Volumn 13, Issue 6, 2011, Pages 1077-1087

Qualitative and quantitative assessment of metal artifacts arising from implantable cardiac pacing devices in oncological PET/CT studies: A phantom study

Author keywords

Attenuation correction; Defibrillator; Metal artifact reduction; Misalignment; Pacemaker; PET CT

Indexed keywords

ARTICLE; ARTIFACT REDUCTION; CARDIAC RESYNCHRONIZATION THERAPY DEVICE; COMPUTER ASSISTED EMISSION TOMOGRAPHY; CONTROLLED STUDY; DEFIBRILLATOR; HEART ASSIST DEVICE; HEART PACING; IMAGE ANALYSIS; IMAGE PROCESSING; IMAGE QUALITY; IMAGE RECONSTRUCTION; PACEMAKER ELECTRODE; PHANTOM; PRIORITY JOURNAL; QUALITATIVE ANALYSIS; QUANTITATIVE ANALYSIS;

EID: 84855675037     PISSN: 15361632     EISSN: 18602002     Source Type: Journal    
DOI: 10.1007/s11307-010-0467-x     Document Type: Article
Times cited : (18)

References (22)
  • 1
    • 0031847790 scopus 로고    scopus 로고
    • Transmission scanning in emission tomography
    • DOI 10.1007/s002590050282
    • Bailey DL (1998) Transmission scanning in emission tomography. Eur J Nucl Med 25:774-787 (Pubitemid 28377042)
    • (1998) European Journal of Nuclear Medicine , vol.25 , Issue.7 , pp. 774-787
    • Bailey, D.L.1
  • 2
    • 0037315301 scopus 로고    scopus 로고
    • Determination of the attenuation map in emission tomography
    • Zaidi H, Hasegawa BH (2003) Determination of the attenuation map in emission tomography. J Nucl Med 44:291-315
    • (2003) J Nucl Med , vol.44 , pp. 291-315
    • Zaidi, H.1    Hasegawa, B.H.2
  • 3
    • 77749292368 scopus 로고    scopus 로고
    • Attenuation-emission alignment in cardiac PET/CT based on consistency conditions
    • Alessio AM, Kinahan PE, Champley KM, Caldwell JH (2010) Attenuation-emission alignment in cardiac PET/CT based on consistency conditions. Med Phys 37:1191-1200
    • (2010) Med Phys , vol.37 , pp. 1191-1200
    • Alessio, A.M.1    Kinahan, P.E.2    Champley, K.M.3    Caldwell, J.H.4
  • 4
    • 52449084002 scopus 로고    scopus 로고
    • Advances in attenuation correction techniques in PET
    • Zaidi H, Montandon M-L, Alavi A (2007) Advances in attenuation correction techniques in PET. PET Clinics 2:191-217
    • (2007) PET Clinics , vol.2 , pp. 191-217
    • Zaidi, H.1    Montandon, M.-L.2    Alavi, A.3
  • 5
    • 0037936886 scopus 로고    scopus 로고
    • Impact of metallic dental implants on CT-based attenuation correction in a combined PET/CT scanner
    • Kamel EM, Burger C, Buck A, von Schulthess GK, Goerres GW (2003) Impact of metallic dental implants on CT-based attenuation correction in a combined PET/CT scanner. Eur Radiol 13:724-728 (Pubitemid 36790553)
    • (2003) European Radiology , vol.13 , Issue.4 , pp. 724-728
    • Kamel, E.M.1    Burger, C.2    Buck, A.3    Von Schulthess, G.K.4    Goerres, G.W.5
  • 6
    • 46749100234 scopus 로고    scopus 로고
    • Does reducing CT artifacts from dental implants influence the PET interpretation in PET/CT studies of oral cancer and head and neck cancer?
    • DOI 10.2967/jnumed.107.049858
    • Nahmias C, Lemmens C, Faul D, Carlson E, Long M, Blodgett T et al (2008) Does reducing CT artifacts from dental implants influence the PET interpretation in PET/CT studies of oral cancer and head and neck cancer? J Nucl Med 49:1047-1052 (Pubitemid 351948011)
