Volumn 28, Issue 6, 2009, Pages 657-666

Repeatability of oculus pentacam metrics derived from corneal topography

Author keywords

Corneal anterior and posterior surface; Corneal radius of curvature; Corneal shape; Corneal topography; Repeatability; Scheimpflug imaging

Indexed keywords

ADULT; AGING; ARTICLE; CAMERA; COMPUTER PROGRAM; CONTROLLED STUDY; DIGITAL IMAGING; EYELID REFLEX; HUMAN; HUMAN EXPERIMENT; KERATOMETRY; NORMAL HUMAN; PRIORITY JOURNAL; REPRODUCIBILITY;

EID: 67650624137     PISSN: 02773740     EISSN: None     Source Type: Journal    
DOI: 10.1097/ICO.0b013e31819b01b5     Document Type: Article
Times cited : (29)

References (66)
  • 1
    • 0034865916 scopus 로고    scopus 로고
    • Precision of three topography instruments in keratoconus subjects
    • McMahon TT, Anderson RJ, Joslin CE, et al. Precision of three topography instruments in keratoconus subjects. Optom Vis Sci. 2001;78:599-604.
    • (2001) Optom Vis Sci , vol.78 , pp. 599-604
    • McMahon, T.T.1    Anderson, R.J.2    Joslin, C.E.3
  • 2
    • 0001008215 scopus 로고    scopus 로고
    • Maeda N, Klyce SD. Videokeratography in contact lens practice. Int Contact Lens Clin. 1994;21:163-169.
    • Maeda N, Klyce SD. Videokeratography in contact lens practice. Int Contact Lens Clin. 1994;21:163-169.
  • 3
    • 0031732528 scopus 로고    scopus 로고
    • The role of computerized corneal topography in rigid gas permeable contact lens fitting
    • Bufidis T, Konstas AGP, Mamtziou E. The role of computerized corneal topography in rigid gas permeable contact lens fitting. CLAO J. 1998;24: 206-209.
    • (1998) CLAO J , vol.24 , pp. 206-209
    • Bufidis, T.1    Konstas, A.G.P.2    Mamtziou, E.3
  • 4
    • 0036782005 scopus 로고    scopus 로고
    • Initial selection of soft contact lenses based on corneal characteristics
    • Douthwaite WA. Initial selection of soft contact lenses based on corneal characteristics. CLAO J. 2002;28:202-205.
    • (2002) CLAO J , vol.28 , pp. 202-205
    • Douthwaite, W.A.1
  • 6
    • 0032938929 scopus 로고    scopus 로고
    • Videokeratography, keratometry, and refraction after penetrating keratoplasty
    • Borderie VM, Touzeau O, Laroche L. Videokeratography, keratometry, and refraction after penetrating keratoplasty. J Refract Surg. 1999; 15:32-37.
    • (1999) J Refract Surg , vol.15 , pp. 32-37
    • Borderie, V.M.1    Touzeau, O.2    Laroche, L.3
  • 7
    • 0036799897 scopus 로고    scopus 로고
    • Contact lens fitting to correct irregular astigmatism after corneal refractive surgery
    • Alio JL, Belda JI, Artola A, et al. Contact lens fitting to correct irregular astigmatism after corneal refractive surgery. J Cataract Refract Surg. 2002;28:1750-1757.
    • (2002) J Cataract Refract Surg , vol.28 , pp. 1750-1757
    • Alio, J.L.1    Belda, J.I.2    Artola, A.3
  • 8
    • 0032168661 scopus 로고    scopus 로고
    • Evaluation of mild, moderate, and advanced keratoconus using ultrasound pachometry and the eyesys videokeratoscope
    • Watters GA, Owens H. Evaluation of mild, moderate, and advanced keratoconus using ultrasound pachometry and the eyesys videokeratoscope. Optom Vis Sci. 1998;75:640-646.
    • (1998) Optom Vis Sci , vol.75 , pp. 640-646
    • Watters, G.A.1    Owens, H.2
  • 9
    • 0036124279 scopus 로고    scopus 로고
    • The performance of four different corneal topographers on normal human corneas and its impact on orthokeratology lens fitting
    • Cho P, Lam AKC, Mountford J, et al. The performance of four different corneal topographers on normal human corneas and its impact on orthokeratology lens fitting. Optom Vis Sci. 2002;79:175-183.
