Volumn 85, Issue 3, 2007, Pages 272-279

Comparisons between Pascal dynamic contour tonometry, the TonoPen, and Goldmann applanation tonometry in patients with glaucoma

Author keywords

Central corneal thickness; Goldmann applanation tonometry; Pascal dynamic contour tonometry; Primary open angle glaucoma; Tonometry; TonoPen

Indexed keywords

ADULT; AGE; AGED; ARTICLE; CONTROLLED STUDY; CORNEA THICKNESS; HUMAN; INTERMETHOD COMPARISON; INTRAOCULAR PRESSURE; LINEAR REGRESSION ANALYSIS; MAJOR CLINICAL STUDY; OCULOPLETHYSMOGRAPHY; OPEN ANGLE GLAUCOMA; PRIORITY JOURNAL; TONOMETER;

EID: 34248162503     PISSN: 13953907     EISSN: 16000420     Source Type: Journal    
DOI: 10.1111/j.1600-0420.2006.00834.x     Document Type: Article
Times cited : (72)

References (41)
  • 1
    • 0035182854 scopus 로고    scopus 로고
    • Clinical comparison of the measurement of the IOP with the ocular blood flow tonometer, the TonoPen XL and the Goldmann applanation tonometer
    • Bafa M, Lambrinakis I, Dayan M & Birch M (2001): Clinical comparison of the measurement of the IOP with the ocular blood flow tonometer, the TonoPen XL and the Goldmann applanation tonometer. Acta Ophthalmol Scand 79: 15-18.
    • (2001) Acta Ophthalmol Scand , vol.79 , pp. 15-18
    • Bafa, M.1    Lambrinakis, I.2    Dayan, M.3    Birch, M.4
  • 2
    • 0036235038 scopus 로고    scopus 로고
    • Effect of corneal thickness on intraocular pressure measurements with the pneumotonometer, Goldmann applanation tonometer, and TonoPen
    • Bhan A, Browning AC, Shah S, Hamilton R, Dave D & Dua HS (2002): Effect of corneal thickness on intraocular pressure measurements with the pneumotonometer, Goldmann applanation tonometer, and TonoPen. Invest Ophthalmol Vis Sci 43: 1389-1392.
    • (2002) Invest Ophthalmol Vis Sci , vol.43 , pp. 1389-1392
    • Bhan, A.1    Browning, A.C.2    Shah, S.3    Hamilton, R.4    Dave, D.5    Dua, H.S.6
  • 3
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • Bland JM & Altman DG (1986): Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1: 307-310.
    • (1986) Lancet , vol.1 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 4
    • 1842504236 scopus 로고    scopus 로고
    • Corneal thickness in glaucoma screening, diagnosis and management
    • Brandt JD (2004): Corneal thickness in glaucoma screening, diagnosis and management. Curr Opin Ophthalmol 15: 85-89.
    • (2004) Curr Opin Ophthalmol , vol.15 , pp. 85-89
    • Brandt, J.D.1
  • 5
    • 0027411154 scopus 로고
    • Tonometry in the general practice setting. (I): TonoPen compared to Goldmann applanation tonometry
    • Christoffensen T, Fors T, Ringberg U & Holtedahl K (1993): Tonometry in the general practice setting. (I): TonoPen compared to Goldmann applanation tonometry. Acta Ophthalmol (Copenh) 71: 103-108.
    • (1993) Acta Ophthalmol (Copenh) , vol.71 , pp. 103-108
    • Christoffensen, T.1    Fors, T.2    Ringberg, U.3    Holtedahl, K.4
  • 6
    • 0032936567 scopus 로고    scopus 로고
    • Corneal thickness in ocular hypertension, primary open-angle glaucoma, and normal tension glaucoma
    • Copt RP, Thomas R & Mermoud A (1999): Corneal thickness in ocular hypertension, primary open-angle glaucoma, and normal tension glaucoma. Arch Ophthalmol 117: 14-16.
