Volumn 36, Issue 11, 2009, Pages 4881-4889

The effect of angular and longitudinal tube current modulations on the estimation of organ and effective doses in x-ray computed tomography

Author keywords

Computed tomography; Monte Carlo; Radiation dosimetry; Tube current modulation; Voxelized phantom

Indexed keywords

DOSIMETRY; MODULATION; MONTE CARLO METHODS; PHANTOMS; SCANNING; TUBES (COMPONENTS);

EID: 70350722097     PISSN: 00942405     EISSN: None     Source Type: Journal    
DOI: 10.1118/1.3231948     Document Type: Article
Times cited : (58)

References (28)
  • 2
    • 0034118614 scopus 로고    scopus 로고
    • Calculation of effective dose
    • DOI 10.1118/1.598948
    • C. H. McCollough and B. A. Schueler, " Calculation of effective dose.," Med. Phys. 0094-2405 27, 828-837 (2000). 10.1118/1.598948 (Pubitemid 30349990)
    • (2000) Medical Physics , vol.27 , Issue.5 , pp. 828-837
    • McCollough, C.H.1    Schueler, B.A.2
  • 3
    • 40949104442 scopus 로고    scopus 로고
    • P103: The 2007 Recommendations of the International Commission on Radiological Protection
    • International Commission on Radiological Protection, "," 0146-6453.
    • International Commission on Radiological Protection, " P103: The 2007 Recommendations of the International Commission on Radiological Protection.," Ann. ICRP 0146-6453 37, 1-332 (2007).
    • (2007) Ann. ICRP , vol.37 , pp. 1-332
  • 4
    • 51549088875 scopus 로고    scopus 로고
    • Converting dose-length product to effective dose at CT
    • 0033-8419,. 10.1148/radiol.2483071964
    • W. Huda, K. M. Ogden, and M. R. Khorasani, " Converting dose-length product to effective dose at CT.," Radiology 0033-8419 248, 995-1003 (2008). 10.1148/radiol.2483071964
    • (2008) Radiology , vol.248 , pp. 995-1003
    • Huda, W.1    Ogden, K.M.2    Khorasani, M.R.3
  • 5
    • 0032744531 scopus 로고    scopus 로고
    • Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studies
    • 0094-2405,. 10.1118/1.598779
    • M. Gies, W. A. Kalender, H. Wolf, and C. Suess, " Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studies.," Med. Phys. 0094-2405 26, 2235-2247 (1999). 10.1118/1.598779
    • (1999) Med. Phys. , vol.26 , pp. 2235-2247
    • Gies, M.1    Kalender, W.A.2    Wolf, H.3    Suess, C.4
  • 6
    • 0032742081 scopus 로고    scopus 로고
    • Dose reduction in CT by anatomically adapted tube current modulation. II. Phantom measurements
    • 0094-2405,. 10.1118/1.598738
    • W. A. Kalender, H. Wolf, and C. Suess, " Dose reduction in CT by anatomically adapted tube current modulation. II. Phantom measurements.," Med. Phys. 0094-2405 26, 2248-2253 (1999). 10.1118/1.598738
    • (1999) Med. Phys. , vol.26 , pp. 2248-2253
    • Kalender, W.A.1    Wolf, H.2    Suess, C.3
  • 8
    • 85027223072 scopus 로고    scopus 로고
    • CT scanner automatic exposure control systems, London.
    • N. Keat, CT scanner automatic exposure control systems, London, 2005.
    • (2005)
    • Keat, N.1
  • 9
    • 4644266911 scopus 로고    scopus 로고
    • New x-ray tube performance in computed tomography by introducing the rotating envelope tube technology
    • DOI 10.1118/1.1783552
