Comparison of Corneal Powers Obtained from 4 Different Devices

Author keywords

[No Author keywords available]

Indexed keywords

ADULT; ANALYZER; ARTICLE; ASTIGMATISM; CONTROLLED STUDY; CORNEA; DEVICE; FEMALE; HUMAN; INTERMETHOD COMPARISON; KERATOMETRY; LENS POWER; MALE; NORMAL HUMAN; PRIORITY JOURNAL; PROSPECTIVE STUDY; REPRODUCIBILITY;

EID: 70349265986     PISSN: 00029394     EISSN: None     Source Type: Journal    
DOI: 10.1016/j.ajo.2009.04.028     Document Type: Article
Times cited : (137)

References (25)
  • 1
    • 0026970445 scopus 로고
    • Comparison of the accuracy and reproducibility of the keratometer and EyeSys Corneal Analysis System Model I
    • Koch D.D., Wakil J.S., Samuelson S.W., and Haft E.A. Comparison of the accuracy and reproducibility of the keratometer and EyeSys Corneal Analysis System Model I. J Cataract Refract Surg 18 (1992) 342-347
    • (1992) J Cataract Refract Surg , vol.18 , pp. 342-347
    • Koch, D.D.1    Wakil, J.S.2    Samuelson, S.W.3    Haft, E.A.4
  • 3
    • 0027179635 scopus 로고
    • Comparison of the accuracy of computerized videokeratography and keratometry for use in intraocular lens calculations
    • Cuaycong M.J., Gay C.A., Emery J., Haft E.A., and Koch D.D. Comparison of the accuracy of computerized videokeratography and keratometry for use in intraocular lens calculations. J Cataract Refract Surg 19 (1993) 178-181
    • (1993) J Cataract Refract Surg , vol.19 , pp. 178-181
    • Cuaycong, M.J.1    Gay, C.A.2    Emery, J.3    Haft, E.A.4    Koch, D.D.5
  • 4
    • 0030781960 scopus 로고    scopus 로고
    • Comparison of portable automated keratometry and manual keratometry for IOL calculation
    • Manning C.A., and Kloess P.M. Comparison of portable automated keratometry and manual keratometry for IOL calculation. J Cataract Refract Surg 23 (1997) 1213-1216
    • (1997) J Cataract Refract Surg , vol.23 , pp. 1213-1216
    • Manning, C.A.1    Kloess, P.M.2
  • 5
    • 0035205647 scopus 로고    scopus 로고
    • Reproducibility of optical biometry using partial coherence interferometry: intraobserver and interobserver reliability
    • Vogel A., Dick H.B., and Krummenauer F. Reproducibility of optical biometry using partial coherence interferometry: intraobserver and interobserver reliability. J Cataract Refract Surg 27 (2001) 1961-1968
    • (2001) J Cataract Refract Surg , vol.27 , pp. 1961-1968
    • Vogel, A.1    Dick, H.B.2    Krummenauer, F.3
  • 6
    • 33845495864 scopus 로고    scopus 로고
    • Comparison of different techniques of anterior chamber depth and keratometric measurements
    • Elbaz U., Barkana Y., Gerber Y., Avni I., and Zadok D. Comparison of different techniques of anterior chamber depth and keratometric measurements. Am J Ophthalmol 143 (2007) 48-53
    • (2007) Am J Ophthalmol , vol.143 , pp. 48-53
    • Elbaz, U.1    Barkana, Y.2    Gerber, Y.3    Avni, I.4    Zadok, D.5
  • 8
    • 1642318773 scopus 로고    scopus 로고
    • Reliability of the IOLMaster in measuring corneal power changes after photorefractive keratectomy
    • Rosa N., Capasso L., Lanza M., Furgiuele D., and Romano A. Reliability of the IOLMaster in measuring corneal power changes after photorefractive keratectomy. J Cataract Refract Surg 30 (2004) 409-413
    • (2004) J Cataract Refract Surg , vol.30 , pp. 409-413
    • Rosa, N.1    Capasso, L.2    Lanza, M.3    Furgiuele, D.4    Romano, A.5
  • 9
    • 0037254713 scopus 로고    scopus 로고
    • Optical and ultrasound measurement of axial length and anterior chamber depth for intraocular lens calculation
    • Nemeth J., Fekete O., and Pesztenlehrer N. Optical and ultrasound measurement of axial length and anterior chamber depth for intraocular lens calculation. J Cataract Refract Surg 29 (2003) 85-88
    • (2003) J Cataract Refract Surg , vol.29 , pp. 85-88
    • Nemeth, J.1    Fekete, O.2    Pesztenlehrer, N.3
  • 11
    • 0037399122 scopus 로고    scopus 로고
    • Clinical evaluation of the Shin-Nippon Nvision-K5001/Grand Seiko WR-5100K autorefractor
    • Davies L.N., Mallen E.A., Wolffsohn J.S., and Gilmartin B. Clinical evaluation of the Shin-Nippon Nvision-K5001/Grand Seiko WR-5100K autorefractor. Optom Vis Sci 80 (2003) 320-324
    • (2003) Optom Vis Sci , vol.80 , pp. 320-324
    • Davies, L.N.1    Mallen, E.A.2    Wolffsohn, J.S.3    Gilmartin, B.4
  • 12
    • 0038068829 scopus 로고    scopus 로고
    • First steps with the Zeiss IOLMaster: a comparison between acoustic contact biometry and non-contact optical biometry
    • Gantenbein C., Lang H.M., Ruprecht K.W., and Georg T. First steps with the Zeiss IOLMaster: a comparison between acoustic contact biometry and non-contact optical biometry. Klin Monatsbl Augenheilkd 220 (2003) 309-314
    • (2003) Klin Monatsbl Augenheilkd , vol.220 , pp. 309-314
    • Gantenbein, C.1    Lang, H.M.2    Ruprecht, K.W.3    Georg, T.4
  • 14
    • 0026652930 scopus 로고