    • (2002) Optom Vis Sci , vol.79 , pp. 175-183
    • Cho, P.1    Lam, A.K.C.2    Mountford, J.3
  • 10
    • 0033485849 scopus 로고    scopus 로고
    • Accuracy of Orbscan total optical power maps in detecting refractive change after myopic laser in situ keratomileusis
    • Srivannaboon S, Reinstein DZ, Sutton HFS, et al. Accuracy of Orbscan total optical power maps in detecting refractive change after myopic laser in situ keratomileusis. J Cataract Refract Surg. 1999;25:1596-1599.
    • (1999) J Cataract Refract Surg , vol.25 , pp. 1596-1599
    • Srivannaboon, S.1    Reinstein, D.Z.2    Sutton, H.F.S.3
  • 11
    • 34547898746 scopus 로고    scopus 로고
    • Repeatability of corneal parameters with Pentacam after laser in situ keratomileusis
    • Jain R, Dilraj G, Grewal SPS. Repeatability of corneal parameters with Pentacam after laser in situ keratomileusis. Indian J Ophthalmol. 2007;5 5: 341-347.
    • (2007) Indian J Ophthalmol , vol.5 , Issue.5 , pp. 341-347
    • Jain, R.1    Dilraj, G.2    Grewal, S.P.S.3
  • 13
    • 0038118687 scopus 로고    scopus 로고
    • Changes in corneal structure with continuous wear of high-Dk soft contact lenses: A pilot study
    • González-Méijome JM, Gonzalez-Perez J, Cervino A, et al. Changes in corneal structure with continuous wear of high-Dk soft contact lenses: A pilot study. Optom Vis Sci. 2003;80:440-46.
    • (2003) Optom Vis Sci , vol.80 , pp. 440-446
    • González-Méijome, J.M.1    Gonzalez-Perez, J.2    Cervino, A.3
  • 14
    • 3242811374 scopus 로고    scopus 로고
    • Accuracy and precision of EyeSys and Orbscan systems on calibrated spherical test surfaces
    • Pérez JG, Cerviño A, Giraldez MJ, et al. Accuracy and precision of EyeSys and Orbscan systems on calibrated spherical test surfaces. Eye Contact Lens. 2004;30:74-78.
    • (2004) Eye Contact Lens , vol.30 , pp. 74-78
    • Pérez, J.G.1    Cerviño, A.2    Giraldez, M.J.3
  • 15
    • 0027416835 scopus 로고
    • A comparison of the TMS-1 and the corneal analysis system for the evaluation of abnormal corneas
    • Antalis JJ, Lembach RG, Carney LG A comparison of the TMS-1 and the corneal analysis system for the evaluation of abnormal corneas. CLAO J. 1993;19:58-63.
    • (1993) CLAO J , vol.19 , pp. 58-63
    • Antalis, J.J.1    Lembach, R.G.2    Carney, L.G.3
  • 16
    • 0025816796 scopus 로고
    • A comparison of the EH-270 corneal topographer with conventional keratometry
    • Davis LJ, Dresner MS. A comparison of the EH-270 corneal topographer with conventional keratometry. CLAO J. 1991;17:191-196.
    • (1991) CLAO J , vol.17 , pp. 191-196
    • Davis, L.J.1    Dresner, M.S.2
  • 17
    • 34249722133 scopus 로고    scopus 로고
    • External factors affecting data acquisition during corneal topography examination
    • González-Méijome JM, Queirós A, Jorge J, et al. External factors affecting data acquisition during corneal topography examination. Eye Contact Lens. 2007;33:91-97.
    • (2007) Eye Contact Lens , vol.33 , pp. 91-97
    • González-Méijome, J.M.1    Queirós, A.2    Jorge, J.3
  • 18
    • 0030238265 scopus 로고    scopus 로고
    • Asymmetric corneal toricity and pseudokeratoconus in videokeratography
    • Mandell RB, Chiang CS, Yee L. Asymmetric corneal toricity and pseudokeratoconus in videokeratography. J Am Optom Assoc. 1996;67: 540-547.
