Attachment systems for mandibular two-implant overdentures: A review of in vitro investigations on retention and wear features

Author keywords

[No Author keywords available]

Indexed keywords

DENTAL SURGERY; DENTURE; HUMAN; MATERIALS TESTING; REVIEW; TOOTH IMPLANTATION; TOOTH PROSTHESIS;

EID: 77649222402     PISSN: 08932174     EISSN: None     Source Type: Journal    
DOI: None     Document Type: Article
Times cited : (111)

References (40)
  • 3
    • 0019571593 scopus 로고
    • Magnetic retention for complete and partial overdenture. Part 1
    • Gillings BR. Magnetic retention for complete and partial overdenture. Part 1. J Prosthet Dent 1981; 45: 484-491.
    • (1981) J Prosthet Dent , vol.45 , pp. 484-491
    • Gillings, B.R.1
  • 4
    • 0017483288 scopus 로고
    • Attachment fixation for overdenture. Part 1
    • Mensor MC Jr. Attachment fixation for overdenture. Part 1. J Prosthet Dent 1977; 37: 366-372.
    • (1977) J Prosthet Dent , vol.37 , pp. 366-372
    • Mensor Jr, M.C.1
  • 6
    • 0343175200 scopus 로고
    • The bar joint mandibular denture
    • Dolder E. The bar joint mandibular denture. J Prosthet Dent 1961; 11: 689-707.
    • (1961) J Prosthet Dent , vol.11 , pp. 689-707
    • Dolder, E.1
  • 7
    • 0042735730 scopus 로고    scopus 로고
    • Overdentures supported by ITI implants
    • Schroeder A, Sutter F, Buser D, Krekeler G (eds), New York: Thieme Medical
    • Mericske-Stern R. Overdentures supported by ITI implants. In: Schroeder A, Sutter F, Buser D, Krekeler G (eds). Oral Implantology: Basics, ITI Hollow Cylinder System. New York: Thieme Medical, 1996: 330-367.
    • (1996) Oral Implantology: Basics, ITI Hollow Cylinder System , pp. 330-367
    • Mericske-Stern, R.1
  • 8
    • 0023412024 scopus 로고
    • A classification of precision attachments
    • Becerra G, MacEntee M. A classification of precision attachments. J Prosthet Dent 1987; 58: 322-327.
    • (1987) J Prosthet Dent , vol.58 , pp. 322-327
    • Becerra, G.1    MacEntee, M.2
  • 9
    • 0030159211 scopus 로고    scopus 로고
    • Retention force changes in implant-supported titanium telescope crowns over long-term use in vitro
    • Besimo CH, Graber G, Flühler M. Retention force changes in implant-supported titanium telescope crowns over long-term use in vitro. J Oral Rehabil 1996; 23: 372-378.
    • (1996) J Oral Rehabil , vol.23 , pp. 372-378
    • Besimo, C.H.1    Graber, G.2    Flühler, M.3
  • 10
    • 0035489302 scopus 로고    scopus 로고
    • Orientation of retentive matrices on spherical attachments independent of implant parallelism
    • Wiemeyer AS, Agar JR, Kazemi RB. Orientation of retentive matrices on spherical attachments independent of implant parallelism. J Prosthet Dent 2001; 86: 434-437.
    • (2001) J Prosthet Dent , vol.86 , pp. 434-437
    • Wiemeyer, A.S.1    Agar, J.R.2    Kazemi, R.B.3
  • 12
    • 84995755085 scopus 로고    scopus 로고
    • Review of the clinical use of magnets in implantretained overdentures
    • Maeda Y, Walmsley AD (eds), Tokyo: Quintessence
    • Walmsley AD. Review of the clinical use of magnets in implantretained overdentures. In: Maeda Y, Walmsley AD (eds). Implant Dentistry with New Generation Magnetic Attachments. Tokyo: Quintessence, 2005: 14-17.
    • (2005) Implant Dentistry with New Generation Magnetic Attachments , pp. 14-17
    • Walmsley, A.D.1
  • 14
    • 0036632312 scopus 로고    scopus 로고
    • The McGill Consensus Statement on Overdentur. Monreal, Quebec, Canada. May 24-25, 2002
    • Feine JS, Carlsson GE, Awad MA, et al. The McGill Consensus Statement on Overdentur. Monreal, Quebec, Canada. May 24-25, 2002. Int J Prosthodont 2002; 15: 413-414.
