Volumn 20, Issue 2, 2010, Pages 207-215

Comparison of ultrasonic Doppler flow monitor, oscillometric, and direct arterial blood pressure measurements in ill dogs

Author keywords

Cardiovascular monitoring; Critical care monitoring; Intensive care medicine; Small animal critical care

Indexed keywords

ANIMAL; ANIMAL DISEASE; ARTICLE; BLOOD PRESSURE; BLOOD PRESSURE MEASUREMENT; COMPARATIVE STUDY; DOG; DOG DISEASE; DOPPLER FLOWMETRY; ECHOGRAPHY; EVALUATION; FEMALE; HOSPITAL; HYPERTENSION; HYPOTENSION; MALE; METHODOLOGY; OSCILLOMETRY; PATHOPHYSIOLOGY; PHYSIOLOGY; PROSPECTIVE STUDY; STANDARD; UNITED STATES; UNIVERSITY HOSPITAL; DOG DISEASES; EVALUATION STUDY; PROCEDURES; STANDARDS; VETERINARY;

EID: 77952796818     PISSN: 14793261     EISSN: 14764431     Source Type: Journal    
DOI: 10.1111/j.1476-4431.2010.00520.x     Document Type: Article
Times cited : (78)

References (32)
  • 1
    • 0030184741 scopus 로고    scopus 로고
    • Comparison of direct and indirect (oscillometric) measurements of arterial blood pressure in conscious dogs
    • Bodey AR, Michell AR, Bovee KC. Comparison of direct and indirect (oscillometric) measurements of arterial blood pressure in conscious dogs. Res Vet Sci 1996, 61((1)):17-21.
    • (1996) Res Vet Sci , vol.61 , Issue.1 , pp. 17-21
    • Bodey, A.R.1    Michell, A.R.2    Bovee, K.C.3    et al4
  • 2
    • 0029331209 scopus 로고
    • Comparison of direct and indirect blood pressure measurements in anesthetized dogs
    • Gains MJ, Grodecki KM, Jacobs RM. Comparison of direct and indirect blood pressure measurements in anesthetized dogs. Can J Vet Res 1995, 59((3)):238-240.
    • (1995) Can J Vet Res , vol.59 , Issue.3 , pp. 238-240
    • Gains, M.J.1    Grodecki, K.M.2    Jacobs, R.M.3    et al4
  • 3
    • 0001108377 scopus 로고    scopus 로고
    • Clinical comparison of three methods to measure blood pressure in nonsedated dogs
    • Stepien RL, Rapoport GS. Clinical comparison of three methods to measure blood pressure in nonsedated dogs. J Am Vet Med Assoc 1999, 215((11)):1623-1628.
    • (1999) J Am Vet Med Assoc , vol.215 , Issue.11 , pp. 1623-1628
    • Stepien, R.L.1    Rapoport, G.S.2
  • 4
    • 0028922101 scopus 로고
    • American National Standard for nonautomated sphygmomanometers. Summary Report
    • Prisant LM, Alpert BS, Robbins CB. American National Standard for nonautomated sphygmomanometers. Summary Report. Am J Hypertens 1995, 8((2)):210-213.
    • (1995) Am J Hypertens , vol.8 , Issue.2 , pp. 210-213
    • Prisant, L.M.1    Alpert, B.S.2    Robbins, C.B.3    et al4
  • 5
    • 34249341125 scopus 로고    scopus 로고
    • Guidelines for the identification, evaluation, and management of systemic hypertension in dogs and cats
    • Brown S, Atkins C, Bagley R. Guidelines for the identification, evaluation, and management of systemic hypertension in dogs and cats. J Vet Intern Med 2007, 21((3)):542-558.
    • (2007) J Vet Intern Med , vol.21 , Issue.3 , pp. 542-558
    • Brown, S.1    Atkins, C.2    Bagley, R.3    et al4
  • 6
    • 0028538907 scopus 로고
    • A comparison of direct and indirect (oscillometric) measurements of arterial blood pressure in anaesthetised dogs, using tail and limb cuffs
    • Bodey AR, Young LE, Bartram DH. A comparison of direct and indirect (oscillometric) measurements of arterial blood pressure in anaesthetised dogs, using tail and limb cuffs. Res Vet Sci 1994, 57((3)):265-269.