    • The repeatability of measurement of the ocular components
    • Zadnik K., Mutti D.O., and Adams A.J. The repeatability of measurement of the ocular components. Invest Ophthalmol Vis Sci 33 (1992) 2325-2333
    • (1992) Invest Ophthalmol Vis Sci , vol.33 , pp. 2325-2333
    • Zadnik, K.1    Mutti, D.O.2    Adams, A.J.3
  • 15
    • 37349056419 scopus 로고    scopus 로고
    • Anterior segment biometry with the Pentacam: comprehensive assessment of repeatability of automated measurements
    • Shankar H., Taranath D., Santhirathelagan C.T., and Pesudovs K. Anterior segment biometry with the Pentacam: comprehensive assessment of repeatability of automated measurements. J Cataract Refract Surg 34 (2008) 103-113
    • (2008) J Cataract Refract Surg , vol.34 , pp. 103-113
    • Shankar, H.1    Taranath, D.2    Santhirathelagan, C.T.3    Pesudovs, K.4
  • 16
    • 0035142284 scopus 로고    scopus 로고
    • Analysis of aggregate surgically induced refractive change, prediction error, and intraocular astigmatism
    • Holladay J.T., Moran J.R., and Kezirian G.M. Analysis of aggregate surgically induced refractive change, prediction error, and intraocular astigmatism. J Cataract Refract Surg 27 (2001) 61-79
    • (2001) J Cataract Refract Surg , vol.27 , pp. 61-79
    • Holladay, J.T.1    Moran, J.R.2    Kezirian, G.M.3
  • 17
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • Bland J.M., and Altman D.G. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1 (1986) 307-310
    • (1986) Lancet , vol.1 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 18
    • 41949113224 scopus 로고    scopus 로고
    • Sources of error in intraocular lens power calculation
    • Norrby S. Sources of error in intraocular lens power calculation. J Cataract Refract Surg 34 (2008) 368-376
    • (2008) J Cataract Refract Surg , vol.34 , pp. 368-376
    • Norrby, S.1
  • 19
    • 0026990947 scopus 로고
    • Sources of error in intraocular lens power calculation
    • Olsen T. Sources of error in intraocular lens power calculation. J Cataract Refract Surg 18 (1992) 125-129
    • (1992) J Cataract Refract Surg , vol.18 , pp. 125-129
    • Olsen, T.1
  • 20
    • 52249090464 scopus 로고    scopus 로고
    • Comparison and reproducibility of corneal thickness and curvature readings obtained by the Galilei and the Orbscan II analysis systems
    • Menassa N., Kaufmann C., Goggin M., Job O.M., Bachmann L.M., and Thiel M.A. Comparison and reproducibility of corneal thickness and curvature readings obtained by the Galilei and the Orbscan II analysis systems. J Cataract Refract Surg 34 (2008) 1742-1747
    • (2008) J Cataract Refract Surg , vol.34 , pp. 1742-1747
    • Menassa, N.1    Kaufmann, C.2    Goggin, M.3    Job, O.M.4    Bachmann, L.M.5    Thiel, M.A.6
  • 21
    • 0036124279 scopus 로고    scopus 로고
    • The performance of four different corneal topographers on normal human corneas and its impact on orthokeratology lens fitting
    • Cho P., Lam A.K., Mountford J., and Ng L. The performance of four different corneal topographers on normal human corneas and its impact on orthokeratology lens fitting. Optom Vis Sci 79 (2002) 175-183
    • (2002) Optom Vis Sci , vol.79 , pp. 175-183
    • Cho, P.1    Lam, A.K.2    Mountford, J.3    Ng, L.4
  • 22
    • 57749191496 scopus 로고    scopus 로고
    • Repeatability, reproducibility, and agreement characteristics of rotating Scheimpflug photography and scanning-slit corneal topography for corneal power measurement
    • Kawamorita T., Uozato H., Kamiya K., et al. Repeatability, reproducibility, and agreement characteristics of rotating Scheimpflug photography and scanning-slit corneal topography for corneal power measurement. J Cataract Refract Surg 35 (2009) 127-133
    • (2009) J Cataract Refract Surg , vol.35 , pp. 127-133
    • Kawamorita, T.1    Uozato, H.2    Kamiya, K.3
  • 23
    • 0018937490 scopus 로고
    • Biometry of 7,500 cataractous eyes
    • Hoffer K.J. Biometry of 7,500 cataractous eyes. Am J Ophthalmol 90 (1980) 360-368
    • (1980) Am J Ophthalmol , vol.90 , pp. 360-368
    • Hoffer, K.J.1
  • 24
    • 0028220914 scopus 로고
    • Cataract patients in a defined Swedish population 1986-1990. II. Preoperative observations
    • Ninn-Pedersen K., Stenevi U., and Ehinger B. Cataract patients in a defined Swedish population 1986-1990. II. Preoperative observations. Acta Ophthalmol (Copenh) 72 (1994) 10-15
    • (1994) Acta Ophthalmol (Copenh) , vol.72 , pp. 10-15
    • Ninn-Pedersen, K.1    Stenevi, U.2    Ehinger, B.3
  • 25
    • 0009516737 scopus 로고
    • Incidence of astigmatism in a cataract practice
    • Buzard K., Shearing S., and Relyea R. Incidence of astigmatism in a cataract practice. J Refract Surg 4 (1988) 173-178
    • (1988) J Refract Surg , vol.4 , pp. 173-178
    • Buzard, K.1    Shearing, S.2    Relyea, R.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.