Taser X26 discharges in swine: Ventricular rhythm capture is dependent on discharge vector

Author keywords

Dysrhythmia; Echocardiography; Electromuscular incapacitation; Stun gun; Swine; Taser; Ventricular fibrillation

Indexed keywords

ANIMAL EXPERIMENT; ARTICLE; CONTROLLED STUDY; DEVICE; ECHOCARDIOGRAPHY; ELECTROCARDIOGRAM; ELECTROMUSCULAR INCAPACITATION DEVICE; HEART RHYTHM; HEART VENTRICLE CONTRACTION; HEART VENTRICLE FIBRILLATION; HEART VENTRICLE TACHYCARDIA; NONHUMAN; PRIORITY JOURNAL; SWINE; VECTORCARDIOGRAPHY; ANIMAL; ECHOGRAPHY; ELECTRIC INJURY; ELECTROCARDIOGRAPHY; HEART INJURY; HEART VENTRICLE; PATHOPHYSIOLOGY; SUDDEN DEATH; WEAPON;

EID: 67650641578     PISSN: 00225282     EISSN: 15298809     Source Type: Journal    
DOI: 10.1097/TA.0b013e31818bc17a     Document Type: Article
Times cited : (35)

References (52)
  • 1
    • 84869515265 scopus 로고    scopus 로고
    • United States of America Excessive and Lethal Force? Amnesty International's concerns about deaths and ill-treatment involving police use of tasers
    • Accessed October 2, 2007
    • Amnesty International. United States of America Excessive and Lethal Force? Amnesty International's concerns about deaths and ill-treatment involving police use of tasers. Available at: http: //www.amnestyusa.org/countries/usa/ Taser-report.pdf. Published 11-30-2004. Accessed October 2, 2007.
    • Available at: http
  • 2
    • 84869541478 scopus 로고    scopus 로고
    • Taser death update
    • Accessed October 2, 2007
    • Ward, C. Taser death update. Available at: http: //www.cameronward.com/ tasers/. Published 819322-2007. Accessed October 2, 2007.
    • Available at: http
    • Ward, C.1
  • 3
    • 24044523583 scopus 로고    scopus 로고
    • Withdrawal of taser electroshock devices: Too much, too soon
    • Bozeman WP. Withdrawal of taser electroshock devices: too much, too soon. Ann Emerg Med. 2005; 46: 300-301.
    • (2005) Ann Emerg Med , vol.46 , pp. 300-301
    • Bozeman, W.P.1
  • 4
    • 84869529729 scopus 로고    scopus 로고
    • Canada: Inappropriate and excessive use of tasers
    • Accessed January 3, 2008
    • Amnesty International. Canada: Inappropriate and excessive use of tasers. Available at: http: //www.amnesty.org/en/alfresco-asset/6c6342e0-a2bf-11dc- 8d74-6f45f39984e5/amr200022007en.pdf. Published 5-23-2007. Accessed January 3, 2008.
    • Available at: http
  • 5
    • 84869536837 scopus 로고    scopus 로고
    • TASER International Website. Available at:, Accessed March 30, 2008
    • TASER International Website. Available at: www.taser.com. Published 3-30-2008. Accessed March 30, 2008.
  • 6
    • 67650593491 scopus 로고    scopus 로고
    • Bleetman A, Steyn R. The advanced Taser: a medical review. Available at: http: //www.taser.com/research/Science/Documents/Bleetman%20TASER%20safety.pdf, 1-30. Published 4-27-2003. TASER International. Accessed January 3, 2008.
    • Bleetman A, Steyn R. The advanced Taser: a medical review. Available at: http: //www.taser.com/research/Science/Documents/Bleetman%20TASER%20safety.pdf, 1-30. Published 4-27-2003. TASER International. Accessed January 3, 2008.
  • 8
    • 1542605562 scopus 로고    scopus 로고
    • Introduction of the Taser into British policing. Implications for UK emergency departments: An overview of electronic weaponry
    • Bleetman A, Steyn R, Lee C. Introduction of the Taser into British policing. Implications for UK emergency departments: an overview of electronic weaponry. Emerg Med J. 2004; 21: 136-140.
