Volumn 27, Issue 5, 2012, Pages 742-749

Scratching Vulnerability of Conventional vs Highly Cross-Linked Polyethylene Liners Because of Large Embedded Third-Body Particles

Author keywords

Highly cross linked polyethylene; Hip arthroplasty; Hip joint motion simulator; Polyethylene; Third body wear

Indexed keywords

CHROMIUM; COBALT; CONVENTIONAL POLYETHYLENE; HIGHLY CROSS LINKED POLYETHYLENE; MOLYBDENUM; POLYETHYLENE; UNCLASSIFIED DRUG;

EID: 84859909659     PISSN: 08835403     EISSN: 15328406     Source Type: Journal    
DOI: 10.1016/j.arth.2011.10.002     Document Type: Article
Times cited : (10)

References (36)
  • 1
    • 20444399895 scopus 로고    scopus 로고
    • Impingement in total hip arthroplasty-a study of retrieved acetabular components
    • Shon W.Y., Baldini T., Peterson M.G., et al. Impingement in total hip arthroplasty-a study of retrieved acetabular components. J Arthroplasty 2005, 20:427.
    • (2005) J Arthroplasty , vol.20 , pp. 427
    • Shon, W.Y.1    Baldini, T.2    Peterson, M.G.3
  • 2
    • 0031008419 scopus 로고    scopus 로고
    • Analysis of particles in acetabular components from patients with osteolysis
    • Bloebaum R.D., Zou L., Bachus K.N., et al. Analysis of particles in acetabular components from patients with osteolysis. Clin Orthop Relat Res 1997, 338:109.
    • (1997) Clin Orthop Relat Res , vol.338 , pp. 109
    • Bloebaum, R.D.1    Zou, L.2    Bachus, K.N.3
  • 4
    • 0029794815 scopus 로고    scopus 로고
    • Wear of polyethylene cups in total hip arthroplasty. A study of specimens retrieved post mortem
    • Sychterz C.J., Moon K.H., Hashimoto Y., et al. Wear of polyethylene cups in total hip arthroplasty. A study of specimens retrieved post mortem. J Bone Joint Surg Am 1996, 78:1193.
    • (1996) J Bone Joint Surg Am , vol.78 , pp. 1193
    • Sychterz, C.J.1    Moon, K.H.2    Hashimoto, Y.3
  • 5
    • 37349066242 scopus 로고    scopus 로고
    • Association of third body embedment with rim damage in retrieved acetabular liners
    • Lundberg H.J., Liu S.S., Callaghan J.J., et al. Association of third body embedment with rim damage in retrieved acetabular liners. Clin Orthop Relat Res 2007, 465:133.
    • (2007) Clin Orthop Relat Res , vol.465 , pp. 133
    • Lundberg, H.J.1    Liu, S.S.2    Callaghan, J.J.3
  • 6
    • 14644414879 scopus 로고    scopus 로고
    • Surface roughness of ceramic femoral heads after in vivo transfer of metal: correlation to polyethylene wear
    • Kim Y.H., Ritchie A., Hardaker C. Surface roughness of ceramic femoral heads after in vivo transfer of metal: correlation to polyethylene wear. J Bone Joint Surg Am 2005, 87:577.
    • (2005) J Bone Joint Surg Am , vol.87 , pp. 577
    • Kim, Y.H.1    Ritchie, A.2    Hardaker, C.3
  • 7
  • 8
    • 0035124730 scopus 로고    scopus 로고
    • Comparative wear and wear debris under three different counterface conditions of crosslinked and non-crosslinked ultra high molecular weight polyethylene
    • Endo M.M., Barbour P.S., Barton D.C., et al. Comparative wear and wear debris under three different counterface conditions of crosslinked and non-crosslinked ultra high molecular weight polyethylene. Biomed Mater Eng 2001, 11:23.
    • (2001) Biomed Mater Eng , vol.11 , pp. 23
    • Endo, M.M.1    Barbour, P.S.2    Barton, D.C.3
  • 9
    • 0032804049 scopus 로고    scopus 로고
    • Wear of gamma-crosslinked polyethylene acetabular cups against roughened femoral balls
    • McKellop H., Shen F.W., DiMaio W., et al. Wear of gamma-crosslinked polyethylene acetabular cups against roughened femoral balls. Clin Orthop Relat Res 1999, 369:73.
