Volumn 30, Issue 3, 2012, Pages 537-542

Requirements for professional office blood pressure monitors

Author keywords

blood pressure measurement; clinic; hypertension; office; professional; validation

Indexed keywords

ANEROID SPHYGMOMANOMETER; AUTOANALYZER; BLOOD PRESSURE MEASUREMENT; BLOOD PRESSURE MONITOR; BLOOD PRESSURE MONITORING; CARDIOVASCULAR EQUIPMENT; CONFERENCE PAPER; DIAGNOSTIC ACCURACY; HEART AUSCULTATION; HUMAN; HYBRID; HYBRID SPHYGMOMANOMETER; HYPERTENSION; MERCURY SPHYGMOMANOMETER; MONITOR; OSCILLOMETRIC SPHYGMOMANOMETER; OSCILLOMETRY; PRIORITY JOURNAL; SPHYGMOMANOMETER; VALIDITY; WHITE COAT HYPERTENSION;

EID: 84857049772     PISSN: 02636352     EISSN: 14735598     Source Type: Journal    
DOI: 10.1097/HJH.0b013e32834fcfa5     Document Type: Conference Paper
Times cited : (57)

References (37)
  • 2
    • 19944429836 scopus 로고    scopus 로고
    • Recommendations for blood pressure measurement in humans and experimental animals. Part 1: Blood pressure measurement in humans: A statement for professionals from the subcommittee of professional and public education of the American Heart Association council on high blood pressure research
    • DOI 10.1161/01.HYP.0000150859.47929.8e
    • Pickering TG, Hall JE, Appel LJ, Falkner BE, Graves J, Hill MN, et al. Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Recommendations for blood pressure measurement in humans and experimental animals: part 1: blood pressure measurement in humans: a statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Hypertension 2005; 45:142-161. (Pubitemid 40066429)
    • (2005) Hypertension , vol.45 , Issue.1 , pp. 142-161
    • Pickering, T.G.1    Hall, J.E.2    Appel, L.J.3    Falkner, B.E.4    Graves, J.5    Hill, M.N.6    Jones, D.W.7    Kurtz, T.8    Sheps, S.G.9    Roccella, E.J.10
  • 3
    • 77954922244 scopus 로고    scopus 로고
    • Setting-up a blood pressure and vascular protection clinic: Requirements of the European Society of Hypertension
    • Stergiou GS, Myers MG, Reid JL, Burnier M, Narkiewicz K, Viigimaa M, Mancia G. Setting-up a blood pressure and vascular protection clinic: requirements of the European Society of Hypertension. J Hypertens 2010; 28:1780-1781.
    • (2010) J Hypertens , vol.28 , pp. 1780-1781
    • Stergiou, G.S.1    Myers, M.G.2    Reid, J.L.3    Burnier, M.4    Narkiewicz, K.5    Viigimaa, M.6    Mancia, G.7
  • 4
    • 84857048612 scopus 로고    scopus 로고
    • Directive 2007/51/EC of the European Parliament and of the Council of 25 September Assessed 12 July 2011
    • Mercury in measuring devices (amendment of Council Directive 76/769/EEC) Directive 2007/51/EC of the European Parliament and of the Council of 25 September 2007. http://ec.europa.eu/enterprise/chemicals/legislation/markrestr/ amendments-en.htm. Assessed 12 July 2011.
    • (2007) Mercury in Measuring Devices (Amendment of Council Directive 76/769/EEC)
  • 5
    • 0035992036 scopus 로고    scopus 로고
    • Has conventional sphygmomanometry ended with the banning of mercury?
