Volumn 43, Issue 9, 2013, Pages 1117-1127

Comparison of radiation dose estimates, image noise, and scan duration in pediatric body imaging for volumetric and helical modes on 320-detector CT and helical mode on 64-detector CT

Author keywords

Children; Multidetector computed tomography; Radiation dosage; Scan time; Volume computed tomography

Indexed keywords

ABDOMEN; ARTICLE; CASE REPORT; CHILD; CLINICAL PROTOCOL; COLLIMATOR; HUMAN; IMAGE ANALYSIS; IMAGE QUALITY; MULTIDETECTOR COMPUTED TOMOGRAPHY; NOISE MEASUREMENT; PELVIS; PHANTOM; PRESCHOOL CHILD; PRIORITY JOURNAL; RADIATION DOSE; SPIRAL COMPUTER ASSISTED TOMOGRAPHY; THORAX; VOLUMETRY;

EID: 84882815990     PISSN: 03010449     EISSN: 14321998     Source Type: Journal    
DOI: 10.1007/s00247-013-2690-5     Document Type: Article
Times cited : (29)

References (26)
  • 1
    • 79956299842 scopus 로고    scopus 로고
    • Rising use of CT in child visits to the emergency department in the United States, 1995-2008
    • 21467249 10.1148/radiol.11101939
    • Larson DB, Johnson LW, Schnell BM et al (2011) Rising use of CT in child visits to the emergency department in the United States, 1995-2008. Radiology 259:793-801
    • (2011) Radiology , vol.259 , pp. 793-801
    • Larson, D.B.1    Johnson, L.W.2    Schnell, B.M.3
  • 2
    • 0035139873 scopus 로고    scopus 로고
    • Estimated risks of radiation-induced fatal cancer from pediatric CT
    • 11159059 10.2214/ajr.176.2.1760289 1:STN:280:DC%2BD3M3isVWktg%3D%3D
    • Brenner D, Elliston C, Hall E et al (2001) Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR Am J Roentgenol 176:289-296
    • (2001) AJR Am J Roentgenol , vol.176 , pp. 289-296
    • Brenner, D.1    Elliston, C.2    Hall, E.3
  • 3
    • 84864776903 scopus 로고    scopus 로고
    • Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: A retrospective cohort study
    • 22681860 10.1016/S0140-6736(12)60815-0
    • Pearce MS, Salotti JA, Little MP et al (2012) Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet 380:499-505
    • (2012) Lancet , vol.380 , pp. 499-505
    • Pearce, M.S.1    Salotti, J.A.2    Little, M.P.3
  • 4
    • 80051953463 scopus 로고    scopus 로고
    • X-rays and unshielded infants
    • Published 27 February 2011: Accessed 17 December 2012
    • Bogdanich W, Rebelo K (2011) X-rays and unshielded infants. New York Times, http://www.nytimes.com/2011/02/28/health/28radiation.html?-r=0. Published 27 February 2011: Accessed 17 December 2012.
    • (2011) New York Times
    • Bogdanich, W.1    Rebelo, K.2
  • 5
    • 73149113560 scopus 로고    scopus 로고
    • Projected cancer risks from computed tomographic scans performed in the United States in 2007
    • 20008689 10.1001/archinternmed.2009.440
    • Berrington de Gonzalez A, Mahesh M, Kim KP et al (2009) Projected cancer risks from computed tomographic scans performed in the United States in 2007. Arch Intern Med 169:2071-2077
    • (2009) Arch Intern Med , vol.169 , pp. 2071-2077
    • Berrington De Gonzalez, A.1    Mahesh, M.2    Kim, K.P.3
  • 6
    • 64449087081 scopus 로고    scopus 로고
    • Recent developments in wide-detector cardiac computed tomography
    • 19255875 10.1007/s10554-009-9443-4
    • Choi SI, George RT, Schuleri KH et al (2009) Recent developments in wide-detector cardiac computed tomography. Int J Cardiovasc Imaging 25:23-29
    • (2009) Int J Cardiovasc Imaging , vol.25 , pp. 23-29
    • Choi, S.I.1    George, R.T.2    Schuleri, K.H.3
  • 7
    • 77951022687 scopus 로고    scopus 로고
    • Scan time and patient dose for thoracic imaging in neonates and small children using axial volumetric 320-detector row CT compared to helical 64-, 32-, and 16- detector row CT acquisitions
    • 19997730 10.1007/s00247-009-1436-x
    • Kroft LJ, Roelofs JJ, Geleijns J (2009) Scan time and patient dose for thoracic imaging in neonates and small children using axial volumetric 320-detector row CT compared to helical 64-, 32-, and 16- detector row CT acquisitions. Pediatr Radiol 40:294-300
    • (2009) Pediatr Radiol , vol.40 , pp. 294-300
    • Kroft, L.J.1    Roelofs, J.J.2    Geleijns, J.3
  • 8
    • 84878354597 scopus 로고    scopus 로고
    • Experience with volumetric (320 rows) pediatric CT
