Volumn 49, Issue 9, 2016, Pages 708-712

Analytical evaluation of the epoc® point-of-care blood analysis system in cardiopulmonary bypass patients

Author keywords

Blood gas analysis; Cardiac surgery; Comparison study; Epoc ; Method evaluation; Point of care testing

Indexed keywords

ARTICLE; BLOOD ANALYSIS; BLOOD GAS; BLOOD GAS ANALYSIS; BLOOD GAS ANALYZER; CARDIOPULMONARY BYPASS; HEART PERFUSION; PH; POINT OF CARE SYSTEM; POINT OF CARE TESTING; PRIORITY JOURNAL; QUALITY CONTROL; ANALYSIS; BLOOD EXAMINATION; BLOOD GLUCOSE MONITORING; GLUCOSE BLOOD LEVEL; HEALTH CARE QUALITY; HUMAN; PROCEDURES; STANDARDS;

EID: 84958281103     PISSN: 00099120     EISSN: 18732933     Source Type: Journal    
DOI: 10.1016/j.clinbiochem.2015.12.015     Document Type: Article
Times cited : (19)

References (26)
  • 1
    • 77949440540 scopus 로고    scopus 로고
    • Market trends in point-of-care testing 2007-2008
    • Winter M.C. Market trends in point-of-care testing 2007-2008. Point of Care 2010, 9:12-20.
    • (2010) Point of Care , vol.9 , pp. 12-20
    • Winter, M.C.1
  • 2
    • 84894601346 scopus 로고    scopus 로고
    • Point-of-care testing - current and emerging quality perspectives
    • O'Kane M.J. Point-of-care testing - current and emerging quality perspectives. Point of Care 2014, 13:1-5.
    • (2014) Point of Care , vol.13 , pp. 1-5
    • O'Kane, M.J.1
  • 3
    • 0026523133 scopus 로고
    • Cardiopulmonary bypass surgery
    • Utley J.R. Cardiopulmonary bypass surgery. Curr. Opin. Cardiol. 1992, 7:267-275.
    • (1992) Curr. Opin. Cardiol. , vol.7 , pp. 267-275
    • Utley, J.R.1
  • 4
    • 0036211291 scopus 로고    scopus 로고
    • Optimal blood gas management during deep hypothermic paediatric cardiac surgery: alpha-stat is easy, but pH-stat may be preferable
    • Laussen P.C. Optimal blood gas management during deep hypothermic paediatric cardiac surgery: alpha-stat is easy, but pH-stat may be preferable. Paediatr. Anaesth. 2002, 12:199-204.
    • (2002) Paediatr. Anaesth. , vol.12 , pp. 199-204
    • Laussen, P.C.1
  • 5
    • 26444475893 scopus 로고    scopus 로고
    • Blood gas strategies and management during pediatric cardiopulmonary bypass
    • Griffin D.A. Blood gas strategies and management during pediatric cardiopulmonary bypass. ASAIO J. 2005, 51:657-658.
    • (2005) ASAIO J. , vol.51 , pp. 657-658
    • Griffin, D.A.1
  • 6
    • 35548993658 scopus 로고    scopus 로고
    • Point of care hematocrit and hemoglobin in cardiac surgery: a review
    • Myers G.J., Browne J. Point of care hematocrit and hemoglobin in cardiac surgery: a review. Perfusion 2007, 22:179-183.
    • (2007) Perfusion , vol.22 , pp. 179-183
    • Myers, G.J.1    Browne, J.2
  • 7
    • 41849091093 scopus 로고    scopus 로고
    • Evaluation of the enterprise point-of-care (EPOC) system for blood gas and electrolyte analysis
    • Nichols J.H., Rajadhyaksha A., Rodriguez M. Evaluation of the enterprise point-of-care (EPOC) system for blood gas and electrolyte analysis. Point of Care 2008, 7:7-11.
    • (2008) Point of Care , vol.7 , pp. 7-11
    • Nichols, J.H.1    Rajadhyaksha, A.2    Rodriguez, M.3
  • 8
    • 84886239789 scopus 로고    scopus 로고
    • Analytical and clinical performance of the epoc blood analysis system: experience at a large tertiary academic medical center
    • Stotler B.A., Kratz A. Analytical and clinical performance of the epoc blood analysis system: experience at a large tertiary academic medical center. Am. J. Clin. Pathol. 2013, 140:715-720.
    • (2013) Am. J. Clin. Pathol. , vol.140 , pp. 715-720
    • Stotler, B.A.1    Kratz, A.2
  • 9
    • 73349083470 scopus 로고    scopus 로고
    • Evaluation of precision performance of quantitative measurement methods; approved guideline
    • CLSI, Wayne, PA, US
    • Clinical Laboratory Standards Institute Evaluation of precision performance of quantitative measurement methods; approved guideline. CLSI Document EP05-A2 2004, CLSI, Wayne, PA, US, (ISBN 1-56238-542-9). Second edition.
    • (2004) CLSI Document EP05-A2
  • 10
    • 84969772605 scopus 로고    scopus 로고
    • User verification of performance for precision and trueness; approved guideline
    • CLSI, Wayne, PA, US
    • Clinical Laboratory Standards Institute User verification of performance for precision and trueness; approved guideline. CLSI Document EP15-A2 2006, CLSI, Wayne, PA, US, (ISBN 1-56238-574-7). Second edition.
    • (2006) CLSI Document EP15-A2
  • 13
    • 84961851177 scopus 로고    scopus 로고
    • Epocal Inc., Ottawa, Ontario, Canada
