Volumn 31, Issue 2, 2005, Pages 143-165

A review of calibration techniques for freehand 3-D ultrasound systems

Author keywords

3 D ultrasound; Freehand acquisition; Navigation; Ultrasound probe calibration

Indexed keywords

CALIBRATION; FEATURE EXTRACTION; IMAGE ANALYSIS; LEAST SQUARES APPROXIMATIONS; MATHEMATICAL TRANSFORMATIONS; MEDICAL APPLICATIONS; OPTIMIZATION; SENSORS; THREE DIMENSIONAL; ULTRASONIC DEVICES;

EID: 13644249894     PISSN: 03015629     EISSN: None     Source Type: Journal    
DOI: 10.1016/j.ultrasmedbio.2004.11.001     Document Type: Article
Times cited : (197)

References (110)
  • 2
    • 0034043228 scopus 로고    scopus 로고
    • The impact of sound speed errors on medical ultrasound imaging
    • M.E. Anderson, M.S. McKeag, G.E. Trahey The impact of sound speed errors on medical ultrasound imaging J Acoust Soc Am 107 2000 3540 3548
    • (2000) J Acoust Soc Am , vol.107 , pp. 3540-3548
    • Anderson, M.E.1    McKeag, M.S.2    Trahey, G.E.3
  • 5
    • 0035670075 scopus 로고    scopus 로고
    • Accuracy of an electromagnetic three-dimensional ultrasound system for carotid artery imaging
    • D.C. Barratt, A.H. Davies, A.D. Hughes Accuracy of an electromagnetic three-dimensional ultrasound system for carotid artery imaging Ultrasound Med Biol 27 2001 1421 1425
    • (2001) Ultrasound Med Biol , vol.27 , pp. 1421-1425
    • Barratt, D.C.1    Davies, A.H.2    Hughes, A.D.3
  • 6
    • 0034907381 scopus 로고    scopus 로고
    • Optimisation and evaluation of an electromagnetic tracking device for high-accuracy three-dimensional ultrasound imaging of the carotid arteries
    • D.C. Barratt, A.H. Davies, A.D. Hughes Optimisation and evaluation of an electromagnetic tracking device for high-accuracy three-dimensional ultrasound imaging of the carotid arteries Ultrasound Med Biol 27 2001 957 968
    • (2001) Ultrasound Med Biol , vol.27 , pp. 957-968
    • Barratt, D.C.1    Davies, A.H.2    Hughes, A.D.3
  • 7
    • 0030853139 scopus 로고    scopus 로고
    • Three-dimensional freehand ultrasound: Image reconstruction and volume analysis
    • C.D. Barry, C.P. Allott, N.W. John Three-dimensional freehand ultrasound Image reconstruction and volume analysis Ultrasound Med Biol 23 1997 1209 1224
    • (1997) Ultrasound Med Biol , vol.23 , pp. 1209-1224
    • Barry, C.D.1    Allott, C.P.2    John, N.W.3
  • 9
    • 0032971037 scopus 로고    scopus 로고
    • Dynamic three-dimensional freehand echocardiography using raw digital ultrasound data
    • S. Berg, H. Torp, D. Martens Dynamic three-dimensional freehand echocardiography using raw digital ultrasound data Ultrasound Med Biol 25 1999 745 753
    • (1999) Ultrasound Med Biol , vol.25 , pp. 745-753
    • Berg, S.1    Torp, H.2    Martens, D.3
  • 11
    • 0027469612 scopus 로고
    • Speed of sound in pure water as a function of temperature
    • N. Bilaniuk, G. Wong Speed of sound in pure water as a function of temperature J Acoust Soc Am 93 1993 1609 1612
    • (1993) J Acoust Soc Am , vol.93 , pp. 1609-1612
    • Bilaniuk, N.1    Wong, G.2
  • 12
    • 0031767122 scopus 로고    scopus 로고
    • Systematic distortions in magnetic position digitizers
    • W. Birkfellner, F. Watzinger, F. Wanschitz Systematic distortions in magnetic position digitizers Med Phys 25 1998 2242 2248
    • (1998) Med Phys , vol.25 , pp. 2242-2248
    • Birkfellner, W.1    Watzinger, F.2    Wanschitz, F.3
  • 14
    • 0041857843 scopus 로고    scopus 로고
    • A rapid calibration method for registration and 3D tracking of ultrasound images using spatial localizer
