Continuous localization technologies for radiotherapy delivery: Report of the American Society for Radiation Oncology Emerging Technology Committee

Author keywords

[No Author keywords available]

Indexed keywords

ABLATION THERAPY; ANDROGEN THERAPY; CANCER RADIOTHERAPY; CLINICAL EFFECTIVENESS; CLINICAL SUPERVISION; COMPUTER ASSISTED TOMOGRAPHY; CONTINUOUS LOCALIZATION TECHNOLOGY; DIAGNOSTIC ACCURACY; EQUIPMENT; HEALTH CARE DELIVERY; HUMAN; IMAGING SYSTEM; MEDICAL SOCIETY; PATIENT POSITIONING; PRIORITY JOURNAL; PROCESS DEVELOPMENT; PROSTATE CANCER; QUALITY CONTROL; RADIATION PROCESSING; RADIOSURGERY; REVIEW; RISK BENEFIT ANALYSIS; SIMULATION; STAFF TRAINING; TREATMENT PLANNING; TUMOR LOCALIZATION;

EID: 84858864241     PISSN: 18798500     EISSN: None     Source Type: Journal    
DOI: 10.1016/j.prro.2011.10.005     Document Type: Review
Times cited : (17)

References (29)
  • 2
    • 34247547835 scopus 로고    scopus 로고
    • Optical tracking technology in stereotactic radiation therapy
    • Wagner TH, Meeks SL, Bova FJ, et al. Optical tracking technology in stereotactic radiation therapy. Med Dosim 2007, 32:111-120.
    • (2007) Med Dosim , vol.32 , pp. 111-120
    • Wagner, T.H.1    Meeks, S.L.2    Bova, F.J.3
  • 3
    • 34447256580 scopus 로고    scopus 로고
    • Positional stability of electromagnetic transponders used for prostate localization and continuous, real-time tracking
    • Litzenberg DW, Willoughby JM, Balter HM, et al. Positional stability of electromagnetic transponders used for prostate localization and continuous, real-time tracking. Int J Radiat Oncol Biol Phys 2007, 68:1199-1206.
    • (2007) Int J Radiat Oncol Biol Phys , vol.68 , pp. 1199-1206
    • Litzenberg, D.W.1    Willoughby, J.M.2    Balter, H.M.3
  • 4
    • 33646487136 scopus 로고    scopus 로고
    • Target localization and real-time tracking using the Calypso 4D localization system in patients with localized prostate cancer
    • Willoughby TR, Kupelian PA, Pouliot J, et al. Target localization and real-time tracking using the Calypso 4D localization system in patients with localized prostate cancer. Int J Radiat Oncol Biol Phys 2006, 65:528-534.
    • (2006) Int J Radiat Oncol Biol Phys , vol.65 , pp. 528-534
    • Willoughby, T.R.1    Kupelian, P.A.2    Pouliot, J.3
  • 5
    • 33847328730 scopus 로고    scopus 로고
    • Multi-institutional clinical experience with the Calypso System in localization and continuous, real-time monitoring of the prostate gland during external radiotherapy
    • Kupelian P, Willoughby T, Mahadevan A, et al. Multi-institutional clinical experience with the Calypso System in localization and continuous, real-time monitoring of the prostate gland during external radiotherapy. Int J Radiat Oncol Biol Phys 2007, 67:1088-1098.
    • (2007) Int J Radiat Oncol Biol Phys , vol.67 , pp. 1088-1098
    • Kupelian, P.1    Willoughby, T.2    Mahadevan, A.3
  • 6
    • 33645368752 scopus 로고    scopus 로고
    • Influence of intrafraction motion on margins for prostate radiotherapy
    • Litzenberg DW, Balter JM, Hadley SW, et al. Influence of intrafraction motion on margins for prostate radiotherapy. Int J Radiat Oncol Biol Phys 2006, 65:548-553.
    • (2006) Int J Radiat Oncol Biol Phys , vol.65 , pp. 548-553
    • Litzenberg, D.W.1    Balter, J.M.2    Hadley, S.W.3
  • 8
    • 84858866997 scopus 로고    scopus 로고
    • Stability of intraprostatic electromagnetic transponders in patients receiving radiation therapy, with and without neoadjuvant and/or concurrent androgen suppression therapy
    • Weinstein G, Jani SK, Kupelian P, et al. Stability of intraprostatic electromagnetic transponders in patients receiving radiation therapy, with and without neoadjuvant and/or concurrent androgen suppression therapy. Int J Radiat Oncol Biol Phys 2006, 66(Suppl 1):S358-S359.