    • epoc® System Manual 2014, Epocal Inc., Ottawa, Ontario, Canada.
    • (2014) epoc® System Manual
  • 16
    • 84904353619 scopus 로고    scopus 로고
    • Comparison of Becton Dickinson vacutainer rapid serum tube with the serum separator tube for routine chemistry and immunoassay tests
    • Yan R., Lou A., Watts G., Tarr H., Smith H., Kinney L., Chen Y. Comparison of Becton Dickinson vacutainer rapid serum tube with the serum separator tube for routine chemistry and immunoassay tests. J. Clin. Pathol. 2014, 67:599-604.
    • (2014) J. Clin. Pathol. , vol.67 , pp. 599-604
    • Yan, R.1    Lou, A.2    Watts, G.3    Tarr, H.4    Smith, H.5    Kinney, L.6    Chen, Y.7
  • 17
    • 84906940169 scopus 로고    scopus 로고
    • Suitability of Becton Dickinson vacutainer rapid serum tube for collecting and storing blood samples for antibiotic and anticonvulsant drug monitoring
    • Yan R., Colantonio D., Wong P.Y., Chen Y. Suitability of Becton Dickinson vacutainer rapid serum tube for collecting and storing blood samples for antibiotic and anticonvulsant drug monitoring. J. Clin. Pathol. 2014, 67:807-810.
    • (2014) J. Clin. Pathol. , vol.67 , pp. 807-810
    • Yan, R.1    Colantonio, D.2    Wong, P.Y.3    Chen, Y.4
  • 18
    • 84969829863 scopus 로고
    • Department of health and human services - regulations implementing the clinical laboratory improvement amendments of 1988 (CLIA) proficiency testing criteria for acceptable analytical performance
    • Department of health and human services - regulations implementing the clinical laboratory improvement amendments of 1988 (CLIA) proficiency testing criteria for acceptable analytical performance. Fed. Regist. 1992, 57:7002-7187.
    • (1992) Fed. Regist. , vol.57 , pp. 7002-7187
  • 19
    • 85000472382 scopus 로고    scopus 로고
    • Linearity analysis and comparison study on the epoc® point-of-care blood analysis system in cardiopulmonary bypass patients
    • Chen J., Gorman M., O'Reilly B., Chen Y. Linearity analysis and comparison study on the epoc® point-of-care blood analysis system in cardiopulmonary bypass patients. Data Brief 2016, 6:847-852.
    • (2016) Data Brief , vol.6 , pp. 847-852
    • Chen, J.1    Gorman, M.2    O'Reilly, B.3    Chen, Y.4
  • 20
    • 0027410804 scopus 로고
    • Evaluation of a novel point-of-care system, the i-STAT protable clinical analyzer
    • Erickson K.A., Wilding P. Evaluation of a novel point-of-care system, the i-STAT protable clinical analyzer. Clin. Chem. 1993, 39:283-287.
    • (1993) Clin. Chem. , vol.39 , pp. 283-287
    • Erickson, K.A.1    Wilding, P.2
  • 21
    • 0027255920 scopus 로고
    • Analytical evaluation of i-STAT portable clinical analyzer and use by nonlaboratory health-care professionals
    • Jacobs E., Vadasdi E., Sarkozi L., Colman N. Analytical evaluation of i-STAT portable clinical analyzer and use by nonlaboratory health-care professionals. Clin. Chem. 1993, 39:1069-1074.
    • (1993) Clin. Chem. , vol.39 , pp. 1069-1074
    • Jacobs, E.1    Vadasdi, E.2    Sarkozi, L.3    Colman, N.4
  • 22
    • 0028985807 scopus 로고
    • Evaluation of the i-STAT system: a portable chemistry analyzer for the measurement of sodium, potassium, chloride, urea, glucose, and hematocrit
    • Mock T., Morrison D., Yatscoff R. Evaluation of the i-STAT system: a portable chemistry analyzer for the measurement of sodium, potassium, chloride, urea, glucose, and hematocrit. Clin. Biochem. 1995, 28:187-192.
    • (1995) Clin. Biochem. , vol.28 , pp. 187-192
    • Mock, T.1    Morrison, D.2    Yatscoff, R.3
  • 23
    • 0025089106 scopus 로고
    • A comparison of conductivity-based hematocrit determinations with conventional laboratory methods in autologous blood transfusions
    • McMahon D.J., Carpenter R.L. A comparison of conductivity-based hematocrit determinations with conventional laboratory methods in autologous blood transfusions. Anesth. Analg. 1990, 71:541-544.
    • (1990) Anesth. Analg. , vol.71 , pp. 541-544
    • McMahon, D.J.1    Carpenter, R.L.2
  • 24
    • 0028827014 scopus 로고
    • A comparison of four bedside methods of hemoglobin assessment during cardiac surgery
    • McNulty S.E., Torjman M., Grodecki W., Marr A., Schieren H. A comparison of four bedside methods of hemoglobin assessment during cardiac surgery. Anesth. Analg. 1995, 81:1197-1202.
    • (1995) Anesth. Analg. , vol.81 , pp. 1197-1202
    • McNulty, S.E.1    Torjman, M.2    Grodecki, W.3    Marr, A.4    Schieren, H.5
  • 25


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.