메뉴 건너뛰기




Volumn 20, Issue 1, 2005, Pages 22-33

Haemodialysis with on-line monitoring equipment: Tools or toys?

Author keywords

Bio feedback; Blood volume control; Blood volume monitoring; Conductivity; Dialysis dose; Haemodialysis; Hypotension; Ionic dialysance; On line monitoring; Sodium balance; Thermal energy balance; Urea monitors; Vascular access monitoring

Indexed keywords

ARTICLE; BLOOD VOLUME; BODY TEMPERATURE; CORRELATION ANALYSIS; DIAGNOSTIC ACCURACY; DIALYSATE; FEEDBACK SYSTEM; HEMODIALYSIS; HEMODYNAMIC MONITORING; HUMAN; HYPOTENSION; ION CONDUCTANCE; KIDNEY BLOOD FLOW; MEDICAL INSTRUMENTATION; ONLINE MONITORING; PATIENT MONITORING; PRIORITY JOURNAL; SENSITIVITY AND SPECIFICITY; SODIUM BALANCE; STATISTICAL ANALYSIS; STATISTICAL SIGNIFICANCE; ULTRAFILTRATION; VALIDATION PROCESS; VASCULAR ACCESS;

EID: 13444274884     PISSN: 09310509     EISSN: None     Source Type: Journal    
DOI: 10.1093/ndt/gfh555     Document Type: Article
Times cited : (114)

