Nuclear Regulatory Commission regulatory status of approved intravascular brachytherapy systems
-
DOI 10.1016/S1522-1865(02)00151-8, PII S1522186502001518
-
G. P. Glasgow, " Nuclear Regulatory Commission regulatory status of approved intravascular brachytherapy systems.," Cardiovasc. Radiat. Med. 3, 60-66 (2002). 10.1016/S1522-1865(02)00151-8 1522-1865 (Pubitemid 35461903)
-
(2002)
Cardiovascular Radiation Medicine
, vol.3
, Issue.1
, pp. 60-66
-
-
Glasgow, G.P.1
-
2
-
-
85027217171
-
-
See, for example, Klein v. Biscup, 673 N.E.2d 225, 231 (Ohio Ct. Ap), appeal denied, 667 N.E.2d 987 (Ohio)
-
See, for example, Klein v. Biscup, 673 N.E.2d 225, 231 (Ohio Ct. App.), appeal denied, 667 N.E.2d 987 (Ohio 1996).
-
(1996)
-
-
-
3
-
-
85027237383
-
-
Note
-
See, e.g., Rhone-Poulenc Rorer Pharm., Inc, v. Marion Merrell Dow, Inc., 93 F.3d 511, 514 n.3 (8th Cir., 1996), Bristol Myers Squibb Co. v. Shalala, 91 F.3d 1493 (D.C. Cir. 1996); Weaver v. Reagen, 886 F.2d 194, 198 (8th Cir. 1989); United States v. An Article of DeviceDiapulse, 768 F.2d 826, 832 (7th Cir. 1985); Schlessing v. United States, 239 F.2d 885, 886 (9th Cir. 1956); Washington Legal Found., 880 F. Supp. at 28, n.1.; United States v. Evers, 453 F. Supp. 1141, 1149-1150 (M.D. Ala. 1978), aff'd, 643 F.2d 1043, 1052-53 (5th Cir. 1981); FTC v. Simeon Management Corp., 391 F. Supp. 697, 706 (N.D. Cal. 1975), aff'd, 532 F.2d 708 (9th Cir. 1976); Upjohn Co. v. MacMurdo, 562 So.2d 680, 683 (Fla. 1990); Jones v. Petland Orlando Store, 622 So.2d 1114, 1115 (Fla Dist. Ct. App. 1993); Haynes v, Baton Rouge Gen. Hosp. 298 So.2d 149, 153 (La. Ct. App. 1974), writ denied, 302 So.2d 33 (La. 1974); Peter Hutt, Regulation of the Practice of Medicine Under the Pure Food & Drug Laws, 33 Food & Drug Officials Q. Bull. 7-11 (1969).
-
-
-
-
5
-
-
0026510050
-
Off-label uses of anticancer drugs. Letter to the Editor
-
JAMAAP 0098-7484. 10.1001/jama.267.18.2473
-
M. D. Abeloff, " Off-label uses of anticancer drugs. Letter to the Editor.," JAMA, J. Am. Med. Assoc. JAMAAP 0098-7484 267, 2473-2474 (1992). 10.1001/jama.267.18.2473
-
(1992)
JAMA, J. Am. Med. Assoc.
, vol.267
, pp. 2473-2474
-
-
Abeloff, M.D.1
-
6
-
-
77952387100
-
-
Proctor v. Davis, 682 N.E.2d 1203, 1213 (Ill. Ap)
-
Proctor v. Davis, 682 N.E.2d 1203, 1213 (Ill. App. 1997).
-
(1997)
-
-
-
7
-
-
0027179396
-
-
See, e.g., Washington Legal Foundation v. Kessler, 880 F. Su26, 28 n. 1(D.D.C.). Off-label can mean many things. "[U]sing an approved drug to treat a disease that is not indicated on its label, but is closely related to an indicated disease, treating unrelated, unindicated diseases, and treating the indicated disease but varying from the indicated dosage, regimen, or patient population may all be considered off-label use.", 48 Food and Drug Law Journal, (1993) (footnotes omitted)]
-
See, e.g., Washington Legal Foundation v. Kessler, 880 F. Supp. 26, 28 n. 1(D.D.C. 1995). Off-label can mean many things. "[U]sing an approved drug to treat a disease that is not indicated on its label, but is closely related to an indicated disease, treating unrelated, unindicated diseases, and treating the indicated disease but varying from the indicated dosage, regimen, or patient population may all be considered off-label use." W. Christopher, " Off-label drug prescription: Filling the regulatory vacuum.," 48 Food and Drug Law Journal, pp. 247-262 (1993) (footnotes omitted)].
-
(1995)
Off-label Drug Prescription: Filling the Regulatory Vacuum
, pp. 247-262
-
-
Christopher, W.1
-
8
-
-
0031923812
-
-
See, e.g., Food and Drug Law Journal 53, Femrite v. Abbot Northwestern Hospital, 568 N.W.2d 535 at 539 n4 (Minn. Ct. A1997); Klein v. Biscup, 673 N.E.2d 225, 231 (Ohio Ct. Ap), appeal denied, 667 N.E.2d 987 (Ohio 1996); In re Orthopedic Bone Screw Products Liability Litigation, MDL No. 1014, 1996 WL 107556 at 3 (E.D. Pa. Mar. 8, 1996) ("the decision whether or not to use a drug for an off-label purpose is a matter of medical judgment, not of regulatory approval") (citation and quotation marks omitted). The FDCA was "not intended as a medical practices act and [did] not interfere with the practice of the healing art." S.R. No. 361, 74th Cong., 1st Sess. 3 (1935)
-
See, e.g., J. M. Beck and E. D. Azari, " FDA: Off-label use, and informed consent: Debunking myths and misconceptions.," Food and Drug Law Journal 53, 71-104 (1998); Femrite v. Abbot Northwestern Hospital, 568 N.W.2d 535 at 539 n4 (Minn. Ct. App. 1997); Klein v. Biscup, 673 N.E.2d 225, 231 (Ohio Ct. App.), appeal denied, 667 N.E.2d 987 (Ohio 1996); In re Orthopedic Bone Screw Products Liability Litigation, MDL No. 1014, 1996 WL 107556 at 3 (E.D. Pa. Mar. 8, 1996) ("the decision whether or not to use a drug for an off-label purpose is a matter of medical judgment, not of regulatory approval") (citation and quotation marks omitted). The FDCA was "not intended as a medical practices act and [did] not interfere with the practice of the healing art." S.R. No. 361, 74th Cong., 1st Sess. 3 (1935).
-
(1998)
FDA: Off-label Use, and Informed Consent: Debunking Myths and Misconceptions
, pp. 71-104
-
-
Beck, J.M.1
Azari, E.D.2
-
9
-
-
85027238809
-
-
Genetic Engineering and Biotechnology News (GEN), 1 June
-
P. S. Spivak, " Off-label promotion under scrutiny by DOJ, The Department of Justice is aggressively pursuing healthcare fraud allegations.," Genetic Engineering and Biotechnology News (GEN), 1 June 2008, Vol. 1, No. 11.