    • E.M. Boctor, A. Jain, M. Choti A rapid calibration method for registration and 3D tracking of ultrasound images using spatial localizer SPIE Proc 5035 2003 521 532
    • (2003) SPIE Proc , vol.5035 , pp. 521-532
    • Boctor, E.M.1    Jain, A.2    Choti, M.3
  • 15
    • 0035125985 scopus 로고    scopus 로고
    • Calibration of three-dimensional ultrasound images for image-guided radiation therapy
    • L.G. Bouchet, S.L. Meeks, G. Goodchild Calibration of three-dimensional ultrasound images for image-guided radiation therapy Phys Med Biol 46 2001 559 577
    • (2001) Phys Med Biol , vol.46 , pp. 559-577
    • Bouchet, L.G.1    Meeks, S.L.2    Goodchild, G.3
  • 16
    • 0036393252 scopus 로고    scopus 로고
    • Registration of 3D CT and ultrasound datasets of the spine using bone structures
    • B. Brendel, S. Winter, A. Rick Registration of 3D CT and ultrasound datasets of the spine using bone structures Comput Aided Surg 7 2002 146 155
    • (2002) Comput Aided Surg , vol.7 , pp. 146-155
    • Brendel, B.1    Winter, S.2    Rick, A.3
  • 17
    • 0018735228 scopus 로고
    • A stereotactic head frame for use with CT body scanners
    • R.A. Brown A stereotactic head frame for use with CT body scanners Invest Radiol 14 1979 300 304
    • (1979) Invest Radiol , vol.14 , pp. 300-304
    • Brown, R.A.1
  • 20
    • 0033727604 scopus 로고    scopus 로고
    • Design of a clinical free-hand 3D ultrasound system
    • J.C. Carr, J.L. Stallkamp, M.M. Fynes Design of a clinical free-hand 3D ultrasound system SPIE Proc 3982 2000 14 25
    • (2000) SPIE Proc , vol.3982 , pp. 14-25
    • Carr, J.C.1    Stallkamp, J.L.2    Fynes, M.M.3
  • 23
    • 0006324061 scopus 로고    scopus 로고
    • Integrated MR and ultrasound imaging for improved image guidance in neurosurgery
    • R.M. Comeau, A. Fenster, T.M. Peters Integrated MR and ultrasound imaging for improved image guidance in neurosurgery SPIE Proc 3338 1998 747 754
    • (1998) SPIE Proc , vol.3338 , pp. 747-754
    • Comeau, R.M.1    Fenster, A.2    Peters, T.M.3
  • 24
    • 0033995330 scopus 로고    scopus 로고
    • Intraoperative ultrasound for guidance and tissue shift correction in image-guided neurosurgery
    • R.M. Comeau, A.F. Sadikot, A. Fenster Intraoperative ultrasound for guidance and tissue shift correction in image-guided neurosurgery Med Phys 27 2000 787 800
    • (2000) Med Phys , vol.27 , pp. 787-800
    • Comeau, R.M.1    Sadikot, A.F.2    Fenster, A.3
  • 25
    • 0031171025 scopus 로고    scopus 로고
    • 3D reconstruction of prostate from ultrasound images
    • D. Crivianu-Gaita, F. Miclea, A. Gaspar 3D reconstruction of prostate from ultrasound images Int J Med Inf 45 1997 43 51
    • (1997) Int J Med Inf , vol.45 , pp. 43-51
    • Crivianu-Gaita, D.1    Miclea, F.2    Gaspar, A.3
  • 28
    • 0028597060 scopus 로고
    • 3D ultrasonic image feature localization based on magnetic scanhead tracking: In vitro calibration and validation
    • P.R. Detmer, G. Bashein, T. Hodges 3D ultrasonic image feature localization based on magnetic scanhead tracking In vitro calibration and validation Ultrasound Med Biol 20 1994 923 936
    • (1994) Ultrasound Med Biol , vol.20 , pp. 923-936
    • Detmer, P.R.1    Bashein, G.2    Hodges, T.3
  • 29
    • 0036007865 scopus 로고    scopus 로고
    • Automatic fusion of freehand endoscopic brain images to three-dimensional surfaces: Creating stereoscopic panoramas
    • D. Dey, D.G. Gobbi, P.J. Slomka Automatic fusion of freehand endoscopic brain images to three-dimensional surfaces Creating stereoscopic panoramas IEEE Trans Med Imaging 21 2002 23 30
