.1    Jemt, T.2    Lekholm, U.3
  • 11
    • 34748907222 scopus 로고    scopus 로고
    • Analysis of the potential association of implant stability, laboratory, and image-based measures used to assess osteotomy sites: early versus delayed loading
    • Tözüm TF, Turkyilmaz I, Yamalik N, Karabulut E, Eratalay K. Analysis of the potential association of implant stability, laboratory, and image-based measures used to assess osteotomy sites: early versus delayed loading. J Periodontol 2007, 78:1675-1682.
    • (2007) J Periodontol , vol.78 , pp. 1675-1682
    • Tözüm, T.F.1    Turkyilmaz, I.2    Yamalik, N.3    Karabulut, E.4    Eratalay, K.5
  • 12
    • 38549135121 scopus 로고    scopus 로고
    • The impact of prosthetic design on the stability, marginal bone loss, peri-implant sulcus fluid volume, and nitric oxide metabolism of conventionally loaded endosseous dental implants: a 12-month clinical study
    • Tözüm TF, Güncü GN, Yamalik N, Turkyilmaz I, Güncü MB. The impact of prosthetic design on the stability, marginal bone loss, peri-implant sulcus fluid volume, and nitric oxide metabolism of conventionally loaded endosseous dental implants: a 12-month clinical study. J Periodontol 2008, 79:55-63.
    • (2008) J Periodontol , vol.79 , pp. 55-63
    • Tözüm, T.F.1    Güncü, G.N.2    Yamalik, N.3    Turkyilmaz, I.4    Güncü, M.B.5
  • 13
    • 33751351701 scopus 로고    scopus 로고
    • Assessment of correlation between computerized tomography values of the bone, and maximum torque and resonance frequency values at dental implant placement
    • Turkyilmaz I, Tözüm TF, Tümer C, Özbek EN. Assessment of correlation between computerized tomography values of the bone, and maximum torque and resonance frequency values at dental implant placement. J Oral Rehabil 2006, 33:881-888.
    • (2006) J Oral Rehabil , vol.33 , pp. 881-888
    • Turkyilmaz, I.1    Tözüm, T.F.2    Tümer, C.3    Özbek, E.N.4
  • 14
    • 34547174147 scopus 로고    scopus 로고
    • Relations between the bone density values from computerized tomography, and implant stability parameters: a clinical study of 230 regular platform implants
    • Turkyilmaz I, Tümer C, Özbek EN, Tözüm TF. Relations between the bone density values from computerized tomography, and implant stability parameters: a clinical study of 230 regular platform implants. J Clin Periodontol 2007, 34:716-722.
    • (2007) J Clin Periodontol , vol.34 , pp. 716-722
    • Turkyilmaz, I.1    Tümer, C.2    Özbek, E.N.3    Tözüm, T.F.4
  • 15
    • 0032061729 scopus 로고    scopus 로고
    • Resonance frequency analysis: measuring implant stability and osseointegration
    • 500, 502
    • Sennerby L, Meredith N. Resonance frequency analysis: measuring implant stability and osseointegration. Compend Contin Educ Dent 1998, 19:493-498. 500, 502
    • (1998) Compend Contin Educ Dent , vol.19 , pp. 493-498
    • Sennerby, L.1    Meredith, N.2
  • 16
    • 38149035045 scopus 로고    scopus 로고
    • Analysis of the possibility of the relationship between various implant-related measures: an 18-month follow-up study
    • Tözüm TF, Turkyilmaz I, Yamalik N, Karabulut E, Türkyilmaz AS, Eratalay K. Analysis of the possibility of the relationship between various implant-related measures: an 18-month follow-up study. J Oral Rehabil 2008, 35:95-104.
    • (2008) J Oral Rehabil , vol.35 , pp. 95-104
    • Tözüm, T.F.1    Turkyilmaz, I.2    Yamalik, N.3    Karabulut, E.4    Türkyilmaz, A.S.5    Eratalay, K.6
  • 17
  • 18
    • 33748314334 scopus 로고    scopus 로고
    • Stability and marginal bone level measurements of unsplinted implants used for mandibular overdentures: a 1-year randomized prospective clinical study comparing early and conventional loading protocols
    • Turkyilmaz I, Sennerby L, Tümer C, Yenigul M, Avci M. Stability and marginal bone level measurements of unsplinted implants used for mandibular overdentures: a 1-year randomized prospective clinical study comparing early and conventional loading protocols. Clin Oral Implants Res 2006, 17:501-505.