-
(2008)
Off-label Promotion under Scrutiny by DOJ, the Department of Justice Is Aggressively Pursuing Healthcare Fraud Allegations
, vol.1
, Issue.11
-
-
Spivak, P.S.1
-
10
-
-
85027241833
-
-
The Wall Street Journal, Front Page, 19 November, Vol. CCLII, No.120
-
The Wall Street Journal, Front Page, 19 November 2008, Vol. CCLII, No.120.
-
(2008)
-
-
-
12
-
-
65949101284
-
Guidance for industry on good reprint practices for the distribution of medical journal articles and medical or scientific reference publications on unapproved new uses of approved drugs and approved or cleared medical devices: Availability
-
Department of Health and Human Services, Food and Drug Administration, Office of the Commissioner, Office of Policy, FEREAC 0097-6326
-
Department of Health and Human Services, Food and Drug Administration, Office of the Commissioner, Office of Policy, " Guidance for industry on good reprint practices for the distribution of medical journal articles and medical or scientific reference publications on unapproved new uses of approved drugs and approved or cleared medical devices: Availability.," Fed. Regist. FEREAC 0097-6326 74 (8), 1694-1695 (2009).
-
(2009)
Fed. Regist.
, vol.74
, Issue.8
, pp. 1694-1695
-
-
-
13
-
-
0036305225
-
Uses of drugs not described in the package insert (off-label uses)
-
DOI 10.1542/peds.110.1.181
-
American Academy of Pediatrics, Committee on Drugs, " Uses of drugs not described in the package insert: Off-label uses.," Pediatrics PEDIAU 0031-4005 110, 181-183 (2002). 10.1542/peds.110.1.181 (Pubitemid 34742660)
-
(2002)
Pediatrics
, vol.110
, Issue.1
, pp. 181-183
-
-
Ward, R.M.1
Bates, B.A.2
Benitz, W.E.3
Burchfield, D.J.4
Ring, J.C.5
Walls, R.P.6
Walson, P.D.7
-
15
-
-
33644685442
-
The major role of clinicians in the discovery of off-label drug therapies
-
DOI 10.1592/phco.26.3.323
-
H. J. DeMonaco, A. Ali, and E. von Hippel, " The major role of clinicians in the discovery of off-label drug therapies.," Pharmacotherapy ZZZZZZ 0277-0008 26, 323-332 (2006). 10.1592/phco.26.3.323 (Pubitemid 43334671)
-
(2006)
Pharmacotherapy
, vol.26
, Issue.3
, pp. 323-332
-
-
DeMonaco, H.J.1
Ali, A.2
Von Hippel, E.3
-
16
-
-
1242293824
-
Verification of Ir-192 near source dosimetry using GAFCHROMIC® film
-
DOI 10.1118/1.1637733
-
® film.," Med. Phys. MPHYA6 0094-2405 31, 201-207 (2004). 10.1118/1.1637733 (Pubitemid 38235390)
-
(2004)
Medical Physics
, vol.31
, Issue.2
, pp. 201-207
-
-
Chiu-Tsao, S.-T.1
Duckworth, T.L.2
Patel, N.S.3
Pisch, J.4
Harrison, L.B.5
-
17
-
-
77952401619
-
Cordis Checkmate: Manually loaded iridium-192 ribbon
-
3rd ed., edited by R. Waksman (Futura, Armonk)
-
S. K. Jani, G. Long, G. B. Huppe, V. Massullo, P. Tripuraneni, and P. S. Teirstein, " Cordis Checkmate: Manually loaded iridium-192 ribbon.," in Vascular Brachytherapy, 3rd ed., edited by, R. Waksman, (Futura, Armonk, 2002), pp. 707-716.
-
(2002)
Vascular Brachytherapy
, pp. 707-716
-
-
Jani, S.K.1
Long, G.2
Huppe, G.B.3
Massullo, V.4
Tripuraneni, P.5
Teirstein, P.S.6
-
18
-
-
0005465728
-
Novoste Beta-Cath system: Intracoronary radiation with strontium 90
-
3rd ed., edited by R. Waksman (Futura, Armonk)
-
K. W. Elliot, R. P. Walsh, and R. diMonda, " Novoste Beta-Cath system: Intracoronary radiation with strontium 90.," in Vascular Brachytherapy, 3rd ed., edited by, R. Waksman, (Futura, Armonk, 2002), pp. 717-728.
-
(2002)
Vascular Brachytherapy
, pp. 717-728
-
-
Elliot, K.W.1
Walsh, R.P.2
Dimonda, R.3
-
19
-
-
0032692758
-
Radiochromic film dosimetry of a high dose rate beta source for intravascular brachytherapy
-
MPHYA6 0094-2405. 10.1118/1.598814
-
D. M. Duggan, C. W. Coffey II, J. L. Lobdell, and M. C. Schell, " Radiochromic film dosimetry of a high dose rate beta source for intravascular brachytherapy.," Med. Phys. MPHYA6 0094-2405 26, 2461-2464 (1999). 10.1118/1.598814
-
(1999)
Med. Phys.
, vol.26
, pp. 2461-2464
-
-
Duggan, D.M.1
Ii C. W, C.2
Lobdell, J.L.3
Schell, M.C.4
-
20
-
-
0035996455
-
90Y intravascular brachytherapy sources used with balloon catheters
-
DOI 10.1118/1.1488604
-
M. Piessens, N. Reynaert, J. Potempa, H. Thierens, W. Wijns, and L. Verbeke, " Dose distributions for Y 90 intravascular brachytherapy sources used with balloon catheters.," Med. Phys. MPHYA6 0094-2405 29, 1562-1571 (2002). 10.1118/1.1488604 (Pubitemid 34774589)
-
(2002)
Medical Physics
, vol.29
, Issue.7
, pp. 1562-1571
-
-
Piessens, M.1
Reynaert, N.2
Potempa, J.3
Thierens, H.4
Wijns, W.5
Verbeke, L.6
-
21
-
-
2442650301
-
90Sr/Y source trains at submillimeter distances
-
DOI 10.1118/1.1697695
-
D. E. Roa, H. Song, N. Yue, F. d'Errico, and R. Nath, " Dosimetric characteristics of the Novoste Beta-Cath S 90 r/Y source trains at submillimeter distances.," Med. Phys. MPHYA6 0094-2405 31, 1269-1276 (2004). 10.1118/1.1697695 (Pubitemid 38656802)
-
(2004)
Medical Physics
, vol.31
, Issue.5
, pp. 1269-1276
-
-
Roa, D.E.1
Song, H.2
Yue, N.3
D'Errico, F.4
Nath, R.5
-
22
-
-
0034967228
-
The water-equivalence of phantom materials for S 90 r- y 90 beta particles
-
MPHYA6 0094-2405. 10.1118/1.1376176
-
L. A. Buckley, B. R. Thomadsen, and L. A. DeWerd, " The water-equivalence of phantom materials for S 90 r- Y 90 beta particles.," Med. Phys. MPHYA6 0094-2405 28, 1010-1015 (2001). 10.1118/1.1376176
-
(2001)
Med. Phys.