    • (1994) Res Vet Sci , vol.57 , Issue.3 , pp. 265-269
    • Bodey, A.R.1    Young, L.E.2    Bartram, D.H.3    et al4
  • 7
    • 38349070676 scopus 로고    scopus 로고
    • Clinical evaluation of the Surgivet V60046, a noninvasive blood pressure monitor in anaesthetized dogs
    • Deflandre CJ, Hellebrekers LJ. Clinical evaluation of the Surgivet V60046, a noninvasive blood pressure monitor in anaesthetized dogs. Vet Anaesth Analg 2008, 35((1)):13-21.
    • (2008) Vet Anaesth Analg , vol.35 , Issue.1 , pp. 13-21
    • Deflandre, C.J.1    Hellebrekers, L.J.2
  • 8
    • 33748743900 scopus 로고    scopus 로고
    • Accuracy of an oscillometric blood pressure monitor during phenylephrine-induced hypertension in dogs
    • McMurphy RM, Stoll MR, McCubrey R. Accuracy of an oscillometric blood pressure monitor during phenylephrine-induced hypertension in dogs. Am J Vet Res 2006, 67((9)):1541-1545.
    • (2006) Am J Vet Res , vol.67 , Issue.9 , pp. 1541-1545
    • McMurphy, R.M.1    Stoll, M.R.2    McCubrey, R.3
  • 9
    • 0030278419 scopus 로고    scopus 로고
    • Comparison of the indirect oscillometric and direct arterial methods for blood pressure measurements in anesthetized dogs
    • Meurs KM, Miller MW, Slater MR. Comparison of the indirect oscillometric and direct arterial methods for blood pressure measurements in anesthetized dogs. J Am Anim Hosp Assoc 1996, 32((6)):471-475.
    • (1996) J Am Anim Hosp Assoc , vol.32 , Issue.6 , pp. 471-475
    • Meurs, K.M.1    Miller, M.W.2    Slater, M.R.3
  • 10
    • 0025783122 scopus 로고
    • Comparison of direct and indirect blood pressure measurement in anesthetized dogs
    • Sawyer DC, Brown M, Striler EL. Comparison of direct and indirect blood pressure measurement in anesthetized dogs. Lab Anim Sci 1991, 41((2)):134-138.
    • (1991) Lab Anim Sci , vol.41 , Issue.2 , pp. 134-138
    • Sawyer, D.C.1    Brown, M.2    Striler, E.L.3    et al4
  • 11
    • 1542426070 scopus 로고    scopus 로고
    • Evaluation of a new oscillometric blood pressure monitor in isoflurane-anesthetized dogs
    • Sawyer DC, Guikema AH, Siegel EM. Evaluation of a new oscillometric blood pressure monitor in isoflurane-anesthetized dogs. Vet Anaesth Analg 2004, 31((1)):27-39.
    • (2004) Vet Anaesth Analg , vol.31 , Issue.1 , pp. 27-39
    • Sawyer, D.C.1    Guikema, A.H.2    Siegel, E.M.3
  • 12
    • 34248569701 scopus 로고    scopus 로고
    • Indirect measurement of blood pressure using a pulse oximeter in isoflurane anesthetized dogs
    • Dyson D. Indirect measurement of blood pressure using a pulse oximeter in isoflurane anesthetized dogs. J Vet Emerg Crit Care 2007, 17((2)):135-142.
    • (2007) J Vet Emerg Crit Care , vol.17 , Issue.2 , pp. 135-142
    • Dyson, D.1
  • 13
    • 33746430032 scopus 로고    scopus 로고
    • Evaluation of oscillometric and Doppler ultrasonic methods of indirect blood pressure estimation in conscious dogs
    • Haberman CE, Kang CW, Morgan JD. Evaluation of oscillometric and Doppler ultrasonic methods of indirect blood pressure estimation in conscious dogs. Can J Vet Res 2006, 70((3)):211-217.
    • (2006) Can J Vet Res , vol.70 , Issue.3 , pp. 211-217
    • Haberman, C.E.1    Kang, C.W.2    Morgan, J.D.3    et al4
  • 15
    • 0031179887 scopus 로고    scopus 로고
    • Evaluation of an oscillometric blood pressure monitor on anesthetized cats and the effect of cuff placement and fur on accuracy