    • (1999) Clin Orthop Relat Res , vol.369 , pp. 73
    • McKellop, H.1    Shen, F.W.2    DiMaio, W.3
  • 10
    • 0032927112 scopus 로고    scopus 로고
    • Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements
    • McKellop H., Shen F.W., Lu B., et al. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements. J Orthop Res 1999, 17:157.
    • (1999) J Orthop Res , vol.17 , pp. 157
    • McKellop, H.1    Shen, F.W.2    Lu, B.3
  • 11
    • 11844282688 scopus 로고    scopus 로고
    • Surface-gradient cross-linked polyethylene acetabular cups: oxidation resistance and wear against smooth and rough femoral balls
    • Shen F.W., McKellop H. Surface-gradient cross-linked polyethylene acetabular cups: oxidation resistance and wear against smooth and rough femoral balls. Clin Orthop Relat Res 2005, 430:80.
    • (2005) Clin Orthop Relat Res , vol.430 , pp. 80
    • Shen, F.W.1    McKellop, H.2
  • 12
    • 0035211829 scopus 로고    scopus 로고
    • A comparison of the wear and physical properties of silane cross-linked polyethylene and ultra-high molecular weight polyethylene
    • Sakoda H., Voice A.M., McEwen H.M., et al. A comparison of the wear and physical properties of silane cross-linked polyethylene and ultra-high molecular weight polyethylene. J Arthroplasty 2001, 16:1018.
    • (2001) J Arthroplasty , vol.16 , pp. 1018
    • Sakoda, H.1    Voice, A.M.2    McEwen, H.M.3
  • 13
    • 33645228500 scopus 로고    scopus 로고
    • Wear of crosslinked polyethylene under different tribological conditions
    • Galvin A., Kang L., Tipper J., et al. Wear of crosslinked polyethylene under different tribological conditions. J Mater Sci Mater Med 2006, 17:235.
    • (2006) J Mater Sci Mater Med , vol.17 , pp. 235
    • Galvin, A.1    Kang, L.2    Tipper, J.3
  • 14
    • 8644239396 scopus 로고    scopus 로고
    • Wear, debris, and biologic activity of cross-linked polyethylene in the knee: benefits and potential concerns
    • Fisher J., McEwen H.M., Tipper J.L., et al. Wear, debris, and biologic activity of cross-linked polyethylene in the knee: benefits and potential concerns. Clin Orthop Relat Res 2004, 428:114.
    • (2004) Clin Orthop Relat Res , vol.428 , pp. 114
    • Fisher, J.1    McEwen, H.M.2    Tipper, J.L.3
  • 15
    • 56349163020 scopus 로고    scopus 로고
    • Sliding direction dependence of polyethylene wear for metal counterface traverse of severe scratches
    • Glennon L.P., Baer T.E., Martin J.A., et al. Sliding direction dependence of polyethylene wear for metal counterface traverse of severe scratches. J Biomech Eng 2008, 130:051006.
    • (2008) J Biomech Eng , vol.130 , pp. 051006
    • Glennon, L.P.1    Baer, T.E.2    Martin, J.A.3
  • 16
    • 0017462028 scopus 로고
    • Reactions of the articular capsule to wear products of artificial joint prostheses
    • Willert H.G., Semlitsch M. Reactions of the articular capsule to wear products of artificial joint prostheses. J Biomed Mater Res 1977, 11:157.
    • (1977) J Biomed Mater Res , vol.11 , pp. 157
    • Willert, H.G.1    Semlitsch, M.2
  • 17
    • 33646023729 scopus 로고    scopus 로고
    • Problematic sites of third body embedment in polyethylene for total hip wear acceleration
    • Lundberg H.J., Stewart K.J., Pedersen D.R., et al. Problematic sites of third body embedment in polyethylene for total hip wear acceleration. J Biomech 2006, 39:1208.
    • (2006) J Biomech , vol.39 , pp. 1208
    • Lundberg, H.J.1    Stewart, K.J.2    Pedersen, D.R.3
  • 18