    • (2002) Int J Prosthodont , vol.15 , pp. 413-414
    • Feine, J.S.1    Carlsson, G.E.2    Awad, M.A.3
  • 15
    • 0033122044 scopus 로고    scopus 로고
    • Mandibular overdentures stabilized by Astra Tech implants with either ball attachments or magnets: 5-year results
    • Davis DM, Packer ME. Mandibular overdentures stabilized by Astra Tech implants with either ball attachments or magnets: 5-year results. Int J Prosthodont 1999; 12: 222-229.
    • (1999) Int J Prosthodont , vol.12 , pp. 222-229
    • Davis, D.M.1    Packer, M.E.2
  • 16
    • 11144233686 scopus 로고    scopus 로고
    • A clinical trial of patient satisfaction and prosthodontic needs with ball and bar attachments for implant-retained complete overdentures: Three-year results
    • MacEntee MI, Walton JN, Glick N. A clinical trial of patient satisfaction and prosthodontic needs with ball and bar attachments for implant-retained complete overdentures: Three-year results. J Prosthet Dent 2005; 93: 28-37.
    • (2005) J Prosthet Dent , vol.93 , pp. 28-37
    • MacEntee, M.I.1    Walton, J.N.2    Glick, N.3
  • 17
    • 0141682755 scopus 로고    scopus 로고
    • A randomized clinical trial comparing two mandibular implant overdenture designs: 3-year prosthetic outcomes using six-field protocol
    • Walton JN. A randomized clinical trial comparing two mandibular implant overdenture designs: 3-year prosthetic outcomes using six-field protocol. Int J Prosthodont 2003; 16: 255-260.
    • (2003) Int J Prosthodont , vol.16 , pp. 255-260
    • Walton, J.N.1
  • 18
    • 34648846775 scopus 로고    scopus 로고
    • Investigation of the wear of prefabricated attachments-An in vitro study of retention forces and fitting tolerances
    • Bayer S, Grüner M, Keilig L, et al. Investigation of the wear of prefabricated attachments-An in vitro study of retention forces and fitting tolerances. Quintessence Int 2007; 38: e229-e237.
    • (2007) Quintessence Int , vol.38 , pp. e229-e237
    • Bayer, S.1    Grüner, M.2    Keilig, L.3
  • 19
    • 0041920814 scopus 로고    scopus 로고
    • In vitro retention force changes of prefabricated attachments for overdentures
    • Besimo CE, Guarneri A. In vitro retention force changes of prefabricated attachments for overdentures. J Oral Rehabil 2003; 30: 671-678.
    • (2003) J Oral Rehabil , vol.30 , pp. 671-678
    • Besimo, C.E.1    Guarneri, A.2
  • 20
    • 0029283713 scopus 로고
    • Prospective clinical evaluation of mandibular implant overdentures. Part II-Patient satisfaction and preference
    • Burns DR, Unger JW, Elswick RK Jr, Gigilo JA. Prospective clinical evaluation of mandibular implant overdentures. Part II-Patient satisfaction and preference. J Prosthet Dent 1995; 73: 364-369.
    • (1995) J Prosthet Dent , vol.73 , pp. 364-369
    • Burns, D.R.1    Unger, J.W.2    Elswick Jr, R.K.3    Gigilo, J.A.4
  • 21
    • 17144363947 scopus 로고    scopus 로고
    • Patient satisfaction and preference with magnet, bar-clip, and ball-socket retained mandibular implant overdentures: A cross-over clinical trial
    • Cune M, van Kampen F, van der Bilt A, Bosman F. Patient satisfaction and preference with magnet, bar-clip, and ball-socket retained mandibular implant overdentures: A cross-over clinical trial. Int J Prosthodont 2005; 18: 99-105.
    • (2005) Int J Prosthodont , vol.18 , pp. 99-105
    • Cune, M.1    van Kampen, F.2    van der Bilt, A.3    Bosman, F.4
  • 22
    • 0037637381 scopus 로고
    • Studies on the retention forces on snapon attachments
    • Lehmann KM, Arnim FV. Studies on the retention forces on snapon attachments. Quintessence Dent Technol 1978; 7: 45-48.
