Evaluation of pulse rate variability obtained by the pulse onsets of the photoplethysmographic signal

Author keywords

HRV; PPG; PRV; pulse onset; variability

Indexed keywords

CORRELATION COEFFICIENT; HEART RATE VARIABILITY; HIGH FREQUENCY HF; LOWER FREQUENCIES; PHOTOPLETHYSMOGRAPHIC; PHOTOPLETHYSMOGRAPHIC SIGNALS; PULSE ONSET; PULSE RATE VARIABILITY; VARIABILITY; VARIABILITY INDEX;

EID: 84873389701     PISSN: 09673334     EISSN: 13616579     Source Type: Journal    
DOI: 10.1088/0967-3334/34/2/179     Document Type: Article
Times cited : (35)

References (48)
  • 2
    • 0019827638 scopus 로고
    • Power spectrum analysis of heart rate fluctuation: A quantitative probe of beat-to-beat cardiovascular control
    • Akselrod S, Gordon D, Ubel F A, Shannon D C, Berger A C and Cohen R J 1981 Power spectrum analysis of heart rate fluctuation: a quantitative probe of beat-to-beat cardiovascular control Science 213 220-2 (Pubitemid 11060568)
    • (1981) Science , vol.213 , Issue.4504 , pp. 220-222
    • Akselrod, S.1    Gordon, D.2    Ubel, F.A.3
  • 3
    • 34247467602 scopus 로고    scopus 로고
    • Photoplethysmography and its application in clinical physiological measurement
    • 0967-3334 R01
    • Allen J 2007 Photoplethysmography and its application in clinical physiological measurement Physiol. Meas. 28 R1-39
    • (2007) Physiol. Meas. , vol.28 , pp. 1-39
    • Allen, J.1
  • 7
    • 84873368330 scopus 로고    scopus 로고
    • An automated algorithm for fast pulse wave detection
    • Bistra N and Ivo I 2010 An automated algorithm for fast pulse wave detection Int. J. BIOautomation 14 203-16
    • (2010) Int. J. BIOautomation , vol.14 , pp. 203-216
    • Bistra, N.1    Ivo, I.2
  • 8
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • PII S0140673686908378
    • Bland J M and Altman D G 1986 Statistical methods for assessing agreement between two methods of clinical measurement Lancet 1 307-10 (Pubitemid 16134762)
    • (1986) Lancet , vol.1 , Issue.8476 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 9
    • 0000461905 scopus 로고
    • Photoelectric plethysmography for estimating cutaneous blood flow
    • Challoner A V J 1979 Photoelectric plethysmography for estimating cutaneous blood flow Non-Invasive Physiological Measurements vol 1 ed P Rolfe (London: Academic) pp 125-51
    • (1979) Non-Invasive Physiological Measurements , vol.1 , pp. 125-151
    • Challoner, A.V.J.1
  • 10
    • 76249091685 scopus 로고    scopus 로고
    • Interchangeability between heart rate and photoplethysmography variabilities during sympathetic stimulations
    • 10.1088/0967-3334/30/12/005 0967-3334
    • Charlot K, Cornolo J, Brugniaux J V, Richalet J P and Pichon A 2009 Interchangeability between heart rate and photoplethysmography variabilities during sympathetic stimulations Physiol. Meas. 30 1357-69
    • (2009) Physiol. Meas. , vol.30 , pp. 1357-1369
    • Charlot, K.1    Cornolo, J.2    Brugniaux, J.V.3    Richalet, J.P.4    Pichon, A.5
  • 11
    • 0025763193 scopus 로고
    • Determination of pulse wave velocities with computerized algorithms
    • 10.1016/0002-8703(91)90153-9 0002-8703
    • Chiu Y C, Arand P W, Shroff S G, Feldman T and Carroll J D 1991 Determination of pulse wave velocities with computerized algorithms Am. Heart J. 121 1460-70
    • (1991) Am. Heart J. , vol.121 , pp. 1460-1470
    • Chiu, Y.C.1    Arand, P.W.2    Shroff, S.G.3    Feldman, T.4    Carroll, J.D.5
  • 14
    • 75049086062 scopus 로고    scopus 로고
    • Beyond blood pressure: Pulse wave analysis a better way of assessing cardiovascular risk?