References (52)
  • 3
    • 0033824757 scopus 로고    scopus 로고
    • Improving the outcome of dialysis - Opinion vs scientific evidence
    • Ledebo I, Lameire N, Charra B et al. Improving the outcome of dialysis - opinion vs scientific evidence. Nephrol Dial Transplant 2000; 15: 1310-1316
    • (2000) Nephrol. Dial. Transplant. , vol.15 , pp. 1310-1316
    • Ledebo, I.1    Lameire, N.2    Charra, B.3
  • 4
    • 0002823792 scopus 로고    scopus 로고
    • Changes in the clinical condition of haemodialysis patients
    • Locatelli F, Manzoni C, Del Vecchio L, Di Filippo S. Changes in the clinical condition of haemodialysis patients. J Nephrol 1999; 12 [Suppl 2]: 82-91
    • (1999) J. Nephrol. , vol.12 , Issue.SUPPL. 2 , pp. 82-91
    • Locatelli, F.1    Manzoni, C.2    Del Vecchio, L.3    Di Filippo, S.4
  • 5
    • 0029562035 scopus 로고
    • Monitoring sodium removal and delivered dialysis by conductivity
    • Locatelli F, Di Filippo S, Manzoni C et al. Monitoring sodium removal and delivered dialysis by conductivity. Int J Artif Organs 1995; 18: 716-721
    • (1995) Int. J. Artif. Organs , vol.18 , pp. 716-721
    • Locatelli, F.1    Di Filippo, S.2    Manzoni, C.3
  • 6
    • 0033154894 scopus 로고    scopus 로고
    • The role of technology in hemodialysis
    • Ronco C, Ghezzi PM, La Greca G. The role of technology in hemodialysis. J Nephrol 1999; 12 [Suppl 2]: S68-S81
    • (1999) J. Nephrol. , vol.12 , Issue.SUPPL. 2
    • Ronco, C.1    Ghezzi, P.M.2    La Greca, G.3
  • 7
    • 0002315488 scopus 로고    scopus 로고
    • Functional organization of the human body and control of the internal environment
    • 9th edn. WB Saunders Philadelphia
    • Guyton AC, Hall JE. Functional organization of the human body and control of the internal environment. In: Textbook of Medical Physiology, 9th edn. WB Saunders, Philadelphia; 1996: 3-9
    • (1996) Textbook of Medical Physiology , pp. 3-9
    • Guyton, A.C.1    Hall, J.E.2
  • 9
    • 0014721985 scopus 로고
    • Blood volume changes and hypotension during hemodialysis
    • Kim K, Neff M, Kohen B et al. Blood volume changes and hypotension during hemodialysis. ASAIO Trans 1970; 16: 508-513
    • (1970) ASAIO Trans. , vol.16 , pp. 508-513
    • Kim, K.1    Neff, M.2    Kohen, B.3
  • 10
    • 0026078098 scopus 로고
    • Dialysis hypotension. A hemodynamic analysis
    • Daugirdas JT. Dialysis hypotension. A hemodynamic analysis. Kidney Int 1991; 39: 233-246
    • (1991) Kidney Int. , vol.39 , pp. 233-246
    • Daugirdas, J.T.1
  • 11
    • 0036895459 scopus 로고    scopus 로고
    • The role of blood volume reduction in the genesis of intradialytic hypotension
    • Andrulli S, Colzani S, Mascia F et al. The role of blood volume reduction in the genesis of intradialytic hypotension. Am J Kidney Dis 2002; 40: 1244-1254
    • (2002) Am. J. Kidney Dis. , vol.40 , pp. 1244-1254
    • Andrulli, S.1    Colzani, S.2    Mascia, F.3
  • 13
    • 18744368078 scopus 로고    scopus 로고
    • Blood volume controlled hemodialysis in hypotension-prone patients: A randomized, multicenter controlled trial
    • Santoro A, Mancini E, Basile C et al. Blood volume controlled hemodialysis in hypotension-prone patients: a randomized, multicenter controlled trial. Kidney Int 2002; 62: 1034-1045
    • (2002) Kidney Int. , vol.62 , pp. 1034-1045
    • Santoro, A.1    Mancini, E.2    Basile, C.3
  • 14
    • 0035145183 scopus 로고    scopus 로고
    • Efficacy and safety of haemodialysis treatment with the Hemocontrol biofeedback system: A prospective medium-term study
    • Basile C, Giordano R, Vernaglione L et al. Efficacy and safety of haemodialysis treatment with the Hemocontrol biofeedback system: a prospective medium-term study. Nephrol Dial Transplant 2001; 16: 328-334
    • (2001) Nephrol. Dial. Transplant. , vol.16 , pp. 328-334
    • Basile, C.1    Giordano, R.2    Vernaglione, L.3
  • 15
    • 0033864894 scopus 로고    scopus 로고
    • Impact of biofeedback-induced cardiovascular stability on hemodialysis tolerance and efficiency
    • Ronco C, Brendolan A, Milan M, Rodeghiero MP, Zanella M, La Greca G. Impact of biofeedback-induced cardiovascular stability on hemodialysis tolerance and efficiency. Kidney Int 2000; 58: 800-808
    • (2000) Kidney Int. , vol.58 , pp. 800-808
    • Ronco, C.1    Brendolan, A.2    Milan, M.3    Rodeghiero, M.P.4    Zanella, M.5    La Greca, G.6
  • 16
    • 0027483770 scopus 로고
    • A model for non-invasive estimation of in vivo dialyzer performances and patient's conductivity during hemodialysis