    • (2006) Clin Oral Implants Res , vol.17 , pp. 501-505
    • Turkyilmaz, I.1    Sennerby, L.2    Tümer, C.3    Yenigul, M.4    Avci, M.5
  • 19
    • 51149100968 scopus 로고    scopus 로고
    • Relationship between dental implant stability determined by resonance frequency analysis measurements and peri-implant vertical defects: an in vitro study
    • Tözüm TF, Turkyilmaz I, McGlumphy EA. Relationship between dental implant stability determined by resonance frequency analysis measurements and peri-implant vertical defects: an in vitro study. J Oral Rehabil 2008, 35:739-744.
    • (2008) J Oral Rehabil , vol.35 , pp. 739-744
    • Tözüm, T.F.1    Turkyilmaz, I.2    McGlumphy, E.A.3
  • 20
    • 33750541744 scopus 로고    scopus 로고
    • Clinical outcome of 802 immediately loaded 2-stage submerged implants with a new grit-blasted and acid-etched surface: 12-month follow-up
    • Degidi M, Piattelli A, Gehrke P, Carinci F. Clinical outcome of 802 immediately loaded 2-stage submerged implants with a new grit-blasted and acid-etched surface: 12-month follow-up. Int J Oral Maxillofac Implants 2006, 21:763-768.
    • (2006) Int J Oral Maxillofac Implants , vol.21 , pp. 763-768
    • Degidi, M.1    Piattelli, A.2    Gehrke, P.3    Carinci, F.4
  • 21
    • 33947110495 scopus 로고    scopus 로고
    • Influence of transducer orientation on Osstell stability measurements of osseointegrated implants
    • Veltri M, Balleri P, Ferrari M. Influence of transducer orientation on Osstell stability measurements of osseointegrated implants. Clin Implant Dent Relat Res 2007, 9:60-64.
    • (2007) Clin Implant Dent Relat Res , vol.9 , pp. 60-64
    • Veltri, M.1    Balleri, P.2    Ferrari, M.3
  • 22
    • 0034574211 scopus 로고    scopus 로고
    • Measurements comparing the initial stability of five designs of dental implants: a human cadaver study
    • O'Sullivan D, Sennerby L, Meredith N. Measurements comparing the initial stability of five designs of dental implants: a human cadaver study. Clin Implant Dent Relat Res 2000, 2:85-92.
    • (2000) Clin Implant Dent Relat Res , vol.2 , pp. 85-92
    • O'Sullivan, D.1    Sennerby, L.2    Meredith, N.3
  • 23
    • 33645051957 scopus 로고    scopus 로고
    • Resonance frequency analysis and damping capacity assessment. Part 2: peri-implant bone loss follow-up. An in vitro study with the Periotest and Osstell instruments
    • Lachmann S, Laval JY, Jager B. Resonance frequency analysis and damping capacity assessment. Part 2: peri-implant bone loss follow-up. An in vitro study with the Periotest and Osstell instruments. Clin Oral Implants Res 2006, 17:80-84.
    • (2006) Clin Oral Implants Res , vol.17 , pp. 80-84
    • Lachmann, S.1    Laval, J.Y.2    Jager, B.3
  • 24
    • 60049095643 scopus 로고    scopus 로고
    • Influence of defect depth on resonance frequency analysis and insertion torque values for implants placed in fresh extraction sockets: a human cadaver study
    • Turkyilmaz I, Sennerby L, Yilmaz B, Bilecenoglu B, Özbek EN. Influence of defect depth on resonance frequency analysis and insertion torque values for implants placed in fresh extraction sockets: a human cadaver study. Clin Implant Dent Relat Res 2009, 11:52-58.
    • (2009) Clin Implant Dent Relat Res , vol.11 , pp. 52-58
    • Turkyilmaz, I.1    Sennerby, L.2    Yilmaz, B.3    Bilecenoglu, B.4    Özbek, E.N.5
  • 25
    • 33748309280 scopus 로고    scopus 로고
    • Validity and clinical significance of biomechanical testing of implant/bone interface
    • Aparicio C, Lang NP, Rangert B. Validity and clinical significance of biomechanical testing of implant/bone interface. Clin Oral Implants Res 2006, 17(Suppl. 2):2-7.
    • (2006) Clin Oral Implants Res , vol.17 , Issue.SUPPL. 2 , pp. 2-7
    • Aparicio, C.1    Lang, N.P.2    Rangert, B.3
  • 26
    • 62849086385 scopus 로고    scopus 로고
    • A comparative study of two noninvasive techniques to evaluate implant stability: Periotest and Osstell Mentor
    • Oh JS, Kim SG, Lim SC, Ong JL. A comparative study of two noninvasive techniques to evaluate implant stability: Periotest and Osstell Mentor. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2009, 107:513-518.