    • (2002) IEEE Trans Med Imaging , vol.21 , pp. 23-30
    • Dey, D.1    Gobbi, D.G.2    Slomka, P.J.3
  • 30
    • 0036866882 scopus 로고    scopus 로고
    • The effect of speed of sound in ultrasound test objects on lateral resolution
    • N.J. Dudley, N.M. Gibson, M.J. Fleckney The effect of speed of sound in ultrasound test objects on lateral resolution Ultrasound Med Biol 28 2002 1561 1564
    • (2002) Ultrasound Med Biol , vol.28 , pp. 1561-1564
    • Dudley, N.J.1    Gibson, N.M.2    Fleckney, M.J.3
  • 31
    • 0030705773 scopus 로고    scopus 로고
    • Estimating 3-D rigid body transformations: A comparison of four major algorithms
    • D.W. Eggert, A. Lorusso, R.B. Fisher Estimating 3-D rigid body transformations A comparison of four major algorithms Machine Vision Applic 9 1997 272 290
    • (1997) Machine Vision Applic , vol.9 , pp. 272-290
    • Eggert, D.W.1    Lorusso, A.2    Fisher, R.B.3
  • 32
    • 0034578101 scopus 로고    scopus 로고
    • Three-dimensional ultrasound imaging
    • A. Fenster, D.B. Downey Three-dimensional ultrasound imaging Annu Rev Biomed Eng 2 2000 457 475
    • (2000) Annu Rev Biomed Eng , vol.2 , pp. 457-475
    • Fenster, A.1    Downey, D.B.2
  • 33
    • 0019574599 scopus 로고
    • Random sample consensus: A paradigm for model fitting with applications to image analysis and automated cartography
    • M.A. Fischler, R.C. Bolles Random sample consensus A paradigm for model fitting with applications to image analysis and automated cartography Comm ACM 24 1981 381 395
    • (1981) Comm ACM , vol.24 , pp. 381-395
    • Fischler, M.A.1    Bolles, R.C.2
  • 36
    • 0034281588 scopus 로고    scopus 로고
    • The effect of acoustic velocity on phantom measurements
    • A. Goldstein The effect of acoustic velocity on phantom measurements Ultrasound Med Biol 26 2000 1133 1143
    • (2000) Ultrasound Med Biol , vol.26 , pp. 1133-1143
    • Goldstein, A.1
  • 37
    • 0018000447 scopus 로고
    • Comprehensive compilation of empirical ultrasonic properties of mammalian tissues
    • S.A. Goss, R.L. Johnston, F. Dunn Comprehensive compilation of empirical ultrasonic properties of mammalian tissues J Acoust Soc Am 64 1978 423 457
    • (1978) J Acoust Soc Am , vol.64 , pp. 423-457
    • Goss, S.A.1    Johnston, R.L.2    Dunn, F.3
  • 38
    • 0033646692 scopus 로고    scopus 로고
    • SonoWand, an ultrasound-based neuronavigation system
    • A. Gronningsaeter, A. Kleven, S. Ommedal SonoWand, an ultrasound-based neuronavigation system Neurosurgery 47 2000 1373 1380
    • (2000) Neurosurgery , vol.47 , pp. 1373-1380
    • Gronningsaeter, A.1    Kleven, A.2    Ommedal, S.3
  • 39
    • 0003262321 scopus 로고    scopus 로고
    • Error analysis for a free-hand three-dimensional ultrasound system for neuronavigation
    • A. Hartov, S.D. Eisner, D.W. Roberts Error analysis for a free-hand three-dimensional ultrasound system for neuronavigation Neurosurg Focus 6 1999
    • (1999) Neurosurg Focus , vol.6
    • Hartov, A.1    Eisner, S.D.2    Roberts, D.W.3
  • 41
    • 0037326107 scopus 로고    scopus 로고
    • Quantification of prostate shrinkage after microwave thermotherapy: A comparison of calculated cell-kill versus 3D transrectal ultrasound planimetry
    • A.L. Hoffmann, M.P. Laguna, J.J. de la Rosette Quantification of prostate shrinkage after microwave thermotherapy A comparison of calculated cell-kill versus 3D transrectal ultrasound planimetry Eur Urol 43 2003 181 187
    • (2003) Eur Urol , vol.43 , pp. 181-187
