Performance of leak compensation in all-age ICU ventilators during volume-targeted neonatal ventilation: A lung model study

Author keywords

Acute care ventilator; Leak compensation; Mechanical ventilation; Neonates; Patient ventilator interaction; Volume targeted ventilation

Indexed keywords

AIRWAY PRESSURE; ALGORITHM; ARTICLE; BODY WEIGHT; HUMAN; LEAK COMPENSATION ALGORITHM; LUNG MODEL; LUNG VENTILATION; MAXIMAL INSPIRATORY PRESSURE; POSITIVE END EXPIRATORY PRESSURE; TIDAL VOLUME; VOLUME TARGETED NEONATAL VENTILATION; ARTIFICIAL VENTILATION; BIOLOGICAL MODEL; COMPARATIVE STUDY; COMPUTER SIMULATION; DEVICE FAILURE; DEVICES; EVALUATION STUDY; INTENSIVE CARE UNIT; LUNG; MANIKIN; MECHANICAL VENTILATOR; NEWBORN; PHYSIOLOGY;

EID: 85027378721     PISSN: 00201324     EISSN: 19433654     Source Type: Journal    
DOI: 10.4187/respcare.05012     Document Type: Article
Times cited : (19)

References (41)
  • 1
    • 84894102129 scopus 로고    scopus 로고
    • Volume-targeted ventilation is more suitable than pressure-limited ventilation for preterm infants: A systematic review and meta-analysis
    • Peng W, Zhu H, Shi H, Liu E. Volume-targeted ventilation is more suitable than pressure-limited ventilation for preterm infants: a systematic review and meta-analysis. Arch Dis Child Fetal Neonatal Ed 2014;99(2):F158-F165.
    • (2014) Arch Dis Child Fetal Neonatal Ed , vol.99 , Issue.2 , pp. F158-F165
    • Peng, W.1    Zhu, H.2    Shi, H.3    Liu, E.4
  • 2
    • 33947494928 scopus 로고    scopus 로고
    • Volume Guarantee Ventilation
    • Keszler M, Abubakar K. Volume Guarantee Ventilation. Clin Perinatol 2007;34(1):107-116, vii.
    • (2007) Clin Perinatol , vol.34 , Issue.1 , pp. 107-116
    • Keszler, M.1    Abubakar, K.2
  • 3
    • 84877109540 scopus 로고    scopus 로고
    • Clinical guidelines for the use of volume guarantee: Practice guidelines for the bedside
    • Keszler M. Clinical guidelines for the use of volume guarantee: practice guidelines for the bedside. Clin Perinatol 2007;34(1):116e1-116e4.
    • (2007) Clin Perinatol , vol.34 , Issue.1 , pp. 116e1-116e4
    • Keszler, M.1
  • 4
    • 80052376585 scopus 로고    scopus 로고
    • A practical guide to neonatal volume guarantee ventilation
    • Klingenberg C, Wheeler KI, Davis PG, Morley CJ. A practical guide to neonatal volume guarantee ventilation. J Perinatol 2011;31(9): 575-585.
    • (2011) J Perinatol , vol.31 , Issue.9 , pp. 575-585
    • Klingenberg, C.1    Wheeler, K.I.2    Davis, P.G.3    Morley, C.J.4
  • 5
    • 79959449010 scopus 로고    scopus 로고
    • Volume-targeted versus pressure-limited ventilation for preterm infants: A systematic review and meta-analysis
    • Wheeler KI, Klingenberg C, Morley CJ, Davis PG. Volume-targeted versus pressure-limited ventilation for preterm infants: a systematic review and meta-analysis. Neonatology 2011;100(3):219-227.
    • (2011) Neonatology , vol.100 , Issue.3 , pp. 219-227
    • Wheeler, K.I.1    Klingenberg, C.2    Morley, C.J.3    Davis, P.G.4
  • 8
    • 84864811811 scopus 로고    scopus 로고
    • Patient-ventilator asynchrony during noninvasive ventilation: A bench and clinical study
    • Carteaux G, Lyazidi A, Cordoba-Izquierdo A, Vignaux L, Jolliet P, Thille AW, et al. Patient-ventilator asynchrony during noninvasive ventilation: a bench and clinical study. Chest 2012;142(2):367-376.