    • (1996) J Am Optom Assoc , vol.67 , pp. 540-547
    • Mandell, R.B.1    Chiang, C.S.2    Yee, L.3
  • 19
    • 29544440497 scopus 로고    scopus 로고
    • Comparison of three methods of measuring corneal thickness and anterior chamber depth
    • Buehl W, Stojanac D, Sacu S, et al. Comparison of three methods of measuring corneal thickness and anterior chamber depth. Am J Ophthalmol. 2006;141:7-12.
    • (2006) Am J Ophthalmol , vol.141 , pp. 7-12
    • Buehl, W.1    Stojanac, D.2    Sacu, S.3
  • 20
    • 25644448981 scopus 로고    scopus 로고
    • Validity and repeatability of anterior chamber depth measurements with Pentacam and Orbscan
    • Lackner B, Schmidinger G, Skorpik C. Validity and repeatability of anterior chamber depth measurements with Pentacam and Orbscan. Optom Vis Sci. 2005;82:858-861.
    • (2005) Optom Vis Sci , vol.82 , pp. 858-861
    • Lackner, B.1    Schmidinger, G.2    Skorpik, C.3
  • 21
    • 0036667901 scopus 로고    scopus 로고
    • Radius and asphericity of the posterior corneal surface determined by corrected Scheimpflug photography
    • Dubbelman M, Weeber HA, Van Der Heijde RGL, et al. Radius and asphericity of the posterior corneal surface determined by corrected Scheimpflug photography. Acta Ophthalmol Scand. 2002;80:379-383.
    • (2002) Acta Ophthalmol Scand , vol.80 , pp. 379-383
    • Dubbelman, M.1    Weeber, H.A.2    Van Der Heijde, R.G.L.3
  • 22
    • 67649877661 scopus 로고    scopus 로고
    • Topography and Scheimpflug Imaging: Pearls for refractive surgery screening and keratoconus detection
    • Belin MW. Topography and Scheimpflug Imaging: pearls for refractive surgery screening and keratoconus detection. Cataract Refract Surg Today. 2006;1:48-50.
    • (2006) Cataract Refract Surg Today , vol.1 , pp. 48-50
    • Belin, M.W.1
  • 23
    • 0002467354 scopus 로고
    • Aspherical and continuous curve contact lenses. Part one
    • Bennett AG. Aspherical and continuous curve contact lenses. Part one. Optom Today. 1988;28:11-14.
    • (1988) Optom Today , vol.28 , pp. 11-14
    • Bennett, A.G.1
  • 24
    • 0037637611 scopus 로고    scopus 로고
    • The asphericity, curvature and tilt of the human cornea measured using a videokeratoscope
    • Douthwaite WA. The asphericity, curvature and tilt of the human cornea measured using a videokeratoscope. Ophthalmic Physiol Opt. 2003;23: 141-150.
    • (2003) Ophthalmic Physiol Opt , vol.23 , pp. 141-150
    • Douthwaite, W.A.1
  • 25
    • 29544435798 scopus 로고    scopus 로고
    • The shape of the anterior and posterior surface of the aging human cornea
    • Dubbelman M, Sicam VADP, Van Der Heijde GL. The shape of the anterior and posterior surface of the aging human cornea. Vision Res. 2006;46:993-1001.
    • (2006) Vision Res , vol.46 , pp. 993-1001
    • Dubbelman, M.1    Sicam, V.A.D.P.2    Van Der Heijde, G.L.3
  • 26
    • 0002915308 scopus 로고    scopus 로고
    • Orthokeratology in low myopia. Part 2: Corneal topographic changes and safety over 100 days
    • Lui WO, Edwards MH. Orthokeratology in low myopia. Part 2: Corneal topographic changes and safety over 100 days. Cont Lens Anterior Eye. 2000;23:90-99.
    • (2000) Cont Lens Anterior Eye , vol.23 , pp. 90-99
    • Lui, W.O.1    Edwards, M.H.2
  • 27
    • 0842310253 scopus 로고    scopus 로고
    • Contribution of the ocular surface to visual optics
    • Courville CB, Smolek MK, Klyce SD. Contribution of the ocular surface to visual optics. Exp Eye Res. 2004;78:417-25.