, vol.28
, pp. 1010-1015
-
-
Buckley, L.A.1
Thomadsen, B.R.2
Dewerd, L.A.3
-
23
-
-
0037665088
-
The Angiorad Gamma wire system
-
3rd ed., edited by R. Waksman (Futura, Armonk)
-
R. Waksman, D. P. Faxon, and S. F. Liprie, " The Angiorad Gamma wire system.," in Vascular Brachytherapy, 3rd ed., edited by, R. Waksman, (Futura, Armonk, 2002), pp. 771-778.
-
(2002)
Vascular Brachytherapy
, pp. 771-778
-
-
Waksman, R.1
Faxon, D.P.2
Liprie, S.F.3
-
24
-
-
77952335710
-
The Guidant coronary source wire system
-
3rd ed., edited by R. Waksman (Futura, Armonk)
-
A. E. Raizner, G. L. Kaluza, R. V. Caffee, and A. J. Bradshaw, " The Guidant coronary source wire system.," in Vascular Brachytherapy, 3rd ed., edited by, R. Waksman, (Futura, Armonk, 2002), pp. 737-752.
-
(2002)
Vascular Brachytherapy
, pp. 737-752
-
-
Raizner, A.E.1
Kaluza, G.L.2
Caffee, R.V.3
Bradshaw, A.J.4
-
25
-
-
0034850033
-
32P intravascular brachytherapy source from Monte Carlo calculations
-
DOI 10.1118/1.1384459
-
T. D. Bohm, F. A. Mourtada, and R. K. Das, " Dose rate table for a P 32 intravascular brachytherapy source from Monte Carlo calculations.," Med. Phys. MPHYA6 0094-2405 28, 1770-1775 (2001). 10.1118/1.1384459 (Pubitemid 32783222)
-
(2001)
Medical Physics
, vol.28
, Issue.8
, pp. 1770-1775
-
-
Bohm, T.D.1
Mourtada, F.A.2
Das, R.K.3
-
26
-
-
0033624388
-
32P catheter-based vascular brachytherapy source wire
-
DOI 10.1118/1.1286551
-
F. A. Mourtada, C. G. Soares, S. M. Seltzer, and S. H. Lott, " Dosimetry characterization of P 32 catheter-based vascular brachytherapy source wire.," Med. Phys. MPHYA6 0094-2405 27, 1770-1776 (2000). 10.1118/1.1286551 (Pubitemid 30666352)
-
(2000)
Medical Physics
, vol.27
, Issue.8
, pp. 1770-1776
-
-
Mourtada, F.A.1
Soares, C.G.2
Seltzer, S.M.3
Lott, S.H.4
-
27
-
-
0037572354
-
32P source wire used for intravascular brachytherapy with automated stepping
-
DOI 10.1118/1.1567832
-
F. Mourtada, C. G. Soares, S. M. Seltzer, P. M. Bergstrom, Jr., J. M. Fernandez-Verea, J. Asenjoa, and S. H. Lott, " Dosimetry characterization of a P 32 source wire used for intravascular brachytherapy with automated stepping.," Med. Phys. MPHYA6 0094-2405 30, 959-971 (2003). 10.1118/1.1567832 (Pubitemid 36578700)
-
(2003)
Medical Physics
, vol.30
, Issue.5
, pp. 959-971
-
-
Mourtada, F.1
Soares, C.G.2
Seltzer, S.M.3
Bergstrom Jr., P.M.4
Fernandez-Verea, J.M.5
Asenjo, J.6
Lott, S.H.7
-
28
-
-
1642332960
-
32P source Monte Carlo model to derive AAPM TG-60 dosimetric parameters used for intravascular brachytherapy
-
DOI 10.1118/1.1646651
-
F. Mourtada, C. G. Soares, and J. L. Horton, " A segmented P 32 source Monte Carlo model to derive AAPM TG-60 dosimetric parameters used for intravascular brachytherapy.," Med. Phys. MPHYA6 0094-2405 31, 602-608 (2004). 10.1118/1.1646651 (Pubitemid 38381024)
-
(2004)
Medical Physics
, vol.31
, Issue.3
, pp. 602-608
-
-
Mourtada, F.1
Soares, C.G.2
Horton, J.L.3
-
29
-
-
1242316317
-
32P intravascular brachytherapy source wires using Monte Carlo codes PENELOPE and GEANT4
-
DOI 10.1118/1.1637970
-
J. Torres, M. J. Buades, J. F. Almansa, R. Guerrero, and A. M. Lallena, " Dosimetry characterization of P 32 intravascular brachytherapy source wires using Monte Carlo codes PENELOPE and GEANT4.," Med. Phys. MPHYA6 0094-2405 31, 296-304 (2004). 10.1118/1.1637970 (Pubitemid 38235399)
-
(2004)
Medical Physics
, vol.31
, Issue.2
, pp. 296-304
-
-
Torres, J.1
Buades, M.J.2
Almansa, J.F.3
Guerrero, R.4
Lallena, A.M.5
-
30
-
-
0034851951
-
32P source for intravascular brachytherapy
-
DOI 10.1118/1.1388222
-
R. Wang and X. Allen Li, " Monte Carlo characterization of a P 32 source for intravascular brachytherapy.," Med. Phys. MPHYA6 0094-2405 28, 1776-1795 (2001). 10.1118/1.1388222 (Pubitemid 32783223)
-
(2001)
Medical Physics
, vol.28
, Issue.8
, pp. 1776-1785
-
-
Wang, R.1
Li, X.A.2
-
31
-
-
0030292806
-
32P source for endovascular irradiation
-
DOI 10.1016/S0360-3016(96)00300-8, PII S0360301696003008
-
Z. Xu, P. R. Almond, and J. O. Deasy, " The dose distribution produced by a P 32 source for endovascular irradiation.," Int. J. Radiat. Oncol., Biol., Phys. IOBPD3 0360-3016 36, 933-939 (1996). 10.1016/S0360-3016(96) 00300-8 (Pubitemid 27008610)
-
(1996)
International Journal of Radiation Oncology Biology Physics
, vol.36
, Issue.4
, pp. 933-939
-
-
Xu, Z.1
Almond, P.R.2
Deasy, J.O.3
-
32
-
-
0030609049
-
32P wire for catheter-based endovascular irradiation
-
DOI 10.1118/1.597944
-
Z. Xu, L. E. Reinstein, G. Yang, S. Pai, G. Gluckman, and P. R. Almond, " The investigation of P 32 wire for catheter-based endovascular irradiation.," Med. Phys. MPHYA6 0094-2405 24, 1788-1792 (1997). 10.1118/1.597944 (Pubitemid 27495915)
-
(1997)
Medical Physics
, vol.24
, Issue.11
, pp. 1788-1792
-
-
Xu, Z.1
Reinstein, L.E.2
Yang, G.3
Pai, S.4
Gluckman, G.5
Almond, P.R.6
-
33
-
-
0028970569
-
Intra-arterial y 90 brachytherapy: Preliminary dosimetric study using a specially modified angioplasty balloon
-
IOBPD3 0360-3016. 10.1016/0360-3016(95)00259-2
-
Y. Popowski, V. Verin, I. Papirov, P. Nouet, M. Rouzaud, E. Grob, M. Schwager, P. Urban, W. Rutishauser, and J. M. Kurtz, " Intra-arterial Y 90 brachytherapy: Preliminary dosimetric study using a specially modified angioplasty balloon.," Int. J. Radiat. Oncol., Biol., Phys. IOBPD3 0360-3016 33, 713-717 (1995). 10.1016/0360-3016(95)00259-2
-
(1995)
Int. J. Radiat. Oncol., Biol., Phys.
