Volumn 93, Issue 4, 2004, Pages 532-539

Comparison of manufacturers' specifications for 44 types of heat and moisture exchanging filters

Author keywords

Anaesthesia; Equipment, heat and moisture exchanging filters; Equipment, specifications; Intensive care

Indexed keywords

ANESTHESIOLOGICAL TECHNIQUES; ARTICLE; ARTIFICIAL VENTILATION; BACILLUS SUBTILIS; CONTROLLED STUDY; ELECTRICITY; FILTER; FILTRATION; HEAT; HEPATITIS C VIRUS; INTERMETHOD COMPARISON; INTERNET; LABORATORY TEST; MECHANICS; MEDICAL DECISION MAKING; MEDICAL INFORMATION; MOISTURE; MYCOBACTERIUM TUBERCULOSIS; NONHUMAN; PERFORMANCE; PRIORITY JOURNAL; STANDARD; VALIDATION PROCESS; COMPARATIVE STUDY; CROSS INFECTION; EQUIPMENT DESIGN; HUMAN; HUMIDITY; INHALATION ANESTHESIA; INSTRUMENTATION;

EID: 5444271763     PISSN: 00070912     EISSN: None     Source Type: Journal    
DOI: 10.1093/bja/aeh239     Document Type: Article
Times cited : (19)

