Volumn 254, Issue 3, 2010, Pages 698-706

Radiation dose from single-heartbeat coronary CT angiography performed with a 320-detector row volume scanner

Author keywords

[No Author keywords available]

Indexed keywords

ANGIOCARDIOGRAPHY; ARTICLE; COMPUTED TOMOGRAPHIC ANGIOGRAPHY; COMPUTED TOMOGRAPHY SCANNER; DOSIMETRY; IMAGE ANALYSIS; MONTE CARLO METHOD; MULTIDETECTOR COMPUTED TOMOGRAPHY; PHANTOM; PRIORITY JOURNAL; RADIATION EXPOSURE; SEMICONDUCTOR; SIMULATION; SPIRAL COMPUTER ASSISTED TOMOGRAPHY;

EID: 77950143210     PISSN: 00338419     EISSN: 15271315     Source Type: Journal    
DOI: 10.1148/radiol.09090779     Document Type: Article
Times cited : (150)

References (23)
  • 1
    • 48449089390 scopus 로고    scopus 로고
    • 64-slice CT for diagnosis of coronary artery disease: A systematic review
    • Stein PD, Yaekoub AY, Matta F, Sostman HD. 64-slice CT for diagnosis of coronary artery disease: a systematic review. Am J Med 2008;121:715-725.
    • (2008) Am J Med , vol.121 , pp. 715-725
    • Stein, P.D.1    Yaekoub, A.Y.2    Matta, F.3    Sostman, H.D.4
  • 2
    • 67650536250 scopus 로고    scopus 로고
    • Ionizing radiation exposure of the population of the United States
    • National Council on Radiation Protection and Measurements, Report no. 160. Bethesda, Md: National Council on Radiation Protection and Measurements
    • National Council on Radiation Protection and Measurements. Ionizing radiation exposure of the population of the United States. Report no. 160. Bethesda, Md: National Council on Radiation Protection and Measurements, 2009.
    • (2009)
  • 3
    • 34447558140 scopus 로고    scopus 로고
    • Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography
    • Einstein AJ, Henzlova MJ, Rajagopalan S. Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA 2007;298:317-323.
    • (2007) JAMA , vol.298 , pp. 317-323
    • Einstein, A.J.1    Henzlova, M.J.2    Rajagopalan, S.3
  • 4
    • 67049116139 scopus 로고    scopus 로고
    • Radiation dose from cardiac computed tomography before and after implementation of radiation dose-reduction techniques
    • Raff GL, Chinnaiyan KM, Share DA, et al. Radiation dose from cardiac computed tomography before and after implementation of radiation dose-reduction techniques. JAMA 2009; 301:2340-2348.
    • (2009) JAMA , vol.301 , pp. 2340-2348
    • Raff, G.L.1    Chinnaiyan, K.M.2    Share, D.A.3
  • 5
    • 49049116757 scopus 로고    scopus 로고
    • Prospective versus retrospective ECG gating for 64-detector CT of the coronary arteries: Comparison of image quality and patient radiation dose
    • Shuman WP, Branch KR, May JM, et al. Prospective versus retrospective ECG gating for 64-detector CT of the coronary arteries: comparison of image quality and patient radiation dose. Radiology 2008;248:431-437.
    • (2008) Radiology , vol.248 , pp. 431-437
    • Shuman, W.P.1    Branch, K.R.2    May, J.M.3
  • 6
    • 40949165279 scopus 로고    scopus 로고
    • Prospectively gated transverse coronary CT angiography versus retrospectively gated helical technique: Improved image quality and reduced radiation dose
    • Earls JP, Berman EL, Urban BA, et al. Prospectively gated transverse coronary CT angiography versus retrospectively gated helical technique: improved image quality and reduced radiation dose. Radiology 2008;246:742-753.
    • (2008) Radiology , vol.246 , pp. 742-753
    • Earls, J.P.1    Berman, E.L.2    Urban, B.A.3
  • 7
    • 56949107203 scopus 로고    scopus 로고
    • Effect of dual-source cardiac computed tomography on patient radiation dose in a clinical setting: Comparison to single-source imaging
    • Halliburton SS, Sola S, Kuzmiak SA, et al. Effect of dual-source cardiac computed tomography on patient radiation dose in a clinical setting: comparison to single-source imaging. J Cardiovasc Comput Tomogr 2008;2:392-400.
    • (2008) J Cardiovasc Comput Tomogr , vol.2 , pp. 392-400
    • Halliburton, S.S.1    Sola, S.2    Kuzmiak, S.A.3
  • 8
    • 77950116874 scopus 로고    scopus 로고
    • The 2007 recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP 2007;37:1-332.
    • The 2007 recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP 2007;37:1-332.
  • 9
    • 0025930020 scopus 로고
    • ICRP publication 60
    • 1990 Recommendations of the International Commission on Radiological Protection
    • 1990 Recommendations of the International Commission on Radiological Protection. ICRP publication 60. Ann ICRP 1991;21:1-201.
    • (1991) Ann ICRP , vol.21 , pp. 1-201
  • 12
    • 59649127716 scopus 로고    scopus 로고
    • Estimated radiation dose associated with cardiac CT angiography