    • 10.1016/j.ejrad.2011.12.001
    • Sorantin E, Riccabona M, Stucklschweiger G et al (2012) Experience with volumetric (320 rows) pediatric CT. Eur J Radiol. doi: 10.1016/j.ejrad.2011.12. 001
    • (2012) Eur J Radiol
    • Sorantin, E.1    Riccabona, M.2    Stucklschweiger, G.3
  • 9
    • 84861229375 scopus 로고    scopus 로고
    • Computed tomography angiography in children with cardiovascular disease: Low dose techniques and image quality
    • 21213049 10.1007/s10554-010-9782-1
    • Greenberg SB, Bhutta S, Braswell L et al (2011) Computed tomography angiography in children with cardiovascular disease: low dose techniques and image quality. Int J Cardiovasc Imaging 28:163-170
    • (2011) Int J Cardiovasc Imaging , vol.28 , pp. 163-170
    • Greenberg, S.B.1    Bhutta, S.2    Braswell, L.3
  • 10
    • 79953673547 scopus 로고    scopus 로고
    • Use of 320-detector computed tomographic angiography for infants and young children with congenital heart disease
    • 21210093 10.1007/s00246-010-9873-8
    • Al-Mousily F, Shifrin RY, Fricker FJ et al (2011) Use of 320-detector computed tomographic angiography for infants and young children with congenital heart disease. Pediatr Cardiol 32:426-432
    • (2011) Pediatr Cardiol , vol.32 , pp. 426-432
    • Al-Mousily, F.1    Shifrin, R.Y.2    Fricker, F.J.3
  • 11
    • 68449092006 scopus 로고    scopus 로고
    • Low-dose coronary artery computed tomography angiogram of an infant with tetralogy of fallot using a 256-slice multidetector computed tomography scanner
    • 19214622 10.1007/s00246-009-9389-2
    • Bardo DM, Asamato J, Mackay CS et al (2009) Low-dose coronary artery computed tomography angiogram of an infant with tetralogy of fallot using a 256-slice multidetector computed tomography scanner. Pediatr Cardiol 30:824-826
    • (2009) Pediatr Cardiol , vol.30 , pp. 824-826
    • Bardo, D.M.1    Asamato, J.2    Mackay, C.S.3
  • 12
    • 84870447572 scopus 로고    scopus 로고
    • Initial experience on the application of 320-row CT angiography with low-dose prospective ECG-triggered in children with congenital heart disease
    • 10.1007/s10554-011-0005-1
    • Zhang T, Wang W, Luo Z et al (2011) Initial experience on the application of 320-row CT angiography with low-dose prospective ECG-triggered in children with congenital heart disease. Int J Cardiovasc Imaging. doi: 10.1007/s10554-011-0005-1
    • (2011) Int J Cardiovasc Imaging
    • Zhang, T.1    Wang, W.2    Luo, Z.3
  • 13
    • 84868032021 scopus 로고    scopus 로고
    • Radiation dose estimation for prospective and retrospective ECG-gated cardiac CT angiography in infants and small children using a 320-MDCT volume scanner
    • 23096189 10.2214/AJR.12.8480
    • Podberesky DJ, Angel E, Yoshizumi TT et al (2012) Radiation dose estimation for prospective and retrospective ECG-gated cardiac CT angiography in infants and small children using a 320-MDCT volume scanner. AJR Am J Roentgenol 199:1129-1135
    • (2012) AJR Am J Roentgenol , vol.199 , pp. 1129-1135
    • Podberesky, D.J.1    Angel, E.2    Yoshizumi, T.T.3
  • 14
    • 40949104442 scopus 로고    scopus 로고
    • The 2007 recommendations of the International Commission on Radiological Protection. ICRP publication 103
    • ICRP
    • ICRP (2007) The 2007 recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP 37:1-332
    • (2007) Ann ICRP , vol.37 , pp. 1-332
  • 15
    • 34247606356 scopus 로고    scopus 로고
    • Validation of metal oxide semiconductor field effect transistor technology for organ dose assessment during CT: Comparison with thermoluminescent dosimetry
    • 17449779 10.2214/AJR.06.0742
    • Yoshizumi TT, Goodman PC, Frush DP et al (2007) Validation of metal oxide semiconductor field effect transistor technology for organ dose assessment during CT: comparison with thermoluminescent dosimetry. AJR Am J Roentgenol 188:1332-1336
    • (2007) AJR Am J Roentgenol , vol.188 , pp. 1332-1336
    • Yoshizumi, T.T.1    Goodman, P.C.2    Frush, D.P.3
  • 17
    • 61349202754 scopus 로고    scopus 로고
    • Variability of surface and center position radiation dose in MDCT: Monte Carlo simulations using CTDI and anthropomorphic phantoms