    • DOI 10.1097/00126097-20020200000007
    • O'Brien E. Has conventional sphygmomanometry ended with the banning of mercury? Blood Press Monit 2002; 7:37-40. (Pubitemid 34304050)
    • (2002) Blood Pressure Monitoring , vol.7 , Issue.1 , pp. 37-40
    • O'Brien, E.1
  • 6
    • 0037300144 scopus 로고    scopus 로고
    • Demise of the mercury sphygmomanometer and the dawning of a new era in blood pressure measurement
    • DOI 10.1097/00126097-20030200000004
    • O'Brien E. Demise of the mercury sphygmomanometer and the dawning of a new era in blood pressure measurement. Blood Press Monit 2003; 8:19-21. (Pubitemid 36343966)
    • (2003) Blood Pressure Monitoring , vol.8 , Issue.1 , pp. 19-21
    • O'Brien, E.1
  • 7
    • 0037301090 scopus 로고    scopus 로고
    • What will replace the mercury sphygmomanometer?
    • DOI 10.1097/00126097-20030200000005
    • Pickering TG. What will replace the mercury sphygmomanometer? Blood Press Monit 2003; 8:23-25. (Pubitemid 36343967)
    • (2003) Blood Pressure Monitoring , vol.8 , Issue.1 , pp. 23-25
    • Pickering, T.G.1
  • 8
    • 29144461692 scopus 로고    scopus 로고
    • How to cope with unreliable office blood pressure measurement?
    • DOI 10.1016/j.amjhyper.2005.07.014, PII S0895706105010952
    • Stergiou GS. How to cope with unreliable office blood pressure measurement? Am J Hypertens 2005; 18:1519-1521. (Pubitemid 41814669)
    • (2005) American Journal of Hypertension , vol.18 , Issue.12 , pp. 1519-1521
    • Stergiou, G.S.1
  • 9
    • 39049184652 scopus 로고    scopus 로고
    • Should doctors still measure blood pressure?
    • Pickering TG. Should doctors still measure blood pressure? J Clin Hypertens 2006; 8:394-396.
    • (2006) J Clin Hypertens , vol.8 , pp. 394-396
    • Pickering, T.G.1
  • 10
    • 40049083842 scopus 로고    scopus 로고
    • Office blood pressure measurement with electronic devices: Has the time come?
    • DOI 10.1038/ajh.2007.78, PII AJH200778
    • Stergiou GS. Office blood pressure measurement with electronic devices: has the time come? Am J Hypertens 2008; 21:246. (Pubitemid 351323284)
    • (2008) American Journal of Hypertension , vol.21 , Issue.3 , pp. 246
    • Stergiou, G.S.1
  • 11
    • 59849115803 scopus 로고    scopus 로고
    • Franz Volhard lecture: Should doctors still measure blood pressure? the missing patients with masked hypertension
    • Pickering TG, Gerin W, Schwartz JE, Spruill TM, Davidson KW. Franz Volhard lecture: should doctors still measure blood pressure? The missing patients with masked hypertension. J Hypertens 2008; 26:2259-2267.
    • (2008) J Hypertens , vol.26 , pp. 2259-2267
    • Pickering, T.G.1    Gerin, W.2    Schwartz, J.E.3    Spruill, T.M.4    Davidson, K.W.5
  • 12
    • 77949391742 scopus 로고    scopus 로고
    • Assessing the validity of the Omron HEM-907XL oscillometric blood pressure measurement device in a national survey environment
    • Ostchega Y, Nwankwo T, Sorlie PD, Wolz M, Zipf G. Assessing the validity of the Omron HEM-907XL oscillometric blood pressure measurement device in a national survey environment. J Clin Hypertens 2010; 12:22-28.
    • (2010) J Clin Hypertens , vol.12 , pp. 22-28
    • Ostchega, Y.1    Nwankwo, T.2    Sorlie, P.D.3    Wolz, M.4    Zipf, G.5
  • 13
    • 80054063638 scopus 로고    scopus 로고
    • Replacing the mercury manometer with an oscillometric device in a hypertension clinic: Implications for clinical decision making
    • Stergiou GS, Lourida P, Tzamouranis D. Replacing the mercury manometer with an oscillometric device in a hypertension clinic: implications for clinical decision making. J Hum Hypertens 2011; 25:692-628.