    • (1999) Arch Ophthalmol , vol.117 , pp. 14-16
    • Copt, R.P.1    Thomas, R.2    Mermoud, A.3
  • 9
    • 0031794457 scopus 로고    scopus 로고
    • Positive correlation between TonoPen intraocular pressure and central corneal thickness
    • Dohadwala AA, Munger R & Damji KF (1998): Positive correlation between TonoPen intraocular pressure and central corneal thickness. Ophthalmology 105: 1849-1854.
    • (1998) Ophthalmology , vol.105 , pp. 1849-1854
    • Dohadwala, A.A.1    Munger, R.2    Damji, K.F.3
  • 10
    • 0034603359 scopus 로고    scopus 로고
    • Human corneal thickness and its impact on intraocular pressure measures: A review and meta-analysis approach
    • Doughty MJ & Zaman ML (2000): Human corneal thickness and its impact on intraocular pressure measures: A review and meta-analysis approach. Surv Ophthalmol 44: 367-408.
    • (2000) Surv Ophthalmol , vol.44 , pp. 367-408
    • Doughty, M.J.1    Zaman, M.L.2
  • 11
    • 22544486553 scopus 로고    scopus 로고
    • Comparison of dynamic contour tonometry with Goldmann applanation tonometry over a wide range of central corneal thickness
    • Doyle A & Lachkar Y (2005): Comparison of dynamic contour tonometry with Goldmann applanation tonometry over a wide range of central corneal thickness. J Glaucoma 14: 288-292.
    • (2005) J Glaucoma , vol.14 , pp. 288-292
    • Doyle, A.1    Lachkar, Y.2
  • 13
    • 0031703801 scopus 로고    scopus 로고
    • Tonometry in adults and children. A manometric evaluation of pneumotonometry, applanation, and TonoPen in vitro and in vivo
    • Eisenberg DL, Sherman BG, McKeown CA & Schuman J (1998): Tonometry in adults and children. A manometric evaluation of pneumotonometry, applanation, and TonoPen in vitro and in vivo. Ophthalmology 105: 1173-1181.
    • (1998) Ophthalmology , vol.105 , pp. 1173-1181
    • Eisenberg, D.L.1    Sherman, B.G.2    McKeown, C.A.3    Schuman, J.4
  • 14
    • 0035163029 scopus 로고    scopus 로고
    • Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings
    • Feltgen N, Leifert D & Funk J (2001): Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings. Br J Ophthalmol 85: 85-87.
    • (2001) Br J Ophthalmol , vol.85 , pp. 85-87
    • Feltgen, N.1    Leifert, D.2    Funk, J.3
  • 15
    • 0023893222 scopus 로고
    • Comparison of the TonoPen to the Goldmann applanation tonometer
    • Frenkel RP, Hong J & Shin DH (1988): Comparison of the TonoPen to the Goldmann applanation tonometer. Arch Ophthalmol 106: 750-753.
    • (1988) Arch Ophthalmol , vol.106 , pp. 750-753
    • Frenkel, R.P.1    Hong, J.2    Shin, D.H.3
  • 17
    • 0343628684 scopus 로고    scopus 로고
    • Influence of refraction on tonometric readings after photorefractive keratectomy and laser assisted in situ keratomileusis
    • Gimeno JA, Munoz LA, Valenzuela LA, Molto F & Rahhal MS (2000): Influence of refraction on tonometric readings after photorefractive keratectomy and laser assisted in situ keratomileusis. Cornea 19: 512-516.
    • (2000) Cornea , vol.19 , pp. 512-516
    • Gimeno, J.A.1    Munoz, L.A.2    Valenzuela, L.A.3    Molto, F.4    Rahhal, M.S.5
  • 18
    • 70449195078 scopus 로고
    • Über Applanationstonometrie
    • Goldmann H & Schmidt T (1957): Über Applanationstonometrie. Ophthalmologica 134: 221-242.
