Effects of electromagnetic interference on implanted cardiac devices and their management

Author keywords

Electromagnetic interference; Implantable cardioverter defibrillator; Pacemakers

Indexed keywords

ANESTHESIST; CARDIOLOGY; CARDIOVERSION; CATHETER ABLATION; CAUTERIZATION; DEFIBRILLATION; DEFIBRILLATOR; DEVICE; ELECTROCONVULSIVE THERAPY; ELECTROMAGNETIC FIELD; EMERGENCY MEDICINE; EXTRACORPOREAL LITHOTRIPSY; HEART RHYTHM; HUMAN; IMPLANT; INTENSIVE CARE; NUCLEAR MAGNETIC RESONANCE IMAGING; PACEMAKER; PATIENT CARE; RADIOTHERAPY; REVIEW; SINUS NODE; TELEPHONE; TRANSCUTANEOUS NERVE STIMULATION; VIDEORECORDING;

EID: 37348999169     PISSN: 10615377     EISSN: None     Source Type: Journal    
DOI: 10.1097/CRD.0b013e31813e0ba9     Document Type: Review
Times cited : (53)

References (27)
  • 1
    • 0022637562 scopus 로고
    • Electrocautery and pacemakers: Management of the paced patient subject to electrocautery
    • Levine PA, Balady GJ, Lazar HL, et al. Electrocautery and pacemakers: management of the paced patient subject to electrocautery. Ann Thorac Surg. 1986;41:313-317.
    • (1986) Ann Thorac Surg , vol.41 , pp. 313-317
    • Levine, P.A.1    Balady, G.J.2    Lazar, H.L.3
  • 2
    • 0024444299 scopus 로고
    • Upper limit ventricular stimulation in respiratory rate responsive pacing due to electrocautery
    • VanHemel NM, Hamerlijnk RP, Pronk KJ, et al. Upper limit ventricular stimulation in respiratory rate responsive pacing due to electrocautery. Pacing Clin Electrophysiol. 1989;12:1720-1723.
    • (1989) Pacing Clin Electrophysiol , vol.12 , pp. 1720-1723
    • VanHemel, N.M.1    Hamerlijnk, R.P.2    Pronk, K.J.3
  • 3
    • 0026592072 scopus 로고
    • Interaction between permanent cardiac pacing and electrocautery. The significance of electrode position
    • Chauvin M, Crenner F, Brechenmacher C. Interaction between permanent cardiac pacing and electrocautery. The significance of electrode position. Pacing Clin Electrophysiol. 1992;15:2028-2033.
    • (1992) Pacing Clin Electrophysiol , vol.15 , pp. 2028-2033
    • Chauvin, M.1    Crenner, F.2    Brechenmacher, C.3
  • 4
    • 0020638202 scopus 로고
    • Adverse acute and chronic effects of electrical defibrillation and cardioversion on implanted unipolar cardiac pacing systems
    • Levine PA, Barold S, Fletcher RD, et al. Adverse acute and chronic effects of electrical defibrillation and cardioversion on implanted unipolar cardiac pacing systems. J Am Coll Cardiol. 1983;1:1413-1422.
    • (1983) J Am Coll Cardiol , vol.1 , pp. 1413-1422
    • Levine, P.A.1    Barold, S.2    Fletcher, R.D.3
  • 5
    • 0031019136 scopus 로고    scopus 로고
    • Selective dysfunction of ventricular electrode-endocardial junction following CD cardioversion in a patient with a dual chamber pacemaker
    • Das G, Staffanson DB. Selective dysfunction of ventricular electrode-endocardial junction following CD cardioversion in a patient with a dual chamber pacemaker. Pacing Clin Electrophysiol. 1997;20:346-365.
    • (1997) Pacing Clin Electrophysiol , vol.20 , pp. 346-365
    • Das, G.1    Staffanson, D.B.2
  • 6
    • 3042766221 scopus 로고    scopus 로고
    • Adverse effects of direct current cardioversion of cardiac pacemakers and electrodes. Is external cardioversion contraindicated in patients with permanent pacing systems?
    • Waller C, Callies F, Langenfeld H. Adverse effects of direct current cardioversion of cardiac pacemakers and electrodes. Is external cardioversion contraindicated in patients with permanent pacing systems? Europace. 2004;6:165-168.
    • (2004) Europace , vol.6 , pp. 165-168
    • Waller, C.1    Callies, F.2    Langenfeld, H.3
  • 7
    • 1842478179 scopus 로고    scopus 로고
    • Magnetic resonance imaging and cardiac pacemaker safety at 1.5 Tesla
    • Martin ET, Coman JA, Shellock FG, et al. Magnetic resonance imaging and cardiac pacemaker safety at 1.5 Tesla. J Am Coll Cardiol. 2004;43:1315-1324.
    • (2004) J Am Coll Cardiol , vol.43 , pp. 1315-1324
    • Martin, E.T.1    Coman, J.A.2    Shellock, F.G.3
  • 8
    • 0038221323 scopus 로고    scopus 로고
    • Cardiac pacemakers, ICDs, and loop recorder: Evaluation of translational attraction using conventional ("long-bore") and "short-bore" 1.5- and 3.0-Tesla MR systems
