Clinical evaluation of the new BMU 40 in-line blood analysis monitor

Author keywords

blood parameter; cardiopulmonary bypass; continuous in line blood parameter monitoring system; laboratory analyser

Indexed keywords

HEMOGLOBIN;

EID: 75649099060     PISSN: 02676591     EISSN: None     Source Type: Journal    
DOI: 10.1177/0267659109350935     Document Type: Article
Times cited : (3)

References (29)
  • 1
    • 0025056994 scopus 로고
    • Continuous blood gas monitoring during cardiopulmonary bypass - How soon will it be the standard of care?
    • Rubsamen DS Continuous blood gas monitoring during cardiopulmonary bypass - how soon will it be the standard of care? J Cardiothorac Anesth 1990 ; 4: 1-4
    • (1990) J Cardiothorac Anesth , vol.4 , pp. 1-4
    • Rubsamen, D.S.1
  • 2
    • 0029858998 scopus 로고    scopus 로고
    • Adverse effects of hyperoxemia during cardiopulmonary bypass
    • Joachimson PO, Sjoberg F., Forsman M., et al. Adverse effects of hyperoxemia during cardiopulmonary bypass. J Thorac Cardiovasc Surg 1996 ; 112: 812-819
    • (1996) J Thorac Cardiovasc Surg , vol.112 , pp. 812-819
    • Joachimson, P.O.1    Sjoberg, F.2    Forsman, M.3
  • 3
    • 0026343648 scopus 로고
    • The effect of hyperoxia during cardiopulmonary bypass on blood cell rheology and post-operative morbidity associated with cardiac surgery
    • Belboul A., Al-khaja N., Ericson C., et al. The effect of hyperoxia during cardiopulmonary bypass on blood cell rheology and post-operative morbidity associated with cardiac surgery. J Extra Corpor Technol 1991 ; 23: 43-48
    • (1991) J Extra Corpor Technol , vol.23 , pp. 43-48
    • Belboul, A.1    Al-Khaja, N.2    Ericson, C.3
  • 4
    • 0026077657 scopus 로고
    • Continuous arterial and venous blood gas monitoring during cardiopulmonary bypass
    • Mark JB, FitzGerald D., Fenton T., et al. Continuous arterial and venous blood gas monitoring during cardiopulmonary bypass. J Thorac Cardiovasc Surg 1991 ; 102: 431-439
    • (1991) J Thorac Cardiovasc Surg , vol.102 , pp. 431-439
    • Mark, J.B.1    Fitzgerald, D.2    Fenton, T.3
  • 5
    • 0032729775 scopus 로고    scopus 로고
    • Clinical evaluation of the CDI-100 in-line hematocrit/saturation monitor
    • Walton HG, Boucher D., Linne D. Clinical evaluation of the CDI-100 in-line hematocrit/saturation monitor. J Extra Corpor Technol 1999 ; 31: 80-83
    • (1999) J Extra Corpor Technol , vol.31 , pp. 80-83
    • Walton, H.G.1    Boucher, D.2    Linne, D.3
  • 6
    • 0033762621 scopus 로고    scopus 로고
    • The effects of continuous blood gas monitoring during cardiopulmonary bypass: A prospective, randomized study-part 2
    • Trowbridge CC, Vasquez M., Stammers AH, et al. The effects of continuous blood gas monitoring during cardiopulmonary bypass: A prospective, randomized study-part 2. J Extra Corpor Technol 2000 ; 32: 129-137
    • (2000) J Extra Corpor Technol , vol.32 , pp. 129-137
    • Trowbridge, C.C.1    Vasquez, M.2    Stammers, A.H.3
  • 7
    • 0027172643 scopus 로고
    • Comparison of blood gas and electrolyte test results from the Gem-Stat and CDI-300 versus a conventional laboratory analyzer
    • Walton HG, Duvall TB, Van den Hout J., et al. Comparison of blood gas and electrolyte test results from the Gem-Stat and CDI-300 versus a conventional laboratory analyzer. J Extra Corpor Technol 1993 ; 25: 53-57
    • (1993) J Extra Corpor Technol , vol.25 , pp. 53-57
    • Walton, H.G.1    Duvall, T.B.2    Van Den Hout, J.3
  • 8
    • 0034060856 scopus 로고    scopus 로고
    • CDI blood parameter monitoring system 500 - A new tool for the clinical perfusionist
    • Fried DW, Leo JJ, Mattioni GJ, et al. CDI blood parameter monitoring system 500 - a new tool for the clinical perfusionist. J Extra Corpor Technol 2000 ; 32: 25-30
    • (2000) J Extra Corpor Technol , vol.32 , pp. 25-30
    • Fried, D.W.1    Leo, J.J.2    Mattioni, G.J.3
  • 9
    • 0032453944 scopus 로고    scopus 로고
    • Clinical evaluation of a new in-line continuous blood gas monitor
    • Southworth R., Sutton R., Mize S., et al. Clinical evaluation of a new in-line continuous blood gas monitor. J Extra Corpor Technol 1998 ; 30: 166-170
