Volumn 59, Issue 3, 2014, Pages 561-577

A volume of intersection approach for on-the-fly system matrix calculation in 3D PET image reconstruction

Author keywords

3D; image reconstruction; list mode; on the fly system matrix calculation; PET; tube of response; volume of intersection

Indexed keywords

3D; GEOMETRICAL MODELING; HIGH SPATIAL RESOLUTION; LISTMODE; PET IMAGE RECONSTRUCTION; RECONSTRUCTION ALGORITHMS; RECONSTRUCTION SOFTWARE; SYSTEM MATRICES;

EID: 84892741218     PISSN: 00319155     EISSN: 13616560     Source Type: Journal    
DOI: 10.1088/0031-9155/59/3/561     Document Type: Article
Times cited : (17)

References (37)
  • 1
    • 3242772275 scopus 로고    scopus 로고
    • Optimization of a fully 3D single scatter simulation algorithm for 3D PET
    • 10.1088/0031-9155/49/12/008 0031-9155 008
    • Accorsi R et al 2004 Optimization of a fully 3D single scatter simulation algorithm for 3D PET Phys. Med. Biol. 49 2577
    • (2004) Phys. Med. Biol. , vol.49 , Issue.12 , pp. 2577
    • Accorsi, R.1
  • 2
    • 78149443566 scopus 로고    scopus 로고
    • Geometrical and Monte Carlo projectors in 3D PET reconstruction
    • 10.1118/1.3501884 0094-2405
    • Aguiar P et al 2010 Geometrical and Monte Carlo projectors in 3D PET reconstruction Med. Phys. 37 5691-702
    • (2010) Med. Phys. , vol.37 , pp. 5691-5702
    • Aguiar, P.1
  • 3
    • 51049121561 scopus 로고    scopus 로고
    • Application of a spatially variant system model for 3-D whole-body pet image reconstruction
    • 10.1109/ISBI.2008.4541246
    • Alessio A and Kinahan P 2008 Application of a spatially variant system model for 3-D whole-body pet image reconstruction 5th IEEE Int. Symp. on Biomed. Imaging p 1315
    • (2008) 5th IEEE Int. Symp. on Biomed. Imaging , pp. 1315
    • Alessio, A.1    Kinahan, P.2
  • 4
    • 77749264192 scopus 로고    scopus 로고
    • Application and evaluation of a measured spatially variant system model for PET image reconstruction
    • 10.1109/TMI.2010.2040188 0278-0062
    • Alessio A et al 2010 Application and evaluation of a measured spatially variant system model for PET image reconstruction IEEE Trans. Med. Imaging 29 938-49
    • (2010) IEEE Trans. Med. Imaging , vol.29 , pp. 938-949
    • Alessio, A.1
  • 6
    • 80053578484 scopus 로고    scopus 로고
    • Physical performance of the new hybrid PET/CT Discovery-690
    • 10.1118/1.3635220 0094-2405
    • Bettinardi V et al 2011 Physical performance of the new hybrid PET/CT Discovery-690 Med. Phys. 38 5394-411
    • (2011) Med. Phys. , vol.38 , pp. 5394-5411
    • Bettinardi, V.1
  • 7
    • 34147126296 scopus 로고    scopus 로고
    • A scatter-corrected list-mode reconstruction and a practical scatter/random approximation technique for dynamic PET imaging
    • 10.1088/0031-9155/52/8/004 0031-9155 004
    • Cheng J-C K et al 2007 A scatter-corrected list-mode reconstruction and a practical scatter/random approximation technique for dynamic PET imaging Phys. Med. Biol. 52 2089
    • (2007) Phys. Med. Biol. , vol.52 , Issue.8 , pp. 2089
    • Cheng, J.-C.K.1
  • 8
    • 78149319138 scopus 로고    scopus 로고
    • Non-Gaussian space-variant resolution modelling for list-mode reconstruction
    • 10.1088/0031-9155/55/17/011 0031-9155 011
    • Cloquet C et al 2010 Non-Gaussian space-variant resolution modelling for list-mode reconstruction Phys. Med. Biol. 55 5045
    • (2010) Phys. Med. Biol. , vol.55 , Issue.17 , pp. 5045
    • Cloquet, C.1
  • 10
    • 76149130763 scopus 로고    scopus 로고