    • (2004) Emerg Med J , vol.21 , pp. 136-140
    • Bleetman, A.1    Steyn, R.2    Lee, C.3
  • 9
    • 0024691736 scopus 로고
    • Tests on a shocking device-the stun gun
    • Roy OZ, Podgorski AS. Tests on a shocking device-the stun gun. Med Biol Eng Comput. 1989; 27: 445-448.
    • (1989) Med Biol Eng Comput , vol.27 , pp. 445-448
    • Roy, O.Z.1    Podgorski, A.S.2
  • 10
    • 56849109241 scopus 로고    scopus 로고
    • Forensic engineering analysis of electro-shock weapon safety
    • Ruggieri JA. Forensic engineering analysis of electro-shock weapon safety. J Natl Acad Foren Engin. 2005; 22: 1-34.
    • (2005) J Natl Acad Foren Engin , vol.22 , pp. 1-34
    • Ruggieri, J.A.1
  • 11
    • 33847730873 scopus 로고    scopus 로고
    • Electromuscular incapacitating devices
    • Webster JG. Electromuscular incapacitating devices. IFBME Proceedings. 2005; 9: 150-151.
    • (2005) IFBME Proceedings , vol.9 , pp. 150-151
    • Webster, J.G.1
  • 12
    • 67650585745 scopus 로고    scopus 로고
    • Loy E. Taser energy beyond specs, jury hears. canoe network. Available at: http: //vancouver.24hrs.ca/News/2007/05/25/pf-4206571.html. Published May 25, 2007.
    • Loy E. Taser energy beyond specs, jury hears. canoe network. Available at: http: //vancouver.24hrs.ca/News/2007/05/25/pf-4206571.html. Published May 25, 2007.
  • 13
    • 84869545139 scopus 로고    scopus 로고
    • TASER ECD distribution
    • Published September 13
    • Panescu D. TASER ECD distribution. Available at: http: //www.taser.com/research/Science/Pages/TASERECDDistrubition.aspx. Published September 13, 2007.
    • (2007) Available at: http
    • Panescu, D.1
  • 14
    • 36949009654 scopus 로고    scopus 로고
    • Electromagnetic modelling of current flow in the heart from TASER devices and the risk of cardiac dysrhythmias
    • Holden SJ, Sheridan RD, Coffey TJ, Scaramuzza RA, Diamantopoulos P. Electromagnetic modelling of current flow in the heart from TASER devices and the risk of cardiac dysrhythmias. Phys Med Biol. 2007; 52: 7193-7209.
    • (2007) Phys Med Biol , vol.52 , pp. 7193-7209
    • Holden, S.J.1    Sheridan, R.D.2    Coffey, T.J.3    Scaramuzza, R.A.4    Diamantopoulos, P.5
  • 15
    • 34548261485 scopus 로고    scopus 로고
    • Can the direct cardiac effects of the electric pulses generated by the TASER X26 cause immediate or delayed sudden cardiac arrest in normal adults?
    • Ideker RE, Dosdall DJ. Can the direct cardiac effects of the electric pulses generated by the TASER X26 cause immediate or delayed sudden cardiac arrest in normal adults? Am J Forensic Med Pathol. 2007; 28: 195-201.
    • (2007) Am J Forensic Med Pathol , vol.28 , pp. 195-201
    • Ideker, R.E.1    Dosdall, D.J.2
  • 16
    • 33746828597 scopus 로고    scopus 로고
    • Cardiac electrophysiological consequences of neuromuscular incapacitating device discharges
    • Nanthakumar K, Billingsley IM, Masse S, et al. Cardiac electrophysiological consequences of neuromuscular incapacitating device discharges. J Am Coll Cardiol. 2006; 48: 798-804.