, vol.33
, pp. 713-717
-
-
Popowski, Y.1
Verin, V.2
Papirov, I.3
Nouet, P.4
Rouzaud, M.5
Grob, E.6
Schwager, M.7
Urban, P.8
Rutishauser, W.9
Kurtz, J.M.10
-
34
-
-
0030293787
-
Endovascular β-irradiation after percutaneous transluminal coronary balloon angioplasty
-
DOI 10.1016/S0360-3016(96)00411-7, PII S0360301696004117
-
Y. Popowski, V. Verin, and P. Urban, " Endovascular Β -irradiation after percutaneous transluminal coronary angioplasty.," Int. J. Radiat. Oncol., Biol., Phys. IOBPD3 0360-3016 36, 841-845 (1996). 10.1016/S0360-3016(96)00411-7 (Pubitemid 27008598)
-
(1996)
International Journal of Radiation Oncology Biology Physics
, vol.36
, Issue.4
, pp. 841-845
-
-
Popowski, Y.1
Verin, V.2
Urban, P.3
-
35
-
-
0030294367
-
A novel system for intracoronary β-irradiation: Description and dosimetric results
-
DOI 10.1016/S0360-3016(96)00409-9, PII S0360301696004099
-
Y. Popowski, V. Verin, M. Schwager, P. Nouet, I. Papirov, M. Rouzaud, P. Urban, J. R. Landis, W. Rutishauser, and J. M. Kurtz, " A novel system for intracornary Β -irradiation: Description and dosimetric results.," Int. J. Radiat. Oncol., Biol., Phys. IOBPD3 0360-3016 36, 923-931 (1996). 10.1016/S0360-3016(96)00409-9 (Pubitemid 27008609)
-
(1996)
International Journal of Radiation Oncology Biology Physics
, vol.36
, Issue.4
, pp. 923-931
-
-
Popowski, Y.1
Verin, V.2
Schwager, M.3
Nouet, P.4
Papirov, I.5
Rouzaud, M.6
Urban, P.7
Landis, J.R.8
Rutishauser, W.9
Kurtz, J.M.10
-
36
-
-
0342906197
-
Dosimetry of a W-188/Re-188 beta line source for endovascular brachytherapy
-
DOI 10.1118/1.598928
-
U. O. Häfeli, W. K. Roberts, D. S. Meier, J. P. Ciezki, G. J. Pauer, E. J. Lee, and M. S. Weinhous, " Dosimetry of a W-188/Re-188 beta line source for endovascular brachytherapy.," Med. Phys. MPHYA6 0094-2405 27, 668-675 (2000). 10.1118/1.598928 (Pubitemid 30227043)
-
(2000)
Medical Physics
, vol.27
, Issue.4
, pp. 668-675
-
-
Hafeli, U.O.1
Roberts, W.K.2
Meier, D.S.3
Ciezki, J.P.4
Pauer, G.J.5
Lee, E.J.6
Weinhous, M.S.7
-
37
-
-
0036852442
-
The scaling method applied to beta particle line sources with a finite diameter
-
DOI 10.1118/1.1513567
-
D. R. Schaart, " The scaling method applied to beta particle line sources with a finite diameter.," Med. Phys. MPHYA6 0094-2405 29, 2682-2686 (2002). 10.1118/1.1513567 (Pubitemid 35375467)
-
(2002)
Medical Physics
, vol.29
, Issue.11
, pp. 2682-2686
-
-
Schaart, D.R.1
-
38
-
-
77952400969
-
The calculation and measurement of radiation dose surrounding radioactive stents: The effects of radionuclide selection and stent design on dosimetric results
-
3rd ed., edited by R. Waksman (Futura, Armonk)
-
C. W. Coffey II and D. M. Duggan, " The calculation and measurement of radiation dose surrounding radioactive stents: The effects of radionuclide selection and stent design on dosimetric results.," in Vascular Brachytherapy, 3rd ed., edited by, R. Waksman, (Futura, Armonk, 2002), pp. 349-378.
-
(2002)
Vascular Brachytherapy
, pp. 349-378
-
-
Ii C. W, C.1
Duggan, D.M.2
-
39
-
-
0030899828
-
Radiation dose from a phosphorous-32 impregnated wire mesh vascular stent
-
DOI 10.1118/1.597910
-
C. Janicki, D. M. Duggan, C. W. Coffey, D. R. Fischell, and T. A. Fischell, " Radiation dose from a phosphorous-32 impregnated wire mesh vascular stent.," Med. Phys. MPHYA6 0094-2405 24, 437-445 (1997). 10.1118/1.597910 (Pubitemid 27133489)
-
(1997)
Medical Physics
, vol.24
, Issue.3
, pp. 437-445
-
-
Janicki, C.1
Duggan, D.M.2
Coffey, C.W.3
Fischell, D.R.4
Fischell, T.A.5
-
40
-
-
0032742628
-
Dose model for a beta-emitting stent in a realistic artery consisting of soft tissue and plaque
-
MPHYA6 0094-2405. 10.1118/1.598813
-
C. Janicki, D. M. Duggan, A. Gonzalez, C. W. Coffey II, and D. A. Rahdert, " Dose model for a beta-emitting stent in a realistic artery consisting of soft tissue and plaque.," Med. Phys. MPHYA6 0094-2405 26, 2451-2460 (1999). 10.1118/1.598813
-
(1999)
Med. Phys.
, vol.26
, pp. 2451-2460
-
-
Janicki, C.1
Duggan, D.M.2
Gonzalez, A.3
Ii C. W, C.4
Rahdert, D.A.5
-
41
-
-
77952362942
-
Liquid-filled balloons for coronary brachytherapy
-
3rd ed., edited by R. Waksman (Futura, Armonk)
-
J. Weinberger and F. F. Knap, Jr., " Liquid-filled balloons for coronary brachytherapy.," in Vascular Brachytherapy, 3rd ed., edited by, R. Waksman, (Futura, Armonk, 2002), pp. 753-770.