    • Effects of cold sphere walls in PET phantom measurements on the volume reproducing threshold
    • 10.1088/0031-9155/55/4/013 0031-9155 013
    • Hofheinz F et al 2010 Effects of cold sphere walls in PET phantom measurements on the volume reproducing threshold Phys. Med. Biol. 55 1099
    • (2010) Phys. Med. Biol. , vol.55 , Issue.4 , pp. 1099
    • Hofheinz, F.1
  • 11
    • 0028700063 scopus 로고
    • Accelerated image reconstruction using ordered subsets of projection data
    • 10.1109/42.363108 0278-0062
    • Hudson H and Larkin R 1994 Accelerated image reconstruction using ordered subsets of projection data IEEE Trans. Med. Imaging 13 601-9
    • (1994) IEEE Trans. Med. Imaging , vol.13 , pp. 601-609
    • Hudson, H.1    Larkin, R.2
  • 12
    • 79953683986 scopus 로고    scopus 로고
    • Physical and clinical performance of the mCT time-of-flight PET/CT scanner
    • 10.1088/0031-9155/56/8/004 0031-9155 004
    • Jakoby B et al 2011 Physical and clinical performance of the mCT time-of-flight PET/CT scanner Phys. Med. Biol. 56 2375
    • (2011) Phys. Med. Biol. , vol.56 , Issue.8 , pp. 2375
    • Jakoby, B.1
  • 13
    • 84873971559 scopus 로고    scopus 로고
    • Influence of PET reconstruction parameters on the TrueX algorithm. A combined phantom and patient study
    • 10.3413/Nukmed-0523-12-07
    • Knäusl B et al 2013 Influence of PET reconstruction parameters on the TrueX algorithm. A combined phantom and patient study Nucl. Med. 52 28-35
    • (2013) Nucl. Med. , vol.52 , pp. 28-35
    • Knäusl, B.1
  • 14
    • 0025444920 scopus 로고
    • Evaluation of the volume of intersection of a sphere with a cylinder by elliptic integrals
    • 10.1016/0010-4655(90)90184-3 0010-4655
    • Lamarche F and Leroy C 1990 Evaluation of the volume of intersection of a sphere with a cylinder by elliptic integrals Comput. Phys. Commun. 59 359-69
    • (1990) Comput. Phys. Commun. , vol.59 , pp. 359-369
    • Lamarche, F.1    Leroy, C.2
  • 15
    • 33646441146 scopus 로고    scopus 로고
    • Optimized list-mode acquisition and data processing procedures for ACS2 based PET systems
    • 10.1078/0939-3889-00294 0939-3889
    • Langner J et al 2006 Optimized list-mode acquisition and data processing procedures for ACS2 based PET systems Z. Med. Phys. 16 75-82
    • (2006) Z. Med. Phys. , vol.16 , pp. 75-82
    • Langner, J.1
  • 16
    • 0035384871 scopus 로고    scopus 로고
    • The design and implementation of COSEN, an iterative algorithm for fully 3-D listmode data
    • 10.1109/42.932747 0278-0062
    • Levkovilz R et al 2001 The design and implementation of COSEN, an iterative algorithm for fully 3-D listmode data IEEE Trans. Med. Imaging 20 633-42
    • (2001) IEEE Trans. Med. Imaging , vol.20 , pp. 633-642
    • Levkovilz, R.1
  • 17
    • 0025499943 scopus 로고
    • Multidimensional digital image representation using generalized Kaiser-Bessel window functions
    • 10.1364/JOSAA.7.001834 0740-3232 A
    • Lewitt R M 1990 Multidimensional digital image representation using generalized Kaiser-Bessel window functions J. Opt. Soc. Am. A 7 1834-46
    • (1990) J. Opt. Soc. Am. , vol.7 , pp. 1834-1846
    • Lewitt, R.M.1
  • 18
    • 0030085857 scopus 로고    scopus 로고
    • Practical considerations for 3-d image reconstruction using spherically symmetric volume elements
    • 10.1109/42.481442 0278-0062
    • Matej S and Lewitt R 1996 Practical considerations for 3-d image reconstruction using spherically symmetric volume elements IEEE Trans. Med. Imaging 15 68-78
