Volumn 8, Issue 4, 2006, Pages 218-222

Correlation between implant stability quotient and bone-implant contact: A retrospective histological and histomorphometrical study of seven titanium implants retrieved from humans

Author keywords

Bone structure; Bone implant contact percentage; Histomorphometry; Resonance frequency analysis

Indexed keywords

TITANIUM;

EID: 33845654011     PISSN: 15230899     EISSN: None     Source Type: Journal    
DOI: 10.1111/j.1708-8208.2006.00022.x     Document Type: Article
Times cited : (97)

References (22)
  • 1
    • 4444312847 scopus 로고    scopus 로고
    • A comparison between cutting torque and resonance frequency in the assessment of primary stability and final torque capacity of standard and TiUnite single-tooth implants under immediate loading
    • Da Cunha HA, Francischone CE, Filho DE, Oliveira RCG. A comparison between cutting torque and resonance frequency in the assessment of primary stability and final torque capacity of standard and TiUnite single-tooth implants under immediate loading. Int J Oral Maxillofac Implants 2004; 19:578-585.
    • (2004) Int J Oral Maxillofac Implants , vol.19 , pp. 578-585
    • Da Cunha, H.A.1    Francischone, C.E.2    Filho, D.E.3    Oliveira, R.C.G.4
  • 2
    • 0142088696 scopus 로고    scopus 로고
    • Resonance frequency measurement of implant stability in vivo on implants with a sandblasted and acid-etched surface
    • Barewal RM, Oates TW, Meredith N, Cochran DL. Resonance frequency measurement of implant stability in vivo on implants with a sandblasted and acid-etched surface. Int J Oral Maxillofac Implants 2003; 18:641-651.
    • (2003) Int J Oral Maxillofac Implants , vol.18 , pp. 641-651
    • Barewal, R.M.1    Oates, T.W.2    Meredith, N.3    Cochran, D.L.4
  • 3
    • 7944234226 scopus 로고    scopus 로고
    • Predicting osseointegration by means of implant primary stability. A resonance-frequency analysis study with delayed and immediately loaded ITI SLA implants
    • Nedir R, Bischof M, Szmukler-Moncler S, Bernard JP, Samson J. Predicting osseointegration by means of implant primary stability. A resonance-frequency analysis study with delayed and immediately loaded ITI SLA implants. Clin Oral Implants Res 2004; 15:520-528.
    • (2004) Clin Oral Implants Res , vol.15 , pp. 520-528
    • Nedir, R.1    Bischof, M.2    Szmukler-Moncler, S.3    Bernard, J.P.4    Samson, J.5
  • 4
    • 18744390029 scopus 로고    scopus 로고
    • Implant stability measurements for implants placed at the time of extraction: A cohort, prospective clinical study
    • Becker W, Sennerby L, Bedrossian E, Becker BE, Lucchini JP. Implant stability measurements for implants placed at the time of extraction: a cohort, prospective clinical study. J Periodontol 2005; 76:391-397.
    • (2005) J Periodontol , vol.76 , pp. 391-397
    • Becker, W.1    Sennerby, L.2    Bedrossian, E.3    Becker, B.E.4    Lucchini, J.P.5
  • 5
    • 27544491195 scopus 로고    scopus 로고
    • Stability measurements of 1-stage implants in the maxilla by means of resonance frequency analysis: A pilot study
    • Zix J, Kessler-Liechti G, Merickse-Stern R. Stability measurements of 1-stage implants in the maxilla by means of resonance frequency analysis: a pilot study. Int J Oral Maxillofac Implants 2005; 20:747-752.
    • (2005) Int J Oral Maxillofac Implants , vol.20 , pp. 747-752
    • Zix, J.1    Kessler-Liechti, G.2    Merickse-Stern, R.3
  • 6
    • 4644230057 scopus 로고    scopus 로고
    • Resonance frequency analysis of implants subjected to immediate or early occlual loading. Successful vs. failing implants
    • Glauser R, Sennerby L, Meredith N, et al. Resonance frequency analysis of implants subjected to immediate or early occlual loading. Successful vs. failing implants. Clin Oral Implants Res 2004; 15:428-434.
    • (2004) Clin Oral Implants Res , vol.15 , pp. 428-434
    • Glauser, R.1    Sennerby, L.2    Meredith, N.3
  • 8
    • 0042513564 scopus 로고    scopus 로고