    • (2008) Journal of Nuclear Medicine , vol.49 , Issue.7 , pp. 1047-1052
    • Nahmias, C.1    Lemmens, C.2    Faul, D.3    Carlson, E.4    Long, M.5    Blodgett, T.6    Nuyts, J.7    Townsend, D.8
  • 8
    • 33744782590 scopus 로고    scopus 로고
    • A knowledge-based method for reducing attenuation artefacts caused by cardiac appliances in myocardial PET/CT
    • DOI 10.1088/0031-9155/51/11/015, PII S0031915506149003, 015
    • Hamill JJ, Brunken RC, Bybel B, DiFilippo FP, Faul DD (2006) A knowledge-based method for reducing attenuation artefacts caused by cardiac appliances in myocardial PET/CT. Phys Med Biol 51:2901-2918 (Pubitemid 43832944)
    • (2006) Physics in Medicine and Biology , vol.51 , Issue.11 , pp. 2901-2918
    • Hamill, J.J.1    Brunken, R.C.2    Bybel, B.3    DiFilippo, F.P.4    Faul, D.D.5
  • 9
    • 74349111841 scopus 로고    scopus 로고
    • A virtual sinogram method to reduce dental metallic implant artefacts in computed tomography-based attenuation correction for PET
    • Abdoli M, Ay MR, Ahmadian A, Zaidi H (2010) A virtual sinogram method to reduce dental metallic implant artefacts in computed tomography-based attenuation correction for PET. Nucl Med Commun 31:22-31
    • (2010) Nucl Med Commun , vol.31 , pp. 22-31
    • Abdoli, M.1    Ay, M.R.2    Ahmadian, A.3    Zaidi, H.4
  • 11
    • 59449107042 scopus 로고    scopus 로고
    • Suppression of metal artifacts in CT using a reconstruction procedure that combines MAP and projection completion
    • Lemmens C, Faul D, Nuyts J (2009) Suppression of metal artifacts in CT using a reconstruction procedure that combines MAP and projection completion. IEEE Trans Med Imaging 28:250-260
    • (2009) IEEE Trans Med Imaging , vol.28 , pp. 250-260
    • Lemmens, C.1    Faul, D.2    Nuyts, J.3
  • 12
    • 0037294323 scopus 로고    scopus 로고
    • Artifacts at PET and PET/CT caused by metallic hip prosthetic material
    • DOI 10.1148/radiol.2262012141
    • Goerres GW, Ziegler SI, Burger C, Berthold T, Von Schulthess GK, Buck A (2003) Artifacts at PET and PET/CT caused by metallic hip prosthetic material. Radiology 226:577-584 (Pubitemid 36313453)
    • (2003) Radiology , vol.226 , Issue.2 , pp. 577-584
    • Goerres, G.W.1    Ziegler, S.I.2    Burger, C.3    Berthold, T.4    Von Schulthess, G.K.5    Buck, A.6
  • 14
    • 37349020985 scopus 로고    scopus 로고
    • Cardiac device implantation in the United States from 1997 through 2004: A population-based analysis
    • DOI 10.1007/s11606-007-0392-0, New Medical Technologies: The Potential and Pitfalls of Innovation, Guest Editors: Jeffrey A. Tice MD, Mitchell D. Feldman MD/MPhil, Deputy Editors: Richard M. Hoffman MD/MPH, Claudia A. SteinerMD/MPH
    • Zhan C, Baine W, Sedrakyan A, Steiner C (2008) Cardiac device implantation in the United States from 1997 through 2004: A population-based analysis. J Gen Intern Med 23(Suppl 1):13-19 (Pubitemid 350295396)
    • (2008) Journal of General Internal Medicine , vol.23 , Issue.1 SUPPL. , pp. 13-19
    • Zhan, C.1    Baine, W.B.2    Sedrakyan, A.3    Steiner, C.4
  • 16