    • (1998) J Extra Corpor Technol , vol.30 , pp. 166-170
    • Southworth, R.1    Sutton, R.2    Mize, S.3
  • 10
    • 0033758493 scopus 로고    scopus 로고
    • The effects of continuous blood gas monitoring during cardiopulmonary bypass: A prospective, randomized study-Part 1
    • Trowbridge CC, Vasquez M., Stammers AH, et al. The effects of continuous blood gas monitoring during cardiopulmonary bypass: a prospective, randomized study-Part 1. J Extra Corpor Technol 2000 ; 32: 120-128
    • (2000) J Extra Corpor Technol , vol.32 , pp. 120-128
    • Trowbridge, C.C.1    Vasquez, M.2    Stammers, A.H.3
  • 11
    • 33645906535 scopus 로고    scopus 로고
    • Use of the CDI blood parameter monitoring system 500 for continuous blood gas measurement during extracorporeal membrane oxygenation simulation
    • Schreur A., Niles S., Ploessl J. Use of the CDI blood parameter monitoring system 500 for continuous blood gas measurement during extracorporeal membrane oxygenation simulation. J Extra Corpor Technol 2005 ; 37: 377-380
    • (2005) J Extra Corpor Technol , vol.37 , pp. 377-380
    • Schreur, A.1    Niles, S.2    Ploessl, J.3
  • 12
    • 0028299124 scopus 로고
    • Clinical evaluation of the CDI® system 400 blood gas monitor
    • Svenmarker S., Lindholm R., HCurrency signggmark S., et al. Clinical evaluation of the CDI System 400 blood gas monitor. Perfusion 1994 ; 9: 71-76 (Pubitemid 24094080)
    • (1994) Perfusion , vol.9 , Issue.1 , pp. 71-76
    • Svenmarker, S.1    Lindholm, R.2    Haggmark, S.3    Jansson, E.4    Benze, S.5
  • 13
    • 0030323015 scopus 로고    scopus 로고
    • Comparison between CDI-100 continuous SO2, Hb and Hct monitoring, intermittent ABL-4 saturation and Hb monitoring, and lab Hb and Hct monitoring
    • Nauwynck M., Driessen J., Mulier J. Comparison between CDI-100 continuous SO2, Hb and Hct monitoring, intermittent ABL-4 saturation and Hb monitoring, and lab Hb and Hct monitoring. Acta Anaesthesiologica Belgica 1996 ; 47: 199-206
    • (1996) Acta Anaesthesiologica Belgica , vol.47 , pp. 199-206
    • Nauwynck, M.1    Driessen, J.2    Mulier, J.3
  • 14
    • 0029130698 scopus 로고
    • Accuracy of in-line venous saturation and hematocrit monitors in pediatric perfusion
    • Yaskulka SM, Burnside J., Bennett D., et al. Accuracy of in-line venous saturation and hematocrit monitors in pediatric perfusion. J Extra Corpor Technol 1995 ; 27: 132-136
    • (1995) J Extra Corpor Technol , vol.27 , pp. 132-136
    • Yaskulka, S.M.1    Burnside, J.2    Bennett, D.3
  • 16
    • 0346334617 scopus 로고    scopus 로고
    • Experiences with continuous intra-arterial blood gas monitoring: Precision and drift of a pure optode-system
    • Menzel M., Soukub J., Henze D., et al. Experiences with continuous intra-arterial blood gas monitoring: precision and drift of a pure optode-system. Intensive Care Med 2003 ; 29: 2180-2186
    • (2003) Intensive Care Med , vol.29 , pp. 2180-2186
    • Menzel, M.1    Soukub, J.2    Henze, D.3
  • 17
    • 0035116012 scopus 로고    scopus 로고
    • Accuracy and utility of a continuous intra-arterial blood gas monitoring system in pediatric patients
    • Coule LW, Truemper EJ, Steinhart CM, et al. Accuracy and utility of a continuous intra-arterial blood gas monitoring system in pediatric patients. Crit Care Med 2001 ; 29: 420-426
    • (2001) Crit Care Med , vol.29 , pp. 420-426
    • Coule, L.W.1    Truemper, E.J.2    Steinhart, C.M.3
  • 18
    • 0023676782 scopus 로고
    • In vitro assessment of a flow-through fluometric blood gas monitor
    • Pino JA, Basheim G., Kenny MA In vitro assessment of a flow-through fluometric blood gas monitor. J Clin Monit 1988 ; 4: 186-194
    • (1988) J Clin Monit , vol.4 , pp. 186-194
    • Pino, J.A.1    Basheim, G.2    Kenny, M.A.3
  • 19
    • 0026572422 scopus 로고
    • Con: Continuous blood gas monitoring should not be a standard during cardiopulmonary bypass
    • Merin RG Con: Continuous blood gas monitoring should not be a standard during cardiopulmonary bypass. J Cardiothorac Vasc Anesth 1992 ; 6: 109-110
    • (1992) J Cardiothorac Vasc Anesth , vol.6 , pp. 109-110
    • Merin, R.G.1
  • 20
    • 14244267544 scopus 로고    scopus 로고