    • Hoffmann, A.L.1    Laguna, M.P.2    De La Rosette, J.J.3
  • 43
    • 84938447945 scopus 로고
    • "direct search" solution of numerical and statistical problems
    • R. Hooke, T.A. Jeeves "Direct search" solution of numerical and statistical problems J Assoc Comput Mach 8 1961 212 229
    • (1961) J Assoc Comput Mach , vol.8 , pp. 212-229
    • Hooke, R.1    Jeeves, T.A.2
  • 44
    • 84975553202 scopus 로고
    • Closed-form solution of absolute orientation using unit quaternions
    • B.K.P. Horn Closed-form solution of absolute orientation using unit quaternions J Opt Soc Am A 4 1987 629 642
    • (1987) J Opt Soc Am a , vol.4 , pp. 629-642
    • Horn, B.K.P.1
  • 45
    • 84975562075 scopus 로고
    • Closed-form solution of absolute orientation using orthonormal matrices
    • B.K.P. Horn, H.M. Hilden, S. Negahdaripour Closed-form solution of absolute orientation using orthonormal matrices J Opt Soc Am A 5 1988 1127 1135
    • (1988) J Opt Soc Am a , vol.5 , pp. 1127-1135
    • Horn, B.K.P.1    Hilden, H.M.2    Negahdaripour, S.3
  • 46
    • 0029937261 scopus 로고    scopus 로고
    • Volume estimation from multiplanar 2D ultrasound images using a remote electromagnetic position and orientation sensor
    • S.W. Hughes, T.J. D'Arcy, D.J. Maxwell Volume estimation from multiplanar 2D ultrasound images using a remote electromagnetic position and orientation sensor Ultrasound Med Biol 22 1996 561 572
    • (1996) Ultrasound Med Biol , vol.22 , pp. 561-572
    • Hughes, S.W.1    D'Arcy, T.J.2    Maxwell, D.J.3
  • 49
    • 0028352215 scopus 로고
    • Three-dimensional US of the fetus. Work in progress
    • I.M. Kelly, J.E. Gardener, A.D. Brett Three-dimensional US of the fetus. Work in progress Radiology 192 1994 253 259
    • (1994) Radiology , vol.192 , pp. 253-259
    • Kelly, I.M.1    Gardener, J.E.2    Brett, A.D.3
  • 50
    • 0034043270 scopus 로고    scopus 로고
    • Comparative tracking error analysis of five different optical tracking systems
    • R. Khadem, C.C. Yeh, M. Sadeghi-Tehrani Comparative tracking error analysis of five different optical tracking systems Comput Aided Surg 5 2000 98 107
    • (2000) Comput Aided Surg , vol.5 , pp. 98-107
    • Khadem, R.1    Yeh, C.C.2    Sadeghi-Tehrani, M.3
  • 51
    • 0242630709 scopus 로고    scopus 로고
    • A survey of electromagnetic position tracker calibration techniques
    • V.V. Kindratenko A survey of electromagnetic position tracker calibration techniques Virtual Reality Res Dev Applic 5 2000 169 182
    • (2000) Virtual Reality Res Dev Applic , vol.5 , pp. 169-182
    • Kindratenko, V.V.1
  • 52
    • 0026065475 scopus 로고
    • Evaluation of in vitro measurement accuracy of a three-dimensional ultrasound scanner
    • D.L. King, D.L. King Jr, M.Y. Shao Evaluation of in vitro measurement accuracy of a three-dimensional ultrasound scanner J Ultrasound Med 10 1991 77 82
    • (1991) J Ultrasound Med , vol.10 , pp. 77-82
    • King, D.L.1    King Jr., D.L.2    Shao, M.Y.3
  • 53
    • 4344624671 scopus 로고    scopus 로고
    • On the development and comparative evaluation of an ultrasound B-mode probe calibration method
    • J. Kowal, C.A. Amstutz, M. Caversaccio On the development and comparative evaluation of an ultrasound B-mode probe calibration method Comput Aided Surg 8 2003 107 119
    • (2003) Comput Aided Surg , vol.8 , pp. 107-119
    • Kowal, J.1    Amstutz, C.A.2    Caversaccio, M.3
  • 55
    • 0032053964 scopus 로고    scopus 로고
    • System for quantitative three-dimensional echocardiography of the left ventricle based on a magnetic-field position and orientation sensing system