    • (2004) Exp Eye Res , vol.78 , pp. 417-425
    • Courville, C.B.1    Smolek, M.K.2    Klyce, S.D.3
  • 28
    • 36549024082 scopus 로고    scopus 로고
    • The influence of chronic diabetes mellitus on the thickness and the shape of the anterior and posterior surface of the cornea
    • Wiemer NGM, Dubbelman M, Kostense PJ, et al. The influence of chronic diabetes mellitus on the thickness and the shape of the anterior and posterior surface of the cornea. Cornea. 2007;26:1165-1170.
    • (2007) Cornea , vol.26 , pp. 1165-1170
    • Wiemer, N.G.M.1    Dubbelman, M.2    Kostense, P.J.3
  • 29
    • 67650596987 scopus 로고    scopus 로고
    • American National Standards Institute. American National Standrad Ophthalmics-Corneal Topography Systems-Standard Terminology. Requirements. ANSI Z80.23-1999. Optical Laboratories Association; 1999.
    • American National Standards Institute. American National Standrad Ophthalmics-Corneal Topography Systems-Standard Terminology. Requirements. ANSI Z80.23-1999. Optical Laboratories Association; 1999.
  • 30
    • 0026065647 scopus 로고
    • Quantitative descriptors of corneal topography. A clinical study
    • Wilson SE, Klyce SD. Quantitative descriptors of corneal topography. A clinical study. Arch Ophthalmol. 1991;109:349-353.
    • (1991) Arch Ophthalmol , vol.109 , pp. 349-353
    • Wilson, S.E.1    Klyce, S.D.2
  • 31
    • 0004077090 scopus 로고    scopus 로고
    • Corneal Topography: A review, new ANSI standards and problems to solve
    • Klein SA. Corneal Topography: A review, new ANSI standards and problems to solve. J Opt Soc Am Vis Sci. 2000;35:286-296.
    • (2000) J Opt Soc Am Vis Sci , vol.35 , pp. 286-296
    • Klein, S.A.1
  • 32
    • 27644542761 scopus 로고    scopus 로고
    • Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound
    • Lackner B, Schmidinger G, Pieh S, et al. Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound. Optom Vis Sci. 2005;82:892-899.
    • (2005) Optom Vis Sci , vol.82 , pp. 892-899
    • Lackner, B.1    Schmidinger, G.2    Pieh, S.3
  • 33
    • 28044436542 scopus 로고    scopus 로고
    • Agreement and repeatability of central thickness measurement in normal corneas using ultrasound pachymetry and the OCULUS Pentacam
    • O'Donnell C, Maldonado-Codina C. Agreement and repeatability of central thickness measurement in normal corneas using ultrasound pachymetry and the OCULUS Pentacam. Cornea. 2005;24:920-924.
    • (2005) Cornea , vol.24 , pp. 920-924
    • O'Donnell, C.1    Maldonado-Codina, C.2
  • 34
    • 27144479245 scopus 로고    scopus 로고
    • Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry
    • Barkana Y, Gerber Y, Elbaz U, et al. Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry. J Cataract Refract Surg. 2005;31:1729-1735.
    • (2005) J Cataract Refract Surg , vol.31 , pp. 1729-1735
    • Barkana, Y.1    Gerber, Y.2    Elbaz, U.3
  • 35
    • 33847125749 scopus 로고    scopus 로고
    • Pentacam pachometry: Comparison with non-contact specular microscopy on the central cornea and inter-session repeatability on the peripheral cornea
    • Lam AK, Chen D. Pentacam pachometry: comparison with non-contact specular microscopy on the central cornea and inter-session repeatability on the peripheral cornea. Clin Exp Optom. 2006;90:108-114.
    • (2006) Clin Exp Optom , vol.90 , pp. 108-114
    • Lam, A.K.1    Chen, D.2
  • 36
    • 4143152915 scopus 로고    scopus 로고
    • Changes in myopia with low-Dk hydrogel and high-Dk silicone hydrogel extended wear
    • Jalbert I, Stretton S, Naduvilath T, et al. Changes in myopia with low-Dk hydrogel and high-Dk silicone hydrogel extended wear. Optom Vis Sci. 2004;81:591-596.