    • (1996) IEEE Trans. Med. Imaging , vol.15 , pp. 68-78
    • Matej, S.1    Lewitt, R.2
  • 20
    • 0030348846 scopus 로고    scopus 로고
    • Maximum likelihood reconstruction in fully 3D PET via the SAGE algorithm
    • 10.1109/NSSMIC.1996.587929 1095-7863
    • Ollinger J M and Goggin A S 1996 Maximum likelihood reconstruction in fully 3D PET via the SAGE algorithm IEEE Nucl. Sci. Symp. Conf. Rec. vol 3 pp 1594-8
    • (1996) IEEE Nucl. Sci. Symp. Conf. Rec. , vol.3 , pp. 1594-1598
    • Ollinger, J.M.1    Goggin, A.S.2
  • 21
    • 33746191812 scopus 로고    scopus 로고
    • Fully 3-D PET reconstruction with system matrix derived from point source measurements
    • 10.1109/TMI.2006.876171 0278-0062
    • Panin V et al 2006 Fully 3-D PET reconstruction with system matrix derived from point source measurements IEEE Trans. Med. Imaging 25 907-21
    • (2006) IEEE Trans. Med. Imaging , vol.25 , pp. 907-921
    • Panin, V.1
  • 22
    • 84878952408 scopus 로고    scopus 로고
    • Resolution modeling in PET imaging: Theory, practice, benefits, and pitfalls
    • 10.1118/1.4800806 0094-2405 064301
    • Rahmim A et al 2013 Resolution modeling in PET imaging: theory, practice, benefits, and pitfalls Med. Phys. 40 064301
    • (2013) Med. Phys. , vol.40
    • Rahmim, A.1
  • 23
    • 77958159834 scopus 로고    scopus 로고
    • Image-based point spread function implementation in a fully 3D OSEM reconstruction algorithm for PET
    • 10.1088/0031-9155/55/14/012 0031-9155 012
    • Rapisarda E et al 2010 Image-based point spread function implementation in a fully 3D OSEM reconstruction algorithm for PET Phys. Med. Biol. 55 4131
    • (2010) Phys. Med. Biol. , vol.55 , Issue.14 , pp. 4131
    • Rapisarda, E.1
  • 24
    • 0142126682 scopus 로고    scopus 로고
    • EM algorithm system modeling by image-space techniques for PET reconstruction
    • 10.1109/TNS.2003.817327 0018-9499
    • Reader A et al 2003 EM algorithm system modeling by image-space techniques for PET reconstruction IEEE Trans. Nucl. Sci. 50 1392-7
    • (2003) IEEE Trans. Nucl. Sci. , vol.50 , pp. 1392-1397
    • Reader, A.1
  • 25
    • 33947277900 scopus 로고    scopus 로고
    • Analytical calculation of volumes-of-intersection for iterative, fully 3-D PET reconstruction
    • 10.1109/TMI.2006.880679 0278-0062
    • Scheins J et al 2006 Analytical calculation of volumes-of-intersection for iterative, fully 3-D PET reconstruction IEEE Trans. Med. Imaging 25 1363-9
    • (2006) IEEE Trans. Med. Imaging , vol.25 , pp. 1363-1369
    • Scheins, J.1
  • 26
    • 79952156824 scopus 로고    scopus 로고
    • Fully-3D PET image reconstruction using scanner-independent, adaptive projection data and highly rotation-symmetric voxel assemblies
    • 10.1109/TMI.2011.2109732 0278-0062
    • Scheins J J et al 2011 Fully-3D PET image reconstruction using scanner-independent, adaptive projection data and highly rotation-symmetric voxel assemblies IEEE Trans. Med. Imaging 30 879-92
    • (2011) IEEE Trans. Med. Imaging , vol.30 , pp. 879-892
    • Scheins, J.J.1
  • 27
    • 0034205094 scopus 로고    scopus 로고
    • Detector response models for statistical iterative image reconstruction in high resolution PET
    • 10.1109/23.856565 0018-9499
    • Selivanov V et al 2000 Detector response models for statistical iterative image reconstruction in high resolution PET IEEE Trans. Nucl. Sci. 47 1168-75
    • (2000) IEEE Trans. Nucl. Sci. , vol.47 , pp. 1168-1175
    • Selivanov, V.1
  • 28
    • 0020194763 scopus 로고
