Volumn 59, Issue 2, 2013, Pages 363-371

External quality assessment of point-of-care methods: Model for combined assessment of method bias and single-participant performance by the use of native patient samples and noncommutable control materials

Author keywords

[No Author keywords available]

Indexed keywords

ARTICLE; HUMAN; INTERNATIONAL NORMALIZED RATIO; PATIENT PARTICIPATION; POINT OF CARE TESTING; PRIMARY HEALTH CARE; QUALITY CONTROL; SYSTEMATIC ERROR;

EID: 84873199537     PISSN: 00099147     EISSN: 15308561     Source Type: Journal    
DOI: 10.1373/clinchem.2012.191957     Document Type: Article
Times cited : (22)

References (29)
  • 1
    • 84857887803 scopus 로고    scopus 로고
    • Quality management
    • Burtis CA, Ashwood ER, Bruns DE, eds, Oxford: Elsevier Saunders
    • Klee GG, Westgard JO. Quality management. In: Burtis CA, Ashwood ER, Bruns DE, eds. Tietz fundamentals of clinical chemistry. 6th ed. Oxford: Elsevier Saunders; 2008. p 249 -62.
    • (2008) Tietz Fundamentals of Clinical Chemistry. 6th Ed , pp. 249-262
    • Klee, G.G.1    Westgard, J.O.2
  • 2
    • 82755167810 scopus 로고    scopus 로고
    • Proficiency testing/external quality assessment: Current challenges and future directions
    • Miller WG, Jones GR, Horowitz GL, Weykamp C. Proficiency testing/external quality assessment: current challenges and future directions. Clin Chem 2011;57:1670-80.
    • (2011) Clin Chem , vol.57 , pp. 1670-1680
    • Miller, W.G.1    Jones, G.R.2    Horowitz, G.L.3    Weykamp, C.4
  • 3
    • 0029813456 scopus 로고    scopus 로고
    • Characterization and classification of external quality assessment schemes (EQA) according to objectives such as evaluation of method and participant bias and standard deviation. External Quality Assessment (EQA) Working Group A on Analytical Goals in Laboratory Medicine
    • Libeer JC, Baadenhuijsen H, Fraser CG, Petersen PH, Ricos C, Stockl D, Thienpont L. Characterization and classification of external quality assessment schemes (EQA) according to objectives such as evaluation of method and participant bias and standard deviation. External Quality Assessment (EQA) Working Group A on Analytical Goals in Laboratory Medicine. Eur J Clin Chem Clin Biochem 1996;34:665-78.
    • (1996) Eur J Clin Chem Clin Biochem , vol.34 , pp. 665-678
    • Libeer, J.C.1    Baadenhuijsen, H.2    Fraser, C.G.3    Petersen, P.H.4    Ricos, C.5    Stockl, D.6    Thienpont, L.7
  • 7
    • 84873123855 scopus 로고    scopus 로고
    • CLSI. Quality management: approaches to reducing errors at the point of care; approved guide-line. CLSI document POCT07-A. Wayne (PA): CLSI; 2010
    • CLSI. Quality management: approaches to reducing errors at the point of care; approved guide-line. CLSI document POCT07-A. Wayne (PA): CLSI; 2010.
  • 8
    • 84873164802 scopus 로고    scopus 로고
    • NCCLS/CLSI. Point-of-care monitoring of anticoagulation therapy; approved guideline. NCCLS/CLSI document H49-A. Wayne (PA): NCCLS/CLSI; 2004
    • NCCLS/CLSI. Point-of-care monitoring of anticoagulation therapy; approved guideline. NCCLS/CLSI document H49-A. Wayne (PA): NCCLS/CLSI; 2004.
  • 12
    • 57149106471 scopus 로고    scopus 로고
    • Problems and practical solutions in the external quality control of point of care devices with respect to the measurement of blood glucose
    • Wood WG. Problems and practical solutions in the external quality control of point of care devices with respect to the measurement of blood glucose. J Diabetes Sci Technol 2007;1:158-63.
    • (2007) J Diabetes Sci Technol , vol.1 , pp. 158-163
    • Wood, W.G.1
  • 14
    • 84857821583 scopus 로고    scopus 로고