-
(2002)
Vascular Brachytherapy
, pp. 753-770
-
-
Weinberger, J.1
Knap Jr., F.F.2
-
42
-
-
0029842002
-
Dosimetry of a radioactive coronary balloon dilitation catheter for treatment of neointimal hyperplasia
-
DOI 10.1118/1.597761
-
H. I. Amols, L. E. Reinstein, and J. Weinberger, " Dosimetry of a radioactive coronary balloon dilation catheter for treatment of neointimal hyperplasia.," Med. Phys. MPHYA6 0094-2405 23, 1783-1788 (1996). 10.1118/1.597761 (Pubitemid 26349791)
-
(1996)
Medical Physics
, vol.23
, Issue.10
, pp. 1783-1788
-
-
Amols, H.I.1
Reinstein, L.E.2
Weinberger, J.3
-
43
-
-
0037320142
-
Dosimetry study of Re-188 liquid balloon for intravascular brachytherapy using polymer gel dosimeters and laser-beam optical CT scanner
-
DOI 10.1118/1.1533749
-
C. -S. Wuu, P. Schiff, M. J. Maryanski, T. Liu, S. Borzillary, and J. Weinberger, " Dosimetry study of Re-188 liquid balloon for intravascular brachytherapy using polymer gel dosimeters and laser-beam optical CT scanner.," Med. Phys. MPHYA6 0094-2405 30, 132-137 (2003). 10.1118/1.1533749 (Pubitemid 36222848)
-
(2003)
Medical Physics
, vol.30
, Issue.2
, pp. 132-137
-
-
Wuu, C.-S.1
Schiff, P.2
Maryanski, M.J.3
Liu, T.4
Borzillary, S.5
Weinberger, J.6
-
44
-
-
77952409837
-
Xenon 133 gas filled balloon
-
3rd ed., edited by R. Waksman (Futura, Armonk)
-
M. Apple and R. Waksman, " Xenon 133 gas filled balloon.," in Vascular Brachytherapy, 3rd ed., edited by, R. Waksman, (Futura, Armonk, 2002), pp. 787-798.
-
(2002)
Vascular Brachytherapy
, pp. 787-798
-
-
Apple, M.1
Waksman, R.2
-
45
-
-
77952344072
-
RDX coronary radiation delivery system: Balloon-based radiation therapy
-
3rd ed., edited by R. Waksman (Futura, Armonk)
-
M. Buchbinder, G. Strathearn, and B. Trauthen, " RDX coronary radiation delivery system: Balloon-based radiation therapy.," in Vascular Brachytherapy, 3rd ed., edited by, R. Waksman, (Futura, Armonk, 2002), pp. 799-810.
-
(2002)
Vascular Brachytherapy
, pp. 799-810
-
-
Buchbinder, M.1
Strathearn, G.2
Trauthen, B.3
-
46
-
-
77952358008
-
Balloon catheters impregnated with radioisotopes for the prevention of restenosis
-
3rd ed., R. Waksman (Futura, Armonk)
-
C. Hehrlein, A. Kovacs, G. K. Wolf, N. Yue, and R. Nath, " Balloon catheters impregnated with radioisotopes for the prevention of restenosis.," in Vascular Brachytherapy, 3rd ed., R. Waksman, (Futura, Armonk, 2002), pp. 811-818.
-
(2002)
Vascular Brachytherapy
, pp. 811-818
-
-
Hehrlein, C.1
Kovacs, A.2
Wolf, G.K.3
Yue, N.4
Nath, R.5
-
47
-
-
77952408777
-
Peripheral vascular brachytherapy using the Nucletron Paris catheter and the microSelectron-HDR afterloader
-
3rd ed., edited by R. Waksman (Futura, Armonk)
-
F. M. van Kireken and E. Loeffler, " Peripheral vascular brachytherapy using the Nucletron Paris catheter and the microSelectron-HDR afterloader.," in Vascular Brachytherapy, 3rd ed., edited by, R. Waksman, (Futura, Armonk, 2002), pp. 729-736.
-
(2002)
Vascular Brachytherapy
, pp. 729-736
-
-
Van Kireken, F.M.1
Loeffler, E.2
-
48
-
-
4243427819
-
External radiation therapy for patients with arteriovenous dialysis shunts
-
3rd ed., edited by R. Waksman (Futura, Armonk)
-
S. Parikh and D. Nori, " External radiation therapy for patients with arteriovenous dialysis shunts.," in Vascular Brachytherapy, 3rd ed., edited by, R. Waksman, (Futura, Armonk, 2002), pp. 505-524.
-
(2002)
Vascular Brachytherapy
, pp. 505-524
-
-
Parikh, S.1
Nori, D.2
-
49
-
-
85027241086
-
-
See (Cordis) for Cordis Checkmate system clearance (last accessed 3 November)
-
See http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfTopic/pma/pma. cfm?num=p990036 (Cordis) for Cordis Checkmate system clearance (last accessed 3 November 2009).
-
(2009)
-
-
-
50
-
-
85027231702
-
-
See (Novoste) for Novost Beta-Cath system clearance (last accessed 3 November)
-
See http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfTopic/pma/pma. cfm?num=p000018 (Novoste) for Novost Beta-Cath system clearance (last accessed 3 November 2009).
-
(2009)
-
-
-
51
-
-
85027221160
-
-
See (Guidant) for GALILEO intravascular radiotherapy system clearance (last accessed 3 November)
-
See http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfTopic/pma/pma. cfm?num=p000052 (Guidant) for GALILEO intravascular radiotherapy system clearance (last accessed 3 November 2009).