    • (2012) Chest , vol.142 , Issue.2 , pp. 367-376
    • Carteaux, G.1    Lyazidi, A.2    Cordoba-Izquierdo, A.3    Vignaux, L.4    Jolliet, P.5    Thille, A.W.6
  • 10
    • 0031033608 scopus 로고    scopus 로고
    • Patient ventilator interaction and inspiratory effort during pressure support ventilation in patients with different pathologies
    • Nava S, Bruschi C, Fracchia C, Braschi A, Rubini F. Patient ventilator interaction and inspiratory effort during pressure support ventilation in patients with different pathologies. Eur Respir J 1997; 10(1):177-183.
    • (1997) Eur Respir J , vol.10 , Issue.1 , pp. 177-183
    • Nava, S.1    Bruschi, C.2    Fracchia, C.3    Braschi, A.4    Rubini, F.5
  • 11
    • 0035161013 scopus 로고    scopus 로고
    • The effect of breath termination criterion on breathing patterns and the work of breathing during pressure support ventilation
    • Tokioka H, Tanaka T, Ishizu T, Fukushima T, Iwaki T, Nakamura Y, Kosogabe Y. The effect of breath termination criterion on breathing patterns and the work of breathing during pressure support ventilation. Anesth Analg 2001;92(1):161-165.
    • (2001) Anesth Analg , vol.92 , Issue.1 , pp. 161-165
    • Tokioka, H.1    Tanaka, T.2    Ishizu, T.3    Fukushima, T.4    Iwaki, T.5    Nakamura, Y.6    Kosogabe, Y.7
  • 12
    • 84888119647 scopus 로고    scopus 로고
    • A comparison of leak compensation in acute care ventilators during non-invasive and invasive ventilation: A lung model study
    • Oto J, Chenelle CT, Marchese AD, Kacmarek RM. A comparison of leak compensation in acute care ventilators during non-invasive and invasive ventilation: a lung model study. Respir Care 2013;58(12): 2027-2037.
    • (2013) Respir Care , vol.58 , Issue.12 , pp. 2027-2037
    • Oto, J.1    Chenelle, C.T.2    Marchese, A.D.3    Kacmarek, R.M.4
  • 13
    • 84893320837 scopus 로고    scopus 로고
    • A comparison of leak compensation during pediatric noninvasive ventilation: A lung model study
    • Oto J, Chenelle CT, Marchese AD, Kacmarek RM. A comparison of leak compensation during pediatric noninvasive ventilation: a lung model study. Respir Care 2014;59(2):241-251.
    • (2014) Respir Care , vol.59 , Issue.2 , pp. 241-251
    • Oto, J.1    Chenelle, C.T.2    Marchese, A.D.3    Kacmarek, R.M.4
  • 14
    • 85030616901 scopus 로고    scopus 로고
    • Effects of leak compensation on patient-ventilator synchrony during premature/neonatal invasive and noninvasive ventilation: A lung model study
    • Itagaki T, Chenelle CT, Bennett DJ, Fisher DF, Kacmarek RM. Effects of leak compensation on patient-ventilator synchrony during premature/neonatal invasive and noninvasive ventilation: a lung model study. Respir Care 2017;62(1):22-33.
    • (2017) Respir Care , vol.62 , Issue.1 , pp. 22-33
    • Itagaki, T.1    Chenelle, C.T.2    Bennett, D.J.3    Fisher, D.F.4    Kacmarek, R.M.5
  • 15
    • 62949225400 scopus 로고    scopus 로고
    • Adult ICU ventilators to provide neonatal ventilation: A lung model study
    • Marchese AD, Chipman D, de la Oliva P, Kacmarek RM. Adult ICU ventilators to provide neonatal ventilation: a lung model study. Intensive Care Med 2009;35(4):631-638.
    • (2009) Intensive Care Med , vol.35 , Issue.4 , pp. 631-638
    • Marchese, A.D.1    Chipman, D.2    de La Oliva, P.3    Kacmarek, R.M.4
  • 16
    • 84941735732 scopus 로고    scopus 로고
    • Neonatal and adult ICU ventilators to provide ventilation in neonates, infants, and children: A bench model study
    • Vignaux L, Piquilloud L, Tourneux P, Jolliet P, Rimensberger PC. Neonatal and adult ICU ventilators to provide ventilation in neonates, infants, and children: a bench model study. Respir Care 2014; 59(10):1463-1475.
    • (2014) Respir Care , vol.59 , Issue.10 , pp. 1463-1475
    • Vignaux, L.1    Piquilloud, L.2    Tourneux, P.3    Jolliet, P.4    Rimensberger, P.C.5
  • 17
    • 84945253273 scopus 로고    scopus 로고
    • Evaluation of a nasal cannula in noninvasive ventilation using a lung simulator
    • Iyer NP, Chatburn R. Evaluation of a nasal cannula in noninvasive ventilation using a lung simulator. Respir Care 2015;60(4):508-512.