    • Petitclerc T, Goux N, Reynier AL, Bene B. A model for non-invasive estimation of in vivo dialyzer performances and patient's conductivity during hemodialysis. Int J Artif Organs 1993; 16: 585-91
    • (1993) Int. J. Artif. Organs , vol.16 , pp. 585-591
    • Petitclerc, T.1    Goux, N.2    Reynier, A.L.3    Bene, B.4
  • 17
    • 0027227636 scopus 로고
    • Automatic non-invasive intradialytic clearance measurements
    • Polaschegg HD. Automatic non-invasive intradialytic clearance measurements. Int J Artif Organs 1993; 16: 185-191
    • (1993) Int. J. Artif. Organs , vol.16 , pp. 185-191
    • Polaschegg, H.D.1
  • 18
    • 0027483770 scopus 로고
    • A model for non-invasive estimation of in-vivo dialyzer performances and patient's conductivity during hemodialysis
    • Petitclerc T, Goux N, Reynier AL, Béne B. A model for non-invasive estimation of in-vivo dialyzer performances and patient's conductivity during hemodialysis. Int J Artif Organs 1993; 16: 585-591
    • (1993) Int. J. Artif. Organs , vol.16 , pp. 585-591
    • Petitclerc, T.1    Goux, N.2    Reynier, A.L.3    Béne, B.4
  • 19
    • 0028915014 scopus 로고
    • Non-invasive monitoring of effective dialysis dose delivered to the haemodialysis patient
    • Petitclerc T, Ben B, Jacobs C, Jaudon MC, Goux N. Non-invasive monitoring of effective dialysis dose delivered to the haemodialysis patient. Nephrol Dial Transplant 1995; 10: 212-216
    • (1995) Nephrol. Dial. Transplant. , vol.10 , pp. 212-216
    • Petitclerc, T.1    Ben, B.2    Jacobs, C.3    Jaudon, M.C.4    Goux, N.5
  • 20
    • 0029958088 scopus 로고    scopus 로고
    • Ionic dialysance as a method for the on-line monitoring of delivered dialysis without blood sampling
    • Manzoni C, Di Filippo S, Corti M, Locatelli F. Ionic dialysance as a method for the on-line monitoring of delivered dialysis without blood sampling. Nephrol Dial Transplant 1996; 11: 2023-2030
    • (1996) Nephrol. Dial. Transplant. , vol.11 , pp. 2023-2030
    • Manzoni, C.1    Di Filippo, S.2    Corti, M.3    Locatelli, F.4
  • 22
    • 0035137881 scopus 로고    scopus 로고
    • How to determine ionic dialysance for the online assessment of delivered dialysis dose
    • Di Filippo S, Manzoni C, Andrulli S et al. How to determine ionic dialysance for the online assessment of delivered dialysis dose. Kidney Int 2001; 59: 774-782
    • (2001) Kidney Int. , vol.59 , pp. 774-782
    • Di Filippo, S.1    Manzoni, C.2    Andrulli, S.3
  • 23
  • 24
    • 0036146385 scopus 로고    scopus 로고
    • Ionic dialysance vs urea clearance in the absence of cardiopulmonary recirculation
    • Mercadal L, Du Montcel ST, Jaudon MC et al. Ionic dialysance vs urea clearance in the absence of cardiopulmonary recirculation. Nephrol Dial Transplant 2002; 17: 106-111
    • (2002) Nephrol. Dial. Transplant. , vol.17 , pp. 106-111
    • Mercadal, L.1    Du Montcel, S.T.2    Jaudon, M.C.3
  • 26
    • 0030337728 scopus 로고    scopus 로고
    • Determining the adequacy of sodium balance in hemodialysis using a kinetic model
    • Di Filippo S, Corti M, Andrulli S, Manzoni C, Locatelli F. Determining the adequacy of sodium balance in hemodialysis using a kinetic model. Blood Purif 1996; 14: 431-436
    • (1996) Blood Purif. , vol.14 , pp. 431-436
    • Di Filippo, S.1    Corti, M.2    Andrulli, S.3    Manzoni, C.4    Locatelli, F.5
  • 28
    • 15444340552 scopus 로고    scopus 로고
    • The effect of on-line conductivity plasma ultrafiltrate kinetic modeling on cardiovascular stability of hemodialysis patients
    • Locatelli F, Andrulli S, Di Filippo S et al. The effect of on-line conductivity plasma ultrafiltrate kinetic modeling on cardiovascular stability of hemodialysis patients. Kidney Int 1998; 53: 1052-1060
    • (1998) Kidney Int. , vol.53 , pp. 1052-1060
    • Locatelli, F.1    Andrulli, S.2    Di Filippo, S.3
  • 29
    • 8044260279 scopus 로고    scopus 로고
    • Optimization of sodium removal in paired filtration dialysis by single pool sodium and conductivity kinetics models
    • Di Filippo S, Corti M, Andrulli S, Pontoriero G, Manzoni C, Locatelli F. Optimization of sodium removal in paired filtration dialysis by single pool sodium and conductivity kinetics models. Blood Purif 1997; 15: 34-44
    • (1997) Blood Purif. , vol.15 , pp. 34-44
    • Di Filippo, S.1    Corti, M.2    Andrulli, S.3    Pontoriero, G.4    Manzoni, C.5    Locatelli, F.6
  • 30
    • 0032777974 scopus 로고    scopus 로고
