Volumn 3, Issue 4, 2007, Pages 389-400

Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension

Author keywords

International Protocol; Omron M1 Plus (HEM 4011C E); Omron M6 Comfort (HEM 7000 E); Self blood pressure measurement; Spengler KP7500 D and Microlife BP A100 plus; Validation

Indexed keywords

ADULT; AGED; ARTICLE; AUTOMATION; BLOOD PRESSURE MEASUREMENT; CONTROLLED STUDY; EQUIPMENT DESIGN; FEMALE; HUMAN; HUMAN EXPERIMENT; HYPERTENSION; INTERMETHOD COMPARISON; MALE; MATHEMATICAL ANALYSIS; MEDICAL SOCIETY; REFERENCE VALUE; SELF CARE; SPHYGMOMANOMETER; STANDARDIZATION; VALIDATION PROCESS; VALIDATION STUDY;

EID: 35148870123     PISSN: 11766344     EISSN: None     Source Type: Journal    
DOI: None     Document Type: Article
Times cited : (77)

References (32)
  • 1
    • 1842531917 scopus 로고    scopus 로고
    • Validation of the Pharma-Smart PS-2000 public use blood pressure monitor
    • Alpert BS. 2004. Validation of the Pharma-Smart PS-2000 public use blood pressure monitor. Blood Press Monit, 9:19-23.
    • (2004) Blood Press Monit , vol.9 , pp. 19-23
    • Alpert, B.S.1
  • 2
    • 0034028579 scopus 로고    scopus 로고
    • Asmar R, Zanchetti A. 2000. Guidelines for the use of self-blood pressure monitoring: a summary report of the first international consensus conference. J Hypertens, 18:493-508.
    • Asmar R, Zanchetti A. 2000. Guidelines for the use of self-blood pressure monitoring: a summary report of the first international consensus conference. J Hypertens, 18:493-508.
  • 5
    • 0032790299 scopus 로고    scopus 로고
    • Validation of the two devices for self-measurement of blood pressure by elderly patients according to the revised British Hypertension Society protocol: The Omron HEM-722C and HEM-735C
    • Bortolotto L, Henry O, Hanon O, et al. 1999. Validation of the two devices for self-measurement of blood pressure by elderly patients according to the revised British Hypertension Society protocol: the Omron HEM-722C and HEM-735C. Blood Press Monit, 4:21-5.
    • (1999) Blood Press Monit , vol.4 , pp. 21-25
    • Bortolotto, L.1    Henry, O.2    Hanon, O.3
  • 6
    • 0031591782 scopus 로고    scopus 로고
    • Validation of a self-measuring blood pressure monitor: The Omron-HM 722C
    • Cordoba R, Fuertes MI, Alvarez A, et al. 1997. Validation of a self-measuring blood pressure monitor: the Omron-HM 722C. Aten Primaria, 20:247-50.
    • (1997) Aten Primaria , vol.20 , pp. 247-250
    • Cordoba, R.1    Fuertes, M.I.2    Alvarez, A.3
  • 7
    • 0036970374 scopus 로고    scopus 로고
    • British Hypertension Society Validation of the Microlife BP 3BTO-A oscillometric blood pressure monitoring device according to a modified British Hypertension Society protocol
    • Cuckson AC, Reinders A, Shabeeh H, et al. 2002. British Hypertension Society Validation of the Microlife BP 3BTO-A oscillometric blood pressure monitoring device according to a modified British Hypertension Society protocol. Blood Press Monit, 7:319-24.
    • (2002) Blood Press Monit , vol.7 , pp. 319-324
    • Cuckson, A.C.1    Reinders, A.2    Shabeeh, H.3
  • 8
    • 0036694384 scopus 로고    scopus 로고
    • Validation of the Omron HEM-907 device for blood pressure measurement
    • El Assaad M, Topouchian J, Damé B, et al. 2002. Validation of the Omron HEM-907 device for blood pressure measurement. Blood Press Monit, 7:237-41.
    • (2002) Blood Press Monit , vol.7 , pp. 237-241
    • El Assaad, M.1    Topouchian, J.2    Damé, B.3
  • 9
    • 0042383300 scopus 로고    scopus 로고