    • (2006) Int J Radiat Oncol Biol Phys , vol.66 , Issue.SUPPL. 1
    • Weinstein, G.1    Jani, S.K.2    Kupelian, P.3
  • 9
    • 84858862442 scopus 로고    scopus 로고
    • Demonstration of minimal impact to radiation beam and portal image quality due to the presence of an electromagnetic array
    • Pouliot J, Werner B, Riley J, et al. Demonstration of minimal impact to radiation beam and portal image quality due to the presence of an electromagnetic array. Med Phys 2003, 30:1473.
    • (2003) Med Phys , vol.30 , pp. 1473
    • Pouliot, J.1    Werner, B.2    Riley, J.3
  • 10
    • 84858866184 scopus 로고    scopus 로고
    • Feasibility of tracking wireless AC electromagnetic transponders in head and neck cancer environment
    • [Abstract]
    • Mate T, Zeller T, Douglas R, et al. Feasibility of tracking wireless AC electromagnetic transponders in head and neck cancer environment. Med Phys 2005, 32:2112. [Abstract].
    • (2005) Med Phys , vol.32 , pp. 2112
    • Mate, T.1    Zeller, T.2    Douglas, R.3
  • 11
    • 84858865491 scopus 로고    scopus 로고
    • Assessment of dental amalgam backscatter with a Beacon Transponder Embedded Mouthpiece for real-time tracking during head and neck IMRT
    • [Abstract]
    • Ye J, Werner B, Mate T, et al. Assessment of dental amalgam backscatter with a Beacon Transponder Embedded Mouthpiece for real-time tracking during head and neck IMRT. Med Phys 2006, 33:2073. [Abstract].
    • (2006) Med Phys , vol.33 , pp. 2073
    • Ye, J.1    Werner, B.2    Mate, T.3
  • 12
    • 0031756389 scopus 로고    scopus 로고
    • Preliminary experience with frameless stereotactic radiotherapy
    • Buatti JM, Bova FJ, Friedman WA, et al. Preliminary experience with frameless stereotactic radiotherapy. Int J Radiat Oncol Biol Phys 1998, 42:591-599.
    • (1998) Int J Radiat Oncol Biol Phys , vol.42 , pp. 591-599
    • Buatti, J.M.1    Bova, F.J.2    Friedman, W.A.3
  • 13
    • 0035889332 scopus 로고    scopus 로고
    • Initial clinical experience with frameless stereotactic radiosurgery: analysis of accuracy and feasibility
    • Ryken TC, Meeks SL, Pennington EC, et al. Initial clinical experience with frameless stereotactic radiosurgery: analysis of accuracy and feasibility. Int J Radiat Oncol Biol Phys 2001, 51:1152-1158.
    • (2001) Int J Radiat Oncol Biol Phys , vol.51 , pp. 1152-1158
    • Ryken, T.C.1    Meeks, S.L.2    Pennington, E.C.3
  • 16
    • 25644432884 scopus 로고    scopus 로고
    • Phantom evaluation of a stereo-vision surface imaging system for radiotherapy patient setup
    • Bert C, Metheany KG, Doppke K, Chen GTA Phantom evaluation of a stereo-vision surface imaging system for radiotherapy patient setup. Med Phys 2005, 32:2753-2762.
    • (2005) Med Phys , vol.32 , pp. 2753-2762
    • Bert, C.1    Metheany, K.G.2    Doppke, K.3    Chen, G.T.A.4
  • 18
    • 33644508623 scopus 로고    scopus 로고
    • Chen GT. Clinical experience with a 3D surface patient setup system for alignment of partial-breast irradiation patients
    • Bert C, Metheany KG, Doppke KP, Taghian AG, Powell SN Chen GT. Clinical experience with a 3D surface patient setup system for alignment of partial-breast irradiation patients. Int J Radiat Oncol Biol Phys 2006, 64:1265-1274.
    • (2006) Int J Radiat Oncol Biol Phys , vol.64 , pp. 1265-1274
    • Bert, C.1    Metheany, K.G.2    Doppke, K.P.3    Taghian, A.G.4    Powell, S.N.5
  • 19
    • 84858861145 scopus 로고    scopus 로고
    • Investigating the correlation between surface and bony anatomy using 3D surface and portal imaging
    • Bidmead M, Corsini L, Lindgren-Turner J, et al. Investigating the correlation between surface and bony anatomy using 3D surface and portal imaging. Radiation & Oncology 2004, 73(Suppl 1):S233.
    • (2004) Radiation & Oncology , vol.73 , Issue.SUPPL. 1
    • Bidmead, M.1    Corsini, L.2    Lindgren-Turner, J.3
