Volumn 33, Issue 12, 2006, Pages 4515-4526

Dosimetric characterization of a novel intracavitary mold applicator for 192Ir high dose rate endorectal brachytherapy treatment

Author keywords

192Ir; GAFCHROMIC EBT film; GEANT4; Intracavitary brachytherapy; Monte Carlo

Indexed keywords

BALLOONS; DOSIMETRY; FLEXIBLE STRUCTURES; IODINE; MOLDS; MONTE CARLO METHODS; ONCOLOGY; PHASE SPACE METHODS; RADIOTHERAPY; SHIELDING; SILICONES; UNCERTAINTY ANALYSIS;

EID: 33845337095     PISSN: 00942405     EISSN: None     Source Type: Journal    
DOI: 10.1118/1.2364054     Document Type: Article
Times cited : (42)

References (36)
  • 1
    • 1842377447 scopus 로고    scopus 로고
    • Improved survival with preoperative radiotherapy in resectable rectal cancer
    • Swedish Rectal Cancer Trial
    • Swedish Rectal Cancer Trial, " Improved survival with preoperative radiotherapy in resectable rectal cancer., " N. Engl. J. Med. 336, 980-987 (1997).
    • (1997) N. Engl. J. Med. , vol.336 , pp. 980-987
  • 2
    • 0035975394 scopus 로고    scopus 로고
    • Preoperative radiotherapy combined with total mesorectal excision for resectable rectal cancer
    • E. Kapiteijn, " Preoperative radiotherapy combined with total mesorectal excision for resectable rectal cancer., " N. Engl. J. Med. 345, 638-646 (2001).
    • (2001) N. Engl. J. Med. , vol.345 , pp. 638-646
    • Kapiteijn, E.1
  • 3
    • 6044226542 scopus 로고    scopus 로고
    • Preoperative versus Postoperative Chemoradiotherapy for Rectal Cancer
    • R. Sauer, " Preoperative versus Postoperative Chemoradiotherapy for Rectal Cancer., " N. Engl. J. Med. 351, 1731-1740 (2004).
    • (2004) N. Engl. J. Med. , vol.351 , pp. 1731-1740
    • Sauer, R.1
  • 5
    • 0035922657 scopus 로고    scopus 로고
    • Adjuvant radiotherapy for rectal cancer: A systematic overview of 8507 patients from 22 randomized trials
    • Colorectal Cancer Collaborative Group, "
    • Colorectal Cancer Collaborative Group, " Adjuvant radiotherapy for rectal cancer: a systematic overview of 8507 patients from 22 randomized trials., " Lancet 358, 1291-1304 (2001).
    • (2001) Lancet , vol.358 , pp. 1291-1304
  • 6
    • 2942750356 scopus 로고    scopus 로고
    • Improved sphincter preservation with high dose preoperative radiotherapy: The Lyon R96-02 randomized trial
    • J. P. Gerard, " Improved sphincter preservation with high dose preoperative radiotherapy: The Lyon R96-02 randomized trial., " J. Clin. Oncol. 22, 2404-2409 (2004).
    • (2004) J. Clin. Oncol. , vol.22 , pp. 2404-2409
    • Gerard, J.P.1
  • 7
    • 25144448513 scopus 로고    scopus 로고
    • High-dose-rate endorectal brachytherapy in the treatment of locally advanced rectal carcinoma: Technical aspects
    • T. Vuong, S. Devic, B. Moftah, M. Evans, and E. B. Podgorsak, " High-dose-rate endorectal brachytherapy in the treatment of locally advanced rectal carcinoma: Technical aspects., " Brachytherapy 4, 230-235 (2005).
    • (2005) Brachytherapy , vol.4 , pp. 230-235
    • Vuong, T.1    Devic, S.2    Moftah, B.3    Evans, M.4    Podgorsak, E.B.5
  • 9
    • 85026339264 scopus 로고    scopus 로고
    • Advantages of inflatable multichannel endorectal applicator in the neo-adjuvant treatment of patients with locally advanced rectal cancer with HDR brachytherapy
    • S. Devic, T. Vuong, and B. Moftah, " Advantages of inflatable multichannel endorectal applicator in the neo-adjuvant treatment of patients with locally advanced rectal cancer with HDR brachytherapy., " J. Appl. Clin. Med. Phys. 6, 44-49 (2005).
    • (2005) J. Appl. Clin. Med. Phys. , vol.6 , pp. 44-49
    • Devic, S.1    Vuong, T.2    Moftah, B.3
  • 10
    • 0028919251 scopus 로고
    • Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM RadiationTherapy Committee Task Group No. 43
    • R. Nath, L. L. Anderson, G. Luxton, K. A. Weaver, J. F. Williamson, and A. S. Meigooni, " Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM RadiationTherapy Committee Task Group No. 43., " Med. Phys. 22, 209-234 (1995).