    • 33750434238 scopus 로고    scopus 로고
    • 18F-FDG-PET/CT imaging of the head and neck: An approach to metal artifact correction
    • Schafers KP, Raupach R, Beyer T (2006) Combined 18 F-FDG-PET/ CT imaging of the head and neck. An approach to metal artifact correction. Nuklearmediziner 45:219-222 (Pubitemid 44645969)
    • (2006) NuklearMedizin , vol.45 , Issue.5 , pp. 219-222
    • Schafers, K.1    Raupach, R.2    Beyer, T.3
  • 17
    • 33747599964 scopus 로고    scopus 로고
    • Attenuation-correction induced artifact in F-18 FDG PET imaging following total knee replacement
    • Heiba SI, Luo J, Sadek S, Macalental E, Cacavio A, Rosen G et al (2000) Attenuation-correction induced artifact in F-18 FDG PET imaging following total knee replacement. Clin Positron Imaging 3:237-239
    • (2000) Clin Positron Imaging , vol.3 , pp. 237-239
    • Heiba, S.I.1    Luo, J.2    Sadek, S.3    Macalental, E.4    Cacavio, A.5    Rosen, G.6
  • 18
    • 0037908633 scopus 로고    scopus 로고
    • 18F]fluoro-D-glucose localization surrounding metallic objects in a PET/CT scanner using CT-based attenuation correction
    • DOI 10.1016/S1536-1632(03)00037-4, PII S1536163203000374
    • Bujenovic S, Mannting F, Chakrabarti R, Ladnier D (2003) Artifactual 2-deoxy-2-[(18)F]fluoro-D-glucose localization surrounding metallic objects in a PET/CT scanner using CT-based attenuation correction. Mol Imaging Biol 5:20-22 (Pubitemid 36596540)
    • (2003) Molecular Imaging and Biology , vol.5 , Issue.1 , pp. 20-22
    • Bujenovic, S.1    Mannting, F.2    Chakrabarti, R.3    Ladnier, D.4
  • 20
    • 17844369745 scopus 로고    scopus 로고
    • Do implanted pacemaker leads and ICD leads cause metal-related artifact in cardiac PET/CT?
    • DiFilippo FP, Brunken RC (2005) Do implanted pacemaker leads and ICD leads cause metal-related artifact in cardiac PET/CT? J Nucl Med 46:436-443 (Pubitemid 43093152)
    • (2005) Journal of Nuclear Medicine , vol.46 , Issue.3 , pp. 436-443
    • DiFilippo, F.P.1    Brunken, R.C.2
  • 21
    • 0027367203 scopus 로고
    • Standardized uptake values of normal tissues at PET with 2-[fluorine-18]- fluoro-2-deoxy-D-glucose: Variations with body weight and a method for correction
    • Zasadny KR, Wahl RL (1993) Standardized uptake values of normal tissues at PET with 2-[fluorine-18]-fluoro-2-deoxy-D-glucose: Variations with body weight and a method for correction. Radiology 189:847-850 (Pubitemid 23354506)
    • (1993) Radiology , vol.189 , Issue.3 , pp. 847-850
    • Zasadny, K.R.1    Wahl, R.L.2
  • 22
    • 78650117233 scopus 로고    scopus 로고
    • Reduction of dental filling metallic artefacts in CT-based attenuation correction of PET data using weighted virtual sinograms optimized by a genetic algorithm
    • Abdoli M, Ay M, Ahmadian A, Dierckx R, Zaidi H. (2010) Reduction of dental filling metallic artefacts in CT-based attenuation correction of PET data using weighted virtual sinograms optimized by a genetic algorithm. Med Phys 37:6166-6177
    • (2010) Med Phys , vol.37 , pp. 6166-6177
    • Abdoli, M.1    Ay, M.2    Ahmadian, A.3    Dierckx, R.4    Zaidi, H.5


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.