    • 19378763 10.1118/1.3078053 1:CAS:528:DC%2BD1MXhtlGhu7rK
    • Zhang D, Savandi AS, Demarco JJ et al (2009) Variability of surface and center position radiation dose in MDCT: Monte Carlo simulations using CTDI and anthropomorphic phantoms. Med Phys 36:1025-1038
    • (2009) Med Phys , vol.36 , pp. 1025-1038
    • Zhang, D.1    Savandi, A.S.2    Demarco, J.J.3
  • 19
    • 25644448940 scopus 로고    scopus 로고
    • Reducing radiation exposure from survey CT scans
    • 16037529 10.2214/ajr.185.2.01850509
    • O'Daniel JC, Stevens DM, Cody DD (2005) Reducing radiation exposure from survey CT scans. AJR Am J Roentgenol 185:509-515
    • (2005) AJR Am J Roentgenol , vol.185 , pp. 509-515
    • O'Daniel, J.C.1    Stevens, D.M.2    Cody, D.D.3
  • 20
    • 80053653277 scopus 로고    scopus 로고
    • Capability of abdominal 320-detector row CT for small vasculature assessment compared with that of 64-detector row CT
    • 20579828 10.1016/j.ejrad.2010.05.014
    • Kitajima K, Maeda T, Ohno Y et al (2010) Capability of abdominal 320-detector row CT for small vasculature assessment compared with that of 64-detector row CT. Eur J Radiol 80:219-223
    • (2010) Eur J Radiol , vol.80 , pp. 219-223
    • Kitajima, K.1    Maeda, T.2    Ohno, Y.3
  • 21
    • 84858798277 scopus 로고    scopus 로고
    • Lung image quality with 320-row wide-volume CT scans: The effect of prospective ECG-gating and comparisons with 64-row helical CT scans
    • 22222025 10.1016/j.acra.2011.12.001
    • Yamashiro T, Miyara T, Takahashi M et al (2012) Lung image quality with 320-row wide-volume CT scans: the effect of prospective ECG-gating and comparisons with 64-row helical CT scans. Acad Radiol 19:380-388
    • (2012) Acad Radiol , vol.19 , pp. 380-388
    • Yamashiro, T.1    Miyara, T.2    Takahashi, M.3
  • 22
    • 84865218038 scopus 로고    scopus 로고
    • Multi-organ perfusion CT in the abdomen using a 320-detector row CT scanner: Preliminary results of perfusion changes in the liver, spleen, and pancreas of cirrhotic patients
    • 22227262 10.1016/j.ejrad.2011.11.054
    • Motosugi U, Ichikawa T, Sou H et al (2012) Multi-organ perfusion CT in the abdomen using a 320-detector row CT scanner: preliminary results of perfusion changes in the liver, spleen, and pancreas of cirrhotic patients. Eur J Radiol 81:2533-2537
    • (2012) Eur J Radiol , vol.81 , pp. 2533-2537
    • Motosugi, U.1    Ichikawa, T.2    Sou, H.3
  • 23
    • 77957699574 scopus 로고    scopus 로고
    • Overranging at multisection CT: An underestimated source of excess radiation exposure
    • 20631368 10.1148/rg.304095167
    • Schilham A, van der Molen AJ, Prokop M et al (2010) Overranging at multisection CT: an underestimated source of excess radiation exposure. Radiographics 30:1057-1067
    • (2010) Radiographics , vol.30 , pp. 1057-1067
    • Schilham, A.1    Van Der Molen, A.J.2    Prokop, M.3
  • 24
    • 21144440833 scopus 로고    scopus 로고
    • The effect of z overscanning on patient effective dose from multidetector helical computed tomography examinations
    • 16013721 10.1118/1.1924309 1:STN:280:DC%2BD2MzlvF2htg%3D%3D
    • Tzedakis A, Damilakis J, Perisinakis K et al (2005) The effect of z overscanning on patient effective dose from multidetector helical computed tomography examinations. Med Phys 32:1621-1629
    • (2005) Med Phys , vol.32 , pp. 1621-1629
    • Tzedakis, A.1    Damilakis, J.2    Perisinakis, K.3
  • 25
    • 58149136841 scopus 로고    scopus 로고
    • Validation of CT dose-reduction simulation
    • 19235386 10.1118/1.3031114
    • Massoumzadeh P, Don S, Hildebolt CF et al (2009) Validation of CT dose-reduction simulation. Med Phys 36:174-189
    • (2009) Med Phys , vol.36 , pp. 174-189
    • Massoumzadeh, P.1    Don, S.2    Hildebolt, C.F.3
  • 26
    • 34948887091 scopus 로고    scopus 로고
    • Effective dose: How should it be applied to medical exposures?
    • 17646189 10.1259/bjr/25922439 1:STN:280:DC%2BD2srhtVSjug%3D%3D
    • Martin CJ (2007) Effective dose: how should it be applied to medical exposures? Br J Radiol 80:639-647
    • (2007) Br J Radiol , vol.80 , pp. 639-647
    • Martin, C.J.1


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.