    • (1957) Ophthalmologica , vol.134 , pp. 221-242
    • Goldmann, H.1    Schmidt, T.2
  • 19
    • 0035091093 scopus 로고    scopus 로고
    • New TonoPen XL: Comparison with the Goldmann tonometer
    • Iester M, Mermoud A, Achache F & Roy S (2001): New TonoPen XL: comparison with the Goldmann tonometer. Eye 15: 52-58.
    • (2001) Eye , vol.15 , pp. 52-58
    • Iester, M.1    Mermoud, A.2    Achache, F.3    Roy, S.4
  • 20
    • 0017840770 scopus 로고
    • Increased corneal thickness simulating elevated intraocular pressure
    • Johnson M, Kass MA, Moses RA & Grodzki WJ (1978): Increased corneal thickness simulating elevated intraocular pressure. Arch Ophthalmol 96: 664-665.
    • (1978) Arch Ophthalmol , vol.96 , pp. 664-665
    • Johnson, M.1    Kass, M.A.2    Moses, R.A.3    Grodzki, W.J.4
  • 21
    • 23744463890 scopus 로고    scopus 로고
    • Dynamic contour tonometry for intraocular pressure measurement
    • Kamppeter BA & Jonas JB (2005): Dynamic contour tonometry for intraocular pressure measurement. Am J Ophthalmol 140: 318-320.
    • (2005) Am J Ophthalmol , vol.140 , pp. 318-320
    • Kamppeter, B.A.1    Jonas, J.B.2
  • 22
    • 0023470773 scopus 로고
    • Clinical comparison of the Oculab TonoPen to the Goldmann applanation tonometer
    • Kao SF, Lichter PR, Bergstrom TJ, Rowe S & Musch DC (1987): Clinical comparison of the Oculab TonoPen to the Goldmann applanation tonometer. Ophthalmology 91: 1541-1544.
    • (1987) Ophthalmology , vol.91 , pp. 1541-1544
    • Kao, S.F.1    Lichter, P.R.2    Bergstrom, T.J.3    Rowe, S.4    Musch, D.C.5
  • 23
    • 0041360373 scopus 로고    scopus 로고
    • Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis
    • Kaufmann C, Bachmann LM & Thiel MA (2003): Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis. Invest Ophthalmol Vis Sci 44: 3790-3794.
    • (2003) Invest Ophthalmol Vis Sci , vol.44 , pp. 3790-3794
    • Kaufmann, C.1    Bachmann, L.M.2    Thiel, M.A.3
  • 24
    • 4344659799 scopus 로고    scopus 로고
    • Comparison of dynamic contour tonometry with Goldmann applanation tonometry
    • Kaufmann C, Bachmann LM & Thiel MA (2004): Comparison of dynamic contour tonometry with Goldmann applanation tonometry. Invest Ophthalmol Vis Sci 45: 3118-3121.
    • (2004) Invest Ophthalmol Vis Sci , vol.45 , pp. 3118-3121
    • Kaufmann, C.1    Bachmann, L.M.2    Thiel, M.A.3
  • 25
    • 20444489648 scopus 로고    scopus 로고
    • Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states
    • Kniestedt C, Nee M & Stamper RL (2005): Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states. Graefes Arch Clin Exp Ophthalmol 243: 359-366.
    • (2005) Graefes Arch Clin Exp Ophthalmol , vol.243 , pp. 359-366
    • Kniestedt, C.1    Nee, M.2    Stamper, R.L.3
  • 26
    • 33745243256 scopus 로고    scopus 로고
    • Varying effect of corneal thickness on intraocular pressure measurements with different tonometers
    • Ko Y-C, Liu CJ-I & Hsu W-M (2004): Varying effect of corneal thickness on intraocular pressure measurements with different tonometers. Eye 1: 1-6.