    • Maximum likelihood reconstruction for emission tomography
    • 10.1109/TMI.1982.4307558 0278-0062
    • Shepp L and Vardi Y 1982 Maximum likelihood reconstruction for emission tomography IEEE Trans. Med. Imaging MI-1 113-22
    • (1982) IEEE Trans. Med. Imaging , vol.MI-1 , pp. 113-122
    • Shepp, L.1    Vardi, Y.2
  • 30
    • 79960301507 scopus 로고    scopus 로고
    • Impact of PSF modelling on the convergence rate and edge behaviour of em images in PET
    • 10.1109/NSSMIC.2010.5874409
    • Thielemans K et al 2010 Impact of PSF modelling on the convergence rate and edge behaviour of EM images in PET NSS/MIC'10: Nucl. Sci. Symp. Conf. Rec. pp 3267-72
    • (2010) NSS/MIC'10: Nucl. Sci. Symp. Conf. Rec. , pp. 3267-3272
    • Thielemans, K.1
  • 31
    • 80054055198 scopus 로고    scopus 로고
    • Properties and mitigation of edge artifacts in PSF-based PET reconstruction
    • 10.1109/TNS.2011.2164579 0018-9499
    • Tong S et al 2011 Properties and mitigation of edge artifacts in PSF-based PET reconstruction IEEE Trans. Nucl. Sci. 58 2264-75
    • (2011) IEEE Trans. Nucl. Sci. , vol.58 , pp. 2264-2275
    • Tong, S.1
  • 32
    • 0010293907 scopus 로고
    • A statistical model for positron emission tomography
    • 10.1080/01621459.1985.10477119 0162-1459
    • Vardi Y et al 1985 A statistical model for positron emission tomography J. Am. Stat. Assoc. 80 8-20
    • (1985) J. Am. Stat. Assoc. , vol.80 , pp. 8-20
    • Vardi, Y.1
  • 33
    • 0034244929 scopus 로고    scopus 로고
    • New, faster, image-based scatter correction for 3D PET
    • 10.1109/23.873020 0018-9499
    • Watson C 2000 New, faster, image-based scatter correction for 3D PET IEEE Trans. Nucl. Sci. 47 1587-94
    • (2000) IEEE Trans. Nucl. Sci. , vol.47 , pp. 1587-1594
    • Watson, C.1
  • 34
    • 35348975173 scopus 로고    scopus 로고
    • Extension of single scatter simulation to scatter correction of time-of-flight PET
    • 10.1109/TNS.2007.901227 0018-9499
    • Watson C 2007 Extension of single scatter simulation to scatter correction of time-of-flight PET IEEE Trans. Nucl. Sci. 54 1679-86
    • (2007) IEEE Trans. Nucl. Sci. , vol.54 , pp. 1679-1686
    • Watson, C.1
  • 35
    • 67349250186 scopus 로고    scopus 로고
    • Fast implementation of the single scatter simulation algorithm and its use in iterative image reconstruction of PET data
    • 10.1109/NSSMIC.1999.842765 1095-7863
    • Werling A et al 1999 Fast implementation of the single scatter simulation algorithm and its use in iterative image reconstruction of PET data IEEE Nucl. Sci. Symp. Conf. Rec. vol 3 pp 1158-62
    • (1999) IEEE Nucl. Sci. Symp. Conf. Rec. , vol.3 , pp. 1158-1162
    • Werling, A.1
  • 36
    • 27644541049 scopus 로고    scopus 로고
    • Transaxial system models for jPET-D4 image reconstruction
    • 10.1088/0031-9155/50/22/009 0031-9155 009
    • Yamaya T et al 2005 Transaxial system models for jPET-D4 image reconstruction Phys. Med. Biol. 50 5339
    • (2005) Phys. Med. Biol. , vol.50 , Issue.22 , pp. 5339
    • Yamaya, T.1
  • 37
    • 79955416013 scopus 로고    scopus 로고
    • Design and performance evaluation of a whole-body Ingenuity TF PET-mRI system
    • 10.1088/0031-9155/56/10/013 0031-9155 013
    • Zaidi H et al 2011 Design and performance evaluation of a whole-body Ingenuity TF PET-mRI system Phys. Med. Biol. 56 3091-106
    • (2011) Phys. Med. Biol. , vol.56 , Issue.10 , pp. 3091-3106
    • Zaidi, H.1


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.