    • Evaluation of two devices for self-measurement of blood pressure according to the international protocol: The Omron M5-I and the Ommn 705IT
    • El Assaad MA, Topouchian JA, Asmar RG. 2003. Evaluation of two devices for self-measurement of blood pressure according to the international protocol: the Omron M5-I and the Ommn 705IT. Blood Press Monit, 8:127-33.
    • (2003) Blood Press Monit , vol.8 , pp. 127-133
    • El Assaad, M.A.1    Topouchian, J.A.2    Asmar, R.G.3
  • 10
    • 0028101528 scopus 로고
    • Accuracy of the Omron HEM 706 portable monitor for home measurement of blood pressure
    • Foster C, McKinlay JM, Cruickshank JM, Coats AJS. 1994. Accuracy of the Omron HEM 706 portable monitor for home measurement of blood pressure. J Hum Hypertens, 8:661-4.
    • (1994) J Hum Hypertens , vol.8 , pp. 661-664
    • Foster, C.1    McKinlay, J.M.2    Cruickshank, J.M.3    Coats, A.J.S.4
  • 11
    • 0020307159 scopus 로고
    • Characterization of the oscillometric method for measuring indirect blood pressure
    • Geddes LA, Voellz M, Combs C, et al. 1982. Characterization of the oscillometric method for measuring indirect blood pressure. Ann Biomed Eng, 10:271-80.
    • (1982) Ann Biomed Eng , vol.10 , pp. 271-280
    • Geddes, L.A.1    Voellz, M.2    Combs, C.3
  • 12
    • 0036966583 scopus 로고    scopus 로고
    • Inflationary oscillometric blood pressure monitoring: Validation of the OMRON-MIT
    • Golara M, Jones C, Randhawa M, et al. 2002. Inflationary oscillometric blood pressure monitoring: validation of the OMRON-MIT. Blood Press Monit, 7:325-8.
    • (2002) Blood Press Monit , vol.7 , pp. 325-328
    • Golara, M.1    Jones, C.2    Randhawa, M.3
  • 13
    • 35148855850 scopus 로고    scopus 로고
    • Groupe d'Evaluation et de Mesure de la Société Française d'Hypertension Artérielle. 1998. La prise de la tension artérielle au cabinet médical. CD ROM. Ran D (ed). Meudon: France.
    • Groupe d'Evaluation et de Mesure de la Société Française d'Hypertension Artérielle. 1998. La prise de la tension artérielle au cabinet médical. CD ROM. Ran D (ed). Meudon: France.
  • 14
    • 0017154547 scopus 로고
    • Improvement of medication compliance in uncontrolled hypertension
    • Haynes R, Sackett D, Gibson E, et al. 1976. Improvement of medication compliance in uncontrolled hypertension. Lancet, 1:1265-8.
    • (1976) Lancet , vol.1 , pp. 1265-1268
    • Haynes, R.1    Sackett, D.2    Gibson, E.3
  • 15
    • 0034892296 scopus 로고    scopus 로고
    • Comparison of the oscillometric blood pressure monitor (BPM-100beta) with the auscultatory mercury sphygmomanometer
    • Mattu G, Perry T, Wright J. 2001. Comparison of the oscillometric blood pressure monitor (BPM-100beta) with the auscultatory mercury sphygmomanometer. Blood Press Monit, 6:153-9.
    • (2001) Blood Press Monit , vol.6 , pp. 153-159
    • Mattu, G.1    Perry, T.2    Wright, J.3
  • 16
    • 0001845560 scopus 로고    scopus 로고
    • Clinical evaluation of the QuietTrak blood pressure recorder according to the protocol of the British Hypertension Society
    • Modesti PA, Costoli A, Cecioni I, et al. 1996. Clinical evaluation of the QuietTrak blood pressure recorder according to the protocol of the British Hypertension Society. Blood Press Monit, 1:63-8.
    • (1996) Blood Press Monit , vol.1 , pp. 63-68
    • Modesti, P.A.1    Costoli, A.2    Cecioni, I.3
  • 17
    • 0033942461 scopus 로고    scopus 로고
    • Accuracy of the Omron M4 automatic blood pressure measuring device