    • 1642499198 scopus 로고    scopus 로고
    • Surface analysis of early retrieved acetabular polyethylene liners: a comparison of conventional and highly crosslinked polyethylenes
    • Muratoglu O.K., Greenbaum E.S., Bragdon C.R., et al. Surface analysis of early retrieved acetabular polyethylene liners: a comparison of conventional and highly crosslinked polyethylenes. J Arthroplasty 2004, 19:68.
    • (2004) J Arthroplasty , vol.19 , pp. 68
    • Muratoglu, O.K.1    Greenbaum, E.S.2    Bragdon, C.R.3
  • 19
    • 84859917447 scopus 로고    scopus 로고
    • ASTM F75-01 Standard specification for cobalt-28 chromium-6 molybdenum alloy castings and casting alloy for surgical implants.
    • ASTM F75-01 Standard specification for cobalt-28 chromium-6 molybdenum alloy castings and casting alloy for surgical implants. 2001.
    • (2001)
  • 20
    • 0034282194 scopus 로고    scopus 로고
    • Modeling of damage to articulating surfaces by third body particles in total joint replacements
    • McNie C.M., Barton D.C., Fisher J., et al. Modeling of damage to articulating surfaces by third body particles in total joint replacements. J Mater Sci Mater Med 2000, 11:569.
    • (2000) J Mater Sci Mater Med , vol.11 , pp. 569
    • McNie, C.M.1    Barton, D.C.2    Fisher, J.3
  • 21
    • 52649174844 scopus 로고    scopus 로고
    • Effects of episodic subluxation events on third body ingress and embedment in the THA bearing surface
    • Heiner A.D., Lundberg H.J., Baer T.E., et al. Effects of episodic subluxation events on third body ingress and embedment in the THA bearing surface. J Biomech 2008, 41:2090.
    • (2008) J Biomech , vol.41 , pp. 2090
    • Heiner, A.D.1    Lundberg, H.J.2    Baer, T.E.3
  • 22
    • 0026228566 scopus 로고
    • Telemeterized in vivo hip joint force data: a report on two patients after total hip surgery
    • Kotzar G.M., Davy D.T., Goldberg V.M., et al. Telemeterized in vivo hip joint force data: a report on two patients after total hip surgery. J Orthop Res 1991, 9:621.
    • (1991) J Orthop Res , vol.9 , pp. 621
    • Kotzar, G.M.1    Davy, D.T.2    Goldberg, V.M.3
  • 23
    • 84859925846 scopus 로고    scopus 로고
    • ASTM F1714-96(2002) Standard guide for gravimetric wear assessment of prosthetic hip-designs in simulator devices.
    • ASTM F1714-96(2002) Standard guide for gravimetric wear assessment of prosthetic hip-designs in simulator devices. 2002.
    • (2002)
  • 25
    • 0027575844 scopus 로고
    • Macroscopic and microscopic wear mechanisms in ultra-high molecular weight polyethylene
    • Cooper J.R., Dowson D., Fisher J. Macroscopic and microscopic wear mechanisms in ultra-high molecular weight polyethylene. Wear 1993, 162-164(Part 1):378.
    • (1993) Wear , vol.162-164 , Issue.PART. 1 , pp. 378
    • Cooper, J.R.1    Dowson, D.2    Fisher, J.3
  • 26
    • 78951490507 scopus 로고    scopus 로고
    • Retrieved highly crosslinked UHMWPE acetabular liners have similar wear damage as conventional UHMWPE
    • Schroder D.T., Kelly N.H., Wright T.M., et al. Retrieved highly crosslinked UHMWPE acetabular liners have similar wear damage as conventional UHMWPE. Clin Orthop Relat Res 2011, 469:387.
    • (2011) Clin Orthop Relat Res , vol.469 , pp. 387
    • Schroder, D.T.1    Kelly, N.H.2    Wright, T.M.3
  • 27
    • 0029818186 scopus 로고    scopus 로고
    • Abrasive three-body wear of polyethylene caused by broken multifilament cables of a total hip prosthesis. A report of these cases