    • (2004) Eye , vol.1 , pp. 1-6
    • Ko, Y.-C.1    Liu, C.J.-I.2    Hsu, W.-M.3
  • 27
    • 33645823534 scopus 로고    scopus 로고
    • Comparison of intraocular pressure measured by Pascal dynamic contour tonometry and Goldmann applanation tonometry
    • Ku JYF, Danesh-Meyer HV, Craig JP, Gamble GD & McGhee CNJ (2006): Comparison of intraocular pressure measured by Pascal dynamic contour tonometry and Goldmann applanation tonometry. Eye 20: 191-198.
    • (2006) Eye , vol.20 , pp. 191-198
    • Ku, J.Y.F.1    Danesh-Meyer, H.V.2    Craig, J.P.3    Gamble, G.D.4    McGhee, C.N.J.5
  • 28
    • 3543128942 scopus 로고    scopus 로고
    • Clinical comparison of the Proview eye pressure monitor with the Goldmann applanation tonometer and the TonoPen
    • Li J, Herndon LW, Asrani SG, Stinnett S & Allingham R (2004): Clinical comparison of the Proview eye pressure monitor with the Goldmann applanation tonometer and the TonoPen. Arch Ophthalmol 122: 1117-1121.
    • (2004) Arch Ophthalmol , vol.122 , pp. 1117-1121
    • Li, J.1    Herndon, L.W.2    Asrani, S.G.3    Stinnett, S.4    Allingham, R.5
  • 29
    • 13844313434 scopus 로고    scopus 로고
    • Influence of corneal biomechanical properties on intraocular pressure measurement: Quantitative analysis
    • Liu J & Roberts CJ (2005): Influence of corneal biomechanical properties on intraocular pressure measurement: Quantitative analysis. J Cataract Refract Surg 31: 146-155.
    • (2005) J Cataract Refract Surg , vol.31 , pp. 146-155
    • Liu, J.1    Roberts, C.J.2
  • 30
    • 0032994224 scopus 로고    scopus 로고
    • TonoPen tonometer and corneal thickness
    • Mok KH, Wong CS & Lee VW (1999): TonoPen tonometer and corneal thickness. Eye 13: 35-37.
    • (1999) Eye , vol.13 , pp. 35-37
    • Mok, K.H.1    Wong, C.S.2    Lee, V.W.3
  • 31
    • 0033021344 scopus 로고    scopus 로고
    • Determination of the true intraocular pressure and modulus of elasticity of the human cornea in vivo
    • Orssengo G & Pye DC (1999): Determination of the true intraocular pressure and modulus of elasticity of the human cornea in vivo. Bull Math Biol 61: 551-572.
    • (1999) Bull Math Biol , vol.61 , pp. 551-572
    • Orssengo, G.1    Pye, D.C.2
  • 32
    • 24944580479 scopus 로고    scopus 로고
    • Dynamic contour tonometry versus Goldmann applanation tonometry: A comparative study
    • Pache M, Wilmsmeyer S, Lautebach S & Funk J (2005): Dynamic contour tonometry versus Goldmann applanation tonometry: A comparative study. Graefes Arch Clin Exp Ophthalmol 243: 763-767.
    • (2005) Graefes Arch Clin Exp Ophthalmol , vol.243 , pp. 763-767
    • Pache, M.1    Wilmsmeyer, S.2    Lautebach, S.3    Funk, J.4
  • 33
    • 0035120736 scopus 로고    scopus 로고
    • Reduction in intraocular pressure after laser in situ keratomileusis
    • Park HJ, Uhm KB & Hong C (2001): Reduction in intraocular pressure after laser in situ keratomileusis. J Cataract Refract Surg 27: 303-309.
    • (2001) J Cataract Refract Surg , vol.27 , pp. 303-309
    • Park, H.J.1    Uhm, K.B.2    Hong, C.3
  • 34
    • 0036172308 scopus 로고    scopus 로고
    • Progression to end-stage glaucoma after laser in situ keratomileusis
    • Shaikh NM, Shaikh S, Singh K & Manche E (2002): Progression to end-stage glaucoma after laser in situ keratomileusis. J Cataract Refract Surg 28: 356-359.