    • Hausleiter J, Meyer T, Hermann F, et al. Estimated radiation dose associated with cardiac CT angiography. JAMA 2009;301:500-507.
    • (2009) JAMA , vol.301 , pp. 500-507
    • Hausleiter, J.1    Meyer, T.2    Hermann, F.3
  • 13
    • 49649101130 scopus 로고    scopus 로고
    • Response functions for computing absorbed dose to skeletal tissues from photon irradiation
    • Eckerman KF, Bolch WE, Zankl M, Petoussi-Henss N. Response functions for computing absorbed dose to skeletal tissues from photon irradiation. Radiat Prot Dosimetry 2007;127:187-191.
    • (2007) Radiat Prot Dosimetry , vol.127 , pp. 187-191
    • Eckerman, K.F.1    Bolch, W.E.2    Zankl, M.3    Petoussi-Henss, N.4
  • 14
    • 77950154352 scopus 로고    scopus 로고
    • International Electrotechnical Commission. Medical electrical equipment: part 2-44 - particular requirements for the basic safety and essential performance of x-ray equipment for computed tomography. International Standard IEC 60601-2-44 edition 3.0. Geneva, Switzerland: International Electrotechnical Commission, 2009.
    • International Electrotechnical Commission. Medical electrical equipment: part 2-44 - particular requirements for the basic safety and essential performance of x-ray equipment for computed tomography. International Standard IEC 60601-2-44 edition 3.0. Geneva, Switzerland: International Electrotechnical Commission, 2009.
  • 16
    • 54349126659 scopus 로고    scopus 로고
    • 64-Slice computed tomography angiography in the diagnosis and assessment of coronary artery disease: Systematic review and meta-analysis
    • Mowatt G, Cook JA, Hillis GS, et al. 64-Slice computed tomography angiography in the diagnosis and assessment of coronary artery disease: systematic review and meta-analysis. Heart 2008;94:1386-1393.
    • (2008) Heart , vol.94 , pp. 1386-1393
    • Mowatt, G.1    Cook, J.A.2    Hillis, G.S.3
  • 17
    • 43449126798 scopus 로고    scopus 로고
    • Initial evaluation of coronary images from 320-detector row computed tomography
    • Rybicki FJ, Otero HJ, Steigner ML, et al. Initial evaluation of coronary images from 320-detector row computed tomography. Int J Cardiovasc Imaging 2008;24:535-546.
    • (2008) Int J Cardiovasc Imaging , vol.24 , pp. 535-546
    • Rybicki, F.J.1    Otero, H.J.2    Steigner, M.L.3
  • 18
    • 33751235554 scopus 로고    scopus 로고
    • Conversion factor for CT dosimetry to assess patient dose using a 256-slice CT scanner
    • Mori S, Nishizawa K, Ohno M, Endo M. Conversion factor for CT dosimetry to assess patient dose using a 256-slice CT scanner. Br J Radiol 2006;79:888-892.
    • (2006) Br J Radiol , vol.79 , pp. 888-892
    • Mori, S.1    Nishizawa, K.2    Ohno, M.3    Endo, M.4
  • 20
    • 47249135481 scopus 로고    scopus 로고
    • Computing effective doses from dose-length product in CT [letter]
    • Huda W, Yoshizumi TT, Hurwitz LM, Goodman PC. Computing effective doses from dose-length product in CT [letter]. Radiology 2008;248:321-322.
    • (2008) Radiology , vol.248 , pp. 321-322
    • Huda, W.1    Yoshizumi, T.T.2    Hurwitz, L.M.3    Goodman, P.C.4
  • 21
    • 67849105498 scopus 로고    scopus 로고
    • Prospectively gated low-dose CCTA:24 months experience in more than 2,000 clinical cases
    • Earls JP, Schrack EC. Prospectively gated low-dose CCTA:24 months experience in more than 2,000 clinical cases. Int J Cardiovasc Imaging 2008;25:177-187.
    • (2008) Int J Cardiovasc Imaging , vol.25 , pp. 177-187
    • Earls, J.P.1    Schrack, E.C.2
  • 22
    • 67949094410 scopus 로고    scopus 로고
    • New radiation dose saving technologies for 256-slice cardiac computed tomography angiography
    • Walker MJ, Olszewski ME, Desai MY, et al. New radiation dose saving technologies for 256-slice cardiac computed tomography angiography. Int J Cardiovasc Imaging 2009;25:189-199.
    • (2009) Int J Cardiovasc Imaging , vol.25 , pp. 189-199
    • Walker, M.J.1    Olszewski, M.E.2    Desai, M.Y.3
  • 23
    • 34247546063 scopus 로고    scopus 로고
    • Coronary artery calcium: A multi-institutional, multimanufacturer international standard for quantification at cardiac CT
    • McCollough CH, Ulzheimer S, Halliburton SS, Shanneik K, White RD, Kalender WA. Coronary artery calcium: a multi-institutional, multimanufacturer international standard for quantification at cardiac CT. Radiology 2007;243:527-538.
    • (2007) Radiology , vol.243 , pp. 527-538
    • McCollough, C.H.1    Ulzheimer, S.2    Halliburton, S.S.3    Shanneik, K.4    White, R.D.5    Kalender, W.A.6


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.