Monte Carlo simulations and radiation dosimetry measurements of peripherally applied HDR 192Ir breast brachytherapy D-shaped applicators

Author keywords

Applicator; Brachytherapy; Dosimetry

Indexed keywords

APPLICATORS; DOSIMETRY; IONIZATION CHAMBERS; ONCOLOGY; POLYSTYRENES; RADIOTHERAPY;

EID: 61349188243     PISSN: 00942405     EISSN: None     Source Type: Journal    
DOI: 10.1118/1.3075818     Document Type: Article
Times cited : (24)

References (21)
  • 1
    • 61349130185 scopus 로고    scopus 로고
    • Stereotactic peripheral brachytherapy and image guidance for the breast
    • M. J. Rivard, R. J. Bricault, C. S. Melhus, and P. Sioshansi, " Stereotactic peripheral brachytherapy and image guidance for the breast.," Med. Phys. 33, 2137 (2006).
    • (2006) Med. Phys. , vol.33 , pp. 2137
    • Rivard, M.J.1    Bricault, R.J.2    Melhus, C.S.3    Sioshansi, P.4
  • 3
    • 61349106041 scopus 로고    scopus 로고
    • Stereotactic breast brachytherapy applied peripherally using a compression applicator and HDR I 192 r
    • M. J. Rivard, R. J. Bricault, C. S. Melhus, J. R. Hiatt, P. Sioshansi, and D. E. Wazer, " Stereotactic breast brachytherapy applied peripherally using a compression applicator and HDR I 192 r.," Brachytherapy 6, 80 (2007).
    • (2007) Brachytherapy , vol.6 , pp. 80
    • Rivard, M.J.1    Bricault, R.J.2    Melhus, C.S.3    Hiatt, J.R.4    Sioshansi, P.5    Wazer, D.E.6
  • 4
    • 61349135744 scopus 로고    scopus 로고
    • Peripheral brachytherapy, dosimetry, and image guidance using the AccuBoost system
    • J, ","
    • M. J. Rivard, R. J. Bricault J, J. R. Hiatt, C. S. Melhus, P. Sioshansi, and D. E. Wazer, " Peripheral brachytherapy, dosimetry, and image guidance using the AccuBoost system.," Med. Phys. 34, 2480 (2007).
    • (2007) Med. Phys. , vol.34 , pp. 2480
    • Rivard, M.J.1    Bricault, R.J.2    Hiatt, J.R.3    Melhus, C.S.4    Sioshansi, P.5    Wazer, D.E.6
  • 6
    • 61349171368 scopus 로고    scopus 로고
    • X-5 Monte Carlo Team, MCNP-A General Monte Carlo N-Particle Transport Code, Version 5, Los Alamos National Laboratory, Los Alamos, NM.
    • X-5 Monte Carlo Team, MCNP-A General Monte Carlo N-Particle Transport Code, Version 5, Los Alamos National Laboratory, Los Alamos, NM, 2003.
    • (2003)
  • 9
    • 85027219388 scopus 로고    scopus 로고
    • NUDAT 2.4, National Nuclear Data Center, Brookhaven National Laboratory, (last accessed 8 February 2009).
    • NUDAT 2.4, National Nuclear Data Center, Brookhaven National Laboratory, http://www.nndc.bnl.gov/nudat2/index.jsp (last accessed 8 February 2009).
  • 10
    • 0031739032 scopus 로고    scopus 로고
    • Monte Carlo-aided dosimetry of a new high dose-rate brachytherapy source
    • 0094-2405 10.1118/1.598418.
    • G. M. Daskalov, E. Löffler, and J. F. Williamson, " Monte Carlo-aided dosimetry of a new high dose-rate brachytherapy source.," Med. Phys. 0094-2405 10.1118/1.598418 25, 2200-2208 (1998).
    • (1998) Med. Phys. , vol.25 , pp. 2200-2208
    • Daskalov, G.M.1    Löffler, E.2    Williamson, J.F.3
  • 13
    • 0033626411 scopus 로고    scopus 로고
    • Validation of a precision radiochromic film dosimetry system for quantitative two-dimensional imaging of acute exposure dose distributions
    • 0094-2405 10.1118/1.1290488.
    • J. F. Dempsey, D. A. Low, S. Mutic, J. Markman, A. S. Kirov, G. H. Nussbaum, and J. F. Williamson, " Validation of a precision radiochromic film dosimetry system for quantitative two-dimensional imaging of acute exposure dose distributions.," Med. Phys. 0094-2405 10.1118/1.1290488 27, 2462-2475 (2000).
    • (2000) Med. Phys. , vol.27 , pp. 2462-2475
    • Dempsey, J.F.1    Low, D.A.2    Mutic, S.3    Markman, J.4    Kirov, A.S.5    Nussbaum, G.H.6    Williamson, J.F.7
  • 14
    • 28044452190 scopus 로고    scopus 로고
    • Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energies
