Physical characterization of a new ct iterative reconstruction method operating in sinogram space

Author keywords

Computed tomography; Image quality; Iterative reconstruction

Indexed keywords


EID: 84880649834     PISSN: None     EISSN: 15269914     Source Type: Journal    
DOI: 10.1120/jacmp.v14i4.4347     Document Type: Article
Times cited : (57)

References (31)
  • 1
    • 75349092506 scopus 로고    scopus 로고
    • Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm - initial clinical experience
    • Marin D, Nelson RC, Schindera ST, et al. Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm - initial clinical experience. Radiology. 2010;254(1):145-53.
    • (2010) Radiology. , vol.254 , Issue.1 , pp. 145-153
    • Marin, D.1    Nelson, R.C.2    Schindera, S.T.3
  • 2
    • 74749087439 scopus 로고    scopus 로고
    • Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm
    • Silva A, Lawder H, Hara A, Kujak J, Pavlicek W. Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm. AJR Am J Roentgenol. 2010;194(1):191-99.
    • (2010) AJR Am J Roentgenol. , vol.194 , Issue.1 , pp. 191-199
    • Silva, A.1    Lawder, H.2    Hara, A.3    Kujak, J.4    Pavlicek, W.5
  • 3
    • 69949096664 scopus 로고    scopus 로고
    • Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study
    • Hara AK, Paden RG, Silva AC, Kujak JL, Lawder HJ, Pavlicek W. Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study. AJR Am J Roentgenol. 2009;193(3):764-71.
    • (2009) AJR Am J Roentgenol. , vol.193 , Issue.3 , pp. 764-771
    • Hara, A.K.1    Paden, R.G.2    Silva, A.C.3    Kujak, J.L.4    Lawder, H.J.5    Pavlicek, W.6
  • 4
    • 77956841482 scopus 로고    scopus 로고
    • Abdominal CT: comparison of low-dose CT with adaptive statistical iterative reconstruction and routine-dose CT with filtered back projection in 53 patients
    • Sagara Y, Hara A, Pavlicek W, Silva A, Paden R, Wu Q. Abdominal CT: comparison of low-dose CT with adaptive statistical iterative reconstruction and routine-dose CT with filtered back projection in 53 patients. AJR Am J Roentgenol. 2010;195(3):713-19.
    • (2010) AJR Am J Roentgenol. , vol.195 , Issue.3 , pp. 713-719
    • Sagara, Y.1    Hara, A.2    Pavlicek, W.3    Silva, A.4    Paden, R.5    Wu, Q.6
  • 5
    • 77956825812 scopus 로고    scopus 로고
    • Adaptive statistical iterative reconstruction: assessment of image noise and image quality in coronary CT angiography
    • Leipsic J, LaBounty TM, Heilbron B, et al. Adaptive statistical iterative reconstruction: assessment of image noise and image quality in coronary CT angiography. AJR Am J Roentgenol. 2010;195(3):649-54.
    • (2010) AJR Am J Roentgenol. , vol.195 , Issue.3 , pp. 649-654
    • Leipsic, J.1    LaBounty, T.M.2    Heilbron, B.3
  • 6
    • 77956832667 scopus 로고    scopus 로고
    • Estimated radiation dose reduction using adaptive statistical iterative reconstruction in coronary CT angiography: the ERASIR study
    • Leipsic J, LaBounty TM, Heilbron B, et al. Estimated radiation dose reduction using adaptive statistical iterative reconstruction in coronary CT angiography: the ERASIR study. AJR Am J Roentgenol. 2010;195(3):655-60.
    • (2010) AJR Am J Roentgenol. , vol.195 , Issue.3 , pp. 655-660
    • Leipsic, J.1    LaBounty, T.M.2    Heilbron, B.3
  • 8
    • 80054984676 scopus 로고    scopus 로고
    • Iterative reconstruction in image space (IRIS) in cardiac computed tomography:initial experience
    • Bittencourt MS, Schmidt B, Seltman M, et al. Iterative reconstruction in image space (IRIS) in cardiac computed tomography:initial experience. Int J Cardiovasc Imaging. 2011;27(7):1081-87.