    • (2004) Optom Vis Sci , vol.81 , pp. 591-596
    • Jalbert, I.1    Stretton, S.2    Naduvilath, T.3
  • 37
    • 0033999943 scopus 로고    scopus 로고
    • The effects of long-term contact lens wear on corneal thickness, curvature, and surface regularity
    • Liu Z, Pflugfelder SC. The effects of long-term contact lens wear on corneal thickness, curvature, and surface regularity. Ophthalmology. 2000; 107:105-111.
    • (2000) Ophthalmology , vol.107 , pp. 105-111
    • Liu, Z.1    Pflugfelder, S.C.2
  • 38
    • 0000898845 scopus 로고
    • An analysis of variance test for normality (complete samples)
    • Shapiro SS, Wilk MB. An analysis of variance test for normality (complete samples). Biometrika. 1965;52:591-611.
    • (1965) Biometrika , vol.52 , pp. 591-611
    • Shapiro, S.S.1    Wilk, M.B.2
  • 39
    • 0032884749 scopus 로고    scopus 로고
    • Measuring agreement in method comparison studies
    • Bland JM, Airman DG Measuring agreement in method comparison studies. Stat Methods Med Res. 1999;8:135-160.
    • (1999) Stat Methods Med Res , vol.8 , pp. 135-160
    • Bland, J.M.1    Airman, D.G.2
  • 40
    • 0003497479 scopus 로고    scopus 로고
    • Bland M, ed, Oxford, UK: Oxford University Press;
    • Bland M, ed. An introduction to medical statistics. Oxford, UK: Oxford University Press; 2000.
    • (2000) An introduction to medical statistics
  • 42
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • Bland JM, Airman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307-310.
    • (1986) Lancet , vol.1 , pp. 307-310
    • Bland, J.M.1    Airman, D.G.2
  • 43
    • 84944811700 scopus 로고
    • The use of ranks to avoid the assumption of normality implicit in the analysis of variance
    • Friedman M. The use of ranks to avoid the assumption of normality implicit in the analysis of variance. J Am Stat Assoc. 1937;32:675-701.
    • (1937) J Am Stat Assoc , vol.32 , pp. 675-701
    • Friedman, M.1
  • 44
    • 0002915307 scopus 로고    scopus 로고
    • Orthokeratology in low myopia. Part 1: Efficacy and predictability
    • Lui WO, Edwards MH. Orthokeratology in low myopia. Part 1: Efficacy and predictability. Cont Lens Anterior Eye. 2000;23:77-89.
    • (2000) Cont Lens Anterior Eye , vol.23 , pp. 77-89
    • Lui, W.O.1    Edwards, M.H.2
  • 45
    • 23044443649 scopus 로고    scopus 로고
    • Contact lenses in myopia reduction: From orthofocus to accelerated orthokeratology
    • Lui WO, Edwards MH, Cho P. Contact lenses in myopia reduction: From orthofocus to accelerated orthokeratology. Cont Lens Anterior Eye. 2000; 23:68-76.
    • (2000) Cont Lens Anterior Eye , vol.23 , pp. 68-76
    • Lui, W.O.1    Edwards, M.H.2    Cho, P.3
  • 46
    • 0028885963 scopus 로고
    • Comparison of methods for detecting keratoconus using videokeratography
    • Maeda N, Klyce SD, Smolek MK. Comparison of methods for detecting keratoconus using videokeratography. Arch Ophthalmol. 1995; 113: 870-874.
    • (1995) Arch Ophthalmol , vol.113 , pp. 870-874
    • Maeda, N.1    Klyce, S.D.2    Smolek, M.K.3
  • 47
    • 0028073952 scopus 로고
    • Keratoconus and contact lens-induced corneal warpage analysis using the keratomorphic diagram
    • Smolek MK, Klyce SD, Maeda N Keratoconus and contact lens-induced corneal warpage analysis using the keratomorphic diagram. Invest Ophthalmol Vis Sci. 1994;35:4192-204.