    • External quality assessment of point-of-care international normalized ratio (INR) testing in Europe
    • Stavelin A, Meijer P, Kitchen D, Sandberg S. External quality assessment of point-of-care international normalized ratio (INR) testing in Europe. Clin Chem Lab Med 2012;50:81-8.
    • (2012) Clin Chem Lab Med , vol.50 , pp. 81-88
    • Stavelin, A.1    Meijer, P.2    Kitchen, D.3    Sandberg, S.4
  • 15
    • 84873198137 scopus 로고    scopus 로고
    • CLSI. Method comparison and bias estimation using patient samples; approved guideline. 2nd ed (interim revision). CLSI document EP09 -A2-IR. Wayne (PA): CLSI 2010
    • CLSI. Method comparison and bias estimation using patient samples; approved guideline. 2nd ed (interim revision). CLSI document EP09 -A2-IR. Wayne (PA): CLSI; 2010.
  • 16
    • 56749154425 scopus 로고    scopus 로고
    • Laboratory equipment for primary health care should be tested by independent authorities. Scandinavian SKUP is a good alternative
    • 3338-9. [Swedish]
    • Sandberg S, Nordin G, Martensson A, Grinsted P, Jensen E, Jacobsen CE, Monsen G. [Laboratory equipment for primary health care should be tested by independent authorities. Scandinavian SKUP is a good alternative]. Lakartidningen 2008; 105:3334-6; 3338-9. [Swedish]
    • (2008) Lakartidningen , vol.105 , pp. 3334-3336
    • Sandberg, S.1    Nordin, G.2    Martensson, A.3    Grinsted, P.4    Jensen, E.5    Jacobsen, C.E.6    Monsen, G.7
  • 17
    • 49749212311 scopus 로고
    • Thrombotest. A new method for controlling anticoagulant therapy
    • Owren PA. Thrombotest. A new method for controlling anticoagulant therapy. Lancet 1959;2:754-8.
    • (1959) Lancet , vol.2 , pp. 754-758
    • Owren, P.A.1
  • 18
    • 3042582967 scopus 로고    scopus 로고
    • INR calibration of Owren-type prothrombin time based on the relationship between PT% and INR utilizing normal plasma samples
    • Lindahl TL, Egberg N, Hillarp A, Odegaard OR, Edlund B, Svensson J, et al. INR calibration of Owren-type prothrombin time based on the relationship between PT% and INR utilizing normal plasma samples. Thromb Haemost 2004;91: 1223-31.
    • (2004) Thromb Haemost , vol.91 , pp. 1223-1231
    • Lindahl, T.L.1    Egberg, N.2    Hillarp, A.3    Odegaard, O.R.4    Edlund, B.5    Svensson, J.6
  • 19
    • 0016585365 scopus 로고
    • Accurate estimation of standard deviations for quantitative methods used in clinical chemistry
    • Burnett RW. Accurate estimation of standard deviations for quantitative methods used in clinical chemistry. Clin Chem 1975;21:1935-8.
    • (1975) Clin Chem , vol.21 , pp. 1935-1938
    • Burnett, R.W.1
  • 21
    • 14544287715 scopus 로고    scopus 로고
    • Preanalytical variables and off-site blood collection: Influences on the results of the prothrombin time/international normalized ratio test and implications for monitoring of oral anticoagulant therapy
    • van Geest-Daalderop JH, Mulder AB, Boonman-de Winter LJ, Hoekstra MM, van den Besselaar AM. Preanalytical variables and off-site blood collection: influences on the results of the prothrombin time/international normalized ratio test and implications for monitoring of oral anticoagulant therapy. Clin Chem 2005;51:561-8.
    • (2005) Clin Chem , vol.51 , pp. 561-568
    • Van Geest-Daalderop, J.H.1    Mulder, A.B.2    Boonman-De Winter, L.J.3    Hoekstra, M.M.4    Van Den Besselaar, A.M.5
  • 22
    • 44949155310 scopus 로고    scopus 로고
    • Stability of coagulation assays performed in plasma from citrated whole blood transported at ambient temperature: Only a part of the story
