Increased rate of subacute lead complications with small-caliber implantable cardioverter-defibrillator leads

Author keywords

Cardiac perforation; Defibrillator; Lead dislodgement; Lead revision; Right ventricular ICD lead

Indexed keywords

ADULT; AGED; ARTICLE; CLINICAL ARTICLE; DEFIBRILLATOR; HEART INJURY; HEART PACING; HEART PERFORATION; HEART RIGHT VENTRICLE; HUMAN; PERFORATION; PRIORITY JOURNAL; RETROSPECTIVE STUDY; RISK ASSESSMENT;

EID: 64649099576     PISSN: 15475271     EISSN: None     Source Type: Journal    
DOI: 10.1016/j.hrthm.2009.02.020     Document Type: Article
Times cited : (44)

References (27)
  • 1
    • 0037453139 scopus 로고    scopus 로고
    • The clinical use of implantable cardioverter defibrillators where are we now? Where should we go?
    • Buxton A.E. The clinical use of implantable cardioverter defibrillators where are we now? Where should we go?. Ann Intern Med 138 (2003) 512-515
    • (2003) Ann Intern Med , vol.138 , pp. 512-515
    • Buxton, A.E.1
  • 2
    • 19944429413 scopus 로고    scopus 로고
    • Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure
    • Bardy G.H., Lee K.L., Mark D.B., et al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med 352 (2005) 225-237
    • (2005) N Engl J Med , vol.352 , pp. 225-237
    • Bardy, G.H.1    Lee, K.L.2    Mark, D.B.3
  • 3
    • 0037149716 scopus 로고    scopus 로고
    • Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction
    • Moss A.J., Zareba W., Hall W.J., et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med 346 (2002) 877-883
    • (2002) N Engl J Med , vol.346 , pp. 877-883
    • Moss, A.J.1    Zareba, W.2    Hall, W.J.3
  • 4
    • 0006918490 scopus 로고    scopus 로고
    • Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia
    • Moss A.J., Hall W.J., Cannom D.S., et al. Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. N Engl J Med 335 (1996) 1933-1940
    • (1996) N Engl J Med , vol.335 , pp. 1933-1940
    • Moss, A.J.1    Hall, W.J.2    Cannom, D.S.3
  • 5
    • 34247186286 scopus 로고    scopus 로고
    • Increased incidence of subacute lead perforation noted with one implantable cardioverter-defibrillator
    • Danik S., Mansour M., Singh J., et al. Increased incidence of subacute lead perforation noted with one implantable cardioverter-defibrillator. Heart Rhythm 4 (2007) 439-442
    • (2007) Heart Rhythm , vol.4 , pp. 439-442
    • Danik, S.1    Mansour, M.2    Singh, J.3
  • 7
    • 34248579318 scopus 로고    scopus 로고
    • Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of >10 years
    • Kleemann T., Becker T., Doenges K., et al. Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of >10 years. Circulation 115 (2007) 2461-2463
    • (2007) Circulation , vol.115 , pp. 2461-2463
    • Kleemann, T.1    Becker, T.2    Doenges, K.3
  • 8
    • 0030008104 scopus 로고    scopus 로고
    • Perforation of the right ventricle by transvenous defibrillator leads: prevention and treatment
    • Molina J.E. Perforation of the right ventricle by transvenous defibrillator leads: prevention and treatment. Pacing Clin Electrophysiol 19 (1996) 288-292
    • (1996) Pacing Clin Electrophysiol , vol.19 , pp. 288-292
    • Molina, J.E.1
  • 9
    • 0028138610 scopus 로고
    • Early complications after dual chamber versus single chamber pacemaker implantation
    • Chauhan A., Grace A.A., Newell S.A., et al. Early complications after dual chamber versus single chamber pacemaker implantation. PACE 17 (1994) 2012-2015
    • (1994) PACE , vol.17 , pp. 2012-2015