    • Significance of continuous blood gas monitoring in cardiac surgery with cardiopulmonary bypass
    • Musat A., Quadirhi Y., Marty JC, Benkhadran S., David M., Girard C. Significance of continuous blood gas monitoring in cardiac surgery with cardiopulmonary bypass. Eur J Anaesthesiol 2004 ; 21: 980-988
    • (2004) Eur J Anaesthesiol , vol.21 , pp. 980-988
    • Musat, A.1    Quadirhi, Y.2    Marty, J.C.3    Benkhadran, S.4    David, M.5    Girard, C.6
  • 21
    • 0025139753 scopus 로고
    • Continuous venous oximetry during cardiopulmonary bypass: Influence of temperature changes, perfusion flow, and hematocrit levels
    • Baraka A., Baroody M., Haroun S., et al. Continuous venous oximetry during cardiopulmonary bypass: influence of temperature changes, perfusion flow, and hematocrit levels. J Cardiothorac Anesth 1990 ; 4: 35-38
    • (1990) J Cardiothorac Anesth , vol.4 , pp. 35-38
    • Baraka, A.1    Baroody, M.2    Haroun, S.3
  • 22
    • 84995176769 scopus 로고
    • Preliminary evaluation of a new continuous intra-arterial blood gas monitoring device
    • Pappert D., Rossaint R., Lewandowski K., et al. Preliminary evaluation of a new continuous intra-arterial blood gas monitoring device. Acta Anaesthesiologica Scandinavia 1995 ; 39 :(Supplementum 107) 67-70
    • (1995) Acta Anaesthesiologica Scandinavia , vol.39 , Issue.107 , pp. 67-70
    • Pappert, D.1    Rossaint, R.2    Lewandowski, K.3
  • 23
    • 0027973811 scopus 로고
    • Multisite evaluation of a continuous intraarterial blood gas monitoring system
    • Larson Jr. CP, Vender J., Seiver A. Multisite evaluation of a continuous intraarterial blood gas monitoring system. Anesthesiology 1994 ; 81: 543-552
    • (1994) Anesthesiology , vol.81 , pp. 543-552
    • Larson Jr., C.P.1    Vender, J.2    Seiver, A.3
  • 24
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • Bland JM, Altman DG Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986 ; 1: 307-310
    • (1986) Lancet , vol.1 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 25
    • 0037339866 scopus 로고    scopus 로고
    • Comparison of blood gas and electrolyte test results from the Gem-Premier and the ABL-70 versus a conventional laboratory analyzer
    • Walton HG, Boucher DM, Marroquin R. Comparison of blood gas and electrolyte test results from the Gem-Premier and the ABL-70 versus a conventional laboratory analyzer. J Extra Corpor Technol 2003 ; 35: 24-27
    • (2003) J Extra Corpor Technol , vol.35 , pp. 24-27
    • Walton, H.G.1    Boucher, D.M.2    Marroquin, R.3
  • 26
    • 33144473663 scopus 로고    scopus 로고
    • Reliability of point-of-care hematocrit, blood gas, electrolyte, lactate and glucose measurement during cardiopulmonary bypass
    • Steinfelder-Visscher J., Weerwind PW, Teerenstra S., et al. Reliability of point-of-care hematocrit, blood gas, electrolyte, lactate and glucose measurement during cardiopulmonary bypass. Perfusion 2006 ; 21: 33-37
    • (2006) Perfusion , vol.21 , pp. 33-37
    • Steinfelder-Visscher, J.1    Weerwind, P.W.2    Teerenstra, S.3
  • 27
    • 0026445856 scopus 로고
    • Standards of care in perfusion: Should not continuous in-line blood gas monitoring be one?
    • Ferries LH Standards of care in perfusion: should not continuous in-line blood gas monitoring be one? J Extra Corpor Technol 1992 ; 24: 45-48
    • (1992) J Extra Corpor Technol , vol.24 , pp. 45-48
    • Ferries, L.H.1
  • 28
    • 33749073889 scopus 로고    scopus 로고
    • Clinical evaluation of the ABL-77 for point-of-care analysis in the cardiovascular operating room
    • Prichard JS, French JS, Alvar N. Clinical evaluation of the ABL-77 for point-of-care analysis in the cardiovascular operating room. J Extra Corpor Technol 2006 ; 38: 128-133
    • (2006) J Extra Corpor Technol , vol.38 , pp. 128-133
    • Prichard, J.S.1    French, J.S.2    Alvar, N.3
  • 29
    • 0031974156 scopus 로고    scopus 로고
    • Monitoring controversies during cardiopulmonary bypass: How far have we come?
    • Stammers AH Monitoring controversies during cardiopulmonary bypass: how far have we come? Perfusion 1998 ; 13: 35-43
    • (1998) Perfusion , vol.13 , pp. 35-43
    • Stammers, A.H.1


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.