    • Branson KR, Wagner-Mann CC, Mann FA. Evaluation of an oscillometric blood pressure monitor on anesthetized cats and the effect of cuff placement and fur on accuracy. Vet Surg 1997, 26((4)):347-353.
    • (1997) Vet Surg , vol.26 , Issue.4 , pp. 347-353
    • Branson, K.R.1    Wagner-Mann, C.C.2    Mann, F.A.3
  • 16
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986, 1((8476)):307-310.
    • (1986) Lancet , vol.1 , Issue.8476 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 17
    • 0010608633 scopus 로고
    • Use of blood pressure monitors
    • Kirk R, Bonagura JD. eds, 11th ed, Philadelphia, WB Saunders Co
    • Podell M. Use of blood pressure monitors. Kirk's Current Veterinary Therapy XI: Small Animal Practice 1992, 834-837. Kirk RBonagura JD. In, eds, 11th ed, Philadelphia, WB Saunders Co, pp
    • (1992) Kirk's Current Veterinary Therapy XI: Small Animal Practice , pp. 834-837
    • Podell, M.1
  • 18
    • 0011435235 scopus 로고
    • Blood pressure measurement
    • Bonagura JD. ed, 12th ed, Philadelphia, WB Saunders Co
    • Hansen B. Blood pressure measurement. Kirk's Current Veterinary Therapy XII: Small Animal Practice 1995, 110-112. Bonagura JD. In, ed, 12th ed, Philadelphia, WB Saunders Co, pp
    • (1995) Kirk's Current Veterinary Therapy XII: Small Animal Practice , pp. 110-112
    • Hansen, B.1
  • 19
    • 0019570413 scopus 로고
    • Comparison of direct and indirect methods of measuring arterial blood pressure, part III
    • Bruner JM, Krenis LJ, Kunsman JM. Comparison of direct and indirect methods of measuring arterial blood pressure, part III. Med Instrum 1981, 15((3)):182-188.
    • (1981) Med Instrum , vol.15 , Issue.3 , pp. 182-188
    • Bruner, J.M.1    Krenis, L.J.2    Kunsman, J.M.3    et al4
  • 20
    • 0019544832 scopus 로고
    • Comparison of direct and indirect methods of measuring arterial blood pressure, part II
    • Bruner JM, Krenis LJ, Kunsman JM. Comparison of direct and indirect methods of measuring arterial blood pressure, part II. Med Instrum 1981, 15((2)):97-101.
    • (1981) Med Instrum , vol.15 , Issue.2 , pp. 97-101
    • Bruner, J.M.1    Krenis, L.J.2    Kunsman, J.M.3    et al4
  • 21
    • 0019445970 scopus 로고
    • Comparison of direct and indirect measuring arterial blood pressure
    • Bruner JM, Krenis LJ, Kunsman JM. Comparison of direct and indirect measuring arterial blood pressure. Med Instrum 1981, 15((1)):11-21.
    • (1981) Med Instrum , vol.15 , Issue.1 , pp. 11-21
    • Bruner, J.M.1    Krenis, L.J.2    Kunsman, J.M.3    et al4
  • 22
    • 0027479012 scopus 로고
    • National standard for measurement of resting and ambulatory blood pressures with automated sphygmomanometers
    • White WB, Berson AS, Robbins C. National standard for measurement of resting and ambulatory blood pressures with automated sphygmomanometers. Hypertension 1993, 21((4)):504-509.
    • (1993) Hypertension , vol.21 , Issue.4 , pp. 504-509
    • White, W.B.1    Berson, A.S.2    Robbins, C.3    et al4
  • 23
    • 0036730658 scopus 로고    scopus 로고
    • Evaluation of an oscillometric blood pressure monitor for use in anesthetized cats
    • Pedersen KM, Butler MA, Ersboll AK. Evaluation of an oscillometric blood pressure monitor for use in anesthetized cats. J Am Vet Med Assoc 2002, 221((5)):646-650.
    • (2002) J Am Vet Med Assoc , vol.221 , Issue.5 , pp. 646-650
    • Pedersen, K.M.1    Butler, M.A.2    Ersboll, A.K.3    et al4