    • (2011) Int J Cardiovasc Imaging. , vol.27 , Issue.7 , pp. 1081-1087
    • Bittencourt, M.S.1    Schmidt, B.2    Seltman, M.3
  • 9
    • 80051517687 scopus 로고    scopus 로고
    • Combination of a low-tube-voltage technique with hybrid iterative reconstruction (iDose) algorithm at coronary computed tomographic angiography
    • Funama Y, Taguchi K, Utsunomiya D, et al. Combination of a low-tube-voltage technique with hybrid iterative reconstruction (iDose) algorithm at coronary computed tomographic angiography. J Comput Assist Tomogr. 2011;35(4):480-85.
    • (2011) J Comput Assist Tomogr. , vol.35 , Issue.4 , pp. 480-485
    • Funama, Y.1    Taguchi, K.2    Utsunomiya, D.3
  • 10
    • 84858293727 scopus 로고    scopus 로고
    • Iterative reconstructions methods in X-ray CT
    • Beister M, Kolditz D, Kalender W. Iterative reconstructions methods in X-ray CT. Phys Med. 2012;28(2):94-108.
    • (2012) Phys Med. , vol.28 , Issue.2 , pp. 94-108
    • Beister, M.1    Kolditz, D.2    Kalender, W.3
  • 11
    • 84858281320 scopus 로고    scopus 로고
    • CT iterative reconstruction in image space: a phantom study
    • Ghetti C, Ortenzia O, Serreli G. CT iterative reconstruction in image space: a phantom study. Phys Med. 2012;28(2):161-65.
    • (2012) Phys Med. , vol.28 , Issue.2 , pp. 161-165
    • Ghetti, C.1    Ortenzia, O.2    Serreli, G.3
  • 12
    • 84870380018 scopus 로고    scopus 로고
    • Low-dose CT of the lung: potential value of iterative reconstructions
    • Baumueller S, Winklehner A, Karlo C, et al. Low-dose CT of the lung: potential value of iterative reconstructions. Eur Radiol. 2012;22(12):2597-606.
    • (2012) Eur Radiol. , vol.22 , Issue.12 , pp. 2597-2606
    • Baumueller, S.1    Winklehner, A.2    Karlo, C.3
  • 13
    • 84871322405 scopus 로고    scopus 로고
    • CT evaluation of coronary artery stents with iterative image reconstruction: improvements in image quality and potential for radiation dose reduction
    • Ebersberger U, Tricarico F, Schoepf UJ, et al. CT evaluation of coronary artery stents with iterative image reconstruction: improvements in image quality and potential for radiation dose reduction. Eur Radiol. 2013;23(1):125-32.
    • (2013) Eur Radiol. , vol.23 , Issue.1 , pp. 125-132
    • Ebersberger, U.1    Tricarico, F.2    Schoepf, U.J.3
  • 14
    • 84863693251 scopus 로고    scopus 로고
    • Contrast-to-noise ratio and low-contrast object resolution on full- and low-dose MDCT: SAFIRE versus filtered back projection in a low-contrast object phantom and in the liver
    • Baker ME, Dong F, Primak A, et al. Contrast-to-noise ratio and low-contrast object resolution on full- and low-dose MDCT: SAFIRE versus filtered back projection in a low-contrast object phantom and in the liver. AJR Am J Roentgenol. 2012;199(1):8-18.
    • (2012) AJR Am J Roentgenol. , vol.199 , Issue.1 , pp. 8-18
    • Baker, M.E.1    Dong, F.2    Primak, A.3
  • 15
    • 84862014130 scopus 로고    scopus 로고
    • Assessment of an iterative reconstruction algorithm (SAFIRE) on image quality in pediatric cardiac CT datasets
    • Han BK, Grant KL, Garberich R, Sedlmair M, Lindberg J, Lesser JR. Assessment of an iterative reconstruction algorithm (SAFIRE) on image quality in pediatric cardiac CT datasets. J Cardiovasc Comput Tomogr. 2012;6(3):200-04.