References (53)
  • 1
    • 5444231884 scopus 로고    scopus 로고
    • Blood borne viruses and anaesthesia: An update, January 1996
    • Association of Anaesthetists of Great Britain and Ireland. London Association of Anaesthetists of Great Britain and Ireland
    • Association of Anaesthetists of Great Britain and Ireland. Blood borne viruses and anaesthesia: an update, January 1996. London: Association of Anaesthetists of Great Britain and Ireland, 1996
    • (1996)
  • 2
    • 0003958923 scopus 로고    scopus 로고
    • Recommandations concernant l'hygiène en anesthésie
    • Société française d'anesthésie et de réanimation. Paris SFAR
    • Société française d'anesthésie et de réanimation. Recommandations concernant l'hygiène en anesthésie. Paris: SFAR, 1997
    • (1997)
  • 3
    • 0033010764 scopus 로고    scopus 로고
    • Humidification for patients with artificial airways
    • Branson RD. Humidification for patients with artificial airways. Respir Care 1999; 44: 630-41
    • (1999) Respir. Care , vol.44 , pp. 630-641
    • Branson, R.D.1
  • 4
    • 5444229739 scopus 로고    scopus 로고
    • Anaesthetic and respiratory equipment. Heat and moisture exchangers (HMEs) for humidifying respired gases in humans. Part I: HMEs for use with minimum tidal volumes of 250 mL (BS EN ISO 9360-1:2000)
    • British Standards Institution. Milton Keynes (UK): British Standards Institution
    • British Standards Institution. Anaesthetic and respiratory equipment. Heat and moisture exchangers (HMEs) for humidifying respired gases in humans. Part I: HMEs for use with minimum tidal volumes of 250 mL (BS EN ISO 9360-1:2000). Milton Keynes (UK): British Standards Institution, 2001
    • (2001)
  • 5
    • 18844408029 scopus 로고    scopus 로고
    • Breathing system filters for anaesthetic and respiratory use. Part I: Salt test method to assess filtration performance (BS EN 13328-1:2001)
    • British Standards Institution. Milton Keynes (UK): British Standards Institution
    • British Standards Institution. Breathing system filters for anaesthetic and respiratory use. Part I: Salt test method to assess filtration performance (BS EN 13328-1:2001). Milton Keynes (UK): British Standards Institution, 2001
    • (2001)
  • 7
    • 0027968422 scopus 로고
    • Heat and moisture exchangers and breathing filters
    • Hedley RM, Allt-Graham J. Heat and moisture exchangers and breathing filters. Br J Anaesth 1994; 73: 227-36
    • (1994) Br. J. Anaesth. , vol.73 , pp. 227-236
    • Hedley, R.M.1    Allt-Graham, J.2
  • 8
    • 0026907566 scopus 로고
    • Clinical practice guideline. Humidification during mechanical ventilation
    • American Association for Respiratory Care
    • American Association for Respiratory Care. Clinical practice guideline. Humidification during mechanical ventilation. Respir Care 1992; 37: 887-90
    • (1992) Respir. Care , vol.37 , pp. 887-890
  • 9
    • 0031920448 scopus 로고    scopus 로고
    • A novel method of evaluation of three heat-moisture exchangers in six different ventilator settings
    • Ünal N, Kanhai JKK, Buijk SLCE, et al. A novel method of evaluation of three heat-moisture exchangers in six different ventilator settings. Intensive Care Med 1998; 24: 138-46
    • (1998) Intensive Care Med. , vol.24 , pp. 138-146
    • Ünal, N.1    Kanhai, J.K.K.2    Buijk, S.L.C.E.3
  • 10
    • 0031897541 scopus 로고    scopus 로고
    • Experimental measurements of particle retention efficiency of filters used to prevent contamination in respiratory devices
    • Becquemin MH, Bouchikhi A, Croix N, Malarbet JL, Bertholon JF, Roy M. Experimental measurements of particle retention efficiency of filters used to prevent contamination in respiratory devices. Intensive Care Med 1998; 24: 81-5
    • (1998) Intensive Care Med. , vol.24 , pp. 81-85
    • Becquemin, M.H.1    Bouchikhi, A.2    Croix, N.3    Malarbet, J.L.4    Bertholon, J.F.5    Roy, M.6
  • 11
    • 0032086343 scopus 로고    scopus 로고
    • Heat and moisture exchangers. Structure and function
    • Branson RD, Peterson BD, Carson KD, eds. Humidification: current therapy and controversy
    • Wilkes AR. Heat and moisture exchangers. Structure and function. In: Branson RD, Peterson BD, Carson KD, eds. Humidification: current therapy and controversy. Respir Care Clin N Am 1998; 4 261-79
    • (1998) Respir. Care Clin. N. Am. , vol.4 , pp. 261-279
    • Wilkes, A.R.1
  • 12
    • 0029760658 scopus 로고    scopus 로고
    • Evaluation of 21 passive humidifiers according to the ISO 9360 standard: Moisture output, dead space, and flow resistance
    • Branson RD, Davis K Jr. Evaluation of 21 passive humidifiers according to the ISO 9360 standard: moisture output, dead space, and flow resistance. Respir Care 1996; 41: 736-43
    • (1996) Respir. Care , vol.41 , pp. 736-743
    • Branson, R.D.1    Davis Jr., K.2
  • 13
    • 84958294736 scopus 로고
    • An apparatus for the study of airborne infections
    • Henderson DW. An apparatus for the study of airborne infections. J Hyg (Lond) 1952; 50: 53-67
    • (1952) J. Hyg. (Lond) , vol.50 , pp. 53-67
    • Henderson, D.W.1
  • 14
    • 0014572961 scopus 로고
    • A mobile form of the Henderson apparatus