  • 24
    • 0030095950 scopus 로고    scopus 로고
    • Epidemiological study of blood pressure in domestic dogs
    • Bodey AR, Michell AR. Epidemiological study of blood pressure in domestic dogs. J Small Anim Pract 1996, 37((3)):116-125.
    • (1996) J Small Anim Pract , vol.37 , Issue.3 , pp. 116-125
    • Bodey, A.R.1    Michell, A.R.2
  • 25
    • 0019011073 scopus 로고
    • Indirect mean arterial pressure in the anesthetized dog
    • Geddes LA, Combs W, Denton W. Indirect mean arterial pressure in the anesthetized dog. Am J Physiol 1980, 238((5)):H664-H666.
    • (1980) Am J Physiol , vol.238 , Issue.5
    • Geddes, L.A.1    Combs, W.2    Denton, W.3    et al4
  • 26
    • 0015291888 scopus 로고
    • Indirect blood pressure determination by the ultrasonic Doppler technique in dogs
    • Freundlich JJ, Detweiler DK, Hance HE. Indirect blood pressure determination by the ultrasonic Doppler technique in dogs. Curr Ther Res Clin Exp 1972, 14((2)):73-80.
    • (1972) Curr Ther Res Clin Exp , vol.14 , Issue.2 , pp. 73-80
    • Freundlich, J.J.1    Detweiler, D.K.2    Hance, H.E.3
  • 27
    • 77957176794 scopus 로고
    • Noninvasive measurement of systemic arterial blood pressure in the conscious beagle dog
    • Pettersen JC, Linartz RR, Hamlin RL. Noninvasive measurement of systemic arterial blood pressure in the conscious beagle dog. Fundam Appl Toxicol 1988, 10((1)):89-97.
    • (1988) Fundam Appl Toxicol , vol.10 , Issue.1 , pp. 89-97
    • Pettersen, J.C.1    Linartz, R.R.2    Hamlin, R.L.3    et al4
  • 28
    • 0029401679 scopus 로고
    • Doppler ultrasonographic, oscillometric sphygmomanometric, and photoplethysmographic techniques for noninvasive blood pressure measurement in anesthetized cats
    • Binns SH, Sisson DD, Buoscio DA. Doppler ultrasonographic, oscillometric sphygmomanometric, and photoplethysmographic techniques for noninvasive blood pressure measurement in anesthetized cats. J Vet Intern Med 1995, 9((6)):405-414.
    • (1995) J Vet Intern Med , vol.9 , Issue.6 , pp. 405-414
    • Binns, S.H.1    Sisson, D.D.2    Buoscio, D.A.3    et al4
  • 30
    • 0026887145 scopus 로고
    • Evaluation of the Doppler ultrasonic method of measuring systolic arterial blood pressure in cats
    • Grandy JL, Dunlop CI, Hodgson DS. Evaluation of the Doppler ultrasonic method of measuring systolic arterial blood pressure in cats. Am J Vet Res 1992, 53((7)):1166-1169.
    • (1992) Am J Vet Res , vol.53 , Issue.7 , pp. 1166-1169
    • Grandy, J.L.1    Dunlop, C.I.2    Hodgson, D.S.3    et al4
  • 31
    • 0032114336 scopus 로고    scopus 로고
    • A comparison of indirect blood pressure monitoring techniques in the anesthetized cat
    • Caulkett NA, Cantwell SL, Houston DM. A comparison of indirect blood pressure monitoring techniques in the anesthetized cat. Vet Surg 1998, 27((4)):370-377.
    • (1998) Vet Surg , vol.27 , Issue.4 , pp. 370-377
    • Caulkett, N.A.1    Cantwell, S.L.2    Houston, D.M.3
  • 32
    • 0014218319 scopus 로고
    • Blood pressure measurement in shock: mechanism of inaccuracy in ausculatory and palpatory methods
    • Cohn JN. Blood pressure measurement in shock: mechanism of inaccuracy in ausculatory and palpatory methods. J Am Med Aassoc 1967, 199((13)):118-122.
    • (1967) J Am Med Aassoc , vol.199 , Issue.13 , pp. 118-122
    • Cohn, J.N.1


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.