    • Measurement of blood access flow rate during hemodialysis from conductivity dialysance
    • Gotch FA, Rosemary B, Panlilio F, Folden T. Measurement of blood access flow rate during hemodialysis from conductivity dialysance. ASAIO J 1999; 45: 139-146
    • (1999) ASAIO J. , vol.45 , pp. 139-146
    • Gotch, F.A.1    Rosemary, B.2    Panlilio, F.3    Folden, T.4
  • 31
    • 0032842362 scopus 로고    scopus 로고
    • Determination of access blood flow from ionic dialysance: Theory and validation
    • Mercadal L, Hamani A, Béne B, Petitclerc T. Determination of access blood flow from ionic dialysance: theory and validation. Kidney Int 1999; 56: 1560-1565
    • (1999) Kidney Int. , vol.56 , pp. 1560-1565
    • Mercadal, L.1    Hamani, A.2    Béne, B.3    Petitclerc, T.4
  • 32
    • 4243519088 scopus 로고    scopus 로고
    • Comparison of the dialysance method and the urea and creatinine methods with the ultrasound dilution technique for the determination of graft access blood flow
    • Colzani S, La Milia V, Andrulli S, Bacchini G, Di Filippo S, Locatelli F. Comparison of the dialysance method and the urea and creatinine methods with the ultrasound dilution technique for the determination of graft access blood flow. Nephrol Dial Transplant 2001; 16: A160
    • (2001) Nephrol. Dial. Transplant. , vol.16
    • Colzani, S.1    La Milia, V.2    Andrulli, S.3    Bacchini, G.4    Di Filippo, S.5    Locatelli, F.6
  • 33
    • 18444373908 scopus 로고    scopus 로고
    • Detection of vascular access stenosis by measurement of access blood flow from ionic dialysance
    • Mercadal L, Challier E, Cluzel P et al. Detection of vascular access stenosis by measurement of access blood flow from ionic dialysance. Blood Purif 2002; 20: 177-181
    • (2002) Blood Purif. , vol.20 , pp. 177-181
    • Mercadal, L.1    Challier, E.2    Cluzel, P.3
  • 34
    • 0035221369 scopus 로고    scopus 로고
    • NKF-K/DOQI Clinical Practice Guidelines for Hemodialysis Adequacy: Update 2000
    • NKF-K/DOQI Clinical Practice Guidelines for Hemodialysis Adequacy: update 2000. Am J Kidney Dis 2001; 37 [Suppl 1]: S7-S64
    • (2001) Am. J. Kidney Dis. , vol.37 , Issue.SUPPL. 1
  • 35
    • 0036036790 scopus 로고    scopus 로고
    • Section II. Haemodialysis adequacy
    • European Best Practice Guidelines Expert Group on Hemodialysis, European Renal Association
    • European Best Practice Guidelines Expert Group on Hemodialysis, European Renal Association. Section II. Haemodialysis adequacy. Nephrol Dial Transplant 2002; 17 [Suppl 7]: 16-31
    • (2002) Nephrol. Dial. Transplant. , vol.17 , Issue.SUPPL. 7 , pp. 16-31
  • 36
    • 0031839030 scopus 로고    scopus 로고
    • A new biosensor for continuous monitoring of the spent dialysate urea level in standard hemodialysis
    • Calzavara P, Calconi G, Da Rin G, Canini E, Paolini F. A new biosensor for continuous monitoring of the spent dialysate urea level in standard hemodialysis. Int J Artif Organs 1998; 21: 147-150
    • (1998) Int. J. Artif. Organs , vol.21 , pp. 147-150
    • Calzavara, P.1    Calconi, G.2    Da Rin, G.3    Canini, E.4    Paolini, F.5
  • 37
    • 8544219678 scopus 로고    scopus 로고
    • Precise quantification of dialysis using continuous sampling of spent dialysate and total dialysate volume measurement
    • Argiles A, Ficheux A, Thomas M et al. Precise quantification of dialysis using continuous sampling of spent dialysate and total dialysate volume measurement. Kidney Int 1997; 52: 530-537
    • (1997) Kidney Int. , vol.52 , pp. 530-537
    • Argiles, A.1    Ficheux, A.2    Thomas, M.3
  • 38
    • 0344625409 scopus 로고    scopus 로고
    • On-line dialysis quantification in acutely ill patients: Preliminary clinical experience with a multipurpose urea sensor monitoring device
    • Canaud B, Bosc JY, Leblanc M et al. On-line dialysis quantification in acutely ill patients: preliminary clinical experience with a multipurpose urea sensor monitoring device. ASAIO J 1998; 44: 184-190
    • (1998) ASAIO J. , vol.44 , pp. 184-190
    • Canaud, B.1    Bosc, J.Y.2    Leblanc, M.3
  • 39
    • 0029869495 scopus 로고    scopus 로고
    • Multicenter clinical validation of an on-line monitor of dialysis adequacy
    • Depner TA, Keshaviah PR, Ebben JP et al. Multicenter clinical validation of an on-line monitor of dialysis adequacy. J Am Soc Nephrol 1996; 7: 464-471
    • (1996) J. Am. Soc. Nephrol. , vol.7 , pp. 464-471
    • Depner, T.A.1    Keshaviah, P.R.2    Ebben, J.P.3
  • 40
    • 0031052571 scopus 로고    scopus 로고