    • (2009) Oral Surg Oral Med Oral Pathol Oral Radiol Endod , vol.107 , pp. 513-518
    • Oh, J.S.1    Kim, S.G.2    Lim, S.C.3    Ong, J.L.4
  • 27
    • 37549021503 scopus 로고    scopus 로고
    • Longevity of teeth and implants - a systematic review
    • Tomasi C, Wennstrom JL, Berglundh T. Longevity of teeth and implants - a systematic review. J Oral Rehabil 2008, 35(Suppl. 1):23-32.
    • (2008) J Oral Rehabil , vol.35 , Issue.SUPPL. 1 , pp. 23-32
    • Tomasi, C.1    Wennstrom, J.L.2    Berglundh, T.3
  • 28
    • 44349157607 scopus 로고    scopus 로고
    • Clinical outcomes of immediate/early loading of dental implants. A literature review of recent controlled prospective clinical studies
    • Sennerby L, Gottlow J. Clinical outcomes of immediate/early loading of dental implants. A literature review of recent controlled prospective clinical studies. Aust Dent J 2008, 53:S82-S88.
    • (2008) Aust Dent J , vol.53
    • Sennerby, L.1    Gottlow, J.2
  • 29
    • 0030224153 scopus 로고    scopus 로고
    • Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis
    • Meredith N, Alleyne D, Cawley P. Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis. Clin Oral Implants Res 1996, 7:261-267.
    • (1996) Clin Oral Implants Res , vol.7 , pp. 261-267
    • Meredith, N.1    Alleyne, D.2    Cawley, P.3
  • 30
    • 0031152150 scopus 로고    scopus 로고
    • Resonance frequency measurements of implant stability in vivo. A cross-sectional and longitudinal study of resonance frequency measurements on implants in the edentulous and partially dentate maxilla
    • Meredith N, Book K, Friberg B, Jemt T, Sennerby L. Resonance frequency measurements of implant stability in vivo. A cross-sectional and longitudinal study of resonance frequency measurements on implants in the edentulous and partially dentate maxilla. Clin Oral Implants Res 1997, 8:226-233.
    • (1997) Clin Oral Implants Res , vol.8 , pp. 226-233
    • Meredith, N.1    Book, K.2    Friberg, B.3    Jemt, T.4    Sennerby, L.5
  • 31
    • 33947322536 scopus 로고    scopus 로고
    • A biomechanical assessment of the relation between the oral implant stability at insertion and subjective bone quality assessment
    • Alsaadi G, Quirynen M, Michiels K, Jacobs R, van Steenberghe D. A biomechanical assessment of the relation between the oral implant stability at insertion and subjective bone quality assessment. J Clin Periodontol 2007, 34:359-366.
    • (2007) J Clin Periodontol , vol.34 , pp. 359-366
    • Alsaadi, G.1    Quirynen, M.2    Michiels, K.3    Jacobs, R.4    van Steenberghe, D.5
  • 32
  • 33
    • 47249142397 scopus 로고    scopus 로고
    • Measurement of dental implant stability by resonance frequency analysis and damping capacity assessment: comparison of both techniques in a clinical trial
    • Zix J, Hug S, Kessler-Liechti G, Mericske-Stern R. Measurement of dental implant stability by resonance frequency analysis and damping capacity assessment: comparison of both techniques in a clinical trial. Int J Oral Maxillofac Implants 2008, 23:525-530.
    • (2008) Int J Oral Maxillofac Implants , vol.23 , pp. 525-530
    • Zix, J.1    Hug, S.2    Kessler-Liechti, G.3    Mericske-Stern, R.4
  • 34
    • 44349141722 scopus 로고    scopus 로고
    • Initial stability measurement of dental implants placed in different anatomical regions of fresh human cadaver jawbone
    • Seong WJ, Holte JE, Holtan JR, Olin PS, Hodges JS, Ko CC. Initial stability measurement of dental implants placed in different anatomical regions of fresh human cadaver jawbone. J Prosthet Dent 2008, 99:425-434.
    • (2008) J Prosthet Dent , vol.99 , pp. 425-434
    • Seong, W.J.1    Holte, J.E.2    Holtan, J.R.3    Olin, P.S.4    Hodges, J.S.5    Ko, C.C.6
  • 35
    • 4644230057 scopus 로고    scopus 로고
    • Resonance frequency analysis of implants subjected to immediate or early functional occlusal loading. Successful vs. failing implants
    • Glauser R, Sennerby L, Meredith N. Resonance frequency analysis of implants subjected to immediate or early functional occlusal loading. Successful vs. failing implants. Clin Oral Implants Res 2004, 15:428-434.
    • (2004) Clin Oral Implants Res , vol.15 , pp. 428-434
    • Glauser, R.1    Sennerby, L.2    Meredith, N.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.