    • (2002) J Cataract Refract Surg , vol.28 , pp. 356-359
    • Shaikh, N.M.1    Shaikh, S.2    Singh, K.3    Manche, E.4
  • 35
    • 13844318981 scopus 로고    scopus 로고
    • Assessment of the Pascal dynamic contour tonometer in monitoring intraocular pressure in unoperated eyes and eyes after LASIK
    • Siganos DS, Papastergiou GI & Moedas C (2004): Assessment of the Pascal dynamic contour tonometer in monitoring intraocular pressure in unoperated eyes and eyes after LASIK. J Cataract Refract Surg 30: 746-751.
    • (2004) J Cataract Refract Surg , vol.30 , pp. 746-751
    • Siganos, D.S.1    Papastergiou, G.I.2    Moedas, C.3
  • 36
    • 21344460937 scopus 로고    scopus 로고
    • The influence of the central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the TonoPen XL, and Goldmann applanation tonometry
    • Tonnu P-A, Ho T, Newson T, El Sheikh A, Sharma K, White E, Bunce C & Garway-Heath D (2005a): The influence of the central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the TonoPen XL, and Goldmann applanation tonometry. Br J Ophthalmol 89: 851-854.
    • (2005) Br J Ophthalmol , vol.89 , pp. 851-854
    • Tonnu, P.-A.1    Ho, T.2    Newson, T.3    El Sheikh, A.4    Sharma, K.5    White, E.6    Bunce, C.7    Garway-Heath, D.8
  • 37
    • 21344456718 scopus 로고    scopus 로고
    • A comparison of four methods of tonometry: Method agreement and interobserver variability
    • Tonnu P-A, Ho T, Sharma K, White E, Bunce C & Garway-Heath D (2005b): A comparison of four methodsMof tonometry: method agreement and interobserver variability. Br J Ophthalmol 89: 847-850.
    • (2005) Br J Ophthalmol , vol.89 , pp. 847-850
    • Tonnu, P.-A.1    Ho, T.2    Sharma, K.3    White, E.4    Bunce, C.5    Garway-Heath, D.6
  • 38
    • 0025614886 scopus 로고
    • Tonometer utilization, accuracy and calibration under field conditions
    • Wessels IF & Oh Y (1990): Tonometer utilization, accuracy and calibration under field conditions. Arch Ophthalmol 108: 1709-1712.
    • (1990) Arch Ophthalmol , vol.108 , pp. 1709-1712
    • Wessels, I.F.1    Oh, Y.2
  • 39
    • 0027291601 scopus 로고
    • Sources of error with use of Goldmann-type tonometers
    • Whitacre MM & Stein R (1993): Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol 38: 1-30.
    • (1993) Surv Ophthalmol , vol.38 , pp. 1-30
    • Whitacre, M.M.1    Stein, R.2
  • 40
    • 0027269723 scopus 로고
    • The effect of corneal thickness on applanation tonometry
    • Whitacre MM, Stein RA & Hassanein K (1993): The effect of corneal thickness on applanation tonometry. Am J Ophthalmol 115: 592-596.
    • (1993) Am J Ophthalmol , vol.115 , pp. 592-596
    • Whitacre, M.M.1    Stein, R.A.2    Hassanein, K.3
  • 41
    • 0030963090 scopus 로고    scopus 로고
    • Distribution of corneal central thickness and its association with intraocular pressure. The Rotterdam Study
    • Wolfs RC, Klaver CC, Vingerling JR, Grobbee DE, Hofman A & de Jong PT (1997): Distribution of corneal central thickness and its association with intraocular pressure. The Rotterdam Study. Am J Ophthalmol 123: 767-772.
    • (1997) Am J Ophthalmol , vol.123 , pp. 767-772
    • Wolfs, R.C.1    Klaver, C.C.2    Vingerling, J.R.3    Grobbee, D.E.4    Hofman, A.5    de Jong, P.T.6


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.