Volumn 59, Issue 3, 2012, Pages 608-618

Electrocardiogram-assisted blood pressure estimation

Author keywords

Blood pressure estimation; cuff pressure (CP); electrocardiogram (ECG); oscillometric analysis; pulse transit time (PTT)

Indexed keywords

BLOOD PRESSURE ESTIMATION; CUFF PRESSURE; ELECTROCARDIOGRAM (ECG); OSCILLOMETRIC ANALYSIS; PULSE TRANSIT TIME;

EID: 84863142760     PISSN: 00189294     EISSN: 15582531     Source Type: Journal    
DOI: 10.1109/TBME.2011.2180019     Document Type: Article
Times cited : (97)

References (38)
  • 1
    • 0037746333 scopus 로고    scopus 로고
    • Baltimore MD: Williams & Wilkins
    • R. E. Klabunde, Cardiovascular Physiology Concepts. Baltimore, MD: Williams & Wilkins, 2005.
    • (2005) Cardiovascular Physiology Concepts
  • 3
    • 0031148823 scopus 로고    scopus 로고
    • Theoretical analysis of non-invasive oscillometric maximum amplitude algorithm for estimating mean blood pressure
    • P. D. Baker, D. R. Westenskow, and K. Kück, "Theoretical analysis of non-invasive oscillometric maximum amplitude algorithm for estimating mean blood pressure," Med. Biol. Eng. Comput., vol. 35, pp. 271-278, 1997. (Pubitemid 27259596)
    • (1997) Medical and Biological Engineering and Computing , vol.35 , Issue.3 , pp. 271-278
    • Baker, P.D.1    Westenskow, D.R.2    Kück, K.3
  • 4
    • 58149359267 scopus 로고    scopus 로고
    • Comparison of noninvasive blood pressure measurements in patients with atrial fibrillation
    • Z. M. Anastas, E. Jimerson, and S. Garolis, "Comparison of noninvasive blood pressure measurements in patients with atrial fibrillation," J. Cardiovasc. Nurs., vol. 23, pp. 519-524, 2008.
    • (2008) J. Cardiovasc. Nurs. , vol.23 , pp. 519-524
    • Anastas, Z.M.1    Jimerson, E.2    Garolis, S.3
  • 5
    • 33745234213 scopus 로고    scopus 로고
    • Blood pressure measurement in patients with rate controlled atrial fibrillation using mercury sphygmomanometer and Omron HEM-750CP deice in the clinic setting
    • DOI 10.1038/sj.jhh.1002016, PII 1002016
    • B. Jani, C. J. Bulpitt, and C. Rajkumar, "Blood pressure measurement in patients with rate controlled atrial fibrillation using mercury sphygmomanometer and Omron HEM-750CP device in the clinic setting," J. Human Hypertens, vol. 20, pp. 543-545, 2006. (Pubitemid 43923127)
    • (2006) Journal of Human Hypertension , vol.20 , Issue.7 , pp. 543-545
    • Jani, B.1    Bulpitt, C.J.2    Rajkumar, C.3
  • 6
    • 0029831681 scopus 로고    scopus 로고
    • A century of confusion: Which bladder for accurate blood pressure measurement
    • E. O'Brien, "A century of confusion: which bladder for accurate blood pressure measurement?" J. Human Hypertens., vol. 10, pp. 565-572, 1996.
    • (1996) J. Human Hypertens. , vol.10 , pp. 565-572
    • O'Brien, E.1
  • 7
    • 2542527785 scopus 로고    scopus 로고
    • Many factors can affect the prevalence of hypertension in obese patients: Role of cuff size and type of obesity
    • M. T. Guagnano, V. P. Palitti, R. Murri, L. Marchione, D. Merlitti, and S. Sensi, "Many factors can affect the prevalence of hypertension in obesepatients: Role of cuff size and type of obesity," Panminerva Med, vol. 40, pp. 22-27, 1998. (Pubitemid 128527502)
    • (1998) Panminerva Medica , vol.40 , Issue.1 , pp. 22-27
    • Guagnano, M.T.1    Palitti, V.P.2    Murri, R.3    Marchione, L.4    Merlitti, D.5    Sensi, S.6
  • 8
    • 85008279049 scopus 로고    scopus 로고
    • ABC of hypertension: Blood pressure measurement-Part I: Sphygmomanometry: Factors common to all techniques
    • G. Beevers, G. Y. H.Lip, and E. O'Brien, "ABC of hypertension: Blood pressure measurement-Part I: Sphygmomanometry: Factors common to all techniques," Brit. Med. J., vol. 322, pp. 981-985, 2001.
    • (2001) Brit. Med. J. , vol.322 , pp. 981-985