    • (2006) J Am Coll Cardiol , vol.48 , pp. 798-804
    • Nanthakumar, K.1    Billingsley, I.M.2    Masse, S.3
  • 17
    • 38449117653 scopus 로고    scopus 로고
    • Acute effects of TASER- X26 discharges in a swine model
    • Dennis AJ, Valentino DJ, Walter RJ, et al. Acute effects of TASER- X26 discharges in a swine model. J Trauma. 2007; 63: 581-590.
    • (2007) J Trauma , vol.63 , pp. 581-590
    • Dennis, A.J.1    Valentino, D.J.2    Walter, R.J.3
  • 18
    • 44449141140 scopus 로고    scopus 로고
    • TASER X26 discharges in swine produce potentially fatal ventricular arrhythmias
    • Walter RJ, Dennis AJ, Valentino DJ, et al. TASER X26 discharges in swine produce potentially fatal ventricular arrhythmias. Acad Emerg Med. 2008; 15: 66-73.
    • (2008) Acad Emerg Med , vol.15 , pp. 66-73
    • Walter, R.J.1    Dennis, A.J.2    Valentino, D.J.3
  • 19
    • 67650637595 scopus 로고    scopus 로고
    • Sun, H. Models of ventricular fibrillation probability and neuromuscular stimulation after TASER use in humans. Available at: http: //ecow.engr.wisc.edu/ cgi-bin/getbig/bme/762/webster/sunthesis-01-21-07.pdf. Published 5-1-2008. University of Wisconsin. Accessed May 1, 2008.
    • Sun, H. Models of ventricular fibrillation probability and neuromuscular stimulation after TASER use in humans. Available at: http: //ecow.engr.wisc.edu/ cgi-bin/getbig/bme/762/webster/sunthesis-01-21-07.pdf. Published 5-1-2008. University of Wisconsin. Accessed May 1, 2008.
  • 21
    • 0025847473 scopus 로고
    • Electronic weaponry-a question of safety
    • O'Brien DJ. Electronic weaponry-a question of safety. Ann Emerg Med. 1991; 20: 583-587.
    • (1991) Ann Emerg Med , vol.20 , pp. 583-587
    • O'Brien, D.J.1
  • 22
    • 0022270483 scopus 로고
    • The Taser weapon: A new emergency medicine problem
    • Koscove EM. The Taser weapon: a new emergency medicine problem. Ann Emerg Med. 1985; 14: 1205-1208.
    • (1985) Ann Emerg Med , vol.14 , pp. 1205-1208
    • Koscove, E.M.1
  • 23
    • 33746858514 scopus 로고    scopus 로고
    • Effects of cocaine intoxication on the threshold for stun gun induction of ventricular fibrillation
    • Lakkireddy D, Wallick D, Ryschon K, et al. Effects of cocaine intoxication on the threshold for stun gun induction of ventricular fibrillation. J Am Coll Cardiol. 2006; 48: 805-811.
    • (2006) J Am Coll Cardiol , vol.48 , pp. 805-811
    • Lakkireddy, D.1    Wallick, D.2    Ryschon, K.3
  • 24
    • 41549112938 scopus 로고    scopus 로고
    • Cardiac effects of electrical stun guns: Does position of barbs contact make a difference?
    • Lakkireddy D, Wallick D, Verma A, et al. Cardiac effects of electrical stun guns: does position of barbs contact make a difference? Pacing Clin Electrophysiol. 2008; 31: 398-408.
    • (2008) Pacing Clin Electrophysiol , vol.31 , pp. 398-408
    • Lakkireddy, D.1    Wallick, D.2    Verma, A.3
  • 25
    • 33646747618 scopus 로고    scopus 로고
    • Cardiovascular and physiologic effects of conducted electrical weapon discharge in resting adults
    • Ho JD, Miner JR, Lakkireddy D, Bultman LL, Heegaard WG. Cardiovascular and physiologic effects of conducted electrical weapon discharge in resting adults. Acad Emerg Med. 2006; 13: 589-595.