-
(2009)
-
-
-
52
-
-
0034806529
-
Beta versus gamma dosimetry close to Ir-192 brachytherapy sources
-
DOI 10.1118/1.1395038
-
D. Baltas, P. Karaiskos, P. Papagiannis, L. Sakelliou, E. Loeffler, and N. Zamboglou, " Beta versus gamma dosimetry close to Ir-192 brachytherapy sources.," Med. Phys. MPHYA6 0094-2405 28, 1875-1882 (2001). 10.1118/1.1395038 (Pubitemid 32899726)
-
(2001)
Medical Physics
, vol.28
, Issue.9
, pp. 1875-1882
-
-
Baltas, D.1
Karaiskos, P.2
Papagiannis, P.3
Sakelliou, L.4
Loeffler, E.5
Zamboglou, N.6
-
53
-
-
0343776958
-
A high-precision, high-resolution and fast dosimetry system for beta sources applied in cardiovascular brachytherapy
-
DOI 10.1118/1.598927
-
M. Bambynek, D. Flühs, U. Quast, D. Wegener, and C. G. Soares, " A high-precision, high-resolution and fast dosimetry system for beta sources applied in cardiovascular brachytherapy.," Med. Phys. MPHYA6 0094-2405 27, 662-667 (2000). 10.1118/1.598927 (Pubitemid 30227042)
-
(2000)
Medical Physics
, vol.27
, Issue.4
, pp. 662-667
-
-
Bambynek, M.1
Fluhs, D.2
Quast, U.3
Wegener, D.4
Soares, C.G.5
-
54
-
-
0037241039
-
IVBTMC, a Monte Carlo dose calculation tool for intravascular brachytherapy
-
DOI 10.1118/1.1528177
-
O. Chibani and X. Allen Li, " IVBTMC: A Monte Carlo dose calculation tool for intravascular brachytherapy.," Med. Phys. MPHYA6 0094-2405 30, 44-51 (2003). 10.1118/1.1528177 (Pubitemid 36106007)
-
(2003)
Medical Physics
, vol.30
, Issue.1
, pp. 44-51
-
-
ChibaniX, O.1
Li, A.2
-
55
-
-
0036771725
-
192Ir sources
-
DOI 10.1118/1.1508378
-
P. Papagiannis, A. Angelopoulos, E. Pantelis, L. Sakelliou, D. Baltas, P. Karaiskos, P. Sandilos, and L. Vlachos, " Dosimetry comparison of I 192 r sources.," Med. Phys. MPHYA6 0094-2405 29, 2239-2246 (2002). 10.1118/1.1508378 (Pubitemid 35204431)
-
(2002)
Medical Physics
, vol.29
, Issue.10
, pp. 2239-2246
-
-
Papagiannis, P.1
Angelopoulos, A.2
Pantelis, E.3
Sakelliou, L.4
Baltas, D.5
Karaiskos, P.6
Sandilos, P.7
Vlachos, L.8
-
56
-
-
0035045004
-
On the applicability of the AAPM TG-60/TG-43 dose calculation formalism to intravascular line sources: Proposal for an adapted formalism
-
DOI 10.1118/1.1357458
-
D. R. Schaart, M. C. Clarijs, and A. J. J. Bos, " On the applicability of the AAPM TG-60/TG-43 dose calculation formalism to intravascular line sources: Proposal for an adapted formalism.," Med. Phys. MPHYA6 0094-2405 28, 638-653 (2001). 10.1118/1.1357458 (Pubitemid 32323316)
-
(2001)
Medical Physics
, vol.28
, Issue.4
, pp. 638-653
-
-
Schaart, D.R.1
Clarijs, M.C.2
Bos, A.J.J.3
-
57
-
-
0031942882
-
Calibration and characterization of beta-particle sources for intravascular brachytherapy
-
DOI 10.1118/1.598204
-
C. G. Soares, D. G. Halpern, and C. -K. Wang, " Calibration and characterization of beta-particle sources for intravascular brachytherapy., " Med. Phys. MPHYA6 0094-2405 25, 339-346 (1998). 10.1118/1.598204 (Pubitemid 28146390)
-
(1998)
Medical Physics
, vol.25
, Issue.3
, pp. 339-346
-
-
Soares, C.G.1
Halpern, D.G.2
Wang, C.-K.3
-
58
-
-
18744400224
-
90Y seed in water and a150: Comment on "calibration and characterization of beta-particle sources for intravascular brachytherapy" [Med. Phys. 25, 339-346 (1998)]
-
DOI 10.1118/1.1510455
-
J. Asenjo, J. M. Fernández-Varea, A. Sánchez-Reyes, and C. -K. Wang, " Radial dose function of a S 90 r- Y 90 seed in water and A150: Comment on Calibration and characterization of beta-particle sources for intravascular brachytherapy' [Med. Phys. 25, 339-346. 1998].," Med. Phys. MPHYA6 0094-2405 29, 2737-2738 (2002). 10.1118/1.1510455 (Pubitemid 35375474)
-
(2002)
Medical Physics
, vol.29
, Issue.11
, pp. 2737-2738
-
-
Asenjo, J.1
Fernandez-Varea, J.M.2
Sanchez-Reyes, A.3
Wang, C.-K.4
-
59
-
-
0034081128
-
Monte Carlo modeling of radiation dose distributions in intravascular radiation therapy
-
DOI 10.1118/1.598973
-
M. G. Stabin, M. Könijnenberg, F. F. Knapp, Jr., and R. H. Spencer, " Monte Carlo modeling of radiation dose distributions in intravascular radiation therapy.," Med. Phys. MPHYA6 0094-2405 27, 1086-1092 (2000). 10.1118/1.598973 (Pubitemid 30350019)
-
(2000)
Medical Physics
, vol.27
, Issue.5
, pp. 1086-1092
-
-
Stabin, M.G.1
Konijnenberg, M.2
Knapp Jr., F.F.3
Spencer, R.H.4
-
60
-
-
0033678807
-
A Monte Carlo calculation of dosimetric parameters of S 90 / y 90 and i 192 r SS sources for intravascular brachytherapy
-
MPHYA6 0094-2405. 10.1118/1.1319374
-
R. Wang and X. Allen Li, " A Monte Carlo calculation of dosimetric parameters of S 90 / Y 90 and I 192 r SS sources for intravascular brachytherapy.," Med. Phys. MPHYA6 0094-2405 27, 2528-2535 (2000). 10.1118/1.1319374
-
(2000)
Med. Phys.