    • (1995) Med. Phys. , vol.22 , pp. 209-234
    • Nath, R.1    Anderson, L.L.2    Luxton, G.3    Weaver, K.A.4    Williamson, J.F.5    Meigooni, A.S.6
  • 12
    • 0031739032 scopus 로고    scopus 로고
    • Monte Carlo-aided dosimetry of a new high dose-rate brachytherapy source
    • G. M. Daskalov, E. Löffler, and J. F. Williamson, " Monte Carlo-aided dosimetry of a new high dose-rate brachytherapy source., " Med. Phys. 25, 2200-2208 (1998).
    • (1998) Med. Phys. , vol.25 , pp. 2200-2208
    • Daskalov, G.M.1    Löffler, E.2    Williamson, J.F.3
  • 13
    • 84864166247 scopus 로고    scopus 로고
    • a web-based database maintained by the National Nuclear Data Center, Brookhaven National Laboratory (Upton, NY)
    • NuDat 2.0 National Nuclear Data Center 2004 Nuclear data from NuDat, a web-based database maintained by the National Nuclear Data Center, Brookhaven National Laboratory (Upton, NY) http://www.nndc.bnl.gov/nudat2.
    • NuDat 2.0 National Nuclear Data Center 2004 Nuclear Data from NuDat
  • 15
    • 0036703411 scopus 로고    scopus 로고
    • Dose characterization in the near-source region for two high dose rate brachytherapy sources
    • R. Wang and X. A. Li, " Dose characterization in the near-source region for two high dose rate brachytherapy sources., " Med. Phys. 29, 1678-1686 (2002).
    • (2002) Med. Phys. , vol.29 , pp. 1678-1686
    • Wang, R.1    Li, X.A.2
  • 16
    • 79960251481 scopus 로고    scopus 로고
    • GEANT4 Collaboration
    • GEANT4 Collaboration," Physics Reference Manual., " (online), available at http://geant4.web.cern.ch/geant4/.
    • Physics Reference Manual
  • 17
    • 14644405544 scopus 로고    scopus 로고
    • Accuracy of the photon and electron physics in GEANT4 for radiotherapy applications
    • E. Poon and F. Verhaegen, " Accuracy of the photon and electron physics in GEANT4 for radiotherapy applications., " Med. Phys. 32, 1696-1711 (2005).
    • (2005) Med. Phys. , vol.32 , pp. 1696-1711
    • Poon, E.1    Verhaegen, F.2
  • 21
    • 0033461145 scopus 로고    scopus 로고
    • Radial dose distribution, dose to water and dose rate constant for monoenergetic photon point sources from 10 keV to 2 MeV: EGS4 Monte Carlo model calculation
    • G. Luxton and G. Jozsef, " Radial dose distribution, dose to water and dose rate constant for monoenergetic photon point sources from 10 keV to 2 MeV: EGS4 Monte Carlo model calculation., Med. Phys. 26, 2531-2538 (1999).
    • (1999) Med. Phys. , vol.26 , pp. 2531-2538
    • Luxton, G.1    Jozsef, G.2
  • 23
    • 0023514509 scopus 로고
    • Monte Carlo evaluation of kerma at a point for photon transport problems
    • J. F. Williamson, " Monte Carlo evaluation of kerma at a point for photon transport problems., " Med. Phys. 14, 567-576 (1987).
    • (1987) Med. Phys. , vol.14 , pp. 567-576
    • Williamson, J.F.1
  • 24
    • 0038715538 scopus 로고    scopus 로고
    • Incorporation of a combinatorial geometry package and improved scoring capabilities in the EGSnrc Monte Carlo code system
    • M. Fragoso, J. Seco, A. E. Nahum, and F. Verhaegen, " Incorporation of a combinatorial geometry package and improved scoring capabilities in the EGSnrc Monte Carlo code system., " Med. Phys. 30, 1076-1085 (2003).
    • (2003) Med. Phys. , vol.30 , pp. 1076-1085
    • Fragoso, M.1    Seco, J.2    Nahum, A.E.3    Verhaegen, F.4
  • 25
    • 0037241039 scopus 로고    scopus 로고
    • IVBTMC, a Monte Carlo dose calculation tool for intravascular brachytherapy
    • O. Chibani and X. A. Li, " IVBTMC, A Monte Carlo dose calculation tool for intravascular brachytherapy., " Med. Phys. 30, 44-51 (2003).
    • (2003) Med. Phys. , vol.30 , pp. 44-51
    • Chibani, O.1    Li, X.A.2
  • 26
    • 33644999267 scopus 로고    scopus 로고
    • MCPI: A sub-minute Monte Carlo dose calculation engine for prostate implants
    • O. Chibani and J. F. Williamson, " MCPI: A sub-minute Monte Carlo dose calculation engine for prostate implants., " Med. Phys. 32, 3688-3698 (2005).