    • (1978) Quintessence Dent Technol , vol.7 , pp. 45-48
    • Lehmann, K.M.1    Arnim, F.V.2
  • 23
    • 0020651579 scopus 로고
    • Retention and wear of precision-type attachments
    • Stewart BL, Edwards RO. Retention and wear of precision-type attachments. J Prosthet Dent 1983; 49: 28-34.
    • (1983) J Prosthet Dent , vol.49 , pp. 28-34
    • Stewart, B.L.1    Edwards, R.O.2
  • 24
    • 0033114158 scopus 로고    scopus 로고
    • In vitro study of the retention and mechanical fatigue behavior of four implant overdenture stud-type attachments
    • Fromentin O, Picard B, Tavernier B. In vitro study of the retention and mechanical fatigue behavior of four implant overdenture stud-type attachments. Pract Periodontics Aesthet Dent 1999; 11: 391-397.
    • (1999) Pract Periodontics Aesthet Dent , vol.11 , pp. 391-397
    • Fromentin, O.1    Picard, B.2    Tavernier, B.3
  • 25
    • 0031088933 scopus 로고    scopus 로고
    • Retention of ERA direct overdenture attachments before and after fatigue loading
    • Gamborena JI, Hazelton LR, NaBadalung D, Brudvik J. Retention of ERA direct overdenture attachments before and after fatigue loading. Int J Prosthodont 1997; 10: 123-130.
    • (1997) Int J Prosthodont , vol.10 , pp. 123-130
    • Gamborena, J.I.1    Hazelton, L.R.2    NaBadalung, D.3    Brudvik, J.4
  • 26
    • 0023455131 scopus 로고
    • The replacement of attachment-retained prostheses
    • Watkinson AC. The replacement of attachment-retained prostheses. Quintessence Int 1987; 18: 759-763.
    • (1987) Quintessence Int , vol.18 , pp. 759-763
    • Watkinson, A.C.1
  • 28
    • 5444257227 scopus 로고    scopus 로고
    • Retention force and fatigue strength of overdenture attachment systems
    • Botega DM, Mesquita MF, Henriques GE, Vaz LG. Retention force and fatigue strength of overdenture attachment systems. J Oral Rehabil 2004; 31: 884-889.
    • (2004) J Oral Rehabil , vol.31 , pp. 884-889
    • Botega, D.M.1    Mesquita, M.F.2    Henriques, G.E.3    Vaz, L.G.4
  • 29
    • 16644369487 scopus 로고    scopus 로고
    • Retention characteristics of attachment systems for implant overdentures
    • Chung KH, Chung CY, Cagna DR, Cronin RJ Jr. Retention characteristics of attachment systems for implant overdentures. J Prosthodont 2004; 13: 221-226.
    • (2004) J Prosthodont , vol.13 , pp. 221-226
    • Chung, K.H.1    Chung, C.Y.2    Cagna, D.R.3    Cronin Jr, R.J.4
  • 30
    • 42449151277 scopus 로고    scopus 로고
    • The influence of interimplant distance and attachment type on the retention characteristics of mandibular overdentures on 2 implants: 6-month fatigue retention values
    • Doukas D, Michelinakis G, Smith PW, Barclay CW. The influence of interimplant distance and attachment type on the retention characteristics of mandibular overdentures on 2 implants: 6-month fatigue retention values. Int J Prosthodont 2008; 21: 152-154.
    • (2008) Int J Prosthodont , vol.21 , pp. 152-154
    • Doukas, D.1    Michelinakis, G.2    Smith, P.W.3    Barclay, C.W.4
  • 31
    • 21244506681 scopus 로고    scopus 로고
    • Effect of implant angulation upon retention of overdenture attachments
    • Gulizio MP, Agar JR, Kelly JR, Taylor TD. Effect of implant angulation upon retention of overdenture attachments. J Prosthodont 2005; 14: 3-11.
    • (2005) J Prosthodont , vol.14 , pp. 3-11
    • Gulizio, M.P.1    Agar, J.R.2    Kelly, J.R.3    Taylor, T.D.4
  • 32
    • 33749337198 scopus 로고    scopus 로고
    • The influence of interimplant distance and attachment type on the retention characteristics of mandibular overdentures on 2 implants: Initial retention values
    • Michelinakis G, Barclay CW, Smith PW. The influence of interimplant distance and attachment type on the retention characteristics of mandibular overdentures on 2 implants: Initial retention values. Int J Prosthodont 2006; 19: 507-512.