    • 10.1517/14796678.1.1.69 1479-6678
    • Davies J I and Struthers A D 2004 Beyond blood pressure: pulse wave analysis a better way of assessing cardiovascular risk? Future Cardiol. 1 69-78
    • (2004) Future Cardiol. , vol.1 , pp. 69-78
    • Davies, J.I.1    Struthers, A.D.2
  • 15
    • 58149085043 scopus 로고    scopus 로고
    • Three section transmission line arterial model for noninvasive assessment of vascular remodelingin primary hypertension
    • 10.1016/j.bspc.2008.07.001 1746-8094
    • de Sá Ferreira A, Filho J B, Cordovil I and de Souza M N 2009 Three section transmission line arterial model for noninvasive assessment of vascular remodelingin primary hypertension Biomed. Signal Process. Control 4 2-6
    • (2009) Biomed. Signal Process. Control , vol.4 , pp. 2-6
    • De Sá Ferreira, A.1    Filho, J.B.2    Cordovil, I.3    De Souza, M.N.4
  • 16
    • 0030666517 scopus 로고    scopus 로고
    • Sympathovagal balance: A critical appraisal
    • Eckberg D L 1997 Sympathovagal balance: a critical appraisal Circulation 96 3224-32 (Pubitemid 27500129)
    • (1997) Circulation , vol.96 , Issue.9 , pp. 3224-3232
    • Eckberg, D.L.1
  • 17
    • 78149474788 scopus 로고    scopus 로고
    • Mathematical methods for determining the foot point of the arterial pulse wave and evaluation of proposed methods
    • Egidijus K R G and Arunas V 2005 Mathematical methods for determining the foot point of the arterial pulse wave and evaluation of proposed methods Inform. Technol. Control 34 29-36
    • (2005) Inform. Technol. Control , vol.34 , pp. 29-36
    • Egidijus, K.R.G.1    Arunas, V.2
  • 18
    • 78149330175 scopus 로고    scopus 로고
    • Photoplethysmography pulse rate variability as a surrogate measurement of heart rate variability during non-stationary conditions
    • 10.1088/0967-3334/31/9/015 0967-3334
    • Gil E, Orini M, Bailón R, Vergara J M, Mainardi L and Laguna P 2010 Photoplethysmography pulse rate variability as a surrogate measurement of heart rate variability during non-stationary conditions Physiol. Meas. 31 1271-90
    • (2010) Physiol. Meas. , vol.31 , pp. 1271-1290
    • Gil, E.1    Orini, M.2    Bailón, R.3    Vergara, J.M.4    Mainardi, L.5    Laguna, P.6
  • 19
    • 27744434850 scopus 로고    scopus 로고
    • Assessment of pulse rate variability by the method of pulse frequency demodulation
    • DOI 10.1186/1475-925X-4-62
    • Hayano J, Barros A K, Kamiya A, Ohte N and Yasuma F 2005 Assessment of pulse rate variability by the method of pulse frequency demodulation Biomed. Eng. Online 4 62 (Pubitemid 41631784)
    • (2005) BioMedical Engineering Online , vol.4 , pp. 62
    • Hayano, J.1    Barros, A.K.2    Kamiya, A.3    Ohte, N.4    Yasuma, F.5
  • 20
    • 42649097909 scopus 로고    scopus 로고
    • An algorithm for extracting intracranial pressure latency relative to electrocardiogram R wave
    • DOI 10.1088/0967-3334/29/4/004, PII S0967333408672585
    • Hu X, Xu P, Lee D, Vespa P and Bergsneider M 2008 An algorithm of extracting intracranial pressure latency relative to electrocardiogram R wave Physiol. Meas. 29 459-71 (Pubitemid 351595093)
    • (2008) Physiological Measurement , vol.29 , Issue.4 , pp. 459-471
    • Hu, X.1    Xu, P.2    Lee, D.J.3    Vespa, P.4    Baldwin, K.5    Bergsneider, M.6
  • 21
    • 0029045838 scopus 로고
    • Impaired low-frequency oscillations of heart rate in patients with prior acute myocardial infarction and life-threatening arrhythmias
    • 10.1016/S0002-9149(99)80801-7 0002-9149
    • Huikuri H V, Koistinen M J, Yli-Mayry S, Airaksinen K E, Seppanen T, Ikaheimo M J and Myerburg R J 1995 Impaired low-frequency oscillations of heart rate in patients with prior acute myocardial infarction and life-threatening arrhythmias Am. J. Cardiol. 76 56-60