    • (2012) J Cardiovasc Comput Tomogr. , vol.6 , Issue.3 , pp. 200-204
    • Han, B.K.1    Grant, K.L.2    Garberich, R.3    Sedlmair, M.4    Lindberg, J.5    Lesser, J.R.6
  • 16
    • 84861502180 scopus 로고    scopus 로고
    • Radiation dose reduction with Sinogram Affirmed Iterative Reconstruction technique for abdominal computed tomography
    • Kalra MK, Woisetschläger M, Dahlström N, et al. Radiation dose reduction with Sinogram Affirmed Iterative Reconstruction technique for abdominal computed tomography. J Comput Assist Tomogr. 2012;36(3):339-46.
    • (2012) J Comput Assist Tomogr. , vol.36 , Issue.3 , pp. 339-346
    • Kalra, M.K.1    Woisetschläger, M.2    Dahlström, N.3
  • 17
    • 84866987989 scopus 로고    scopus 로고
    • Image quality and radiation dose of low dose coronary CT angiography in obese patients: sinogram affirmed iterative reconstruction versus filtered back projection
    • Wang R, Schoepf UJ, Wu R. Image quality and radiation dose of low dose coronary CT angiography in obese patients: sinogram affirmed iterative reconstruction versus filtered back projection. Eur J Radiol. 2012;81(11):3141-45.
    • (2012) Eur J Radiol. , vol.81 , Issue.11 , pp. 3141-3145
    • Wang, R.1    Schoepf, U.J.2    Wu, R.3
  • 18
    • 80053922582 scopus 로고    scopus 로고
    • Coronary CT angiography: image quality, diagnostic accuracy, and potential for radiation dose reduction using a novel iterative image reconstruction technique - comparison with traditional filtered back projection
    • Moscariello A, Takx R, Schoepf UJ, et al. Coronary CT angiography: image quality, diagnostic accuracy, and potential for radiation dose reduction using a novel iterative image reconstruction technique - comparison with traditional filtered back projection. Eur Radiol. 2011;21(10):2130-38.
    • (2011) Eur Radiol. , vol.21 , Issue.10 , pp. 2130-2138
    • Moscariello, A.1    Takx, R.2    Schoepf, U.J.3
  • 19
    • 84873524416 scopus 로고    scopus 로고
    • Clinical evaluation and potential radiation dose reduction of the novel sinogram-affirmed iterative reconstruction technique (SAFIRE) in abdominal computed tomography angiography
    • [Epub ahead of print
    • Schabel C, Fenchel M, Schmidt B, et al. Clinical evaluation and potential radiation dose reduction of the novel sinogram-affirmed iterative reconstruction technique (SAFIRE) in abdominal computed tomography angiography. Acad Radiol. 2013;20(2):165-72. [Epub ahead of print]
    • (2013) Acad Radiol. , vol.20 , Issue.2 , pp. 165-172
    • Schabel, C.1    Fenchel, M.2    Schmidt, B.3
  • 20
    • 84866673170 scopus 로고    scopus 로고
    • Validation of dual-source single-tube reconstruction as a method to obtain half-dose images to evaluate radiation dose and noise reduction: phantom and human assessment using CT colonography and sinogram-affirmed iterative reconstruction (SAFIRE)
    • Fletcher JG, Grant LL, Fidler JL, et al. Validation of dual-source single-tube reconstruction as a method to obtain half-dose images to evaluate radiation dose and noise reduction: phantom and human assessment using CT colonography and sinogram-affirmed iterative reconstruction (SAFIRE). J Comput Assist Tomogr. 2012;36(5):560-69.