    • 0094-2405 10.1118/1.2065467.
    • S. Chiu-Tsao, Y. Ho, R. Shank, L. Wang, and L. B. Harrison, " Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energies.," Med. Phys. 0094-2405 10.1118/1.2065467 32, 3350-3354 (2005).
    • (2005) Med. Phys. , vol.32 , pp. 3350-3354
    • Chiu-Tsao, S.1    Ho, Y.2    Shank, R.3    Wang, L.4    Harrison, L.B.5
  • 15
    • 26244462962 scopus 로고    scopus 로고
    • Post-irradiation colouration of Gafchromic EBT radiochromic film
    • 0031-9155 10.1088/0031-9155/50/20/N04.
    • T. Cheung, M. J. Butson, and P. K. Yu, " Post-irradiation colouration of Gafchromic EBT radiochromic film.," Phys. Med. Biol. 0031-9155 10.1088/0031-9155/50/20/N04 50, N281-N285 (2005).
    • (2005) Phys. Med. Biol. , vol.50
    • Cheung, T.1    Butson, M.J.2    Yu, P.K.3
  • 16
    • 33845337095 scopus 로고    scopus 로고
    • Dosimetric characterization of a novel intracavitary mold applicator for I 192 r high dose rate endorectal brachytherapy treatment
    • 0094-2405 10.1118/1.2364054.
    • E. Poon, B. Reniers, and S. Devic, " Dosimetric characterization of a novel intracavitary mold applicator for I 192 r high dose rate endorectal brachytherapy treatment.," Med. Phys. 0094-2405 10.1118/1.2364054 33, 4515-4527 (2006).
    • (2006) Med. Phys. , vol.33 , pp. 4515-4527
    • Poon, E.1    Reniers, B.2    Devic, S.3
  • 17
    • 38849173037 scopus 로고    scopus 로고
    • Accurate dosimetry with Gafchromic™ EBT film of a 6 MV photon beam in water: What level is achievable?
    • 0094-2405 10.1118/1.2828196.
    • L. J. van Battum, D. Hoffmans, H. Piersma, and S. Heukelom, " Accurate dosimetry with Gafchromic™ EBT film of a 6 MV photon beam in water: What level is achievable? " Med. Phys. 0094-2405 10.1118/1.2828196 35, 704-716 (2008).
    • (2008) Med. Phys. , vol.35 , pp. 704-716
    • Van Battum, L.J.1    Hoffmans, D.2    Piersma, H.3    Heukelom, S.4
  • 18
    • 48349092548 scopus 로고    scopus 로고
    • The use of new GAFCHROMIC® EBT film for I 125 seed dosimetry
    • 0094-2405 10.1118/1.2955746.
    • S. Chiu-Tsao, D. Medich, and J. J. Munro III, " The use of new GAFCHROMIC® EBT film for I 125 seed dosimetry.," Med. Phys. 0094-2405 10.1118/1.2955746 35, 3787-3799 (2008).
    • (2008) Med. Phys. , vol.35 , pp. 3787-3799
    • Chiu-Tsao, S.1    Medich, D.2    Munro III, J.J.3
  • 19
    • 33846592289 scopus 로고    scopus 로고
    • Brachytherapy dosimetry parameters calculated for a C 131 s source
    • 0094-2405 10.1118/1.2432162.
    • M. J. Rivard, " Brachytherapy dosimetry parameters calculated for a C 131 s source.," Med. Phys. 0094-2405 10.1118/1.2432162 34, 754-762 (2007).
    • (2007) Med. Phys. , vol.34 , pp. 754-762
    • Rivard, M.J.1
  • 20
    • 0033743964 scopus 로고    scopus 로고
    • Dosimetric characteristics of a new I 125 brachytherapy source
    • 0094-2405 10.1118/1.1290486.
    • D. M. Gearheart, A. Drogin, K. Sowards, A. S. Meigooni, and G. S. Ibbott, " Dosimetric characteristics of a new I 125 brachytherapy source.," Med. Phys. 0094-2405 10.1118/1.1290486 27, 2278-2285 (2000).
    • (2000) Med. Phys. , vol.27 , pp. 2278-2285
    • Gearheart, D.M.1    Drogin, A.2    Sowards, K.3    Meigooni, A.S.4    Ibbott, G.S.5
  • 21
    • 1142304255 scopus 로고    scopus 로고
    • Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations
    • 0094-2405 10.1118/1.1646040.
    • M. J. Rivard, B. M. Coursey, L. A. DeWerd, W. F. Hanson, M. S. Huq, G. S. Ibbott, M. G. Mitch, R. Nath, and J. F. Williamson, " Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations.," Med. Phys. 0094-2405 10.1118/1.1646040 31, 633-674 (2004).
    • (2004) Med. Phys. , vol.31 , pp. 633-674
    • Rivard, M.J.1    Coursey, B.M.2    Dewerd, L.A.3    Hanson, W.F.4    Huq, M.S.5    Ibbott, G.S.6    Mitch, M.G.7    Nath, R.8    Williamson, J.F.9


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.