    • Beevers, G.1    Lip, G.Y.H.2    O'Brien, E.3
  • 9
  • 12
    • 0032897226 scopus 로고    scopus 로고
    • Pulse transit time: An appraisal of potential clinical applications
    • R. P. Smith, J. Argod, J.-L. Pépin, and P. A. Lévy, "Pulse transit time: An appraisal of potential clinical applications," Thorax, vol. 54, pp. 452-457, 1999. (Pubitemid 29199788)
    • (1999) Thorax , vol.54 , Issue.5 , pp. 452-457
    • Smith, R.P.1    Argod, J.2    Pépin, J.-L.3    Lévy, P.A.4
  • 13
    • 33646751684 scopus 로고    scopus 로고
    • Pulse transit time as an indirect marker for variations in cardiovascular related reactivity
    • J. Y. A. Foo and C. S. Lim, "Pulse transit time as an indirect marker for variations in cardiovascular related reactivity," Technol. Health Care, vol. 14, pp. 97-108, 2006.
    • (2006) Technol. Health Care , vol.14 , pp. 97-108
    • Foo, J.Y.A.1    Lim, C.S.2
  • 15
    • 0000147387 scopus 로고
    • The effect of blood pressure alteration on the pulse wave velocity
    • E. R. Nye, "The effect of blood pressure alteration on the pulse wave velocity," Brit. Heart J., vol. 26, pp. 261-265, 1964.
    • (1964) Brit. Heart J. , vol.26 , pp. 261-265
    • Nye, E.R.1    Nye, R.E.2
  • 17
    • 77958021629 scopus 로고    scopus 로고
    • Pulse wave velocity in continuous blood pressure measurements
    • K. Meigas, J. Lass, D. Karai, R. Karai, and J. Kaik, "Pulse wave velocity in continuous blood pressure measurements," IFMBE Proc., vol. 14, pp. 626-629, 2007.
    • (2007) IFMBE Proc. , vol.14 , pp. 626-629
    • Meigas, K.1    Lass, J.2    Karai, D.3    Karai, R.4    Kaik, J.5
  • 19
    • 41149157083 scopus 로고    scopus 로고
    • Pulse Wave Velocity Is an Independent Predictor of the Longitudinal Increase in Systolic Blood Pressure and of Incident Hypertension in the Baltimore Longitudinal Study of Aging
    • DOI 10.1016/j.jacc.2007.10.065, PII S073510970800329X
    • S. S. Najjar, A. Scuteri, V. Shetty, J. G. Wright, D. C. Muller, J. L. Fleg, H.P. Spurgeon, L. Ferrucci, and E. G. Lakatta, "Pulse wave velocity is an independent predictor of the longitudinal increase in systolic blood pressure and of incident hypertension in the Baltimore longitudinal study of aging," J. Amer. Coll. Cardiol., vol. 51, pp. 1377-1383, 2008. (Pubitemid 351443076)
    • (2008) Journal of the American College of Cardiology , vol.51 , Issue.14 , pp. 1377-1383
    • Najjar, S.S.1    Scuteri, A.2    Shetty, V.3    Wright, J.G.4    Muller, D.C.5    Fleg, J.L.6    Spurgeon, H.P.7    Ferrucci, L.8    Lakatta, E.G.9
  • 20
    • 67650689838 scopus 로고    scopus 로고
    • Non-constrained blood pressure monitoring using ECG and PPG for personal healthcare
    • Y. Yoon, J. H.Cho, and G. Yoon, "Non-constrained blood pressure monitoring using ECG and PPG for personal healthcare," J. Med. Syst., vol. 33, pp. 261-266, 2009.
    • (2009) J. Med. Syst. , vol.33 , pp. 261-266
    • Yoon, Y.1    Cho, J.H.2    Yoon, G.3
  • 22
    • 0020684665 scopus 로고
    • Pulse transit time and blood pressure: An intensive analysis
    • DOI 10.1111/j.1469-8986.1983.tb00899.x
    • J. D. Lane, L. Greenstadt, D. Shapiro, and R. Rubinstein, "Pulse transit time and blood pressure: An intensive analysis," Psychophysiology, vol. 20, pp. 45-49, 1983. (Pubitemid 13224564)
    • (1983) Psychophysiology , vol.20 , Issue.1 , pp. 45-49
    • Lane, J.D.1    Greenstadt, L.2    Shapiro, D.3    Rubinstein, R.4
  • 23
    • 0022385238 scopus 로고
    • A correlation analysis between pulse transit time and instantaneous blood pressure measured indirectly by the vascular unloading method
    • DOI 10.1016/0301-0511(85)90049-3
    • Y. Sawada and K. Yamakoshi, "A correlation analysis between pulse transit time and instantaneous blood pressure measured indirectly by the vascular unloading method," Biol. Psychol., vol. 21, pp. 1-9, 1985. (Pubitemid 16197600)
