Model observer detectability as a substitute for contrast detail analysis in routine digital mammography quality control

Author keywords

Contrast detail analysis; Detectability index; FFDM; Mean glandular dose; Quality control

Indexed keywords

COMPUTED RADIOGRAPHY; CONTRAST DETAIL ANALYSIS; DETECTABILITY INDEX; DIGITAL MAMMOGRAPHY; FFDM; INTEGRATED DETECTORS; MEAN GLANDULAR DOSE; PEARSON CORRELATION;

EID: 84878323870     PISSN: 16057422     EISSN: None     Source Type: Conference Proceeding    
DOI: 10.1117/12.2002408     Document Type: Conference Paper
Times cited : (3)

References (9)
  • 2
    • 79960374186 scopus 로고    scopus 로고
    • Image quality assessment in digital mammography: Part ii. NPWE as a validated alternative for contrast detail analysis
    • Monnin, P., Marshall, N.W., Bosmans, H., Bochud, F.O., and Verdun, F.R., "Image quality assessment in digital mammography: part II. NPWE as a validated alternative for contrast detail analysis., " Phys. Med. Biol. 56(14), 4221-4238 (2011).
    • (2011) Phys. Med. Biol , vol.56 , Issue.14 , pp. 4221-4238
    • Monnin, P.1    Marshall, N.W.2    Bosmans, H.3    Bochud, F.O.4    Verdun, F.R.5
  • 3
    • 84878281135 scopus 로고    scopus 로고
    • www.artinis.com.
  • 4
    • 84878311310 scopus 로고    scopus 로고
    • www.qaelum.com.
  • 5
    • 33745416524 scopus 로고    scopus 로고
    • Comparison of software and human observers in reading images of the cdmam test object to assess digital mammography systems
    • 614206-614206-13
    • Young, K.C., Cook, J.J.H., Oduko, J.M., and Bosmans, H., "Comparison of software and human observers in reading images of the CDMAM test object to assess digital mammography systems, " Proceedings of SPIE 6142, 614206-614206-13 (2006).
    • (2006) Proceedings of SPIE , vol.6142
    • Young, K.C.1    Cook, J.J.H.2    Oduko, J.M.3    Bosmans, H.4
  • 6
    • 0018524989 scopus 로고
    • Motion and vision. Ii. Stabilized spatio-temporal threshold surface
    • Kelly, D.H., "Motion and vision. II. Stabilized spatio-temporal threshold surface., " J. Opt. Soc. Am. 69(10), 1340-1349 (1979).
    • (1979) J. Opt. Soc. Am. , vol.69 , Issue.10 , pp. 1340-1349
    • Kelly, D.H.1
  • 7
    • 0031950889 scopus 로고    scopus 로고
    • A method for measuring the presampled mtf of digital radiographic systems using an edge test device
    • Samei, E., Flynn, M.J., and Reimann, D.A., "A method for measuring the presampled MTF of digital radiographic systems using an edge test device., " Med. Phys. 25(1), 102-13 (1998).
    • (1998) Med. Phys. , vol.25 , Issue.1 , pp. 102-113
    • Samei, E.1    Flynn, M.J.2    Reimann, D.A.3
  • 8
    • 33646485572 scopus 로고    scopus 로고
    • A comparison between objective and subjective image quality measurements for a full field digital mammography system
    • Marshall, N.W., "A comparison between objective and subjective image quality measurements for a full field digital mammography system., " Phys. Med. Biol. 51(10), 2441-63 (2006).
    • (2006) Phys. Med. Biol , vol.51 , Issue.10 , pp. 2441-2463
    • Marshall, N.W.1
  • 9
    • 0033778108 scopus 로고    scopus 로고
    • Additional factors for the estimation of mean glandular breast dose using the uk mammography dosimetry protocol
    • Dance, D.R., Skinner, C.L., Young, K.C., Beckett, J.R., and Kotre, C.J., "Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol., " Phys. Med. Biol. 45(11), 3225-3240 (2000).
    • (2000) Phys. Med. Biol. , vol.45 , Issue.11 , pp. 3225-3240
    • Dance, D.R.1    Skinner, C.L.2    Young, K.C.3    Beckett, J.R.4    Kotre, C.J.5


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.