    • P. Schardt, J. Deuringer, J. Freudenberger, E. Hell, W. Knupfer, D. Mattern, and M. Schild, " New x-ray tube performance in computed tomography by introducing the rotating envelope tube technology.," Med. Phys. 0094-2405 31, 2699-2706 (2004). 10.1118/1.1783552 (Pubitemid 39287887)
    • (2004) Medical Physics , vol.31 , Issue.9 , pp. 2699-2706
    • Schardt, P.1    Deuringer, J.2    Freudenberger, J.3    Hell, E.4    Knupfer, W.5    Mattern, D.6    Schild, M.7
  • 10
    • 0017653997 scopus 로고
    • Dosimeter placement in the Rando phantom
    • DOI 10.1118/1.594320
    • B. R. Archer, S. Glaze, L. B. North, and S. C. Bushong, " Dosimeter placement in the Rando phantom.," Med. Phys. 0094-2405 4, 315-318 (1977). 10.1118/1.594320 (Pubitemid 8147902)
    • (1977) Medical Physics , vol.4 , Issue.4 , pp. 315-318
    • Archer, B.R.1    Glaze, S.2    North, L.B.3    Bushong, S.C.4
  • 11
    • 0032604360 scopus 로고    scopus 로고
    • A PC program for estimating organ dose and effective dose values in computed tomography
    • 0938-7994,. 10.1007/s003300050709
    • W. A. Kalender, B. Schmidt, M. Zankl, and M. Schmidt, " A PC program for estimating organ dose and effective dose values in computed tomography.," Eur. Radiol. 0938-7994 9, 555-562 (1999). 10.1007/s003300050709
    • (1999) Eur. Radiol. , vol.9 , pp. 555-562
    • Kalender, W.A.1    Schmidt, B.2    Zankl, M.3    Schmidt, M.4
  • 12
    • 41349095522 scopus 로고    scopus 로고
    • Validation of a Monte Carlo tool for patient-specific dose simulations in multi-slice computed tomography
    • 0938-7994,. 10.1007/s00330-007-0815-7
    • P. Deak, M. van Straten, P. C. Shrimpton, M. Zankl, and W. A. Kalender, " Validation of a Monte Carlo tool for patient-specific dose simulations in multi-slice computed tomography.," Eur. Radiol. 0938-7994 18, 759-772 (2008). 10.1007/s00330-007-0815-7
    • (2008) Eur. Radiol. , vol.18 , pp. 759-772
    • Deak, P.1    Van Straten, M.2    Shrimpton, P.C.3    Zankl, M.4    Kalender, W.A.5
  • 14
    • 0032767841 scopus 로고    scopus 로고
    • X-ray spectra estimation using attenuation measurements from 25 kVp to 18 MV
    • DOI 10.1118/1.598622
    • R. G. Waggener, M. M. Blough, J. A. Terry, D. Chen, N. E. Lee, S. Zhang, and W. D. McDavid, " X-ray spectra estimation using attenuation measurements from 25 kVp to 18 MV.," Med. Phys. 0094-2405 26, 1269-1278 (1999). 10.1118/1.598622 (Pubitemid 29334795)
    • (1999) Medical Physics , vol.26 , Issue.7 , pp. 1269-1278
    • Waggener, R.G.1    Blough, M.M.2    Terry, J.A.3    Chen, D.4    Lee, N.E.5    Zhang, S.6    McDavid, W.D.7
  • 15
    • 0027958158 scopus 로고
    • Computerized three-dimensional segmented human anatomy
    • DOI 10.1118/1.597290
    • I. G. Zubal, C. R. Harrell, E. O. Smith, Z. Rattner, G. Gindi, and P. B. Hoffer, " Computerized three-dimensional segmented human anatomy.," Med. Phys. 0094-2405 21, 299-302 (1994). 10.1118/1.597290 (Pubitemid 24068796)
    • (1994) Medical Physics , vol.21 , Issue.2 , pp. 299-302
    • Zubal, I.G.1    Harrell, C.R.2    Smith, E.O.3    Rattner, Z.4    Gindi, G.5    Hoffer, P.B.6
  • 16
    • 0027933729 scopus 로고
    • Automatic three-dimensional correlation of CT-CT, CT-MRI, and CT-SPECT using chamfer matching