    • (1985) Biological Psychology , vol.21 , Issue.1 , pp. 1-9
    • Sawada, Y.1    Yamakoshi, K.2
  • 24
    • 33644822671 scopus 로고    scopus 로고
    • Pulse transit time measured from the ECG: An unreliable marker of beat-to-beat blood pressure
    • DOI 10.1152/japplphysiol.00657.2005
    • R. A. Payne, C. N. Symeonides, D. J.Webb, and S. R. J. Maxwell, "Pulse transit time measured from the ECG: An unreliable marker of beat-to-beat blood pressure," J. Appl. Physiol., vol. 100, pp. 136-141, 2006. (Pubitemid 43671100)
    • (2006) Journal of Applied Physiology , vol.100 , Issue.1 , pp. 136-141
    • Payne, R.A.1    Symeonides, C.N.2    Webb, D.J.3    Maxwell, S.R.J.4
  • 25
    • 0019620689 scopus 로고
    • Pulse arrival time as a method of obtaining systolic and diastolic blood pressure indirectly
    • L. A. Geddes, M. Voelz, S. James, and D. Reiner, "Pulse arrival time as a method of obtaining systolic and diastolic blood pressure indirectly," Technical Note, Med. Biol. Eng. Comput., vol. 19, pp. 671-672, 1981. (Pubitemid 12248223)
    • (1981) Medical and Biological Engineering and Computing , vol.19 , Issue.5 , pp. 671-672
    • Geddes, L.A.1    Voelz, M.2    James, S.3    Reiner, D.4
  • 26
    • 0027464195 scopus 로고
    • Validation of a method for noninvasive measurement of central arterial pressure
    • T. Sharir, A. Marmor, C. T. Ting, J. W. Chen, C. P. Liu, M. S. Chang, F. C. Yin, and D. A. Kass, "Validation of a method for noninvasive measurement of central arterial pressure," Hypertension, vol. 21, pp. 74-82, 1993. (Pubitemid 23006455)
    • (1993) Hypertension , vol.21 , Issue.1 , pp. 74-82
    • Sharir, T.1    Marmor, A.2    Ting, C.T.3    Chen, J.W.4    Liu, C.P.5    Chang, M.S.6    Yin, F.C.7    Kass, D.A.8
  • 27
    • 41049090878 scopus 로고    scopus 로고
    • Accuracy comparison of oscillometric and electronic palpation blood pressure measuring methods using intra-arterial method as a reference
    • H.Sorvoja, R. Myllylä, P. Kärjä-Koskenkari, J. Koskenkari, M. Lilja, and Y. A. Kesäniemi, "Accuracy comparison of oscillometric and electronic palpation blood pressure measuring methods using intra-arterial method as a reference," Molecular Quantum Acoust., vol. 26, pp. 235-260, 2005.
    • (2005) Molecular Quantum Acoust. , vol.26 , pp. 235-260
    • Sorvoja, H.1    Myllylä, R.2    Kärjä-Koskenkari, P.3    Koskenkari, J.4    Lilja, M.5    Kesäniemi, Y.A.6
  • 28
    • 0030315989 scopus 로고    scopus 로고
    • Non-invasive blood pressure data acquisition employing pulse transit time detection
    • Amsterdam, TheNetherlands
    • J. Kerola, V. Kontra, and R. Sepponen, "Non-invasive blood pressure data acquisition employing pulse transit time detection," in Proc. 18th Annu. Int. Conf. IEEE Eng. Med. Biol. Soc., vol. 3,Amsterdam, TheNetherlands, 1996, pp. 1308-1309.
    • (1996) Proc. 18th Annu. Int. Conf. IEEE Eng. Med. Biol. Soc. , vol.3 , pp. 1308-1309
    • Kerola, J.1    Kontra, V.2    Sepponen, R.3
  • 29
    • 77957840176 scopus 로고    scopus 로고
    • A prototype of an integrated blood pressure and electrocardiogram device for multi-parameter physiologic monitoring
    • Austin, TX
    • S. Ahmad, S. Chen, K. Soueidan, I. Batkin, M. Bolic, H.Dajani, and V. Groza, "A prototype of an integrated blood pressure and electrocardiogram device for multi-parameter physiologic monitoring," in Proc. IEEE Instrum. Meas. Technol. Conf., Austin, TX, 2010, pp. 1244-1249.
    • (2010) Proc. IEEE Instrum. Meas. Technol. Conf. , pp. 1244-1249
    • Ahmad, S.1    Chen, S.2    Soueidan, K.3    Batkin, I.4    Bolic, M.5    Dajani, H.6    Groza, V.7
  • 31
    • 34248396362 scopus 로고    scopus 로고
    • Validation of the OMRON M7 (HEM-780-E) blood pressure measuring device in a population requiring large cuff use according to the International Protocol of the European Society of Hypertension