    • Shellock FG, Tkach JA, Ruggieri PM, et al. Cardiac pacemakers, ICDs, and loop recorder: evaluation of translational attraction using conventional ("long-bore") and "short-bore" 1.5- and 3.0-Tesla MR systems. J Cardiovasc Magn Reson. 2003;5:387-397.
    • (2003) J Cardiovasc Magn Reson , vol.5 , pp. 387-397
    • Shellock, F.G.1    Tkach, J.A.2    Ruggieri, P.M.3
  • 9
    • 0029068565 scopus 로고
    • Shockwave lithotripsy and pacemakers: Experience with 20 cases
    • Albers DD, Lybrand FE III, Axton JC, et al. Shockwave lithotripsy and pacemakers: experience with 20 cases. J Endourol. 1995;9:301-303.
    • (1995) J Endourol , vol.9 , pp. 301-303
    • Albers, D.D.1    Lybrand III, F.E.2    Axton, J.C.3
  • 10
    • 0033031888 scopus 로고    scopus 로고
    • Effects of extracorporeal shock wave lithotripsy on tiered therapy implantable cardioverter defibrillators
    • Chung MK, Streem SB, Ching E, et al. Effects of extracorporeal shock wave lithotripsy on tiered therapy implantable cardioverter defibrillators. Pacing Clin Electrophysiol. 1999;22:738-742.
    • (1999) Pacing Clin Electrophysiol , vol.22 , pp. 738-742
    • Chung, M.K.1    Streem, S.B.2    Ching, E.3
  • 11
    • 4644247851 scopus 로고    scopus 로고
    • Electroconvulsive therapy in patients with cardiac pacemakers and implantable cardioverter defibrillators
    • Dolenc TJ, Barnes RD, Hayes DL, et al. Electroconvulsive therapy in patients with cardiac pacemakers and implantable cardioverter defibrillators. Pacing Clin Electrophysiol. 2004;27:1257-1263.
    • (2004) Pacing Clin Electrophysiol , vol.27 , pp. 1257-1263
    • Dolenc, T.J.1    Barnes, R.D.2    Hayes, D.L.3
  • 12
    • 27544509139 scopus 로고    scopus 로고
    • Radiotherapy to patients with artificial cardiac pacemakers
    • Sundar S, Symonds RP, Deehan C. Radiotherapy to patients with artificial cardiac pacemakers. Cancer Treat Rev. 2005;31:474-486.
    • (2005) Cancer Treat Rev , vol.31 , pp. 474-486
    • Sundar, S.1    Symonds, R.P.2    Deehan, C.3
  • 13
    • 23244453213 scopus 로고    scopus 로고
    • Influence of radiotherapy on the latest generation of pacemakers
    • Hurkmans CW, Scheepers E, Springorum BG, et al. Influence of radiotherapy on the latest generation of pacemakers. Radiother Oncol. 2005;76:93-98.
    • (2005) Radiother Oncol , vol.76 , pp. 93-98
    • Hurkmans, C.W.1    Scheepers, E.2    Springorum, B.G.3
  • 14
    • 26944493433 scopus 로고    scopus 로고
    • Influence of radiotherapy on the latest generation of implantable cardioverter-defibrillators
    • Hurkmans CW, Scheepers E, Springorum BG, et al. Influence of radiotherapy on the latest generation of implantable cardioverter-defibrillators. Int J Radiat Oncol Biol Phys. 2005;63:282-289.
    • (2005) Int J Radiat Oncol Biol Phys , vol.63 , pp. 282-289
    • Hurkmans, C.W.1    Scheepers, E.2    Springorum, B.G.3
  • 15
    • 0031682609 scopus 로고    scopus 로고
    • Inappropriate shocks delivered by an ICD as a result of sensed potentials from a transcutaneous electronic nerve stimulation unit
    • Philbin DM, Marieb MA, Aithal KH, et al. Inappropriate shocks delivered by an ICD as a result of sensed potentials from a transcutaneous electronic nerve stimulation unit. Pacing Clin Electrophysiol. 1998;21:2010-2011.
    • (1998) Pacing Clin Electrophysiol , vol.21 , pp. 2010-2011
    • Philbin, D.M.1    Marieb, M.A.2    Aithal, K.H.3
  • 16
    • 0037247985 scopus 로고    scopus 로고
    • The simultaneous use of a biventricular implantable cardioverter defibrillator (ICD) and transcutaneous electrical nerve stimulation (TENS) unit: Implications for device interaction
    • Pyatt JR, Trenbath D, Chester M, et al. The simultaneous use of a biventricular implantable cardioverter defibrillator (ICD) and transcutaneous electrical nerve stimulation (TENS) unit: implications for device interaction. Europace. 2003;5:91-93.
    • (2003) Europace , vol.5 , pp. 91-93
    • Pyatt, J.R.1    Trenbath, D.2    Chester, M.3
  • 17
    • 13844307290 scopus 로고    scopus 로고
    • Electromagnetic interference with implantable cardiac pacemakers by video capsule
    • Dubner S, Dubner Y, Gallino S, et al. Electromagnetic interference with implantable cardiac pacemakers by video capsule. Gastrointest Endosc. 2005;61:250-254.