    • (2011) J Hum Hypertens , vol.25 , pp. 692-628
    • Stergiou, G.S.1    Lourida, P.2    Tzamouranis, D.3
  • 14
    • 0028943089 scopus 로고
    • The accuracy of automated blood pressure measuring device in patients with controlled atrial fibrillation
    • Stewart MJ, Gough K, Padfield PL. The accuracy of automated blood pressure measuring device in patients with controlled atrial fibrillation. J Hypertens 1995; 13:297-300.
    • (1995) J Hypertens , vol.13 , pp. 297-300
    • Stewart, M.J.1    Gough, K.2    Padfield, P.L.3
  • 15
    • 33745234213 scopus 로고    scopus 로고
    • The accuracy of blood pressure measurement in atrial fibrillation
    • Jani B, Bulpitt CJ, Rajkumar C. The accuracy of blood pressure measurement in atrial fibrillation. J Hum Hypertens 2006; 20:543-545.
    • (2006) J Hum Hypertens , vol.20 , pp. 543-545
    • Jani, B.1    Bulpitt, C.J.2    Rajkumar, C.3
  • 16
    • 33747644391 scopus 로고    scopus 로고
    • Blood pressure measurement in atrial fibrillation: Goodbye mercury?
    • DOI 10.1038/sj.jhh.1002031, PII 1002031
    • Watson T, Lip GYH. Blood pressure measurement in atrial fibrillation: goodbye mercury? J Hum Hypertens 2006; 20:638-640. (Pubitemid 44269853)
    • (2006) Journal of Human Hypertension , vol.20 , Issue.9 , pp. 638-640
    • Watson, T.1    Lip, G.Y.H.2
  • 18
    • 70749154760 scopus 로고    scopus 로고
    • Unreliable oscillometric blood pressure measurement: Prevalence, repeatability and characteristics of the phenomenon
    • Stergiou GS, Lourida P, Tzamouranis D, Baibas NM. Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon. J Hum Hypertens 2009; 23:794-800.
    • (2009) J Hum Hypertens , vol.23 , pp. 794-800
    • Stergiou, G.S.1    Lourida, P.2    Tzamouranis, D.3    Baibas, N.M.4
  • 19
    • 77149175979 scopus 로고    scopus 로고
    • Validation of three professional devices measuring office blood pressure according to three different methods: The Omron BP10, the Omron HBP T105 and the Pic Indolor Professional
    • Asmar R, Khabouth J, Mattar J, Pecchioli V, Germano G. Validation of three professional devices measuring office blood pressure according to three different methods: the Omron BP10, the Omron HBP T105 and the Pic Indolor Professional. J Hypertens 2010; 28:452-458.
    • (2010) J Hypertens , vol.28 , pp. 452-458
    • Asmar, R.1    Khabouth, J.2    Mattar, J.3    Pecchioli, V.4    Germano, G.5
  • 20
    • 74949121981 scopus 로고    scopus 로고
    • Measurement of blood pressure in the office: Recognizing the problem and proposing the solution
    • Myers M, Godwin M, Dawes M, Kiss A, Tobe SW, Kaczorowski J. Measurement of blood pressure in the office: recognizing the problem and proposing the solution. Hypertension 2010; 55:195-200.
    • (2010) Hypertension , vol.55 , pp. 195-200
    • Myers, M.1    Godwin, M.2    Dawes, M.3    Kiss, A.4    Tobe, S.W.5    Kaczorowski, J.6
  • 22
    • 38349063257 scopus 로고    scopus 로고
    • Validation of the A&D UM-101 professional hybrid device for office blood pressure measurement according to the International Protocol
    • Stergiou GS, Giovas PP, Gkinos CP, Tzamouranis DG. Validation of the A&D UM-101 professional hybrid device for office blood pressure measurement according to the International Protocol. Blood Press Monit 2008; 13:37-42.