    • Immediate occlusal loading of Branemark TiUnite™ implants placed predominantly in soft bone: 1-Year results of a prospective clinical study
    • Glauser R, Lundgren AK, Gotlow J, et al. Immediate occlusal loading of Branemark TiUnite™ implants placed predominantly in soft bone: 1-year results of a prospective clinical study. Clin Implant Dent Relat Res 2003; 5(Suppl 1):47-56.
    • (2003) Clin Implant Dent Relat Res , vol.5 , Issue.SUPPL. 1 , pp. 47-56
    • Glauser, R.1    Lundgren, A.K.2    Gotlow, J.3
  • 9
    • 12244296400 scopus 로고    scopus 로고
    • Diagnostic parameters for monitoring peri-implant conditions
    • Salvi GE, Lang NP. Diagnostic parameters for monitoring peri-implant conditions. Int J Oral Maxillofac Implants 2004; 19(Suppl):116-127.
    • (2004) Int J Oral Maxillofac Implants , vol.19 , Issue.SUPPL. , pp. 116-127
    • Salvi, G.E.1    Lang, N.P.2
  • 10
    • 23044454177 scopus 로고    scopus 로고
    • Resonance frequency analysis of one-stage dental implant stability during the osseointegration period
    • Ersanli S, Karabuda C, Beck F, Leblebicioglu B. Resonance frequency analysis of one-stage dental implant stability during the osseointegration period. J Periodontol 2005; 76:1066-1071.
    • (2005) J Periodontol , vol.76 , pp. 1066-1071
    • Ersanli, S.1    Karabuda, C.2    Beck, F.3    Leblebicioglu, B.4
  • 11
    • 0036481346 scopus 로고    scopus 로고
    • Resonance frequency assessment of dental implant stability with various bone qualities: A numerical approach
    • Huang HM, Lee SY, Yeh CY, Lin CT. Resonance frequency assessment of dental implant stability with various bone qualities: a numerical approach. Clin Oral Implants Res 2002; 13:65-74.
    • (2002) Clin Oral Implants Res , vol.13 , pp. 65-74
    • Huang, H.M.1    Lee, S.Y.2    Yeh, C.Y.3    Lin, C.T.4
  • 12
    • 11844280921 scopus 로고    scopus 로고
    • Monitoring of implant stability in grafted bone using resonance frequency analysis. A clinical study from implant placement to 6 months of loading
    • Sjostrom M, Lundgren S, Nilson H, Sennerby L. Monitoring of implant stability in grafted bone using resonance frequency analysis. A clinical study from implant placement to 6 months of loading. Int J Oral Maxillofac Surg 2005; 34:45-51.
    • (2005) Int J Oral Maxillofac Surg , vol.34 , pp. 45-51
    • Sjostrom, M.1    Lundgren, S.2    Nilson, H.3    Sennerby, L.4
  • 13
    • 0043014563 scopus 로고    scopus 로고
    • Histology of retrieved immediately and early loaded oxidized implants: Light microscopic observations after 5 to 9 months of loading in the posterior mandible
    • Rocci A, Martignoni M, Burgos PM, Gottlow J, Sennerby L. Histology of retrieved immediately and early loaded oxidized implants: light microscopic observations after 5 to 9 months of loading in the posterior mandible. Clin Implant Dent Relat Res 2003; 5(Suppl 1):88-98.
    • (2003) Clin Implant Dent Relat Res , vol.5 , Issue.SUPPL. 1 , pp. 88-98
    • Rocci, A.1    Martignoni, M.2    Burgos, P.M.3    Gottlow, J.4    Sennerby, L.5
  • 14
    • 33645084797 scopus 로고    scopus 로고
    • Resonance frequency analysis and damping capacity assessment. Part I: Reliability and a method of comparison in the determination of primary dental implant stability
    • Lachmann S, Jager B, Axmann D, Gomez-Roman G, Groten M, Weber H. Resonance frequency analysis and damping capacity assessment. Part I: reliability and a method of comparison in the determination of primary dental implant stability. Clin Oral Implants Res 2006; 17:75-79.
    • (2006) Clin Oral Implants Res , vol.17 , pp. 75-79
    • Lachmann, S.1    Jager, B.2    Axmann, D.3    Gomez-Roman, G.4    Groten, M.5    Weber, H.6
  • 15
    • 0031128359 scopus 로고    scopus 로고
    • High-precision, cost-effective system for producing thin sections of oral tissues containing dental implants