    • (1995) Am. J. Cardiol. , vol.76 , pp. 56-60
    • Huikuri, H.V.1    Koistinen, M.J.2    Yli-Mayry, S.3    Airaksinen, K.E.4    Seppanen, T.5    Ikaheimo, M.J.6    Myerburg, R.J.7
  • 22
    • 0023100673 scopus 로고
    • Decreased heart rate variability and its association with increased mortality after acute myocardial infarction
    • DOI 10.1016/0002-9149(87)90795-8
    • Kleiger R E, Miller J P, Bigger J T Jr and Moss A J 1987 Decreased heart rate variability and its association with increased mortality after acute myocardial infarction Am. J. Cardiol. 59 256-62 (Pubitemid 17036947)
    • (1987) American Journal of Cardiology , vol.59 , Issue.4 , pp. 256-262
    • Kleiger, R.E.1    Miller, J.P.2    Bigger Jr., J.T.3
  • 23
    • 0024521543 scopus 로고
    • A concordance correlation coefficient to evaluate reproducibility
    • DOI 10.2307/2532051
    • Lin L I 1989 A concordance correlation coefficient to evaluate reproducibility Biometrics 45 255-68 (Pubitemid 19109172)
    • (1989) Biometrics , vol.45 , Issue.1 , pp. 255-268
    • I-Kuei Lin, L.1
  • 24
    • 70349306240 scopus 로고    scopus 로고
    • Limitations of oximetry to measure heart rate variability measures
    • 10.1007/s10558-009-9082-3 1567-8822
    • Lu G and Yang F 2009 Limitations of oximetry to measure heart rate variability measures Cardiovasc. Eng. 9 119-25
    • (2009) Cardiovasc. Eng. , vol.9 , pp. 119-125
    • Lu, G.1    Yang, F.2
  • 25
    • 70350439728 scopus 로고    scopus 로고
    • A comparison of photoplethysmography and ECG recording to analyse heart rate variability in healthy subjects
    • 10.3109/03091900903150998 0309-1902
    • Lu G, Yang F, Taylor J A and Stein J F 2009 A comparison of photoplethysmography and ECG recording to analyse heart rate variability in healthy subjects J. Med. Eng. Technol. 33 634-41
    • (2009) J. Med. Eng. Technol. , vol.33 , pp. 634-641
    • Lu, G.1    Yang, F.2    Taylor, J.A.3    Stein, J.F.4
  • 26
    • 39949085790 scopus 로고    scopus 로고
    • Can photoplethysmography variability serve as an alternative approach to obtain heart rate variability information?
    • 10.1007/s10877-007-9103-y 1387-1307
    • Lu S, Zhao H, Ju K, Shin K, Lee M, Shelley K and Chon K 2008 Can photoplethysmography variability serve as an alternative approach to obtain heart rate variability information? J. Clin. Monit. Comput. 22 23-9
    • (2008) J. Clin. Monit. Comput. , vol.22 , pp. 23-29
    • Lu, S.1    Zhao, H.2    Ju, K.3    Shin, K.4    Lee, M.5    Shelley, K.6    Chon, K.7
  • 27
    • 0242585020 scopus 로고    scopus 로고
    • ECG artifact due to deep brain stimulation
    • DOI 10.1016/S0140-6736(03)13136-4
    • Martin W A, Camenzind E and Burkhard P R 2003 ECG artifact due to deep brain stimulation Lancet 361 1431 (Pubitemid 36515044)
    • (2003) Lancet , vol.361 , Issue.9367 , pp. 1431
    • Martin, W.A.1    Camenzind, E.2    Burkhard, P.R.3
  • 32
    • 0033139915 scopus 로고    scopus 로고
    • Isolated systolic hypertension, pulse pressure, and arterial stiffness as risk factors for cardiovascular disease
    • 10.1007/s11906-999-0022-9 1522-6417
    • O'Rourke M F 1999 Isolated systolic hypertension, pulse pressure, and arterial stiffness as risk factors for cardiovascular disease Curr. Hypertens. Rep. 1 204-11
    • (1999) Curr. Hypertens. Rep. , vol.1 , pp. 204-211
    • O'Rourke, M.F.1
  • 34
    • 0021912061 scopus 로고
    • Assessment of autonomic function in humans by heart rate spectral analysis
    • 0002-9513
    • Pomeranz B et al 1985 Assessment of autonomic function in humans by heart rate spectral analysis Am. J. Physiol. 248 H151-3
    • (1985) Am. J. Physiol. , vol.248
    • Pomeranz, B.1
  • 35
    • 58149202537 scopus 로고    scopus 로고