    • Chauhan, A.1    Grace, A.A.2    Newell, S.A.3
  • 10
    • 0028819419 scopus 로고
    • Low incidence of lead related complications associated with nonthoracotomy implantable cardioverter defibrillator systems
    • Fahy G.J., Kleman J.M., Wilkoff B.L., et al. Low incidence of lead related complications associated with nonthoracotomy implantable cardioverter defibrillator systems. Pacing Clin Electrophysiol 18 (1995) 172-178
    • (1995) Pacing Clin Electrophysiol , vol.18 , pp. 172-178
    • Fahy, G.J.1    Kleman, J.M.2    Wilkoff, B.L.3
  • 11
    • 0035167747 scopus 로고    scopus 로고
    • The management of surgical complications of pacemaker and implantable cardioverter-defibrillators
    • Pavia S., and Wilkoff B. The management of surgical complications of pacemaker and implantable cardioverter-defibrillators. Curr Opin Cardiol 16 (2001) 66-71
    • (2001) Curr Opin Cardiol , vol.16 , pp. 66-71
    • Pavia, S.1    Wilkoff, B.2
  • 12
    • 33644693959 scopus 로고    scopus 로고
    • Complications of implantable cardioverter defibrillator therapy in 440 consecutive patients
    • Alter P., Waldhans S., Plachta E., et al. Complications of implantable cardioverter defibrillator therapy in 440 consecutive patients. Pacing Clin Electrophysiol 2 (2005) 926-932
    • (2005) Pacing Clin Electrophysiol , vol.2 , pp. 926-932
    • Alter, P.1    Waldhans, S.2    Plachta, E.3
  • 13
    • 0032953658 scopus 로고    scopus 로고
    • Complications of third-generation implantable cardioverter defibrillator therapy
    • Grimm W., Menz V., Hoffmann J., et al. Complications of third-generation implantable cardioverter defibrillator therapy. Pacing Clin Electrophysiol 22 (1999) 206-211
    • (1999) Pacing Clin Electrophysiol , vol.22 , pp. 206-211
    • Grimm, W.1    Menz, V.2    Hoffmann, J.3
  • 14
    • 0036023341 scopus 로고    scopus 로고
    • High incidence of device related and lead related complications in the dual chamber implantable cardioverter defibrillator compared with the single chamber version
    • Takahashi T., Bhandari A., Watanuki M., et al. High incidence of device related and lead related complications in the dual chamber implantable cardioverter defibrillator compared with the single chamber version. Circ J 66 (2002) 746-750
    • (2002) Circ J , vol.66 , pp. 746-750
    • Takahashi, T.1    Bhandari, A.2    Watanuki, M.3
  • 15
    • 0035110248 scopus 로고    scopus 로고
    • Recurrent pericardial chest pain: a case of late right ventricular perforation after implantation of a transvenous active-fixation ICD lead
    • Kautzner J., and Bytesnik J. Recurrent pericardial chest pain: a case of late right ventricular perforation after implantation of a transvenous active-fixation ICD lead. PACE 24 (2001) 116-118
    • (2001) PACE , vol.24 , pp. 116-118
    • Kautzner, J.1    Bytesnik, J.2
  • 16
    • 0033048909 scopus 로고    scopus 로고
    • Complications related to permanent pacemaker therapy
    • Kiviniemi M.S., Pirnes M.A., Eranen J.K., et al. Complications related to permanent pacemaker therapy. PACE 22 (1999) 711-720
    • (1999) PACE , vol.22 , pp. 711-720
    • Kiviniemi, M.S.1    Pirnes, M.A.2    Eranen, J.K.3
  • 17
    • 0032082605 scopus 로고    scopus 로고
    • Complications of dual chamber pacemaker implantation in the elderly
    • Link M.S., Estes III N.A.M., Griffin J.J., et al. Complications of dual chamber pacemaker implantation in the elderly. J Interv Card Electrophysiol 2 (1998) 175-179
    • (1998) J Interv Card Electrophysiol , vol.2 , pp. 175-179