    • M.E. Legget, D.F. Leotta, E.L. Bolson System for quantitative three-dimensional echocardiography of the left ventricle based on a magnetic-field position and orientation sensing system IEEE Trans Biomed Eng 45 1998 494 504
    • (1998) IEEE Trans Biomed Eng , vol.45 , pp. 494-504
    • Legget, M.E.1    Leotta, D.F.2    Bolson, E.L.3
  • 56
    • 4143049309 scopus 로고    scopus 로고
    • An efficient calibration method for freehand 3-D ultrasound imaging systems
    • D.F. Leotta An efficient calibration method for freehand 3-D ultrasound imaging systems Ultrasound Med Biol 30 2004 999 1008
    • (2004) Ultrasound Med Biol , vol.30 , pp. 999-1008
    • Leotta, D.F.1
  • 57
    • 0029454361 scopus 로고
    • Three-dimensional ultrasound imaging using multiple magnetic tracking systems and miniature magnetic sensors
    • D.F. Leotta, P.R. Detmer, O.H. Gilja Three-dimensional ultrasound imaging using multiple magnetic tracking systems and miniature magnetic sensors IEEE Ultrason Sympos 1995 1415 1418
    • (1995) IEEE Ultrason Sympos , pp. 1415-1418
    • Leotta, D.F.1    Detmer, P.R.2    Gilja, O.H.3
  • 58
    • 0031006893 scopus 로고    scopus 로고
    • Performance of a miniature magnetic position sensor for three-dimensional ultrasound imaging
    • D.F. Leotta, P.R. Detmer, R.W. Martin Performance of a miniature magnetic position sensor for three-dimensional ultrasound imaging Ultrasound Med Biol 23 1997 597 609
    • (1997) Ultrasound Med Biol , vol.23 , pp. 597-609
    • Leotta, D.F.1    Detmer, P.R.2    Martin, R.W.3
  • 59
    • 0035985841 scopus 로고    scopus 로고
    • Tissue motion and elevational speckle decorrelation in freehand 3D ultrasound
    • P.C. Li, C.Y. Li, W.C. Yeh Tissue motion and elevational speckle decorrelation in freehand 3D ultrasound Ultrason Imaging 24 2002 1 12
    • (2002) Ultrason Imaging , vol.24 , pp. 1-12
    • Li, P.C.1    Li, C.Y.2    Yeh, W.C.3
  • 60
    • 0031816144 scopus 로고    scopus 로고
    • Progress in two-dimensional arrays for real-time volumetric imaging
    • E.D. Light, R.E. Davidsen, J.O. Fiering Progress in two-dimensional arrays for real-time volumetric imaging Ultrason Imaging 20 1998 1 15
    • (1998) Ultrason Imaging , vol.20 , pp. 1-15
    • Light, E.D.1    Davidsen, R.E.2    Fiering, J.O.3
  • 62
    • 0242364812 scopus 로고    scopus 로고
    • A robust and automatic method for evaluating accuracy in 3-D ultrasound-based navigation
    • F. Lindseth, J. Bang, T. Langø A robust and automatic method for evaluating accuracy in 3-D ultrasound-based navigation Ultrasound Med Biol 29 2003 1439 1452
    • (2003) Ultrasound Med Biol , vol.29 , pp. 1439-1452
    • Lindseth, F.1    Bang, J.2    Langø, T.3
  • 63
    • 2442557217 scopus 로고    scopus 로고
    • Multimodal image fusion in ultrasound-based neuronavigation: Improving overview and interpretation by integrating preoperative MRI with intraoperative 3D ultrasound
    • F. Lindseth, J.H. Kaspersen, S. Ommedal Multimodal image fusion in ultrasound-based neuronavigation Improving overview and interpretation by integrating preoperative MRI with intraoperative 3D ultrasound Comput Aided Surg 8 2003 49 69
    • (2003) Comput Aided Surg , vol.8 , pp. 49-69
    • Lindseth, F.1    Kaspersen, J.H.2    Ommedal, S.3
  • 64
    • 0036449406 scopus 로고    scopus 로고
    • Accuracy evaluation of a 3D ultrasound-based neuronavigation system
    • F. Lindseth, T. Langø, J. Bang Accuracy evaluation of a 3D ultrasound-based neuronavigation system Comput Aided Surg 7 2002 197 222