    • (2008) Blood Press Monit , vol.13 , pp. 37-42
    • Stergiou, G.S.1    Giovas, P.P.2    Gkinos, C.P.3    Tzamouranis, D.G.4
  • 23
    • 77955918547 scopus 로고    scopus 로고
    • Development and validation of a blinded hybrid device according to the European Hypertension Society protocol: Nissei DM-3000
    • Tasker F, De Greeff A, Shennan AH. Development and validation of a blinded hybrid device according to the European Hypertension Society protocol: Nissei DM-3000. J Hum Hypertens 2010; 24:609-616.
    • (2010) J Hum Hypertens , vol.24 , pp. 609-616
    • Tasker, F.1    De Greeff, A.2    Shennan, A.H.3
  • 24
    • 1842531924 scopus 로고    scopus 로고
    • The Accoson Greenlight 300™, the first non-automated mercury-free blood pressure measurement device to pass the International Protocol for blood pressure measuring devices in adults
    • DOI 10.1097/00126097-20040200000004
    • Graves JW, Tibor M, Murtagh B, Klein L, Sheps SG. The Accoson Greenlight 300, the first nonautomated mercury-free blood pressure measurement device to pass the International Protocol for blood pressure measuring devices in adults. Blood Press Monit 2004; 9:13-17. (Pubitemid 38429428)
    • (2004) Blood Pressure Monitoring , vol.9 , Issue.1 , pp. 13-17
    • Graves, J.W.1    Tibor, M.2    Murtagh, B.3    Klein, L.4    Sheps, S.G.5
  • 26
    • 34548433551 scopus 로고    scopus 로고
    • Validation of the Omron M5-I, R5-I and HEM-907 automated blood pressure monitors in elderly individuals according to the international protocol of the European society of hypertension
    • DOI 10.1097/MBP.0b013e32813fa386, PII 0012609720070800000005
    • Omboni S, Riva I, Giglio A, Caldara G, Groppelli A, Parati G. Validation of the Omron M5-I, R5-I and HEM-907 automated blood pressure monitors in elderly individuals according to the International Protocol of the European Society of Hypertension. Blood Press Monit 2007; 12:233-242. (Pubitemid 47356409)
    • (2007) Blood Pressure Monitoring , vol.12 , Issue.4 , pp. 233-242
    • Omboni, S.1    Riva, I.2    Giglio, A.3    Caldara, G.4    Groppelli, A.5    Parati, G.6
  • 27
    • 53549101912 scopus 로고    scopus 로고
    • Validation of the Microlife Watch BP Office professional device for office blood pressure measurement according to the International protocol
    • Stergiou GS, Tzamouranis D, Protogerou A, Nasothimiou E, Kapralos C. Validation of the Microlife Watch BP Office professional device for office blood pressure measurement according to the International protocol. Blood Press Monit 2008; 13:299-303.
    • (2008) Blood Press Monit , vol.13 , pp. 299-303
    • Stergiou, G.S.1    Tzamouranis, D.2    Protogerou, A.3    Nasothimiou, E.4    Kapralos, C.5
  • 28
    • 76449087188 scopus 로고    scopus 로고
    • European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults
    • Working Group on Blood Pressure Monitoring of the European Society of Hypertension
    • O'Brien E, Atkins N, Stergiou G, Karpettas N, Parati G, Asmar R, et al. Working Group on Blood Pressure Monitoring of the European Society of Hypertension. European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults. Blood Press Monit 2010; 15:23-38.
    • (2010) Blood Press Monit , vol.15 , pp. 23-38
    • O'Brien, E.1    Atkins, N.2    Stergiou, G.3    Karpettas, N.4    Parati, G.5    Asmar, R.6
  • 29
    • 0027218269 scopus 로고
    • The British Hypertension Society Protocol for the evaluation of blood pressure measuring devices
    • O'Brien E, Petrie J, Littler WA, de Swiet M, Padfield P, Altman D, et al. The British Hypertension Society Protocol for the evaluation of blood pressure measuring devices. J Hypertens 1993; 11 (Suppl 2): S43-S63.