  • 21
    • 84858860681 scopus 로고    scopus 로고
    • Feasibility of tracking head position under an obscuring immobilization mask using a bite block and a 3-D surface imaging system
    • Drzymala R, Wood R Feasibility of tracking head position under an obscuring immobilization mask using a bite block and a 3-D surface imaging system. Med Phys 2006, 33:1992.
    • (2006) Med Phys , vol.33 , pp. 1992
    • Drzymala, R.1    Wood, R.2
  • 22
    • 39749161503 scopus 로고    scopus 로고
    • Comparison of target registration errors for multiple modalities in image-guided partial breast irradiation
    • Gierga DP, Riboldi M, Turcotte JC, et al. Comparison of target registration errors for multiple modalities in image-guided partial breast irradiation. Int J Radiat Oncol Biol Phys 2008, 70:1239-1246.
    • (2008) Int J Radiat Oncol Biol Phys , vol.70 , pp. 1239-1246
    • Gierga, D.P.1    Riboldi, M.2    Turcotte, J.C.3
  • 23
    • 39749094282 scopus 로고    scopus 로고
    • Target registration ERROR with three-dimensional surface imaging in setup of image-guided partial breast irradiation
    • Gierga DP, Turcotte JC, Sharp G, et al. Target registration ERROR with three-dimensional surface imaging in setup of image-guided partial breast irradiation. Int J Radiat Oncol Biol Phys 2005, 63(Suppl 1):S536-S537.
    • (2005) Int J Radiat Oncol Biol Phys , vol.63 , Issue.SUPPL. 1
    • Gierga, D.P.1    Turcotte, J.C.2    Sharp, G.3
  • 24
    • 25644452191 scopus 로고    scopus 로고
    • Real time 3D surface imaging for the analysis of respiratory motion during radiotherapy
    • Johnson U, Deehan C, Landau D Real time 3D surface imaging for the analysis of respiratory motion during radiotherapy. Int J Radiat Oncol Biol Phys 2004, 60(Suppl):S603-S604.
    • (2004) Int J Radiat Oncol Biol Phys , vol.60 , Issue.SUPPL.
    • Johnson, U.1    Deehan, C.2    Landau, D.3
  • 26
    • 79958229549 scopus 로고    scopus 로고
    • Evaluation of repositioning accuracy of patiens with breast cancer using a 3D surface imaging system
    • Schöffel PJ, Harms W, Karger CP Evaluation of repositioning accuracy of patiens with breast cancer using a 3D surface imaging system. Radiother Oncol 2006, 81(Suppl 1):S195.
    • (2006) Radiother Oncol , vol.81 , Issue.SUPPL. 1
    • Schöffel, P.J.1    Harms, W.2    Karger, C.P.3
  • 27
    • 34250335308 scopus 로고    scopus 로고
    • Accuracy of a commercial optical 3D surface imaging system for realignment of patients for radiotherapy of the thorax
    • Schöffel PJ, Harms W, Sroka-Perez G, Schlegel W, Karger CP Accuracy of a commercial optical 3D surface imaging system for realignment of patients for radiotherapy of the thorax. Phys Med Biol 2007, 52:3949-3963.
    • (2007) Phys Med Biol , vol.52 , pp. 3949-3963
    • Schöffel, P.J.1    Harms, W.2    Sroka-Perez, G.3    Schlegel, W.4    Karger, C.P.5
  • 28
    • 25644447010 scopus 로고    scopus 로고
    • Real-time 3D surface imaging for patient positioning in radiotherapy
    • Smith N, Meir I, Hale G, et al. Real-time 3D surface imaging for patient positioning in radiotherapy. Int J Radiat Oncol Biol Phys 2003, 57:S187.
    • (2003) Int J Radiat Oncol Biol Phys , vol.57
    • Smith, N.1    Meir, I.2    Hale, G.3
  • 29
    • 44949217178 scopus 로고    scopus 로고
    • A non-contact method for the acquisition of breathing signals that enable distinction between abdominal and thoracic breathing
    • Tarte S, McClelland J, Hughes S, Blackall J, Landau D, Hawkes D A non-contact method for the acquisition of breathing signals that enable distinction between abdominal and thoracic breathing. Radiother Oncol 2006, 81:S209.
    • (2006) Radiother Oncol , vol.81
    • Tarte, S.1    McClelland, J.2    Hughes, S.3    Blackall, J.4    Landau, D.5    Hawkes, D.6


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.