    • (2005) Gastrointest Endosc , vol.61 , pp. 250-254
    • Dubner, S.1    Dubner, Y.2    Gallino, S.3
  • 18
    • 28844448757 scopus 로고    scopus 로고
    • Effects of capsule endoscopy on cardiac pacemakers
    • Payeras G, Piqueras J, Moreno VJ, et al. Effects of capsule endoscopy on cardiac pacemakers. Endoscopy. 2005;37:1181-1185.
    • (2005) Endoscopy , vol.37 , pp. 1181-1185
    • Payeras, G.1    Piqueras, J.2    Moreno, V.J.3
  • 19
    • 23044479016 scopus 로고    scopus 로고
    • The effects of mobile phones on pacemaker function
    • Tandogan I, Temizhan A, Yetkin E, et al. The effects of mobile phones on pacemaker function. Int J Cardiol. 2005;103:51-58.
    • (2005) Int J Cardiol , vol.103 , pp. 51-58
    • Tandogan, I.1    Temizhan, A.2    Yetkin, E.3
  • 20
    • 33645701412 scopus 로고    scopus 로고
    • Evaluation of electromagnetic interference of GSM mobile phones with pacemakers featuring remote monitoring functions
    • Calcagnini G, Censi F, Floris M, et al. Evaluation of electromagnetic interference of GSM mobile phones with pacemakers featuring remote monitoring functions. Pacing Clin Electrophysiol. 2006;29:380-385.
    • (2006) Pacing Clin Electrophysiol , vol.29 , pp. 380-385
    • Calcagnini, G.1    Censi, F.2    Floris, M.3
  • 21
    • 0032923717 scopus 로고    scopus 로고
    • Electromagnetic interference of digital and analog cellular telephones with implantable cardioverter defibrillators: In vitro and in vivo studies
    • Barbaro V, Bellocci F, Caruso F, et al. Electromagnetic interference of digital and analog cellular telephones with implantable cardioverter defibrillators: in vitro and in vivo studies. Pacing Clin Electrophysiol. 1999;22:626-634.
    • (1999) Pacing Clin Electrophysiol , vol.22 , pp. 626-634
    • Barbaro, V.1    Bellocci, F.2    Caruso, F.3
  • 22
    • 33644790159 scopus 로고    scopus 로고
    • Effects of mobile telephones on the function of implantable cardioverter defibrillators
    • Tandogan I, Ozin B, Bozbas H, et al. Effects of mobile telephones on the function of implantable cardioverter defibrillators. Ann Noninvasive Electrocardiol. 2005;10:409-413.
    • (2005) Ann Noninvasive Electrocardiol , vol.10 , pp. 409-413
    • Tandogan, I.1    Ozin, B.2    Bozbas, H.3
  • 23
    • 0037530899 scopus 로고    scopus 로고
    • Implantable cardioverter defibrillators and cellular telephones: Is there any interference?
    • Occhetta E, Plebani L, Bortnik M, et al. Implantable cardioverter defibrillators and cellular telephones: is there any interference? Pacing Clin Electrophysiol. 1999;22:983-989.
    • (1999) Pacing Clin Electrophysiol , vol.22 , pp. 983-989
    • Occhetta, E.1    Plebani, L.2    Bortnik, M.3
  • 24
    • 0023780458 scopus 로고    scopus 로고
    • The effect of metal detector gates on implanted permanent pacemakers
    • Copperman Y, Zarfati D, Laniado S. The effect of metal detector gates on implanted permanent pacemakers. Pacing Clin Electrophysiol. 1998;11:1386-1387.
    • (1998) Pacing Clin Electrophysiol , vol.11 , pp. 1386-1387
    • Copperman, Y.1    Zarfati, D.2    Laniado, S.3
  • 25
    • 0033609514 scopus 로고    scopus 로고
    • Interactions between electronic article surveillance systems and implantable cardioverterdefibrillators
    • Groh WJ, Boschee SA, Engelstein ED, et al. Interactions between electronic article surveillance systems and implantable cardioverterdefibrillators. Circulation. 1999;100:387-392.
    • (1999) Circulation , vol.100 , pp. 387-392
    • Groh, W.J.1    Boschee, S.A.2    Engelstein, E.D.3
  • 26
    • 0032487943 scopus 로고    scopus 로고
    • Interference with an implantable defibrillator by an electronic antitheft surveillance device
    • Santucci PA, Haw J, Trohman RG, et al. Interference with an implantable defibrillator by an electronic antitheft surveillance device. N Engl J Med. 1998;339:1371-1374.
    • (1998) N Engl J Med , vol.339 , pp. 1371-1374
    • Santucci, P.A.1    Haw, J.2    Trohman, R.G.3
  • 27
    • 7744228148 scopus 로고    scopus 로고
    • Understanding electromagnetic interference
    • Sweesy MW. Understanding electromagnetic interference. Heart Rhythm. 2004;1:523-524.
    • (2004) Heart Rhythm , vol.1 , pp. 523-524
    • Sweesy, M.W.1


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.