    • (2006) Acad Emerg Med , vol.13 , pp. 589-595
    • Ho, J.D.1    Miner, J.R.2    Lakkireddy, D.3    Bultman, L.L.4    Heegaard, W.G.5
  • 26
    • 35349018401 scopus 로고    scopus 로고
    • Physiological effects of a conducted electrical weapon on human subjects
    • Vilke GM, Sloane CM, Bouton KD, et al. Physiological effects of a conducted electrical weapon on human subjects. Ann Emerg Med. 2007; 50: 569-575.
    • (2007) Ann Emerg Med , vol.50 , pp. 569-575
    • Vilke, G.M.1    Sloane, C.M.2    Bouton, K.D.3
  • 28
    • 34248334997 scopus 로고    scopus 로고
    • Can TASERS directly cause ventricular fibrillation
    • Webster JG, Will JA, Sun H, et al. Can TASERS directly cause ventricular fibrillation. IFBME Proc. 2006; 14: 3307-3310.
    • (2006) IFBME Proc , vol.14 , pp. 3307-3310
    • Webster, J.G.1    Will, J.A.2    Sun, H.3
  • 29
    • 34447633733 scopus 로고    scopus 로고
    • Taser-induced rapid ventricular myocardial capture demonstrated by pacemaker intracardiac electrograms
    • Cao M, Shinbane JS, Gillberg JM, Saxon LA. Taser-induced rapid ventricular myocardial capture demonstrated by pacemaker intracardiac electrograms. J Cardiovasc Electrophysiol. 2007; 18: 876-879.
    • (2007) J Cardiovasc Electrophysiol , vol.18 , pp. 876-879
    • Cao, M.1    Shinbane, J.S.2    Gillberg, J.M.3    Saxon, L.A.4
  • 30
    • 34249007454 scopus 로고    scopus 로고
    • Duration of discharge of neuromuscular incapacitating device and inappropriate implantable cardioverter-defibrillator detections
    • Calton R, Cameron D, Masse S, Nanthakumar K. Duration of discharge of neuromuscular incapacitating device and inappropriate implantable cardioverter-defibrillator detections. Circulation. 2007; 115: 472-474.
    • (2007) Circulation , vol.115 , pp. 472-474
    • Calton, R.1    Cameron, D.2    Masse, S.3    Nanthakumar, K.4
  • 31
    • 34547881084 scopus 로고    scopus 로고
    • Do electrical stun guns (TASER-X26) affect the functional integrity of implantable pacemakers and defibrillators?
    • Lakkireddy D, Khasnis A, Antenacci J, et al. Do electrical stun guns (TASER-X26) affect the functional integrity of implantable pacemakers and defibrillators? Europace. 2007; 9: 551-556.
    • (2007) Europace , vol.9 , pp. 551-556
    • Lakkireddy, D.1    Khasnis, A.2    Antenacci, J.3
  • 32
    • 34248384420 scopus 로고    scopus 로고
    • Repeated thoracic discharges from a stun device
    • Valentino DJ, Walter RJ, Nagy K, et al. Repeated thoracic discharges from a stun device. J Trauma. 2007; 62: 1134-1142.
    • (2007) J Trauma , vol.62 , pp. 1134-1142
    • Valentino, D.J.1    Walter, R.J.2    Nagy, K.3
  • 33
    • 35248849404 scopus 로고    scopus 로고
    • Neuromuscular effects of stun device discharges
    • Valentino DJ, Walter RJ, Dennis AJ, et al. Neuromuscular effects of stun device discharges. J Surg Res. 2007; 143: 78-87.
    • (2007) J Surg Res , vol.143 , pp. 78-87
    • Valentino, D.J.1    Walter, R.J.2    Dennis, A.J.3
  • 34
    • 33746321350 scopus 로고    scopus 로고
    • Acidosis, lactate, electrolytes, muscle enzymes, and other factors in the blood of Sus scrofa following repeated TASER((R)) exposures
    • Sci Int
    • Jauchem JR, Sherry CJ, Fines DA, Cook MC. Acidosis, lactate, electrolytes, muscle enzymes, and other factors in the blood of Sus scrofa following repeated TASER((R)) exposures. Forensic Sci Int. 2006; 161: 20-30.