    • (1994) Invest Ophthalmol Vis Sci , vol.35 , pp. 4192-4204
    • Smolek, M.K.1    Klyce, S.D.2    Maeda, N.3
  • 48
    • 0000952703 scopus 로고    scopus 로고
    • Effects on visual function of approximations of the corneal-ablation profile during refractive surgery
    • Jimenez JR, Anera RG, Del Barco LJ. Effects on visual function of approximations of the corneal-ablation profile during refractive surgery. Appl Optics. 2001;40:2200-2205.
    • (2001) Appl Optics , vol.40 , pp. 2200-2205
    • Jimenez, J.R.1    Anera, R.G.2    Del Barco, L.J.3
  • 49
    • 0034691924 scopus 로고    scopus 로고
    • Retinal image quality in myopic subjects after refractive surgery
    • Jiménez JR, Anera RG, Jiménez Del Barco L, et al. Retinal image quality in myopic subjects after refractive surgery. J Mod Opt. 2000;47:1587-1598.
    • (2000) J Mod Opt , vol.47 , pp. 1587-1598
    • Jiménez, J.R.1    Anera, R.G.2    Jiménez Del Barco, L.3
  • 51
    • 0035022540 scopus 로고    scopus 로고
    • The shape of the ageing human lens: Curvature, equivalent refractive index and the lens paradox
    • Dubbelman M, Van Der Heijde GL. The shape of the ageing human lens: curvature, equivalent refractive index and the lens paradox. Vision Res. 2001;41:1867-1877.
    • (2001) Vision Res , vol.41 , pp. 1867-1877
    • Dubbelman, M.1    Van Der Heijde, G.L.2
  • 52
    • 9644276888 scopus 로고    scopus 로고
    • Dubbelman M, Van Der Heijde GL, Weeber HA. Change in shape of the aging human crystalline lens with accommodation. Vision Res. 2005;45: 117-132.
    • Dubbelman M, Van Der Heijde GL, Weeber HA. Change in shape of the aging human crystalline lens with accommodation. Vision Res. 2005;45: 117-132.
  • 53
    • 0024412191 scopus 로고
    • Improved biometry of the anterior eye segment
    • Kampfer T, Wegener A, Dragomirescu V et al. Improved biometry of the anterior eye segment. Ophthalmic Res. 1989;21:239-248.
    • (1989) Ophthalmic Res , vol.21 , pp. 239-248
    • Kampfer, T.1    Wegener, A.2    Dragomirescu, V.3
  • 54
    • 0032191303 scopus 로고    scopus 로고
    • Regular and irregular refractive powers of the front and back surfaces of the cornea
    • Oshika T, Tomidokoro A, Tsuji H. Regular and irregular refractive powers of the front and back surfaces of the cornea. Exp Eye Res. 1998;67: 443-47.
    • (1998) Exp Eye Res , vol.67 , pp. 443-447
    • Oshika, T.1    Tomidokoro, A.2    Tsuji, H.3
  • 55
    • 0347358049 scopus 로고    scopus 로고
    • Test-retest reliability of clinical Shack-Hartmann measurements
    • Cheng X, Himebaugh NL, Kollbaum PS, et al. Test-retest reliability of clinical Shack-Hartmann measurements. Invest Ophthalmol Vis Sci. 2004; 45:351-360.
    • (2004) Invest Ophthalmol Vis Sci , vol.45 , pp. 351-360
    • Cheng, X.1    Himebaugh, N.L.2    Kollbaum, P.S.3
  • 56
    • 0031608408 scopus 로고    scopus 로고
    • Interferometric measurement of tear film thickness by use of spectral oscillations
    • Fogt N, King-Smith PE, Tuell G. Interferometric measurement of tear film thickness by use of spectral oscillations. J Opt Soc Am Assoc. 1998;15: 268-275.
    • (1998) J Opt Soc Am Assoc , vol.15 , pp. 268-275
    • Fogt, N.1    King-Smith, P.E.2    Tuell, G.3
  • 57
    • 0037809437 scopus 로고    scopus 로고
    • Precorneal and pre- and postlens tear film thickness measured indirectly with optical coherence tomography
    • Wang J, Fonn D, Simpson TL, et al. Precorneal and pre- and postlens tear film thickness measured indirectly with optical coherence tomography. Invest Ophthalmol Vis Sci. 2003;44:2524-2528.