    • Naschitz JE, Loewenstein L, Lewis R, et al. 2000. Accuracy of the Omron M4 automatic blood pressure measuring device. J Hum Hypertens, 14:423-7.
    • (2000) J Hum Hypertens , vol.14 , pp. 423-427
    • Naschitz, J.E.1    Loewenstein, L.2    Lewis, R.3
  • 18
    • 1842479864 scopus 로고    scopus 로고
    • Home blood pressure measurement: Validation of the Braun BP 2550 (UG) monitor according to the ESH International Protocol
    • Nolly H, Romero M, Nolly A, et al. 2004. Home blood pressure measurement: validation of the Braun BP 2550 (UG) monitor according to the ESH International Protocol. Blood Press Monit, 9:53-8.
    • (2004) Blood Press Monit , vol.9 , pp. 53-58
    • Nolly, H.1    Romero, M.2    Nolly, A.3
  • 19
    • 20244382998 scopus 로고    scopus 로고
    • European Society of Hypertension Working Group on Blood Pressure Monitoring. Practice guidelines of the European Society of Hypertension for clinic, ambulatory and self blood pressure measurement
    • O'Brien E, Asmar R, Beilin L, et al. 2005. European Society of Hypertension Working Group on Blood Pressure Monitoring. Practice guidelines of the European Society of Hypertension for clinic, ambulatory and self blood pressure measurement. J Hypertens, 23:697-701.
    • (2005) J Hypertens , vol.23 , pp. 697-701
    • O'Brien, E.1    Asmar, R.2    Beilin, L.3
  • 20
    • 0027194686 scopus 로고
    • Comparative accuracy of six ambulatory devices according to blood pressure levels
    • O'Brien E, Atkins N, Mee F, et al. 1993a. Comparative accuracy of six ambulatory devices according to blood pressure levels. J Hypertens, 11:673-5.
    • (1993) J Hypertens , vol.11 , pp. 673-675
    • O'Brien, E.1    Atkins, N.2    Mee, F.3
  • 21
    • 1842426223 scopus 로고    scopus 로고
    • Evaluation of three devices for sel-measurement of blood pressure according to the revised British Hypertension Society Protocol: The Omron HEM-705CP, Philips HP5332, and Neissei DS-175
    • O'Brien E, Mee F, Atkins N, et al. 1996. Evaluation of three devices for sel-measurement of blood pressure according to the revised British Hypertension Society Protocol: the Omron HEM-705CP, Philips HP5332, and Neissei DS-175. Blood Press Monit, 1:55-61.
    • (1996) Blood Press Monit , vol.1 , pp. 55-61
    • O'Brien, E.1    Mee, F.2    Atkins, N.3
  • 22
    • 0025071770 scopus 로고
    • The British Hypertension Society Protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems
    • O'Brien E, Petrie J, Littler WA, et al. 1990. The British Hypertension Society Protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems. J Hypertens, 8: 607-19.
    • (1990) J Hypertens , vol.8 , pp. 607-619
    • O'Brien, E.1    Petrie, J.2    Littler, W.A.3
  • 23
    • 0027218269 scopus 로고
    • The British Hypertension Society Protocol for the evaluation of blood pressure measuring devices
    • O'Brien E, Petrie J, Littler WA, et al. 1993b. The British Hypertension Society Protocol for the evaluation of blood pressure measuring devices. J Hypertens, 11(Suppl 2):S43-S62.
    • (1993) J Hypertens , vol.11 , Issue.SUPPL. 2
    • O'Brien, E.1    Petrie, J.2    Littler, W.A.3
  • 24
    • 0035992025 scopus 로고    scopus 로고
    • Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults
    • O'Brien E, Pickering T, Asmar R, et al. 2002. Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults. Blood Press Monit, 7:3-17.
    • (2002) Blood Press Monit , vol.7 , pp. 3-17
    • O'Brien, E.1    Pickering, T.2    Asmar, R.3
  • 25
    • 0030066412 scopus 로고    scopus 로고