, vol.27
, pp. 2528-2535
-
-
Wang, R.1
Allen Li, X.2
-
61
-
-
0035117289
-
Monte Carlo dose calculations of beta-emitting sources for intravascular brachytherapy: A comparison between EGS4, EGSnrc, and MCNP
-
DOI 10.1118/1.1339880
-
R. Wang and X. Allen Li, " Monte Carlo dose calculations of beta-emitting sources for intravascular brachytherapy: A comparison between EGS4, EGSnrc, and MCNP.," Med. Phys. MPHYA6 0094-2405 28, 134-141 (2001). 10.1118/1.1339880 (Pubitemid 32165748)
-
(2001)
Medical Physics
, vol.28
, Issue.2
, pp. 134-141
-
-
Wang, R.1
Li, X.A.2
-
62
-
-
0036703411
-
Dose characterization in the near-source region for two high dose rate brachytherapy sources
-
DOI 10.1118/1.1493780
-
R. Wang and X. Allen Li, " Dose characterization in the near-source region for two high dose rate brachytherapy sources.," Med. Phys. MPHYA6 0094-2405 29, 1678-1686 (2002). 10.1118/1.1493780 (Pubitemid 35364202)
-
(2002)
Medical Physics
, vol.29
, Issue.8
, pp. 1678-1686
-
-
Wang, R.1
Li, X.A.2
-
63
-
-
0037341169
-
Dosimetric characteristics of a linear array of γ or β-emitting seeds in intravascular irradiation: Monte Carlo studies for the AAPM TG-43/60 formalism
-
DOI 10.1118/1.1538229
-
S. -J. Ye, E. I. Parsai, and J. J. Feldmeier, " Dosimetric characteristics of a linear array of γ or Β -emitting seeds in intravascular irradiation: Monte Carlo studies for the AAPM TG-43/60 formalism.," Med. Phys. MPHYA6 0094-2405 30, 403-414 (2003). 10.1118/1.1538229 (Pubitemid 36337107)
-
(2003)
Medical Physics
, vol.30
, Issue.3
, pp. 403-414
-
-
Ye, S.-J.1
Parsai, E.I.2
Feldmeier, J.J.3
-
64
-
-
0038372095
-
Dose distributions in bifurcated coronary vessels treated with catheter-based photon and beta emitters in intravascular brachytherapy
-
DOI 10.1118/1.1582813
-
N. Yue, K. B. Roberts, S. E. Pfau, and R. Nath, " Dose distributions in bifurcated coronary vessels treated with catheter based photon and beta emitters in intravascular brachytherapy.," Med. Phys. MPHYA6 0094-2405 30, 1628-1636 (2003). 10.1118/1.1582813 (Pubitemid 36885780)
-
(2003)
Medical Physics
, vol.30
, Issue.7
, pp. 1628-1636
-
-
Yue, N.1
Roberts, K.B.2
Pfau, S.E.3
Nath, R.4
-
65
-
-
0034760771
-
Dosimetric effect of source centering and residual plaque for β-emitting catheter based intravascular brachytherapy sources
-
DOI 10.1118/1.1406520
-
V. Sehgal, Z. Li, J. R. Palta, and W. E. Bolch, " Dosimetric effect of source centering and residual plaque for Β -emitting catheter based intravascular brachytherapy sources.," Med. Phys. MPHYA6 0094-2405 28, 2162-2171 (2001). 10.1118/1.1406520 (Pubitemid 32977563)
-
(2001)
Medical Physics
, vol.28
, Issue.10
, pp. 2162-2171
-
-
Sehgal, V.1
Li, Z.2
Palta, J.R.3
Bolch, W.E.4
-
66
-
-
0035996463
-
Application of imaging-derived parameters to dosimetry of intravascular brachytherapy sources: Perturbation effects of residual plaque burden
-
DOI 10.1118/1.1485053
-
V. Sehgal, Z. Li, J. R. Palta, K. M. Smith, and W. E. Bolch, " Application of imaging-derived parameters to dosimetry of intravascular brachytherapy sources: Perturbation effects of residual plaque burden.," Med. Phys. MPHYA6 0094-2405 29, 1580-1589 (2002). 10.1118/1.1485053 (Pubitemid 34774591)
-
(2002)
Medical Physics
, vol.29
, Issue.7
, pp. 1580-1589
-
-
Sehgal, V.1
Li, Z.2
Palta, J.R.3
Smith, K.M.4
Bolch, W.E.5
-
67
-
-
0032956067
-
188Re
-
DOI 10.1118/1.598586
-
R. A. Fox and P. W. Henson, " The effect of contrast medium and balloon shape on dosimetry for arterial irradiation with R 188 e.," Med. Phys. MPHYA6 0094-2405 26, 771-776 (1999). 10.1118/1.598586 (Pubitemid 29227128)
-
(1999)
Medical Physics
, vol.26
, Issue.5
, pp. 771-776
-
-
Fox, R.A.1
Henson, P.W.2
-
68
-
-
0342468208
-
Dosimetric effects of contrast media for catheter-based intravascular brachytherapy
-
DOI 10.1118/1.1367279
-
X. A. Li, " Dosimetric effects of contrast media for catheter-based intravascular brachytherapy.," Med. Phys. MPHYA6 0094-2405 28, 757-763 (2001). 10.1118/1.1367279 (Pubitemid 32467253)
-
(2001)
Medical Physics
, vol.28
, Issue.5
, pp. 757-763
-
-
Li, X.A.1
-
69
-
-
0036118007
-
Dosimetric effects of source-offset in intravascular brachytherapy
-
DOI 10.1118/1.1461373
-
O. Chibani and X. Allen Li, " Dosimetric effects of source-offset in intravascular brachytherapy.," Med. Phys. MPHYA6 0094-2405 29, 530-537 (2002). 10.1118/1.1461373 (Pubitemid 34408766)
-
(2002)
Medical Physics
, vol.29
, Issue.4
, pp. 530-537
-
-
Chibani, O.1
Allen Li, X.2
-
70
-
-
0037243003
-
90Y source with balloon for intravascular brachytherapy
-
DOI 10.1118/1.1527673
-
R. Wang, X. A. Li, and J. Lobdell, " Monte Carlo dose characterization of a new S 90 r/ Y 90 source with balloon for intravascular brachytherapy.," Med. Phys. MPHYA6 0094-2405 30, 27-33 (2003). 10.1118/1.1527673 (Pubitemid 36106004)
-
(2003)
Medical Physics
, vol.30
, Issue.1
, pp. 27-33
-
-
Wang, R.1
Allen Li, X.2
Lobdell, J.3
-
71
-
-
1142274256
-
Treatment of in-stent restenosis for saphenous vein grafts using intravascular brachytherapy: Regulatory challenges and clinical application
-
1042-3931
-
M. J. Rivard, H. D. Zinkin, J. E. Mignano, and C. D. Kimmelstiel, " Treatment of in-stent restenosis for saphenous vein grafts using intravascular brachytherapy: Regulatory challenges and clinical application.," J. Invasive Cardiol. 16, 96-98 (2004). 1042-3931
-
(2004)