    • Favaloro EJ. Stability of coagulation assays performed in plasma from citrated whole blood transported at ambient temperature: only a part of the story. Thromb Haemost 2008;99:1122-3.
    • (2008) Thromb Haemost , vol.99 , pp. 1122-1123
    • Favaloro, E.J.1
  • 23
    • 84873103651 scopus 로고    scopus 로고
    • Regionernes Lønnings-og Takstnævn (RLTN) Og Praktiserende Lægers Organisation (PLO). Kvalitetssikring Og Kvalitetskrav Til Laboratoriemedicinske Aktiviteter I Almen Praksis. Konsensusdokument Danmark. (Accessed November 2012)
    • Regionernes Lønnings-og Takstnævn (RLTN) og Praktiserende Lægers Organisation (PLO). Kvalitetssikring og kvalitetskrav til laboratoriemedicinske aktiviteter i almen praksis. Konsensusdokument, Danmark; 2010. http://skup.dk/files/15/2010-kvalitetssikring-og-kvalitetskrav.pdf. (Accessed November 2012).
    • (2010)
  • 24
    • 0031855475 scopus 로고    scopus 로고
    • The accuracy of laboratory measurements in clinical chemistry: A study of 11 routine chemistry analytes in the College of American Pathologists Chemistry Survey with fresh frozen serum, definitive methods, and reference methods
    • Ross JW, Miller WG, Myers GL, Praestgaard J. The accuracy of laboratory measurements in clinical chemistry: a study of 11 routine chemistry analytes in the College of American Pathologists Chemistry Survey with fresh frozen serum, definitive methods, and reference methods. Arch Pathol Lab Med 1998;122:587-608.
    • (1998) Arch Pathol Lab Med , vol.122 , pp. 587-608
    • Ross, J.W.1    Miller, W.G.2    Myers, G.L.3    Praestgaard, J.4
  • 25
    • 0038129689 scopus 로고    scopus 로고
    • Trueness verification in European external quality assessment schemes: Time to care about the quality of the samples
    • Thienpont LM, Stockl D, Friedecky B, Kratochvila J, Budina M. Trueness verification in European external quality assessment schemes: time to care about the quality of the samples. Scand J Clin Lab Invest 2003;63:195-201.
    • (2003) Scand J Clin Lab Invest , vol.63 , pp. 195-201
    • Thienpont, L.M.1    Stockl, D.2    Friedecky, B.3    Kratochvila, J.4    Budina, M.5
  • 26
    • 14544272725 scopus 로고    scopus 로고
    • Poor agreement among prothrombin time international normalized ratio methods: Comparison of seven commercial reagents
    • Horsti J, Uppa H, Vilpo JA. Poor agreement among prothrombin time international normalized ratio methods: comparison of seven commercial reagents. Clin Chem 2005;51:553-60.
    • (2005) Clin Chem , vol.51 , pp. 553-560
    • Horsti, J.1    Uppa, H.2    Vilpo, J.A.3
  • 27
    • 77957731819 scopus 로고    scopus 로고
    • Discrepancies in international normalized ratio results between instruments: A model to split the variation into subcomponents
    • Solvik UO, Petersen PH, Monsen G, Stavelin AV, Sandberg S. Discrepancies in international normalized ratio results between instruments: a model to split the variation into subcomponents. Clin Chem 2010;56:1618 -26.
    • (2010) Clin Chem , vol.56 , pp. 1618-1626
    • Solvik, U.O.1    Petersen, P.H.2    Monsen, G.3    Stavelin, A.V.4    Sandberg, S.5
  • 29
    • 24144435337 scopus 로고    scopus 로고
    • Between-lot variation in external quality assessment of glucose: Clinical importance and effect on participant performance evaluation
    • Kristensen GB, Christensen NG, Thue G, Sandberg S. Between-lot variation in external quality assessment of glucose: clinical importance and effect on participant performance evaluation. Clin Chem 2005;51:1632-6.
    • (2005) Clin Chem , vol.51 , pp. 1632-1636
    • Kristensen, G.B.1    Christensen, N.G.2    Thue, G.3    Sandberg, S.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.