    • DOI 10.1118/1.597344
    • M. van Herk and H. M. Kooy, " Automatic three-dimensional correlation of CT-CT, CT-MRI, and CT-SPECT using chamfer matching.," Med. Phys. 0094-2405 21, 1163-1178 (1994). 10.1118/1.597344 (Pubitemid 24233845)
    • (1994) Medical Physics , vol.21 , Issue.7 , pp. 1163-1178
    • Van Herk, M.1    Kooy, H.M.2
  • 17
    • 33846883307 scopus 로고    scopus 로고
    • Skeletal dosimetry in the MAX06 and the FAX06 phantoms for external exposure to photons based on vertebral 3D-microCT images
    • DOI 10.1088/0031-9155/51/24/001, PII S0031915506283076, 001
    • R. Kramer, H. J. Khoury, J. W. Vieira, and I. Kawrakow, " Skeletal dosimetry in the MAX06 and the FAX06 phantoms for external exposure to photons based on vertebral 3D-microCT images.," Phys. Med. Biol. 0031-9155 51, 6265-6289 (2006). 10.1088/0031-9155/51/24/001 (Pubitemid 46217622)
    • (2006) Physics in Medicine and Biology , vol.51 , Issue.24 , pp. 6265-6289
    • Kramer, R.1    Khoury, H.J.2    Vieira, J.W.3    Kawrakow, I.4
  • 19
    • 0025930020 scopus 로고
    • P60: The 1990 Recommendations of the International Commission on Radiological Protection
    • International Commission on Radiological Protection, "," 0146-6453.
    • International Commission on Radiological Protection, " P60: The 1990 Recommendations of the International Commission on Radiological Protection.," Ann. ICRP 0146-6453 21, 1-201 (1991).
    • (1991) Ann. ICRP , vol.21 , pp. 1-201
  • 20
    • 0032603548 scopus 로고    scopus 로고
    • Dose reduction in CT by on-line tube current control: Principles and validation on phantoms and cadavers
    • 0938-7994,. 10.1007/s003300050674
    • W. A. Kalender, H. Wolf, C. Suess, M. Gies, H. Greess, and W. A. Bautz, " Dose reduction in CT by on-line tube current control: Principles and validation on phantoms and cadavers.," Eur. Radiol. 0938-7994 9, 323-328 (1999). 10.1007/s003300050674
    • (1999) Eur. Radiol. , vol.9 , pp. 323-328
    • Kalender, W.A.1    Wolf, H.2    Suess, C.3    Gies, M.4    Greess, H.5    Bautz, W.A.6
  • 21
    • 34347385840 scopus 로고    scopus 로고
    • Angular on-line tube current modulation in multidetector CT examinations of children and adults: The influence of different scanning parameters on dose reduction
    • DOI 10.1118/1.2747048
    • A. E. Papadakis, K. Perisinakis, and J. Damilakis, " Angular on-line tube current modulation in multidetector CT examinations of children and adults: The influence of different scanning parameters on dose reduction.," Med. Phys. 0094-2405 34, 2864-2874 (2007). 10.1118/1.2747048 (Pubitemid 47019807)
    • (2007) Medical Physics , vol.34 , Issue.7 , pp. 2864-2874
    • Papadakis, A.E.1    Perisinakis, K.2    Damilakis, J.3
  • 23
    • 33751167428 scopus 로고    scopus 로고
    • On the need to revise the arm structure in stylized anthropomorphic phantoms in lateral photon irradiation geometry
    • DOI 10.1088/0031-9155/51/21/N03, PII S0031915506236285, N03
    • C. Lee and J. K. Lee, " On the need to revise the arm structure in stylized anthropomorphic phantoms in lateral photon irradiation geometry.," Phys. Med. Biol. 0031-9155 51, N393-N402 (2006). 10.1088/0031-9155/51/21/N03 (Pubitemid 44773320)