    • (1993) J Hypertens , vol.11 , Issue.SUPPL. 2
    • O'Brien, E.1    Petrie, J.2    Littler, W.A.3    De Swiet, M.4    Padfield, P.5    Altman, D.6
  • 30
    • 85042454629 scopus 로고    scopus 로고
    • Noninvasive sphygmomanometers: Clinical validation of automated measurement type
    • Assessed 12 July 2011
    • Noninvasive sphygmomanometers: clinical validation of automated measurement type. International Organization for Standardization (ISO) 81060-2, 2009. www.iso.org. Assessed 12 July 2011.
    • (2009) International Organization for Standardization (ISO) 81060-2
  • 31
    • 0033651061 scopus 로고    scopus 로고
    • The results of the campaign for evaluating sphygmomanometers accuracy and their physical conditions
    • Mion Júnior D, Pierin AM, Alavarce DC, Vasconcellos JH. The results of the campaign for evaluating sphygmomanometers accuracy and their physical conditions. Arq Bras Cardiol 2000; 74:31-38.
    • (2000) Arq Bras Cardiol , vol.74 , pp. 31-38
    • Mion Júnior, D.1    Pierin, A.M.2    Alavarce, D.C.3    Vasconcellos, J.H.4
  • 32
    • 0034956706 scopus 로고    scopus 로고
    • Accuracy of aneroid sphygmomanometers in clinical usage: University of Michigan experience
    • DOI 10.1097/00126097-20010400000006
    • Yarows SA, Qian K. Accuracy of aneroid sphygmomanometers in clinical usage: University of Michigan experience. Blood Press Monit 2001; 6:101-106. (Pubitemid 32587667)
    • (2001) Blood Pressure Monitoring , vol.6 , Issue.2 , pp. 101-106
    • Yarows, S.A.1    Qian, K.2
  • 33
    • 84857055542 scopus 로고    scopus 로고
    • Measurement of blood pressure. Part II. Conventional sphygmomanometry
    • Beevers DG, Lip GH, O'Brien E, editors 5th ed. Oxford UK: BMJ Blackwell Publishing
    • O'Brien E. Measurement of blood pressure. Part II. Conventional sphygmomanometry. In Beevers DG, Lip GH, O'Brien E, editors. ABC of hypertension, 5th ed. Oxford UK: BMJ Blackwell Publishing; 2007. pp.17-32.
    • (2007) ABC of Hypertension , pp. 17-32
    • O'Brien, E.1
  • 34
    • 76449109838 scopus 로고    scopus 로고
    • European Society of Hypertension International Protocol for the validation of blood pressure monitors: A critical review of its application and rationale for revision
    • Stergiou GS, Karpettas N, Atkins N, O'Brien E. European Society of Hypertension International Protocol for the validation of blood pressure monitors: a critical review of its application and rationale for revision. Blood Press Monit 2010; 15:39-48.
    • (2010) Blood Press Monit , vol.15 , pp. 39-48
    • Stergiou, G.S.1    Karpettas, N.2    Atkins, N.3    O'Brien, E.4
  • 35
    • 79954632013 scopus 로고    scopus 로고
    • Impact of applying the more stringent validation criteria of the revised European Society of Hypertension International Protocol 2010 on earlier validation studies
    • Stergiou GS, Karpettas N, Atkins N, O'Brien E. Impact of applying the more stringent validation criteria of the revised European Society of Hypertension International Protocol 2010 on earlier validation studies. Blood Press Monit 2011; 16:67-73.
    • (2011) Blood Press Monit , vol.16 , pp. 67-73
    • Stergiou, G.S.1    Karpettas, N.2    Atkins, N.3    O'Brien, E.4
  • 36
    • 34548398159 scopus 로고    scopus 로고
    • The dabl educational trust device equivalence procedure
    • DOI 10.1097/MBP.0b013e32809efa27
    • Atkins N, O'Brien E. The Dabl Educational Trust device equivalence procedure. Blood Press Monit 2007; 12:245-249. (Pubitemid 47356411)
    • (2007) Blood Pressure Monitoring , vol.12 , Issue.4 , pp. 245-249
    • Atkins, N.1    O'Brien, E.2


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.