    • (2002) Comput Aided Surg , vol.7 , pp. 197-222
    • Lindseth, F.1    Langø, T.2    Bang, J.3
  • 66
    • 0008580784 scopus 로고    scopus 로고
    • A new calibration method in 3D ultrasonic imaging system
    • J. Liu, X. Gao, Z. Zhang A new calibration method in 3D ultrasonic imaging system 20th Annual Conference IEEE Eng Med Biol 20 1998 839 841
    • (1998) IEEE Eng Med Biol , vol.20 , pp. 839-841
    • Liu, J.1    Gao, X.2    Zhang, Z.3
  • 68
    • 0035045307 scopus 로고    scopus 로고
    • Measurement of the speed of sound in ethanol/water mixtures
    • K. Martin, D. Spinks Measurement of the speed of sound in ethanol/water mixtures Ultrasound Med Biol 27 2001 289 291
    • (2001) Ultrasound Med Biol , vol.27 , pp. 289-291
    • Martin, K.1    Spinks, D.2
  • 69
    • 0034056419 scopus 로고    scopus 로고
    • Reconstruction and visualization of irregularly sampled three- and four-dimensional ultrasound data for cerebrovascular applications
    • S. Meairs, J. Beyer, M. Hennerici Reconstruction and visualization of irregularly sampled three- and four-dimensional ultrasound data for cerebrovascular applications Ultrasound Med Biol 26 2000 263 272
    • (2000) Ultrasound Med Biol , vol.26 , pp. 263-272
    • Meairs, S.1    Beyer, J.2    Hennerici, M.3
  • 72
    • 0035661105 scopus 로고    scopus 로고
    • Beam calibration without a phantom for creating a 3-D freehand ultrasound system
    • D.M. Muratore, R.L. Galloway Beam calibration without a phantom for creating a 3-D freehand ultrasound system Ultrasound Med Biol 27 2001 1557 1566
    • (2001) Ultrasound Med Biol , vol.27 , pp. 1557-1566
    • Muratore, D.M.1    Galloway, R.L.2
  • 74
    • 0030785991 scopus 로고    scopus 로고
    • Interactive acquisition, analysis and visualization of sonographic volume data
    • T.R. Nelson, D.H. Pretorius Interactive acquisition, analysis and visualization of sonographic volume data Int J Imag Syst Tech 8 1997 26 37
    • (1997) Int J Imag Syst Tech , vol.8 , pp. 26-37
    • Nelson, T.R.1    Pretorius, D.H.2
  • 76
    • 0032402751 scopus 로고    scopus 로고
    • Calibration and validation of free-hand 3D ultrasound systems based on DC magnetic tracking
    • N. Pagoulatos, W.S. Edwards, D.R. Haynor Calibration and validation of free-hand 3D ultrasound systems based on DC magnetic tracking SPIE Proc 3335 1998 59 71
    • (1998) SPIE Proc , vol.3335 , pp. 59-71
    • Pagoulatos, N.1    Edwards, W.S.2    Haynor, D.R.3
  • 77
    • 0033313805 scopus 로고    scopus 로고
    • Fast calibration for 3D ultrasound imaging and multimodality image
    • N. Pagoulatos, D.R. Haynor, Y. Kim Fast calibration for 3D ultrasound imaging and multimodality image First Joint BMES/EMBS Conference (Serving Humanity, Advancing Technology), Atlanta, GA, 1999 Proc IEEE 1999 1065
    • (1999) Proc IEEE , pp. 1065
    • Pagoulatos, N.1    Haynor, D.R.2    Kim, Y.3
  • 78
    • 0034804115 scopus 로고    scopus 로고
    • A fast calibration method for 3-D tracking of ultrasound images using a spatial localizer
    • N. Pagoulatos, D.R. Haynor, Y. Kim A fast calibration method for 3-D tracking of ultrasound images using a spatial localizer Ultrasound Med Biol 27 2001 1219 1229
    • (2001) Ultrasound Med Biol , vol.27 , pp. 1219-1229
    • Pagoulatos, N.1    Haynor, D.R.2    Kim, Y.3
  • 79
    • 0037290154 scopus 로고    scopus 로고
    • Tracking brain deformation in time sequences of 3D US images
    • X. Pennec, P. Cachier, N. Ayache Tracking brain deformation in time sequences of 3D US images Pattern Recog Lett 24 2003 801 813