    • (2006) Int J Prosthodont , vol.19 , pp. 507-512
    • Michelinakis, G.1    Barclay, C.W.2    Smith, P.W.3
  • 33
    • 0036634487 scopus 로고    scopus 로고
    • Maximum dislodging forces of implant overdenture stud attachments
    • Petropoulos VC, Smith W. Maximum dislodging forces of implant overdenture stud attachments. Int J Oral Maxillofac Implants 2002; 17: 526-535.
    • (2002) Int J Oral Maxillofac Implants , vol.17 , pp. 526-535
    • Petropoulos, V.C.1    Smith, W.2
  • 34
    • 0031086172 scopus 로고    scopus 로고
    • Comparison of retention and release periods for implant overdenture attachments
    • Petropoulos VC, Smith W, Kousvelari E. Comparison of retention and release periods for implant overdenture attachments. Int J Oral Maxillofac Implants 1997; 12: 176-185.
    • (1997) Int J Oral Maxillofac Implants , vol.12 , pp. 176-185
    • Petropoulos, V.C.1    Smith, W.2    Kousvelari, E.3
  • 35
    • 33845748140 scopus 로고    scopus 로고
    • Influence of attachment wear on retention of mandibular overdenture
    • Rutkunas V, Mizutani H, Takahashi H. Influence of attachment wear on retention of mandibular overdenture. J Oral Rehabil 2007; 34: 41-51.
    • (2007) J Oral Rehabil , vol.34 , pp. 41-51
    • Rutkunas, V.1    Mizutani, H.2    Takahashi, H.3
  • 36
    • 0032158735 scopus 로고    scopus 로고
    • Retention of prefabricated attachments for implant stabilized overdentures in the edentulous mandible: An in vitro study
    • Setz I, Lee SH, Engel E. Retention of prefabricated attachments for implant stabilized overdentures in the edentulous mandible: An in vitro study. J Prosthet Dent 1998; 80: 323-329.
    • (1998) J Prosthet Dent , vol.80 , pp. 323-329
    • Setz, I.1    Lee, S.H.2    Engel, E.3
  • 37
    • 2442606259 scopus 로고    scopus 로고
    • Comparative study of retentive anchor systems for overdentures
    • Svetlize CA, Bodereau EF Jr. Comparative study of retentive anchor systems for overdentures. Quintessence Int 2004; 35: 443-448.
    • (2004) Quintessence Int , vol.35 , pp. 443-448
    • Svetlize, C.A.1    Bodereau Jr, E.F.2
  • 38
    • 0033182770 scopus 로고    scopus 로고
    • Wear behavior of precision attachments
    • Wichmann MG, Kuntze W. Wear behavior of precision attachments. Int J Prosthodont 1999; 12: 409-414.
    • (1999) Int J Prosthodont , vol.12 , pp. 409-414
    • Wichmann, M.G.1    Kuntze, W.2
  • 39
    • 0031217052 scopus 로고    scopus 로고
    • The reliability of implant-retained hinging overdentures for the fully edentulous mandibl. An up to 9-year longitudinal study
    • Naert IE, Hooghe M, Quirynen M, van Steenberghe D. The reliability of implant-retained hinging overdentures for the fully edentulous mandibl. An up to 9-year longitudinal study. Clin Oral Investig 1997; 1: 119-124.
    • (1997) Clin Oral Investig , vol.1 , pp. 119-124
    • Naert, I.E.1    Hooghe, M.2    Quirynen, M.3    van Steenberghe, D.4
  • 40
    • 0036008952 scopus 로고    scopus 로고
    • The influence of various attachment types in mandibular implant-retained overdentures on maximum bite force and EMG
    • van Kampen F, van der Bilt A, Cune MS, Bosman F. The influence of various attachment types in mandibular implant-retained overdentures on maximum bite force and EMG. J Dent Res 2002; 81: 170-173.
    • (2002) J Dent Res , vol.81 , pp. 170-173
    • van Kampen, F.1    van der Bilt, A.2    Cune, M.S.3    Bosman, F.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.