    • (2006) Forensic , vol.161 , pp. 20-30
    • Jauchem, J.R.1    Sherry, C.J.2    Fines, D.A.3    Cook, M.C.4
  • 36
    • 0031229448 scopus 로고    scopus 로고
    • Injury by electrical forces: Pathophysiology, manifestations, and therapy
    • Lee RC. Injury by electrical forces: pathophysiology, manifestations, and therapy. Curr Probl Surg. 1997; 34: 677-764.
    • (1997) Curr Probl Surg , vol.34 , pp. 677-764
    • Lee, R.C.1
  • 37
    • 0034575127 scopus 로고    scopus 로고
    • Biophysical injury mechanisms in electrical shock trauma
    • Lee RC, Zhang D, Hannig J. Biophysical injury mechanisms in electrical shock trauma. Annu Rev Biomed Eng. 2000; 2: 477-509.
    • (2000) Annu Rev Biomed Eng , vol.2 , pp. 477-509
    • Lee, R.C.1    Zhang, D.2    Hannig, J.3
  • 39
    • 33847660910 scopus 로고    scopus 로고
    • Comparative analysis of heart functions in micropigs and conventional pigs using echocardiography and radiography
    • Lee MY, Lee SH, Lee SG, et al. Comparative analysis of heart functions in micropigs and conventional pigs using echocardiography and radiography. J Vet Sci. 2007; 8: 7-14.
    • (2007) J Vet Sci , vol.8 , pp. 7-14
    • Lee, M.Y.1    Lee, S.H.2    Lee, S.G.3
  • 40
    • 0031705780 scopus 로고    scopus 로고
    • Anatomy of the pig heart: Comparisons with normal human cardiac structure
    • Crick SJ, Sheppard MN, Ho SY, Gebstein L, Anderson RH. Anatomy of the pig heart: comparisons with normal human cardiac structure. J Anat. 1998; 193(Pt 1): 105-119.
    • (1998) J Anat , vol.193 , Issue.PART 1 , pp. 105-119
    • Crick, S.J.1    Sheppard, M.N.2    Ho, S.Y.3    Gebstein, L.4    Anderson, R.H.5
  • 41
    • 84869526462 scopus 로고    scopus 로고
    • Stun gun fallacy: How the lack of Taser regulation endangers lives
    • Accessed July 14, 2007
    • Schlosberg M, Levin J, Batliwalla S, et al. Stun gun fallacy: How the lack of Taser regulation endangers lives. Available at: http: //www.aclunc.org/issues/criminal-justice/police-practices/special-report-stun- gun-fallacy.shtml. Accessed July 14, 2007.
    • Available at: http
    • Schlosberg, M.1    Levin, J.2    Batliwalla, S.3
  • 42
    • 36849003466 scopus 로고    scopus 로고
    • Twelve-lead electrocardiogram monitoring of subjects before and after voluntary exposure to the Taser X26
    • Vilke GM, Sloane C, Levine S, Neuman T, Castillo E, Chan TC. Twelve-lead electrocardiogram monitoring of subjects before and after voluntary exposure to the Taser X26. Am J Emerg Med. 2008; 26: 1-4.
    • (2008) Am J Emerg Med , vol.26 , pp. 1-4
    • Vilke, G.M.1    Sloane, C.2    Levine, S.3    Neuman, T.4    Castillo, E.5    Chan, T.C.6
  • 43
    • 84869537215 scopus 로고    scopus 로고
    • Taser dart-to-heart distance that causes ventricular fibrillation in pigs
    • Published, Accessed October 2, 2007
    • Wu J-Y, Sun H, O'Rourke AP, Webster JG. Taser dart-to-heart distance that causes ventricular fibrillation in pigs. Available at: http: //www.engr.wisc.edu/bme/faculty/webster-john/EB2006Final.pdf. Published 2006. Accessed October 2, 2007.
    • (2006) Available at: http
    • Wu, J.-Y.1    Sun, H.2    O'Rourke, A.P.3    Webster, J.G.4
  • 44
    • 36048969874 scopus 로고    scopus 로고
    • Estimating neuromuscular stimulation within the human torso with Taser(R) stimulus
    • Sun H, Webster JG. Estimating neuromuscular stimulation within the human torso with Taser(R) stimulus. Phys Med Biol. 2007; 52: 6401-6411.