    • (2003) Pattern Recog Lett , vol.24 , pp. 801-813
    • Pennec, X.1    Cachier, P.2    Ayache, N.3
  • 80
    • 84956858951 scopus 로고
    • Using a 3D position sensor for registration of SPECT and US images of the kidney
    • O. Péria, L. Chevalier, A. François-Joubert Using a 3D position sensor for registration of SPECT and US images of the kidney Lecture Notes Comput Sci 905 1995 23 29
    • (1995) Lecture Notes Comput Sci , vol.905 , pp. 23-29
    • Péria, O.1    Chevalier, L.2    François-Joubert, A.3
  • 81
    • 0033138697 scopus 로고    scopus 로고
    • Stradx: Real-time acquisition and visualization of freehand three-dimensional ultrasound
    • R.W. Prager, A. Gee, L. Berman Stradx Real-time acquisition and visualization of freehand three-dimensional ultrasound Med Image Anal 3 1998 129 140
    • (1998) Med Image Anal , vol.3 , pp. 129-140
    • Prager, R.W.1    Gee, A.2    Berman, L.3
  • 82
    • 0037361601 scopus 로고    scopus 로고
    • Sensorless freehand 3-D ultrasound using regression of the echo intensity
    • R.W. Prager, A.H. Gee, G.M. Treece Sensorless freehand 3-D ultrasound using regression of the echo intensity Ultrasound Med Biol 29 2003 437 446
    • (2003) Ultrasound Med Biol , vol.29 , pp. 437-446
    • Prager, R.W.1    Gee, A.H.2    Treece, G.M.3
  • 85
    • 0029422063 scopus 로고
    • Comparison of relative accuracy between a mechanical and an optical position tracker for image-guided neurosurgery
    • R. Rohling, P. Munger, J.M. Hollerbach Comparison of relative accuracy between a mechanical and an optical position tracker for image-guided neurosurgery J Image Guided Surg 1 1995 30 34
    • (1995) J Image Guided Surg , vol.1 , pp. 30-34
    • Rohling, R.1    Munger, P.2    Hollerbach, J.M.3
  • 86
    • 0031107438 scopus 로고    scopus 로고
    • Three-dimensional spatial compounding of ultrasound images
    • R.N. Rohling, A.H. Gee, L. Berman Three-dimensional spatial compounding of ultrasound images Med Image Anal 1 1997 177 193
    • (1997) Med Image Anal , vol.1 , pp. 177-193
    • Rohling, R.N.1    Gee, A.H.2    Berman, L.3
  • 88
    • 13644257332 scopus 로고    scopus 로고
    • Quality assessment of electromagnetic localizers in the context of 3D ultrasound
    • F. Rousseau, C. Barillot Quality assessment of electromagnetic localizers in the context of 3D ultrasound INRIA 2002
    • (2002) INRIA
    • Rousseau, F.1    Barillot, C.2
  • 92
    • 0032173358 scopus 로고    scopus 로고
    • Image guidance of breast cancer surgery using 3-D ultrasound images and augmented reality visualization
    • Y. Sato, M. Nakamoto, Y. Tamaki Image guidance of breast cancer surgery using 3-D ultrasound images and augmented reality visualization IEEE Trans Med Imaging 17 1998 681 693
    • (1998) IEEE Trans Med Imaging , vol.17 , pp. 681-693
    • Sato, Y.1    Nakamoto, M.2    Tamaki, Y.3
  • 93
    • 0035201925 scopus 로고    scopus 로고
    • Accuracy evaluation of a CAS system: Laboratory protocol and results with 6D localizers, and clinical experiences in otorhinolaryngology
    • S. Schmerber, F. Chassat Accuracy evaluation of a CAS system Laboratory protocol and results with 6D localizers, and clinical experiences in otorhinolaryngology Comput Aided Surg 6 2001 1 13
    • (2001) Comput Aided Surg , vol.6 , pp. 1-13
    • Schmerber, S.1    Chassat, F.2
  • 94
    • 0022095776 scopus 로고
    • Animating rotations with quaternion curves
    • K. Shoemake Animating rotations with quaternion curves ACM SIGGRAPH Comput Graph 19 1985 245 254
    • (1985) ACM SIGGRAPH Comput Graph , vol.19 , pp. 245-254