    • Piattelli A, Scarano A, Quaranta M. High-precision, cost-effective system for producing thin sections of oral tissues containing dental implants. Biomaterials 1997; 18:577-579.
    • (1997) Biomaterials , vol.18 , pp. 577-579
    • Piattelli, A.1    Scarano, A.2    Quaranta, M.3
  • 16
    • 0032151752 scopus 로고    scopus 로고
    • Assessment of implant stability as a prognostic determinant
    • Meredith N. Assessment of implant stability as a prognostic determinant. Int J Prosthodont 1998; 11:491-501.
    • (1998) Int J Prosthodont , vol.11 , pp. 491-501
    • Meredith, N.1
  • 17
    • 0031152189 scopus 로고    scopus 로고
    • The application of resonance frequency measurements to study the stability of titanium implants during healing in the rabbit tibia
    • Meredith N, Shagaldi F, Alleyne D, Sennerby L, Cawley P. The application of resonance frequency measurements to study the stability of titanium implants during healing in the rabbit tibia. Clin Oral Implants Res 1997; 8:234-243.
    • (1997) Clin Oral Implants Res , vol.8 , pp. 234-243
    • Meredith, N.1    Shagaldi, F.2    Alleyne, D.3    Sennerby, L.4    Cawley, P.5
  • 18
    • 0031152150 scopus 로고    scopus 로고
    • Resonance frequency measurements of implant stability in vivo. A cross sectional and longitudinal study of resonance frequency measurements on implants in the edentulous and partially dentate maxilla
    • Meredith N, Book K, Friberg B, Jemt T, Sennerby L. Resonance frequency measurements of implant stability in vivo. A cross sectional and longitudinal study of resonance frequency measurements on implants in the edentulous and partially dentate maxilla. Clin Oral Implants Res 1997; 8:226-233.
    • (1997) Clin Oral Implants Res , vol.8 , pp. 226-233
    • Meredith, N.1    Book, K.2    Friberg, B.3    Jemt, T.4    Sennerby, L.5
  • 19
    • 17444397440 scopus 로고    scopus 로고
    • Biological and biomechanical evaluation of bone remodelling and implant stability after using an osteotome technique
    • Buchter A, Kleinheinz J, Wiesmann HP, et al. Biological and biomechanical evaluation of bone remodelling and implant stability after using an osteotome technique. Clin Oral Implants Res 2005; 16:1-8.
    • (2005) Clin Oral Implants Res , vol.16 , pp. 1-8
    • Buchter, A.1    Kleinheinz, J.2    Wiesmann, H.P.3
  • 20
    • 26944498359 scopus 로고    scopus 로고
    • An evaluation of resonance frequency analysis for the determination of the primary stability of orthodontic palatal implants. A study in human cadavers
    • Gedrange T, Hietschold V, Mai R, Wolf P, Nicklish M, Harzer W. An evaluation of resonance frequency analysis for the determination of the primary stability of orthodontic palatal implants. A study in human cadavers. Clin Oral Implants Res 2005; 16:25-31.
    • (2005) Clin Oral Implants Res , vol.16 , pp. 25-31
    • Gedrange, T.1    Hietschold, V.2    Mai, R.3    Wolf, P.4    Nicklish, M.5    Harzer, W.6
  • 21
    • 0032086907 scopus 로고    scopus 로고
    • Stability assessment and histology of titanium implants placed simultaneously with autogenous onlay bone in the rabbit tibia
    • Rasmusson L, Meredith N, Kahnberg KE, Sennerby L. Stability assessment and histology of titanium implants placed simultaneously with autogenous onlay bone in the rabbit tibia. Int J Oral Maxillofac Surg 1998; 27:229-235.
    • (1998) Int J Oral Maxillofac Surg , vol.27 , pp. 229-235
    • Rasmusson, L.1    Meredith, N.2    Kahnberg, K.E.3    Sennerby, L.4
  • 22
    • 18744396684 scopus 로고    scopus 로고
    • Implant design and intraosseous stability of immediately placed implants: A human cadaver study
    • Accocaoglu M, Uysal S, Tekaemir I, Akca K, Cehreli MC. Implant design and intraosseous stability of immediately placed implants: a human cadaver study. Clin Oral Implants Res 2005; 16:202-209.
    • (2005) Clin Oral Implants Res , vol.16 , pp. 202-209
    • Accocaoglu, M.1    Uysal, S.2    Tekaemir, I.3    Akca, K.4    Cehreli, M.C.5


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.