    • Comparison of time-domain short-term heart interval variability analysis using a wrist-worn heart rate monitor and the conventional electrocardiogram
    • 10.1111/j.1540-8159.2009.02175.x 0147-8389
    • Porto L G and Junqueira L F Jr 2009 Comparison of time-domain short-term heart interval variability analysis using a wrist-worn heart rate monitor and the conventional electrocardiogram Pacing Clin. Electrophysiol. 32 43-51
    • (2009) Pacing Clin. Electrophysiol. , vol.32 , pp. 43-51
    • Porto, L.G.1    Junqueira Jr., L.F.2
  • 36
    • 17144455895 scopus 로고    scopus 로고
    • A comparison between systolic and diastolic pulse contour analysis in the evaluation of arterial stiffness
    • 10.1161/01.HYP.37.6.e15 0194-911X
    • Rietzschel E R, Boeykens E, Buyzere M L D, Duprez D A and Clement D L 2001 A comparison between systolic and diastolic pulse contour analysis in the evaluation of arterial stiffness Hypertension 37 e15-22
    • (2001) Hypertension , vol.37
    • Rietzschel, E.R.1    Boeykens, E.2    Buyzere, M.L.D.3    Duprez, D.A.4    Clement, D.L.5
  • 37
    • 0031106763 scopus 로고    scopus 로고
    • A real-time microprocessor QRS detector system with a 1-ms timing accuracy for the measurement of ambulatory HRV
    • DOI 10.1109/10.554762
    • Ruha A, Sallinen S and Nissila S 1997 A real-time microprocessor QRS detector system with a 1 ms timing accuracy for the measurement of ambulatory HRV IEEE Trans. Biomed. Eng. 44 159-67 (Pubitemid 27084201)
    • (1997) IEEE Transactions on Biomedical Engineering , vol.44 , Issue.3 , pp. 159-167
    • Ruha, A.1    Sallinen, S.2    Nissila, S.3
  • 38
    • 84877733859 scopus 로고    scopus 로고
    • How accurate is pulse rate variability as an estimate of heart rate variability? A review on studies comparing photoplethysmographic technology with an electrocardiogram
    • 0167-5273
    • Schäfer A and Vagedes J 2012 How accurate is pulse rate variability as an estimate of heart rate variability? A review on studies comparing photoplethysmographic technology with an electrocardiogram Int J. Cardiol. doi:10.1016/j.ijcard.2012.03.119
    • (2012) Int J. Cardiol.
    • Schäfer, A.1    Vagedes, J.2
  • 39
    • 56249102489 scopus 로고    scopus 로고
    • Assessment of heart rate variability derived from finger-tip photoplethysmography as compared to electrocardiography
    • 10.1080/03091900701781317 0309-1902
    • Selvaraj N, Jaryal A K, Santhosh J, Deepak K K and Anand S 2008 Assessment of heart rate variability derived from finger-tip photoplethysmography as compared to electrocardiography J. Med. Eng. Technol. 32 479-84
    • (2008) J. Med. Eng. Technol. , vol.32 , pp. 479-484
    • Selvaraj, N.1    Jaryal, A.K.2    Santhosh, J.3    Deepak, K.K.4    Anand, S.5
  • 40
    • 71549152549 scopus 로고    scopus 로고
    • Adaptive threshold method for the peak detection of photoplethysmographic waveform
    • 10.1016/j.compbiomed.2009.10.006 0010-4825
    • Shin H S, Lee C and Lee M 2009 Adaptive threshold method for the peak detection of photoplethysmographic waveform Comput. Biol. Med. 39 1145-52
    • (2009) Comput. Biol. Med. , vol.39 , pp. 1145-1152
    • Shin, H.S.1    Lee, C.2    Lee, M.3
  • 41
    • 38149133210 scopus 로고    scopus 로고
    • A new heart rate variability-based method for the estimation of oxygen consumption without individual laboratory calibration: Application example on postal workers
    • 10.1016/j.apergo.2007.09.001 0003-6870
    • Smolander J, Juuti T, Kinnunen M L, Laine K, Louhevaara V, Mannikko K and Rusko H 2008 A new heart rate variability-based method for the estimation of oxygen consumption without individual laboratory calibration: application example on postal workers Appl. Ergon. 39 325-31