    • DOI 10.1097/MBP.0b013e3280b08367, PII 0012609720070600000006
    • R. N. Feghali, J. A. Topouchaian, B. M. Pannier, H.A El Assaad, and R. G. Asmar, "Validation of the Omron M7 (HEM-780-E) blood pressure measuring device in a population requiring large cuff use according to the International Protocol of the European Society of Hypertension," Blood Pressure Monitoring, vol. 12, pp. 173-178, 2007. (Pubitemid 46743721)
    • (2007) Blood Pressure Monitoring , vol.12 , Issue.3 , pp. 173-178
    • Feghali, R.N.1    Topouchaian, J.A.2    Pannier, B.M.3    El Assaad, H.A.4    Asmar, R.G.5
  • 32
    • 38349044223 scopus 로고    scopus 로고
    • Validation of the Omron M7 HEM-780-E oscillometric blood pressure monitoring device according to the british hypertension society protocol
    • A. Coleman, S. Steel, P. Freeman, A. de Greeff, and A. Shennan, "Validation of the Omron M7 (HEM-780-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol," Blood Pressure Monitoring, vol. 13, pp. 49-54, 2008.
    • (2008) Blood Pressure Monitoring , vol.13 , pp. 49-54
    • Coleman, A.1    Steel, S.2    Freeman, P.3    De Greeff, A.4    Shennan, A.5
  • 33
    • 0032790299 scopus 로고    scopus 로고
    • Validation of two devices for self-measurement of blood pressure by elderly patients according to the revised British Hypertension Society protocol: The Omron HEM-722C and HEM-735C
    • L. A. Bortolotto, O. Henry, O. Hanon, P. Sikias, J. J. Mourad, and X. Girerd, "Validation of two devices for self-measurement of blood pressure by elderly patients according to the revised British Hypertension Society Protocol: TheOmronHEM-722C andHEM-735C," Blood PressMonitoring, vol. 4, pp. 21-25, 1998. (Pubitemid 29346978)
    • (1999) Blood Pressure Monitoring , vol.4 , Issue.1 , pp. 21-25
    • Bortolotto, L.A.1    Henry, O.2    Hanon, O.3    Sikias, P.4    Mourad, J.J.5    Girerd, X.6
  • 34
    • 84863175581 scopus 로고    scopus 로고
    • QRS onset detector. Aug.Online].Available:
    • QRS onset detector. (Aug. 2010). Harvard-MIT Health Sci Technology [Online].Available: http://www.mit.edu/~gari/CODE/ECGtools/ecgBag/ sqrs.m
    • (2010) Harvard-MIT Health Sci Technology
  • 35
    • 84863117853 scopus 로고    scopus 로고
    • Single channel QRS detector
    • Aug.Online]. Available
    • Single channel QRS detector. (Aug. 2010). PhysioNET [Online]. Available: http://www.physionet.org/physiotools/wfdb/app/sqrs.c.
    • (2010) PhysioNET
  • 36
    • 0034620112 scopus 로고    scopus 로고
    • Fail in mean arterial pressure and fetal growth restriction in pregnancy hypertension: A meta-analysis
    • P. Dadelszen, M. P. Ornstein, S. B. Bull, A. G. Logan, G. Koren, and L. A. Magee, "Fall in mean arterial pressure and fetal growth restriction in pregnancy hypertension: A meta-analysis," Lancet, vol. 355, no. 9198, pp. 87-92, 2000. (Pubitemid 30068551)
    • (2000) Lancet , vol.355 , Issue.9198 , pp. 87-92
    • Dadelszen, P.1    Ornstein, M.P.2    Bull, S.B.3    Logan, A.G.4    Koren, G.5    Magee, L.A.6
  • 37
    • 48549092221 scopus 로고    scopus 로고
    • Design analysis and interpretation of methodcomparison studies
    • S. K. Hanneman, "Design, analysis, and interpretation of methodcomparison studies," AACN Adv. Critical Care, vol. 19, pp. 223-234, 2008.
    • (2008) AACN Adv. Critical Care , vol.19 , pp. 223-234
    • Hanneman, S.K.1
  • 38
    • 84952503562 scopus 로고
    • Thirteen ways to look at the correlation coefficient
    • J. L. Rodgers and W. A. Nicewander, "Thirteen ways to look at the correlation coefficient," The Amer. Statistician, vol. 42, pp. 59-66, 1988.
    • (1988) Amer. Statistician , vol.42 , pp. 59-66
    • Rodgers, J.L.1    Nicewander, W.A.2


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.