    • (2003) Invest Ophthalmol Vis Sci , vol.44 , pp. 2524-2528
    • Wang, J.1    Fonn, D.2    Simpson, T.L.3
  • 58
    • 0028051752 scopus 로고
    • Measurement of the precorneal tear film thickness with a non-contact optical interferometry film thickness measurement system
    • Danjo Y, Nakamura M, Hamano T. Measurement of the precorneal tear film thickness with a non-contact optical interferometry film thickness measurement system. Jap J Ophthalmol. 1994;38:260-266.
    • (1994) Jap J Ophthalmol , vol.38 , pp. 260-266
    • Danjo, Y.1    Nakamura, M.2    Hamano, T.3
  • 59
    • 0026650519 scopus 로고
    • Study of human precorneal tear film thickness and structure using laser interferometry
    • Prydal JI, Artal P, Woon H, et al. Study of human precorneal tear film thickness and structure using laser interferometry. Invest Ophthalmol Vis Sci. 1992;33:2006-2011.
    • (1992) Invest Ophthalmol Vis Sci , vol.33 , pp. 2006-2011
    • Prydal, J.I.1    Artal, P.2    Woon, H.3
  • 60
    • 0007528493 scopus 로고
    • The thickness of the precorneal tear film
    • Ehlers N. The thickness of the precorneal tear film. Acta Ophthalmol (Copenh). 1965;8:92.
    • (1965) Acta Ophthalmol (Copenh) , vol.8 , pp. 92
    • Ehlers, N.1
  • 61
    • 0016119232 scopus 로고
    • The precorneal tear film: Breakup time and location in normal subjects
    • Rengstorff RH. The precorneal tear film: breakup time and location in normal subjects. Am J Optom Physiol Opt. 1974;51:765-769.
    • (1974) Am J Optom Physiol Opt , vol.51 , pp. 765-769
    • Rengstorff, R.H.1
  • 62
    • 0014628412 scopus 로고
    • Desiccation of the precorneal film. I. Corneal wetting-time
    • Norn MS. Desiccation of the precorneal film. I. Corneal wetting-time. Acta Ophthalmol. 1969;47:865-880.
    • (1969) Acta Ophthalmol , vol.47 , pp. 865-880
    • Norn, M.S.1
  • 63
    • 0036257336 scopus 로고    scopus 로고
    • Central corneal thickness, radius of the corneal curvature and intraocular pressure in normal subjects using non-contact techniques: Reykjavik eye study
    • Eysteinsson T, Jonasson F, Sasaki H, et al. Central corneal thickness, radius of the corneal curvature and intraocular pressure in normal subjects using non-contact techniques: Reykjavik eye study. Acta Ophthalmol Scand. 2002;80:11-15.
    • (2002) Acta Ophthalmol Scand , vol.80 , pp. 11-15
    • Eysteinsson, T.1    Jonasson, F.2    Sasaki, H.3
  • 64
    • 0035210450 scopus 로고    scopus 로고
    • Anterior chamber depth estimation by Scheimpflug photography
    • Holmen JB, Ekesten B, Lundgren B. Anterior chamber depth estimation by Scheimpflug photography. Acta Ophthalmol Scand. 2001;79:576-579.
    • (2001) Acta Ophthalmol Scand , vol.79 , pp. 576-579
    • Holmen, J.B.1    Ekesten, B.2    Lundgren, B.3
  • 65
    • 0032999776 scopus 로고    scopus 로고
    • Necessary sample size for method comparison studies based on regression analysis
    • Linnet K. Necessary sample size for method comparison studies based on regression analysis. Clin Chem. 1999;45:882-894.
    • (1999) Clin Chem , vol.45 , pp. 882-894
    • Linnet, K.1
  • 66
    • 0036140949 scopus 로고    scopus 로고
    • Statistics for laboratory. Method comparison studies
    • Magari RT. Statistics for laboratory. Method comparison studies. BioPharm. 2002;15:28-32.
    • (2002) BioPharm , vol.15 , pp. 28-32
    • Magari, R.T.1


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.