    • (2006) Physics in Medicine and Biology , vol.51 , Issue.21
    • Lee, C.1    Lee, C.2    Lee, J.-K.3
  • 24
    • 34249087431 scopus 로고    scopus 로고
    • Estimating radiation doses from multidetector CT using Monte Carlo simulations: Effects of different size voxelized patient models on magnitudes of organ and effective dose
    • DOI 10.1088/0031-9155/52/9/017, PII S0031915507337925, 017
    • J. J. DeMarco, C. H. Cagnon, D. D. Cody, D. M. Stevens, C. H. McCollough, M. Zankl, E. Angel, and M. F. McNitt-Gray, " Estimating radiation doses from multidetector CT using Monte Carlo simulations: Effects of different size voxelized patient models on magnitudes of organ and effective dose.," Phys. Med. Biol. 0031-9155 52, 2583-2597 (2007). 10.1088/0031-9155/52/9/017 (Pubitemid 46778528)
    • (2007) Physics in Medicine and Biology , vol.52 , Issue.9 , pp. 2583-2597
    • DeMarco, J.J.1    Cagnon, C.H.2    Cody, D.D.3    Stevens, D.M.4    McCollough, C.H.5    Zankl, M.6    Angel, E.7    McNitt-Gray, M.F.8
  • 25
    • 0036142799 scopus 로고    scopus 로고
    • The GSF family of voxel phantoms
    • DOI 10.1088/0031-9155/47/1/307, PII S0031915502276113
    • N. Petoussi-Henss, M. Zanki, U. Fill, and D. Regulla, " The GSF family of voxel phantoms.," Phys. Med. Biol. 0031-9155 47, 89-106 (2002). 10.1088/0031-9155/47/1/307 (Pubitemid 34081083)
    • (2002) Physics in Medicine and Biology , vol.47 , Issue.1 , pp. 89-106
    • Petoussi-Henss, N.1    Zankl, M.2    Fill, U.3    Regulla, D.4
  • 26
    • 0037151130 scopus 로고    scopus 로고
    • Organ dose conversion coefficients for external photon irradiation of male and female voxel models
    • DOI 10.1088/0031-9155/47/14/301, PII S0031915502351832
    • M. Zankl, U. Fill, N. Petoussi-Henss, and D. Regulla, " Organ dose conversion coefficients for external photon irradiation of male and female voxel models.," Phys. Med. Biol. 0031-9155 47, 2367-2385 (2002). 10.1088/0031-9155/47/14/301 (Pubitemid 34823317)
    • (2002) Physics in Medicine and Biology , vol.47 , Issue.14 , pp. 2367-2385
    • Zankl, M.1    Fill, U.2    Petoussi-Henss, N.3    Regulla, D.4
  • 27
    • 34247488445 scopus 로고    scopus 로고
    • Consideration of the ICRP 2006 revised tissue weighting factors on age-dependent values of the effective dose for external photons
    • DOI 10.1088/0031-9155/52/1/004, PII S003191550726124X, 004
    • C. Lee, E. Y. Han, and W. E. Bolch, " Consideration of the ICRP 2006 revised tissue weighting factors on age-dependent values of the effective dose for external photons.," Phys. Med. Biol. 0031-9155 52, 41-58 (2007). 10.1088/0031-9155/52/1/004 (Pubitemid 46646265)
    • (2007) Physics in Medicine and Biology , vol.52 , Issue.1 , pp. 41-58
    • Lee, C.1    Lee, C.2    Han, E.Y.3    Bolch, W.E.4
  • 28
    • 34147107446 scopus 로고    scopus 로고
    • 100
    • DOI 10.1118/1.2713240
    • J. M. Boone, " The trouble with CTD100.," Med. Phys. 0094-2405 34, 1364-1371 (2007). 10.1118/1.2713240 (Pubitemid 46555766)
    • (2007) Medical Physics , vol.34 , Issue.4 , pp. 1364-1371
    • Boone, J.M.1


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.