    • (2015) Respir Care , vol.60 , Issue.4 , pp. 508-512
    • Iyer, N.P.1    Chatburn, R.2
  • 18
    • 84876408866 scopus 로고    scopus 로고
    • Effect of HFNC flow rate, cannula size, and nares diameter on generated airway pressures: An in vitro study
    • Sivieri EM, Gerdes JS, Abbasi S. Effect of HFNC flow rate, cannula size, and nares diameter on generated airway pressures: an in vitro study. Pediatr Pulmonol 2013;48(5):506-514.
    • (2013) Pediatr Pulmonol , vol.48 , Issue.5 , pp. 506-514
    • Sivieri, E.M.1    Gerdes, J.S.2    Abbasi, S.3
  • 19
    • 82555164566 scopus 로고    scopus 로고
    • High flow through a nasal cannula and CPAP effect in a simulated infant model
    • Volsko TA, Fedor K, Amadei J, Chatburn RL. High flow through a nasal cannula and CPAP effect in a simulated infant model. Respir Care 2011;56(12):1893-1900.
    • (2011) Respir Care , vol.56 , Issue.12 , pp. 1893-1900
    • Volsko, T.A.1    Fedor, K.2    Amadei, J.3    Chatburn, R.L.4
  • 20
    • 33645280587 scopus 로고    scopus 로고
    • Lung function tests in neonates and infants with chronic lung disease: Lung and chest-wall mechanics
    • Gappa M, Pillow JJ, Allen J, Mayer O, Stocks J. Lung function tests in neonates and infants with chronic lung disease: lung and chest-wall mechanics. Pediatr Pulmonol 2006;41(4):291-317.
    • (2006) Pediatr Pulmonol , vol.41 , Issue.4 , pp. 291-317
    • Gappa, M.1    Pillow, J.J.2    Allen, J.3    Mayer, O.4    Stocks, J.5
  • 21
    • 3142683136 scopus 로고    scopus 로고
    • Respiratory outcome in preterm ventilated infants: Importance of early respiratory system resistance
    • Snepvangers Y, de Winter JP, Burger H, Brouwers H, van der Ent CK. Respiratory outcome in preterm ventilated infants: importance of early respiratory system resistance. Eur J Pediatr 2004;163(7): 378-384.
    • (2004) Eur J Pediatr , vol.163 , Issue.7 , pp. 378-384
    • Snepvangers, Y.1    de Winter, J.P.2    Burger, H.3    Brouwers, H.4    van Der Ent, C.K.5
  • 22
    • 0033800120 scopus 로고    scopus 로고
    • Early changes in respiratory compliance and resistance during the development of bronchopulmonary dysplasia in the era of surfactant therapy
    • Lui K, Lloyd J, Ang E, Rynn M, Gupta JM. Early changes in respiratory compliance and resistance during the development of bronchopulmonary dysplasia in the era of surfactant therapy. Pediatr Pulmonol 2000;30(4):282-290.
    • (2000) Pediatr Pulmonol , vol.30 , Issue.4 , pp. 282-290
    • Lui, K.1    Lloyd, J.2    Ang, E.3    Rynn, M.4    Gupta, J.M.5
  • 24
    • 0038757613 scopus 로고    scopus 로고
    • Effect of posture on oxygenation, lung volume, and respiratory mechanics in premature infants studied before discharge
    • Bhat RY, Leipälä JA, Singh NR, Rafferty GF, Hannam S, Greenough A. Effect of posture on oxygenation, lung volume, and respiratory mechanics in premature infants studied before discharge. Pediatrics 2003;112(1 Pt 1):29-32.
    • (2003) Pediatrics , vol.112 , Issue.1 , pp. 29-32
    • Bhat, R.Y.1    Leipälä, J.A.2    Singh, N.R.3    Rafferty, G.F.4    Hannam, S.5    Greenough, A.6
  • 25
    • 0035203421 scopus 로고    scopus 로고
    • Respiratory mechanics in very low birth weight infants during continuous versus intermittent gavage feeds
    • Brar G, Geiss D, Brion LP, Rios A. Respiratory mechanics in very low birth weight infants during continuous versus intermittent gavage feeds. Pediatr Pulmonol 2001;32(6):442-446.