    • Urea rebound and delivered Kt/V determination with a continuous urea sensor
    • Garred LJ, Canaud B, Bosc JY, Tetta C. Urea rebound and delivered Kt/V determination with a continuous urea sensor. Nephrol Dial Transplant 1997; 12: 535-542
    • (1997) Nephrol. Dial. Transplant. , vol.12 , pp. 535-542
    • Garred, L.J.1    Canaud, B.2    Bosc, J.Y.3    Tetta, C.4
  • 41
    • 0028884195 scopus 로고
    • Biochemical aspects and clinical perspectives of continuous urea monitoring in plasma ultrafiltrate. Preliminary results of a multicenter study
    • Colasanti G, Arrigo G, Santoro A et al. Biochemical aspects and clinical perspectives of continuous urea monitoring in plasma ultrafiltrate. Preliminary results of a multicenter study. Int J Artif Organs 1995; 18: 544-547
    • (1995) Int. J. Artif. Organs , vol.18 , pp. 544-547
    • Colasanti, G.1    Arrigo, G.2    Santoro, A.3
  • 42
  • 43
    • 0030964205 scopus 로고    scopus 로고
    • A simple and accurate method to determine equilibrated post-dialysis urea concentration
    • Canaud B, Bosc JY, Leblanc M et al. A simple and accurate method to determine equilibrated post-dialysis urea concentration. Kidney Int 1997; 51: 2000-2005
    • (1997) Kidney Int. , vol.51 , pp. 2000-2005
    • Canaud, B.1    Bosc, J.Y.2    Leblanc, M.3
  • 44
    • 0033865301 scopus 로고    scopus 로고
    • Urea as a marker of adequacy in hemodialysis: Lesson from in vivo urea dynamics monitoring
    • Canaud B, Bosc JY, Cabrol L et al. Urea as a marker of adequacy in hemodialysis: lesson from in vivo urea dynamics monitoring. Kidney Int 2000 [Suppl 76]: S28-S40
    • (2000) Kidney Int. , Issue.SUPPL. 76
    • Canaud, B.1    Bosc, J.Y.2    Cabrol, L.3
  • 45
    • 0038079282 scopus 로고    scopus 로고
    • Estimation of delivered dialysis dose by on-line monitoring of the ultraviolet absorbance in the spent dialysate
    • Uhlin F, Fridolin I, Lindberg LG, Magnusson M. Estimation of delivered dialysis dose by on-line monitoring of the ultraviolet absorbance in the spent dialysate. Am J Kidney Dis 2003; 41: 1026-1036
    • (2003) Am. J. Kidney Dis. , vol.41 , pp. 1026-1036
    • Uhlin, F.1    Fridolin, I.2    Lindberg, L.G.3    Magnusson, M.4
  • 47
    • 0022000994 scopus 로고
    • Amelioration of hemodialysis-associated hypotension by the use of cool dialysate
    • Sherman RA, Rubin MP, Cody RP, Eisinger RP. Amelioration of hemodialysis-associated hypotension by the use of cool dialysate. Am J Kidney Dis 1985; 5: 124-127
    • (1985) Am. J. Kidney Dis. , vol.5 , pp. 124-127
    • Sherman, R.A.1    Rubin, M.P.2    Cody, R.P.3    Eisinger, R.P.4
  • 49
    • 0031959646 scopus 로고    scopus 로고
    • Effects of controlled blood cooling on hemodynamic stability and urea kinetics during high-efficiency hemodialysis
    • Kaufman AM, Morris AT, Lavarias VA et al. Effects of controlled blood cooling on hemodynamic stability and urea kinetics during high-efficiency hemodialysis. J Am Soc Nephrol 1998; 9: 877-883
    • (1998) J. Am. Soc. Nephrol. , vol.9 , pp. 877-883
    • Kaufman, A.M.1    Morris, A.T.2    Lavarias, V.3
  • 50
    • 0036326468 scopus 로고    scopus 로고
    • The effects of control of thermal balance on vascular stability in hemodialysis patients: Results of the European randomized clinical trial
    • Maggiore Q, Pizzarelli F, Santoro A et al. The effects of control of thermal balance on vascular stability in hemodialysis patients: results of the European randomized clinical trial. Am J Kidney Dis 2002; 40: 280-290
    • (2002) Am. J. Kidney Dis. , vol.40 , pp. 280-290
    • Maggiore, Q.1    Pizzarelli, F.2    Santoro, A.3
  • 51
    • 0032919646 scopus 로고    scopus 로고
    • Validation of haemodialysis recirculation and access blood flow measured by thermodilution
    • Schneditz D, Wang E, Levin NW. Validation of haemodialysis recirculation and access blood flow measured by thermodilution. Nephrol Dial Transplant 1999; 14: 376-383
    • (1999) Nephrol. Dial. Transplant. , vol.14 , pp. 376-383
    • Schneditz, D.1    Wang, E.2    Levin, N.W.3
  • 52
    • 0035160459 scopus 로고    scopus 로고
    • Technology: Tools or toys, is it economically feasible with current reimbursement? The case in favor
    • Tattersall J. Technology: tools or toys, is it economically feasible with current reimbursement? The case in favor. Blood Purif 2001; 19: 185-188
    • (2001) Blood Purif. , vol.19 , pp. 185-188
    • Tattersall, J.1


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.