    • (2012) J Comput Assist Tomogr. , vol.36 , Issue.5 , pp. 560-569
    • Fletcher, J.G.1    Grant, L.L.2    Fidler, J.L.3
  • 21
    • 84871927937 scopus 로고    scopus 로고
    • Coronary CT angiography: comparison of a novel iterative reconstruction with filtered back projection for reconstruction of low-dose CT-Initial experience
    • [Epub ahead of print
    • Takx RA, Schoepf UJ, Moscariello A, et al. Coronary CT angiography: comparison of a novel iterative reconstruction with filtered back projection for reconstruction of low-dose CT-Initial experience. Eur J Radiol. 2013;82(2):275-80. [Epub ahead of print]
    • (2013) Eur J Radiol. , vol.82 , Issue.2 , pp. 275-280
    • Takx, R.A.1    Schoepf, U.J.2    Moscariello, A.3
  • 22
    • 84877154860 scopus 로고    scopus 로고
    • Cardiovascular CT angiography in neonates and children: image quality and potential for radiation dose reduction with iterative image reconstruction techniques
    • [Epub ahead of print
    • Tricarico F, Hlavacek AM, Schoepf UJ, et al. Cardiovascular CT angiography in neonates and children: image quality and potential for radiation dose reduction with iterative image reconstruction techniques. Eur Radiol. 2013;23(5):1306-15. [Epub ahead of print]
    • (2013) Eur Radiol. , vol.23 , Issue.5 , pp. 1306-1315
    • Tricarico, F.1    Hlavacek, A.M.2    Schoepf, U.J.3
  • 24
    • 84880691681 scopus 로고    scopus 로고
    • Catphan 600 user manual
    • The Phantom Laboratory.Salem, NY: The Phantom Laboratory
    • The Phantom Laboratory. Catphan 600 user manual. Salem, NY: The Phantom Laboratory; 2012.
    • (2012)
  • 25
    • 84880672889 scopus 로고    scopus 로고
    • Quality Assurance in Radiology and Medicine [website]. Available at:
    • Quality Assurance in Radiology and Medicine [website]. Available at: www.qrm.de
  • 27
    • 84871759658 scopus 로고    scopus 로고
    • Iterative reconstruction methods in two different MDCT scanners: physical metrics and 4-alternative forced-choice detectability experiments - a phantom approach
    • [Epub ahead of print
    • Miéville FA, Gudinchet F, Brunelle F, Bochud FO, Verdun FR. Iterative reconstruction methods in two different MDCT scanners: physical metrics and 4-alternative forced-choice detectability experiments - a phantom approach. Phys Med. 2013;29(1):99-110. [Epub ahead of print]
    • (2013) Phys Med. , vol.29 , Issue.1 , pp. 99-110
    • Miéville, F.A.1    Gudinchet, F.2    Brunelle, F.3    Bochud, F.O.4    Verdun, F.R.5
  • 29
    • 84867302075 scopus 로고    scopus 로고
    • Quantitative comparison of noise texture across CT scanners from different manufacturers
    • Solomon JB, Christianson O, Samei E. Quantitative comparison of noise texture across CT scanners from different manufacturers. Med Phys. 2012;39(10):6048-55.
    • (2012) Med Phys. , vol.39 , Issue.10 , pp. 6048-6055
    • Solomon, J.B.1    Christianson, O.2    Samei, E.3
  • 30
    • 0036236361 scopus 로고    scopus 로고
    • Detection of low-contrast objects: experimental comparison of single- and multi-detector row CT with a phantom
    • Verdun FR, Denys A, Valley JF, Schnyder P, Meuli RA. Detection of low-contrast objects: experimental comparison of single- and multi-detector row CT with a phantom. Radiology. 2002;223(2):426-31.
    • (2002) Radiology. , vol.223 , Issue.2 , pp. 426-431
    • Verdun, F.R.1    Denys, A.2    Valley, J.F.3    Schnyder, P.4    Meuli, R.A.5
  • 31
    • 77953333396 scopus 로고    scopus 로고
    • Computed tomography commissioning programmes: how to obtain a reliable MTF with an automatic approach?
    • Miéville F, Beaumont S, Torfeh T, Gudinchet F, Verdun FR. Computed tomography commissioning programmes: how to obtain a reliable MTF with an automatic approach? Radiat Prot Dosimetry. 2010;139(1-3):443-48.
    • (2010) Radiat Prot Dosimetry , vol.139 , Issue.1-3 , pp. 443-448
    • Miéville, F.1    Beaumont, S.2    Torfeh, T.3    Gudinchet, F.4    Verdun, F.R.5


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.