    • Link, M.S.1    Estes III, N.A.M.2    Griffin, J.J.3
  • 18
    • 59749088353 scopus 로고    scopus 로고
    • Timing of delayed perforation with the St Jude Riata lead: a single-center experience and a review of the literature
    • Danik S., Mansour M., Heist K.E., et al. Timing of delayed perforation with the St Jude Riata lead: a single-center experience and a review of the literature. Heart Rhythm 5 (2008) 1667-1672
    • (2008) Heart Rhythm , vol.5 , pp. 1667-1672
    • Danik, S.1    Mansour, M.2    Heist, K.E.3
  • 19
    • 59149089389 scopus 로고    scopus 로고
    • Bad lead? Good lead? What should we believe?
    • Healey J.S., and Krahn A. Bad lead? Good lead? What should we believe?. Heart Rhythm 6 (2009) 210-211
    • (2009) Heart Rhythm , vol.6 , pp. 210-211
    • Healey, J.S.1    Krahn, A.2
  • 20
    • 58849112108 scopus 로고    scopus 로고
    • Performance of the St. Jude Medical Riata leads
    • Epstein A.E., Baker J., Beau S., et al. Performance of the St. Jude Medical Riata leads. Heart Rhythm 6 (2009) 204-209
    • (2009) Heart Rhythm , vol.6 , pp. 204-209
    • Epstein, A.E.1    Baker, J.2    Beau, S.3
  • 21
    • 44349112346 scopus 로고    scopus 로고
    • Does size really matter?. A comparison of the Riata lead family based on size and its relation to performance
    • Corbisiero R., and Armbruster R. Does size really matter?. A comparison of the Riata lead family based on size and its relation to performance. PACE 31 (2008) 722-726
    • (2008) PACE , vol.31 , pp. 722-726
    • Corbisiero, R.1    Armbruster, R.2
  • 22
    • 0029074982 scopus 로고
    • Early complications of permanent pacemaker implantation no difference between dual and single chamber systems
    • Aggarwal R.K., Connelly D.T., Ray S.G., et al. Early complications of permanent pacemaker implantation no difference between dual and single chamber systems. Br Heart J 73 (1995) 571-575
    • (1995) Br Heart J , vol.73 , pp. 571-575
    • Aggarwal, R.K.1    Connelly, D.T.2    Ray, S.G.3
  • 24
    • 18044386589 scopus 로고    scopus 로고
    • Late perforation of the heart, pericardium and diaphragm by an active-fixation ventricular lead
    • Akyol A., Aydin A., Erdliner I., et al. Late perforation of the heart, pericardium and diaphragm by an active-fixation ventricular lead. PACE 28 (2005) 350-351
    • (2005) PACE , vol.28 , pp. 350-351
    • Akyol, A.1    Aydin, A.2    Erdliner, I.3
  • 25
    • 15344339217 scopus 로고    scopus 로고
    • Delayed lead perforation: a disturbing trend
    • Khan M.N., Joseph G., Khaykin Y., et al. Delayed lead perforation: a disturbing trend. PACE 28 (2005) 251-253
    • (2005) PACE , vol.28 , pp. 251-253
    • Khan, M.N.1    Joseph, G.2    Khaykin, Y.3
  • 26
    • 25444441460 scopus 로고    scopus 로고
    • Incidence and predictors of cardiac perforation after permanent pacemaker placement
    • Mahapatra S., Bybee K.A., Bunch T.J., et al. Incidence and predictors of cardiac perforation after permanent pacemaker placement. Heart Rhythm 2 (2005) 907-911
    • (2005) Heart Rhythm , vol.2 , pp. 907-911
    • Mahapatra, S.1    Bybee, K.A.2    Bunch, T.J.3
  • 27
    • 33644693962 scopus 로고    scopus 로고
    • High pacing impedances: are you overtorquing your leads?
    • Laske T.G., Vieau S.A., Skadsberg N.D., et al. High pacing impedances: are you overtorquing your leads?. PACE 28 (2005) 883-891
    • (2005) PACE , vol.28 , pp. 883-891
    • Laske, T.G.1    Vieau, S.A.2    Skadsberg, N.D.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.