    • Bauer T.W., Ming J., D'Antonio J.A., et al. Abrasive three-body wear of polyethylene caused by broken multifilament cables of a total hip prosthesis. A report of these cases. J Bone Joint Surg Am 1996, 78:1244.
    • (1996) J Bone Joint Surg Am , vol.78 , pp. 1244
    • Bauer, T.W.1    Ming, J.2    D'Antonio, J.A.3
  • 28
    • 0033967283 scopus 로고    scopus 로고
    • Surface roughness of retrieved CoCrMo alloy femoral components from PCA artificial total knee joints
    • Que L., Topoleski L.D., Parks N.L. Surface roughness of retrieved CoCrMo alloy femoral components from PCA artificial total knee joints. J Biomed Mater Res 2000, 53:111.
    • (2000) J Biomed Mater Res , vol.53 , pp. 111
    • Que, L.1    Topoleski, L.D.2    Parks, N.L.3
  • 29
    • 0034953264 scopus 로고    scopus 로고
    • Quantitative evaluation of the prosthetic head damage induced by microscopic third-body particles in total hip replacement
    • Raimondi M.T., Vena P., Pietrabissa R. Quantitative evaluation of the prosthetic head damage induced by microscopic third-body particles in total hip replacement. J Biomed Mater Res 2001, 58:436.
    • (2001) J Biomed Mater Res , vol.58 , pp. 436
    • Raimondi, M.T.1    Vena, P.2    Pietrabissa, R.3
  • 30
    • 58249092213 scopus 로고    scopus 로고
    • The effect of entrapped bone particles on the surface morphology and wear of polyethylene
    • Mimnaugh K.D., Yao J.Q., Laurent M.P., et al. The effect of entrapped bone particles on the surface morphology and wear of polyethylene. J Arthroplasty 2009, 24:303.
    • (2009) J Arthroplasty , vol.24 , pp. 303
    • Mimnaugh, K.D.1    Yao, J.Q.2    Laurent, M.P.3
  • 31
    • 0141889873 scopus 로고    scopus 로고
    • Optical analysis of surface changes on early retrievals of highly cross-linked and conventional polyethylene tibial inserts
    • Muratoglu O.K., Ruberti J., Melotti S., et al. Optical analysis of surface changes on early retrievals of highly cross-linked and conventional polyethylene tibial inserts. J Arthroplasty 2003, 18(7 Suppl 1):42.
    • (2003) J Arthroplasty , vol.18 , Issue.7 SUPPL. 1 , pp. 42
    • Muratoglu, O.K.1    Ruberti, J.2    Melotti, S.3
  • 32
    • 37349051414 scopus 로고    scopus 로고
    • Remelting of highly cross-linked polyethylene worn under laboratory conditions
    • Lazzarini A.M., Cottrell J.M., Padgett D.E., et al. Remelting of highly cross-linked polyethylene worn under laboratory conditions. Clin Orthop Relat Res 2007, 465:128.
    • (2007) Clin Orthop Relat Res , vol.465 , pp. 128
    • Lazzarini, A.M.1    Cottrell, J.M.2    Padgett, D.E.3
  • 33
    • 24344470495 scopus 로고    scopus 로고
    • Ex vivo wear of conventional and cross-linked polyethylene acetabular liners
    • Muratoglu O.K., Wannomae K., Christensen S., et al. Ex vivo wear of conventional and cross-linked polyethylene acetabular liners. Clin Orthop Relat Res 2005, 438:158.
    • (2005) Clin Orthop Relat Res , vol.438 , pp. 158
    • Muratoglu, O.K.1    Wannomae, K.2    Christensen, S.3
  • 34
    • 79953664312 scopus 로고    scopus 로고
    • Medicines and Healthcare products Regulatory Agency
    • Medical Device Alert: All metal-on-metal (MoM) hip replacements (MDA/2010/033).
    • Medicines and Healthcare products Regulatory Agency. Medical Device Alert: All metal-on-metal (MoM) hip replacements (MDA/2010/033). 2010.
    • (2010)
  • 36
    • 84859917445 scopus 로고    scopus 로고
    • May 9, 2011
    • May 9, 2011. http://www.fda.gov/MedicalDevices/ProductsandMedicalProcedures/ImplantsandProsthetics/MetalonMetalHipImplants/ucm241769.htm.


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.