J. Invasive Cardiol.
, vol.16
, pp. 96-98
-
-
Rivard, M.J.1
Zinkin, H.D.2
Mignano, J.E.3
Kimmelstiel, C.D.4
-
72
-
-
1142292138
-
90Sr for saphenous vein grafts having diameters ranging from 2.0-5.0 mm
-
DOI 10.1016/j.brachy.2004.05.004, PII S1538472104000662
-
J. E. Mignano and M. J. Rivard, " Intravascular brachytherapy using S 90 r for saphenous vein grafts having diameters ranging from 2.0-5.0 mm.," Brachytherapy BRACC4 1538-4721 3, 173-178 (2004). 10.1016/j.brachy.2004.05.004 (Pubitemid 39472683)
-
(2004)
Brachytherapy
, vol.3
, Issue.3
, pp. 173-178
-
-
Mignano, J.E.1
Rivard, M.J.2
-
73
-
-
10744222554
-
Virtual 3D IVUS vessel model for intravascular brachytherapy planning. I. 3D segmentation, reconstruction, and visualization of coronary artery architecture and orientation
-
DOI 10.1118/1.1603964
-
F. Weichert, H. Müller, U. Quast, A. Kraushaar, P. Spilles, M. Heintz, C. Wilke, C. von Birgelen, R. Erbel, and D. Wegener, " Virtual 3D IVUS vessel model for intravascular brachytherapy planning I: 3D segmentation, reconstruction, and visualization of coronary artery architecture and orientation.," Med. Phys. MPHYA6 0094-2405 30, 2530-2536 (2003). 10.1118/1.1603964 (Pubitemid 37129759)
-
(2003)
Medical Physics
, vol.30
, Issue.9
, pp. 2530-2536
-
-
Weichert, F.1
Muller, H.2
Quast, U.3
Kraushaar, A.4
Spilles, P.5
Heintz, M.6
Wilke, C.7
Von Birgelen, C.8
Erbel, R.9
Wegener, D.10
-
74
-
-
0038715627
-
GAF film dosimetry of a tandem positioned β-emitting intravascular brachytherapy source train
-
DOI 10.1118/1.1573206
-
I. Iftimia, P. M. Devlin, L. M. Chin, J. M. Baron, and R. A. Cormack, " GAF film dosimetry of a tandem positioned b-emitting intravascular brachytherapy source train.," Med. Phys. MPHYA6 0094-2405 30, 1004-1012 (2003). 10.1118/1.1573206 (Pubitemid 36735964)
-
(2003)
Medical Physics
, vol.30
, Issue.6
, pp. 1004-1012
-
-
Iftimia, I.1
Devlin, P.M.2
Chin, L.M.3
Baron, J.M.4
Cormack, R.A.5
-
75
-
-
0036918330
-
90Y radiation source train stepping for intravascular brachytherapy
-
DOI 10.1118/1.1524169
-
M. M. Miften, S. K. Das, T. D. Shafman, and L. B. Marks, " Optimization of a S 90 r/ Y 90 radiation source train stepping for intravascular brachytherapy.," Med. Phys. MPHYA6 0094-2405 29, 2891-2896 (2002). 10.1118/1.1524169 (Pubitemid 36008485)
-
(2002)
Medical Physics
, vol.29
, Issue.12
, pp. 2891-2896
-
-
Miften, M.M.1
Das, S.K.2
Shafman, T.D.3
Marks, L.B.4
-
76
-
-
3142751126
-
90Y
-
DOI 10.1118/1.1753431
-
P. M. Ortego, C. Prieto, and E. Vano, " Monte Carlo parametric study of stent impact on dose for catheter-based intravascular brachytherapy with S 90 r/ Y 90.," Med. Phys. MPHYA6 0094-2405 31, 1964-1971 (2004). 10.1118/1.1753431 (Pubitemid 38924815)
-
(2004)
Medical Physics
, vol.31
, Issue.7
, pp. 1964-1971
-
-
Ortego, P.M.1
Prieto, C.2
Vano, E.3
-
77
-
-
0036122336
-
Monte Carlo dosimetry of a tandem positioned beta-emitting intravascular brachytherapy source train
-
DOI 10.1118/1.1461845
-
S. A. Wallace, W. Schumer, and M. Horrigan, " Monte Carlo dosimetry of a tandem positioned beta-emitting intravascular brachytherapy source train.," Med. Phys. MPHYA6 0094-2405 29, 544-549 (2002). 10.1118/1.1461845 (Pubitemid 34408768)
-
(2002)
Medical Physics
, vol.29
, Issue.4
, pp. 544-549
-
-
Wallace, S.A.1
Schumer, W.2
Horrigan, M.3
-
78
-
-
0037148683
-
Intravascular brachytherapy of the coronary arteries
-
DOI 10.1088/0031-9155/47/4/201, PII S003191550220531X
-
R. A. Fox, " Intravascular brachytherapy of the coronary arteries.," Phys. Med. Biol. PHMBA7 0031-9155 47, R1-R30 (2002). 10.1088/0031-9155/47/4/201 (Pubitemid 34232652)
-
(2002)
Physics in Medicine and Biology
, vol.47
, Issue.4
-
-
Fox, R.A.1
-
79
-
-
0036851669
-
192Ir line sources using a standard length well chamber
-
DOI 10.1118/1.1517046
-
I. A. Brezovich, R. A. Popple, J. Duan, S. Shen, and P. N. Pareek, " Assaying I 192 r line sources using a standard length well chamber.," Med. Phys. MPHYA6 0094-2405 29, 2692-2697 (2002). 10.1118/1.1517046 (Pubitemid 35375469)
-
(2002)
Medical Physics
, vol.29
, Issue.11
, pp. 2692-2697
-
-
Brezovich, I.A.1
Popple, R.A.2
Duan, J.3
Shen, S.4
Pareek, P.N.5
-
80
-
-
0032692241
-
Assessment of the strength of individual i 192 r seeds in ribbons
-
MPHYA6 0094-2405. 10.1118/1.598766
-
B. Thomadsen, L. DeWerd, T. McNutt, S. DeWerd, and D. Schmidt, " Assessment of the strength of individual I 192 r seeds in ribbons.," Med. Phys. MPHYA6 0094-2405 26, 2471-2475 (1999). 10.1118/1.598766
-
(1999)
Med. Phys.
, vol.26
, pp. 2471-2475
-
-
Thomadsen, B.1
Dewerd, L.2
McNutt, T.3
Dewerd, S.4
Schmidt, D.5
-
81
-
-
0036918526
-
A quality management program in intravascular brachytherapy
-
DOI 10.1118/1.1524167
-
A. Chakri and B. Thomadsen, " A quality management program for intravascular brachytherapy.," Med. Phys. MPHYA6 0094-2405 29, 2850-2860 (2002). 10.1118/1.1524167 (Pubitemid 36008481)
-
(2002)
Medical Physics
, vol.29
, Issue.12
, pp. 2850-2860
-
-
Chakri, A.1
Thomadsen, B.2
-
82
-
-
85027235235
-
-
See for TheraSphere's clearance (last accessed 3 November)
-
See http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cftopic/pma/pma. cfm?num=H980006 for TheraSphere's clearance (last accessed 3 November 2009).
-
(2009)
-
-
-
83
-
-
85027218541
-
-
See for SIR-Spheres clearance (last accessed 3 November)
-
See http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfTopic/pma/pma. cfm?num=p990065 for SIR-Spheres clearance (last accessed 3 November 2009).
-
(2009)
-
-
-
84
-
-
85027224222
-
-
See for TheraSphere and SIRSpheres Yttrium-90 Microspheres, September (last accessed 13 November 2009)
-
See http://www.nrc.gov/materials/miau/med-use-toolkit.html for TheraSphere and SIRSpheres Yttrium-90 Microspheres, September 2008 (last accessed 13 November 2009).
-
(2008)
-
-
-
85
-
-
85027226247
-
-
See for FDA Medical Device Database. August (last accessed 3 November 2009)
-
See http://www.fda.gov/cdrh/databases.html.24 for FDA Medical Device Database. August 2008 (last accessed 3 November 2009).
-
(2008)
-
-
|