    • (2001) Pediatr Pulmonol , vol.32 , Issue.6 , pp. 442-446
    • Brar, G.1    Geiss, D.2    Brion, L.P.3    Rios, A.4
  • 26
    • 84887253314 scopus 로고    scopus 로고
    • Differential effects of immaturity and neonatal lung disease on the lung function of very low birth weight infants at 48-52 postconceptional weeks
    • Schmalisch G, Wilitzki S, Roehr CC, Proquitté H, Bührer C. Differential effects of immaturity and neonatal lung disease on the lung function of very low birth weight infants at 48-52 postconceptional weeks. Pediatr Pulmonol 2013;48(12):1214-1223.
    • (2013) Pediatr Pulmonol , vol.48 , Issue.12 , pp. 1214-1223
    • Schmalisch, G.1    Wilitzki, S.2    Roehr, C.C.3    Proquitté, H.4    Bührer, C.5
  • 28
    • 57349110823 scopus 로고    scopus 로고
    • Measurement of maximal inspiratory pressure in ventilated children
    • Harikumar G, Moxham J, Greenough A, Rafferty GF. Measurement of maximal inspiratory pressure in ventilated children. Pediatr Pulmonol 2008;43(11):1085-1091.
    • (2008) Pediatr Pulmonol , vol.43 , Issue.11 , pp. 1085-1091
    • Harikumar, G.1    Moxham, J.2    Greenough, A.3    Rafferty, G.F.4
  • 29
    • 0023772273 scopus 로고
    • Development of sleep states in normal premature and full-term newborns
    • Curzi-Dascalova L, Peirano P, Morel-Kahn F. Development of sleep states in normal premature and full-term newborns. Dev Psychobiol 1988;21(5):431-444.
    • (1988) Dev Psychobiol , vol.21 , Issue.5 , pp. 431-444
    • Curzi-Dascalova, L.1    Peirano, P.2    Morel-Kahn, F.3
  • 30
    • 19444386654 scopus 로고    scopus 로고
    • Effort and work of breathing in neonates during assisted patient-triggered ventilation
    • Kapasi M, Fujino Y, Kirmse M, Catlin EA, Kacmarek RM. Effort and work of breathing in neonates during assisted patient-triggered ventilation. Pediatr Crit Care Med 2001;2(1):9-16.
    • (2001) Pediatr Crit Care Med , vol.2 , Issue.1 , pp. 9-16
    • Kapasi, M.1    Fujino, Y.2    Kirmse, M.3    Catlin, E.A.4    Kacmarek, R.M.5
  • 31
    • 0023409239 scopus 로고
    • Pulmonary mechanics in normal infants and young children during first 5 years of life
    • Gerhardt T, Hehre D, Feller R, Reifenberg L, Bancalari E. Pulmonary mechanics in normal infants and young children during first 5 years of life. Pediatr Pulmonol 1987;3(5):309-316.
    • (1987) Pediatr Pulmonol , vol.3 , Issue.5 , pp. 309-316
    • Gerhardt, T.1    Hehre, D.2    Feller, R.3    Reifenberg, L.4    Bancalari, E.5
  • 32
    • 0028824792 scopus 로고
    • Airway leak size in neonates and autocycling of three flow-triggered ventilators
    • Bernstein G, Knodel E, Heldt GP. Airway leak size in neonates and autocycling of three flow-triggered ventilators. Crit Care Med 1995; 23(10):1739-1744.
    • (1995) Crit Care Med , vol.23 , Issue.10 , pp. 1739-1744
    • Bernstein, G.1    Knodel, E.2    Heldt, G.P.3
  • 33
    • 79953102791 scopus 로고    scopus 로고
    • Tracheal tube airleak in clinical practice and impact on tidal volume measurement in ventilated neonates
    • Mahmoud RA, Proquitté H, Fawzy N, Bührer C, Schmalisch G. Tracheal tube airleak in clinical practice and impact on tidal volume measurement in ventilated neonates. Pediatr Crit Care Med 2011; 12(2):197-202.
    • (2011) Pediatr Crit Care Med , vol.12 , Issue.2 , pp. 197-202
    • Mahmoud, R.A.1    Proquitté, H.2    Fawzy, N.3    Bührer, C.4    Schmalisch, G.5
  • 34
    • 84903681567 scopus 로고    scopus 로고
    • Optimizing patient-ventilator synchrony during invasive ventilator assist in children and infants remains a difficult task
    • Vignaux L, Grazioli S, Piquilloud L, Bochaton N, Karam O, Jaecklin T, et al. Optimizing patient-ventilator synchrony during invasive ventilator assist in children and infants remains a difficult task. Pediatr Crit Care Med 2013;14(7):e316-e325.