    • (2007) Phys Med Biol , vol.52 , pp. 6401-6411
    • Sun, H.1    Webster, J.G.2
  • 45
    • 67650581055 scopus 로고    scopus 로고
    • Mechanisms of electrostimulation: Application to electromagnetic field exposure standards at frequencies below 100 kHz
    • New York: Springer;
    • Reilly JP. Mechanisms of electrostimulation: application to electromagnetic field exposure standards at frequencies below 100 kHz. Applied Bioelectricity: From Electrical Stimulation to Electropathology. New York: Springer; 1998.
    • (1998) Applied Bioelectricity: From Electrical Stimulation to Electropathology
    • Reilly, J.P.1
  • 46
    • 67650619949 scopus 로고    scopus 로고
    • Reilly JP. Mechanisms of electrostimulation: application to electromagnetic field exposure standards at frequencies below 100 kHz. In: Chadwick P, Gabriel CH, eds. EMF Dosimetry Handbook. Texas: USAF, AFSC, Brooks AF; 2003. Available at: www.emfdosimetry.org.
    • Reilly JP. Mechanisms of electrostimulation: application to electromagnetic field exposure standards at frequencies below 100 kHz. In: Chadwick P, Gabriel CH, eds. EMF Dosimetry Handbook. Texas: USAF, AFSC, Brooks AF; 2003. Available at: www.emfdosimetry.org.
  • 47
    • 56849131220 scopus 로고    scopus 로고
    • Pharmacology
    • CA: Current Clinical Strategies Publ
    • Ezekiel MR. Pharmacology. Handbook of Anesthesiology. CA: Current Clinical Strategies Publ.; 2005: 68-71.
    • (2005) Handbook of Anesthesiology , pp. 68-71
    • Ezekiel, M.R.1
  • 48
    • 0025822718 scopus 로고
    • The conduction system of the swine heart
    • Bharati S, Levine M, Huang SK, et al. The conduction system of the swine heart. Chest. 1991; 100: 207-212.
    • (1991) Chest , vol.100 , pp. 207-212
    • Bharati, S.1    Levine, M.2    Huang, S.K.3
  • 49
    • 0030953698 scopus 로고    scopus 로고
    • Toward optimizing a preshaped catheter and system parameters to achieve single lead DDD pacing
    • Brownlee RR, Swindle MM, Bertolet R, Neff P. Toward optimizing a preshaped catheter and system parameters to achieve single lead DDD pacing. Pacing Clin Electrophysiol. 1997; 20: 1354-1358.
    • (1997) Pacing Clin Electrophysiol , vol.20 , pp. 1354-1358
    • Brownlee, R.R.1    Swindle, M.M.2    Bertolet, R.3    Neff, P.4
  • 50
    • 0030985815 scopus 로고    scopus 로고
    • A technique for conducting noninvasive cardiac electrophysiology studies in conscious swine
    • Smith AC, Knick B, Gillette PC, Swindle MM. A technique for conducting noninvasive cardiac electrophysiology studies in conscious swine. J Invest Surg. 1997; 10: 25-29.
    • (1997) J Invest Surg , vol.10 , pp. 25-29
    • Smith, A.C.1    Knick, B.2    Gillette, P.C.3    Swindle, M.M.4
  • 51
    • 0022851169 scopus 로고
    • Systemic and regional hemodynamic, antiarrhythmic and antiischemic effects of bevantolol in anesthetized pigs
    • Verdouw PD, Hartog JM, Saxena PR, Hugenholtz PG. Systemic and regional hemodynamic, antiarrhythmic and antiischemic effects of bevantolol in anesthetized pigs. Am J Cardiol. 1986; 58: 8E-16E.
    • (1986) Am J Cardiol , vol.58
    • Verdouw, P.D.1    Hartog, J.M.2    Saxena, P.R.3    Hugenholtz, P.G.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.