    • (2005) Med. Phys. , vol.32 , pp. 3688-3698
    • Chibani, O.1    Williamson, J.F.2
  • 27
    • 33644624376 scopus 로고    scopus 로고
    • Impact of interseed attenuation and tissue composition for permanent prostate implants
    • J. F. Carrier, L. Beaulieu, F. Therriault-Proulx, and R. Roy, " Impact of interseed attenuation and tissue composition for permanent prostate implants., " Med. Phys. 33, 595-604 (2006).
    • (2006) Med. Phys. , vol.33 , pp. 595-604
    • Carrier, J.F.1    Beaulieu, L.2    Therriault-Proulx, F.3    Roy, R.4
  • 28
    • 27644471150 scopus 로고    scopus 로고
    • Weak energy dependence of EBT GAFCHROMIC film dose response in the 50 kVp - 10 MVp x-ray range
    • M. J. Butson, T. Cheung, and P. K. N. Yu, " Weak energy dependence of EBT GAFCHROMIC film dose response in the 50 kVp - 10 MVp x-ray range., " Appl. Radiat. Isot. 64, 60-62 (2006).
    • (2006) Appl. Radiat. Isot. , vol.64 , pp. 60-62
    • Butson, M.J.1    Cheung, T.2    Yu, P.K.N.3
  • 30
    • 33845298833 scopus 로고    scopus 로고
    • Sensitivity of document scanners used for radiochromic film dosimetry
    • Y. Wang, D. Lewis, E. Poon, F. Verhaegen, E. Podgorsak, and S. Devic, " Sensitivity of document scanners used for radiochromic film dosimetry., " Med. Phys. 33, 2135 (2006).
    • (2006) Med. Phys. , vol.33 , pp. 2135
    • Wang, Y.1    Lewis, D.2    Poon, E.3    Verhaegen, F.4    Podgorsak, E.5    Devic, S.6
  • 31
    • 0024157010 scopus 로고
    • On the choice of phantom material for the dosimetry of Ir192 sources
    • J. A. Meli, A. S. Meigooni, and R. Nath, " On the choice of phantom material for the dosimetry of Ir192 sources., " Int. J. Radiat. Oncol., Biol., Phys. 14, 587-594 (1988).
    • (1988) Int. J. Radiat. Oncol., Biol., Phys. , vol.14 , pp. 587-594
    • Meli, J.A.1    Meigooni, A.S.2    Nath, R.3
  • 32
    • 0036852599 scopus 로고    scopus 로고
    • Evaluation and characterization of parallel plate microchamber's functionalities in small beam dosimetry
    • H. R. Lee, M. Pankuch, J. C. Chu, and J. J. Spokas, " Evaluation and characterization of parallel plate microchamber's functionalities in small beam dosimetry., " Med. Phys. 29, 2489-2496 (2002).
    • (2002) Med. Phys. , vol.29 , pp. 2489-2496
    • Lee, H.R.1    Pankuch, M.2    Chu, J.C.3    Spokas, J.J.4
  • 33
    • 0029944194 scopus 로고    scopus 로고
    • Measurement and calculation of heterogeneity correction factors for an Ir-192 high dose-rate brachytherapy source behind tungsten alloy and steel shields
    • A. S. Kirov, J. F. Williamson, A. S. Meigooni, and Y. Zhu, " Measurement and calculation of heterogeneity correction factors for an Ir-192 high dose-rate brachytherapy source behind tungsten alloy and steel shields., " Med. Phys. 23, 911-919 (1996).
    • (1996) Med. Phys. , vol.23 , pp. 911-919
    • Kirov, A.S.1    Williamson, J.F.2    Meigooni, A.S.3    Zhu, Y.4
  • 34
    • 0028158549 scopus 로고
    • Dose distributions produced by a shielded vaginal cylinder using a high-activity iridium-192 source
    • F. M. Waterman and D. E. Holcomb, " Dose distributions produced by a shielded vaginal cylinder using a high-activity iridium-192 source., " Med. Phys. 21, 101-106 (1994).
    • (1994) Med. Phys. , vol.21 , pp. 101-106
    • Waterman, F.M.1    Holcomb, D.E.2
  • 36
    • 0028320064 scopus 로고
    • Activity determination of Ir192 solid sources by ionization chamber measurements using calculated corrections for self-absorption
    • L. Büermann, H. M. Kramer, H. Schrader, and H. J. Selbach, " Activity determination of Ir192 solid sources by ionization chamber measurements using calculated corrections for self-absorption., " Nucl. Instrum. Methods Phys. Res. A 339, 369-376 (1994).
    • (1994) Nucl. Instrum. Methods Phys. Res. a , vol.339 , pp. 369-376
    • Büermann, L.1    Kramer, H.M.2    Schrader, H.3    Selbach, H.J.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.