    • Recommendations for the use of home (self) and ambulatory blood pressure monitoring
    • Pickering T, Kaplan N, Krakoff L, et al. 1996. Recommendations for the use of home (self) and ambulatory blood pressure monitoring. Am J Hypertens, 9:1-11.
    • (1996) Am J Hypertens , vol.9 , pp. 1-11
    • Pickering, T.1    Kaplan, N.2    Krakoff, L.3
  • 26
    • 0036967315 scopus 로고    scopus 로고
    • Ploin D, Baguet JP, Pierre H, et al. 2002. Clinical evaluation of a self blood pressure monitor according to the First International Consensus Conference on Self Blood Pressure Measurement. Blood Press Monit, 7:335-41.
    • Ploin D, Baguet JP, Pierre H, et al. 2002. Clinical evaluation of a self blood pressure monitor according to the First International Consensus Conference on Self Blood Pressure Measurement. Blood Press Monit, 7:335-41.
  • 27
    • 33746112932 scopus 로고    scopus 로고
    • Validation of the Microlife BPA100 Plus device for self-home blood pressure measurement according to the International Protocol
    • Stergiou GS, Giovas P, Neofytou MS, et al. 2006. Validation of the Microlife BPA100 Plus device for self-home blood pressure measurement according to the International Protocol. Blood Press Monit, 11:157-60.
    • (2006) Blood Press Monit , vol.11 , pp. 157-160
    • Stergiou, G.S.1    Giovas, P.2    Neofytou, M.S.3
  • 28
    • 0032498690 scopus 로고    scopus 로고
    • Reference values for self-recorded blood pressure. A recta-analysis of summary data
    • Thijs L, Staessen JA, Celis H, et al. 1998. Reference values for self-recorded blood pressure. A recta-analysis of summary data. Arch Intern Med, 158:481-8.
    • (1998) Arch Intern Med , vol.158 , pp. 481-488
    • Thijs, L.1    Staessen, J.A.2    Celis, H.3
  • 29
    • 33746151925 scopus 로고    scopus 로고
    • Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: The Omron M6 (HEM-7001-E) and the Omron R7 (HEM-637-IT)
    • Topouchian JA, El Assaad MA, Orobinskaia LY, et al. 2006. Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM-637-IT). Blood Press Monit, 11:165-71.
    • (2006) Blood Press Monit , vol.11 , pp. 165-171
    • Topouchian, J.A.1    El Assaad, M.A.2    Orobinskaia, L.Y.3
  • 30
    • 4344566251 scopus 로고    scopus 로고
    • Validation of the A&D UA-774 (UA-767Plus) device for self-measurement of blood pressure
    • Verdecchia P, Angeli F, Poeta F, et al. 2004. Validation of the A&D UA-774 (UA-767Plus) device for self-measurement of blood pressure. Blood Press Monit, 9:225-9.
    • (2004) Blood Press Monit , vol.9 , pp. 225-229
    • Verdecchia, P.1    Angeli, F.2    Poeta, F.3
  • 31
    • 0002354124 scopus 로고    scopus 로고
    • Out-of-office blood pressure monitoring
    • Izzo J, Black HR eds, 2nd ed. Dallas: Council for High Blood pressure Research, American Heart Association. pp
    • White W. 1998. Out-of-office blood pressure monitoring. In Izzo J, Black HR (eds). Hypertension primer. 2nd ed. Dallas: Council for High Blood pressure Research, American Heart Association. pp. 302-6.
    • (1998) Hypertension primer , pp. 302-306
    • White, W.1
  • 32
    • 0034959116 scopus 로고    scopus 로고
    • Evaluation of the overall efficacy of the Omron office digital blood pressure HEM-907 monitor in adults
    • White W, Anwar Y. 2001. Evaluation of the overall efficacy of the Omron office digital blood pressure HEM-907 monitor in adults. Blood Press Monit, 6:107-10.
    • (2001) Blood Press Monit , vol.6 , pp. 107-110
    • White, W.1    Anwar, Y.2


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.