    • Shoemake, K.1
  • 95
    • 0028714281 scopus 로고
    • Case study: Observing a volume rendered fetus within a pregnant patient
    • IEEE Computer Society Press Los Alamitos, CA
    • A. State, D.T. Chen, A. Brandt Case study Observing a volume rendered fetus within a pregnant patient Proceedings of IEEE Visualization 1994 IEEE Computer Society Press Los Alamitos, CA 364 368
    • (1994) Proceedings of IEEE Visualization , pp. 364-368
    • State, A.1    Chen, D.T.2    Brandt, A.3
  • 97
  • 99
    • 0031728262 scopus 로고    scopus 로고
    • Automated three-dimensional US frame positioning computed from elevational speckle decorrelation
    • T.A. Tuthill, J.F. Krucker, J.B. Fowlkes Automated three-dimensional US frame positioning computed from elevational speckle decorrelation Radiology 209 1998 575 582
    • (1998) Radiology , vol.209 , pp. 575-582
    • Tuthill, T.A.1    Krucker, J.F.2    Fowlkes, J.B.3
  • 100
    • 0026137964 scopus 로고
    • Least-squares estimation of transformation parameters between two points patterns
    • S. Umeyama Least-squares estimation of transformation parameters between two points patterns IEEE Trans Pattern Anal Machine Intell 13 1991 376 380
    • (1991) IEEE Trans Pattern Anal Machine Intell , vol.13 , pp. 376-380
    • Umeyama, S.1
  • 101
    • 0036556305 scopus 로고    scopus 로고
    • Neuronavigation by intraoperative three-dimensional ultrasound: Initial experience during brain tumor resection
    • G. Unsgaard, S. Ommedal, T. Muller Neuronavigation by intraoperative three-dimensional ultrasound Initial experience during brain tumor resection Neurosurgery 50 2002 804 812
    • (2002) Neurosurgery , vol.50 , pp. 804-812
    • Unsgaard, G.1    Ommedal, S.2    Muller, T.3
  • 102
    • 0026255934 scopus 로고
    • Estimating 3-D location parameters using dual number quaternions
    • M.W. Walker, L. Shao, R.A. Volz Estimating 3-D location parameters using dual number quaternions CVGIP Image Understanding 54 1991 358 367
    • (1991) CVGIP: Image Understanding , vol.54 , pp. 358-367
    • Walker, M.W.1    Shao, L.2    Volz, R.A.3
  • 106
    • 0034579599 scopus 로고    scopus 로고
    • A real-time freehand 3D ultrasound system for image-guided surgery
    • N. Welch, J.A. Johnson, M.R. Bax A real-time freehand 3D ultrasound system for image-guided surgery Ultrasonic Symposium 2 2000 1601 1604
    • (2000) Ultrasonic Symposium , vol.2 , pp. 1601-1604
    • Welch, N.1    Johnson, J.A.2    Bax, M.R.3
  • 107
    • 2542476140 scopus 로고    scopus 로고
    • Designing optically tracked instruments for image-guided surgery
    • J.B. West, C.R. Maurer Jr Designing optically tracked instruments for image-guided surgery IEEE Trans Med Imaging 23 2004 533 545
    • (2004) IEEE Trans Med Imaging , vol.23 , pp. 533-545
    • West, J.B.1    Maurer Jr., C.R.2
  • 108
    • 5644267915 scopus 로고    scopus 로고
    • Accuracy assessment and interpretation for optical tracking systems
    • A.D. Wiles, D.G. Thompson, D.D. Frantz Accuracy assessment and interpretation for optical tracking systems SPIE Proc 2004 5367
    • (2004) SPIE Proc , pp. 5367
    • Wiles, A.D.1    Thompson, D.G.2    Frantz, D.D.3
  • 109
    • 0036448877 scopus 로고    scopus 로고
    • Direct surface extraction from 3D freehand ultrasound images
    • Y. Zhang, R. Rohling, D.K. Pai Direct surface extraction from 3D freehand ultrasound images IEEE Visualization 2002 45 52
    • (2002) IEEE Visualization , pp. 45-52
    • Zhang, Y.1    Rohling, R.2    Pai, D.K.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.