    • (2008) Appl. Ergon. , vol.39 , pp. 325-331
    • Smolander, J.1    Juuti, T.2    Kinnunen, M.L.3    Laine, K.4    Louhevaara, V.5    Mannikko, K.6    Rusko, H.7
  • 42
    • 0031294929 scopus 로고    scopus 로고
    • Development of QRS detection method for realtime ambulatory cardiac monitor
    • So H H and Chan K L 1997 Development of QRS detection method for realtime ambulatory cardiac monitor Proc. IEEE EMBS (Chicago USA,) pp 289-92
    • (1997) Proc. IEEE EMBS , pp. 289-292
    • So, H.H.1    Chan, K.L.2
  • 43
    • 0029988782 scopus 로고    scopus 로고
    • Task Force of the European Society of Cardiology and the North American Society of Pacing and Electrophysiology. Heart rate variability: Standards of measurement, physiological interpretation, and clinical use
    • 10.1093/oxfordjournals.eurheartj.a014868 0195-668X
    • Task Force 1996 Task Force of The European Society of Cardiology and The North American Society of Pacing and Electrophysiology. Heart rate variability: standards of measurement, physiological interpretation, and clinical use Eur. Heart J. 17 354-81
    • (1996) Eur. Heart J. , vol.17 , pp. 354-381
  • 44
    • 79955041195 scopus 로고    scopus 로고
    • BioSig: The free and open source software library for biomedical signal processing
    • 10.1155/2011/935364 1687-5265 935364
    • Vidaurre C, Sander T H and Schlögl A 2011 BioSig: the free and open source software library for biomedical signal processing Comput. Intell. Neurosci. 2011 935364 Available at: http://www.pubmedcentral.nih.gov/ articlerender.fcgi?artid=3061298&tool=pmcentrez&rendertype=abstract
    • (2011) Comput. Intell. Neurosci. , vol.2011
    • Vidaurre, C.1    Sander, T.H.2    Schlögl, A.3
  • 45
    • 33644863316 scopus 로고    scopus 로고
    • Prognostic value of aortic pulse wave velocity as index of arterial stiffness in the general population
    • 10.1161/CIRCULATIONAHA.105.579342 0009-7322
    • Willum-Hansen T, Staessen J A, Torp-Pedersen C, Rasmussen S, Thijs L, Ibsen H and Jeppesen J 2006 Prognostic value of aortic pulse wave velocity as index of arterial stiffness in the general population Circulation 113 664-70
    • (2006) Circulation , vol.113 , pp. 664-670
    • Willum-Hansen, T.1    Staessen, J.A.2    Torp-Pedersen, C.3    Rasmussen, S.4    Thijs, L.5    Ibsen, H.6    Jeppesen, J.7
  • 46
    • 84863095122 scopus 로고    scopus 로고
    • A comparative study of pulse rate variability and heart rate variability in healthy subjects
    • 10.1007/s10877-012-9340-6 1387-1307
    • Wong J S, Lu W A, Wu K T, Liu M, Chen G Y and Kuo C D 2012 A comparative study of pulse rate variability and heart rate variability in healthy subjects J. Clin. Monit. Comput. 26 107-14
    • (2012) J. Clin. Monit. Comput. , vol.26 , pp. 107-114
    • Wong, J.S.1    Lu, W.A.2    Wu, K.T.3    Liu, M.4    Chen, G.Y.5    Kuo, C.D.6
  • 47
    • 60749117508 scopus 로고    scopus 로고
    • Pulse onset detection using neighbor pulse-based signal enhancement
    • 10.1016/j.medengphy.2008.06.005 1350-4533
    • Xu P, Bergsneider M and Hu X 2009 Pulse onset detection using neighbor pulse-based signal enhancement Med. Eng. Phys. 31 337-45
    • (2009) Med. Eng. Phys. , vol.31 , pp. 337-345
    • Xu, P.1    Bergsneider, M.2    Hu, X.3
  • 48
    • 4143146627 scopus 로고    scopus 로고
    • An open-source algorithm to detect onset of arterial blood pressure pulses
    • 0276-6574
    • Zong W, Heldt T, Moody G B and Mark R G 2003 An open-source algorithm to detect onset of arterial blood pressure pulses Comput. Cardiol. 30 259-62
    • (2003) Comput. Cardiol. , vol.30 , pp. 259-262
    • Zong, W.1    Heldt, T.2    Moody, G.B.3    Mark, R.G.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.