    • (2013) Pediatr Crit Care Med , vol.14 , Issue.7 , pp. e316-e325
    • Vignaux, L.1    Grazioli, S.2    Piquilloud, L.3    Bochaton, N.4    Karam, O.5    Jaecklin, T.6
  • 35
    • 84885656037 scopus 로고    scopus 로고
    • Patient-ventilator asynchrony during noninvasive pressure support ventilation and neurally adjusted ventilatory assist in infants and children
    • Vignaux L, Grazioli S, Piquilloud L, Bochaton N, Karam O, Levy-Jamet Y, et al. Patient-ventilator asynchrony during noninvasive pressure support ventilation and neurally adjusted ventilatory assist in infants and children. Pediatr Crit Care Med 2013; 14(8):e357-e364.
    • (2013) Pediatr Crit Care Med , vol.14 , Issue.8 , pp. e357-e364
    • Vignaux, L.1    Grazioli, S.2    Piquilloud, L.3    Bochaton, N.4    Karam, O.5    Levy-Jamet, Y.6
  • 36
    • 67349181016 scopus 로고    scopus 로고
    • Patient-ventilator asynchrony during non-invasive ventilation for acute respiratory failure: A multicenter study
    • Vignaux L, Vargas F, Roeseler J, Tassaux D, Thille AW, Kossowsky MP, et al. Patient-ventilator asynchrony during non-invasive ventilation for acute respiratory failure: a multicenter study. Intensive Care Med 2009;35(5):840-846.
    • (2009) Intensive Care Med , vol.35 , Issue.5 , pp. 840-846
    • Vignaux, L.1    Vargas, F.2    Roeseler, J.3    Tassaux, D.4    Thille, A.W.5    Kossowsky, M.P.6
  • 37
    • 33846858840 scopus 로고    scopus 로고
    • Volume-targeted modes of modern neonatal ventilators: How stable is the delivered tidal volume?
    • Jaecklin T, Morel DR, Rimensberger PC. Volume-targeted modes of modern neonatal ventilators: how stable is the delivered tidal volume? Intensive Care Med 2007;33(2):326-335.
    • (2007) Intensive Care Med , vol.33 , Issue.2 , pp. 326-335
    • Jaecklin, T.1    Morel, D.R.2    Rimensberger, P.C.3
  • 38
    • 33847196138 scopus 로고    scopus 로고
    • Performance of neonatal ventilators in volume targeted ventilation mode
    • Sharma A, Milner AD, Greenough A. Performance of neonatal ventilators in volume targeted ventilation mode. Acta Paediatr 2007; 96(2):176-180.
    • (2007) Acta Paediatr , vol.96 , Issue.2 , pp. 176-180
    • Sharma, A.1    Milner, A.D.2    Greenough, A.3
  • 39
    • 82255163772 scopus 로고    scopus 로고
    • Mechanical ventilation of the premature neonate
    • Brown MK, DiBlasi RM. Mechanical ventilation of the premature neonate. Respir Care 2011;56(9):1298-1311; discussion 1311-1313.
    • (2011) Respir Care , vol.56 , Issue.9 , pp. 1298-1311
    • Brown, M.K.1    Diblasi, R.M.2
  • 40
    • 68849129113 scopus 로고    scopus 로고
    • Bilevel vs ICU ventilators providing noninvasive ventilation: Effect of system leaks: A COPD lung model comparison
    • Ferreira JC, Chipman DW, Hill NS, Kacmarek RM. Bilevel vs ICU ventilators providing noninvasive ventilation: effect of system leaks: a COPD lung model comparison. Chest 2009;136(2):448-456.
    • (2009) Chest , vol.136 , Issue.2 , pp. 448-456
    • Ferreira, J.C.1    Chipman, D.W.2    Hill, N.S.3    Kacmarek, R.M.4
  • 41
    • 0031739744 scopus 로고    scopus 로고
    • A new method to analyze lung compliance when pressure-volume relationship is non-linear
    • Nikischin W, Gerhardt T, Everett R, Bancalari E. A new method to analyze lung compliance when pressure-volume relationship is non-linear. Am J Respir Crit Care Med 1998;158(4):1052-1060.
    • (1998) Am J Respir Crit Care Med , vol.158 , Issue.4 , pp. 1052-1060
    • Nikischin, W.1    Gerhardt, T.2    Everett, R.3    Bancalari, E.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.