    • Druett HA. A mobile form of the Henderson apparatus. J Hyg (Lond) 1969; 67: 437-48
    • (1969) J. Hyg. (Lond) , vol.67 , pp. 437-448
    • Druett, H.A.1
  • 15
    • 0036168438 scopus 로고    scopus 로고
    • Measuring the filtration performance of breathing system filters using sodium chloride particles
    • Wilkes AR. Measuring the filtration performance of breathing system filters using sodium chloride particles. Anaesthesia 2002; 57: 162-8
    • (2002) Anaesthesia , vol.57 , pp. 162-168
    • Wilkes, A.R.1
  • 17
    • 0002046246 scopus 로고
    • Transmission of hepatitis C via anesthetic tubings?
    • [letter]
    • Ragg M. Transmission of hepatitis C via anesthetic tubings? [letter] Lancet 1994; 343: 1419
    • (1994) Lancet , vol.343 , pp. 1419
    • Ragg, M.1
  • 19
    • 0026026102 scopus 로고
    • The global tuberculosis situation and the new control strategy of the World Health Organization
    • Kochi A. The global tuberculosis situation and the new control strategy of the World Health Organization. Tubercle 1991; 72: 1-6
    • (1991) Tubercle , vol.72 , pp. 1-6
    • Kochi, A.1
  • 21
    • 5444261805 scopus 로고
    • Changing heat and moisture exchangers every 48 h rather than 24 h does not affect their efficacy and the incidence of nosocomial pneumonia
    • Diedaini K, Billiard M, Mier L, Le Bourdelles G, Brun P, Markowicz P. Changing heat and moisture exchangers every 48 h rather than 24 h does not affect their efficacy and the incidence of nosocomial pneumonia. Am Respir J Crit Care Med 1995; 152: 156-69
    • (1995) Am. Respir. J. Crit. Care Med. , vol.152 , pp. 156-169
    • Diedaini, K.1    Billiard, M.2    Mier, L.3    Le Bourdelles, G.4    Brun, P.5    Markowicz, P.6
  • 22
    • 0023267534 scopus 로고
    • Contamination control in long-term ventilation: A clinical study using a heat- and moisture-exchanging filter
    • Gallagher J, Strangeways JEM, Allt-Graham J. Contamination control in long-term ventilation: a clinical study using a heat- and moisture-exchanging filter. Anaesthesia 1987; 42: 476-81
    • (1987) Anaesthesia , vol.42 , pp. 476-481
    • Gallagher, J.1    Strangeways, J.E.M.2    Allt-Graham, J.3
  • 23
    • 0031936853 scopus 로고    scopus 로고
    • Efficacy of heat and moisture exchangers after changing every 48 hours rather than 24 hours
    • Thomachot L, Vialet R, Viguier JM, Sidier B, Roulier P, Martin C. Efficacy of heat and moisture exchangers after changing every 48 hours rather than 24 hours. Crit Care Med 1998; 26: 477-81
    • (1998) Crit. Care Med. , vol.26 , pp. 477-481
    • Thomachot, L.1    Vialet, R.2    Viguier, J.M.3    Sidier, B.4    Roulier, P.5    Martin, C.6
  • 24
    • 0034022538 scopus 로고    scopus 로고
    • Changing heat and moisture exchangers after 96 hours rather than after 24 hours: A clinical and microbiological evaluation
    • Thomachot L, Boisson C, Arnaud S, Michelet P, Cambon S, Martin C. Changing heat and moisture exchangers after 96 hours rather than after 24 hours: a clinical and microbiological evaluation. Crit Care Med 2000; 28: 714-20
    • (2000) Crit. Care Med. , vol.28 , pp. 714-720
    • Thomachot, L.1    Boisson, C.2    Arnaud, S.3    Michelet, P.4    Cambon, S.5    Martin, C.6
  • 25
    • 0034112596 scopus 로고    scopus 로고
    • Prolonged use of heat and moisture exchangers does not affect device efficiency or frequency rate of nosocomial pneumonia
    • Davis KJ, Evans S, Campbell R, et al. Prolonged use of heat and moisture exchangers does not affect device efficiency or frequency rate of nosocomial pneumonia. Crit Care Med 2000; 28 1412-8
    • (2000) Crit. Care Med. , vol.28 , pp. 1412-1418
    • Davis, K.J.1    Evans, S.2    Campbell, R.3
  • 26
    • 0031888824 scopus 로고    scopus 로고
    • A randomized clinical trial comparing an extended-use hygroscopic condenser humidifier with heated-water humidification in mechanically ventilated patients
    • Kollef MH, Shapiro SD, Boyd V, et al. A randomized clinical trial comparing an extended-use hygroscopic condenser humidifier with heated-water humidification in mechanically ventilated patients. Chest 1998; 113: 759-67
    • (1998) Chest , vol.113 , pp. 759-767
    • Kollef, M.H.1    Shapiro, S.D.2    Boyd, V.3
  • 27
    • 0027232943 scopus 로고
    • Detection of expiratory antibacterial filter occlusion
    • Barton RM. Detection of expiratory antibacterial filter occlusion. Anesth Analg 1993; 77: 197
    • (1993) Anesth. Analg. , vol.77 , pp. 197
    • Barton, R.M.1
  • 28
    • 0021241622 scopus 로고
    • Increase in resistance of in-line breathing filters in humidified air
    • Buckley PM. Increase in resistance of in-line breathing filters in humidified air. Br J Anaesth 1984; 56: 637-43
    • (1984) Br. J. Anaesth. , vol.56 , pp. 637-643
    • Buckley, P.M.1
  • 29
    • 0030669261 scopus 로고    scopus 로고
    • Adverse anesthetic outcomes arising from gas delivery equipment: A closed claims analysis
    • Caplan R, Vistica MF, Posner KL, Cheney FW. Adverse anesthetic outcomes arising from gas delivery equipment: a closed claims analysis. Anesthesiology 1997; 87: 741-8
    • (1997) Anesthesiology , vol.87 , pp. 741-748
    • Caplan, R.1    Vistica, M.F.2    Posner, K.L.3    Cheney, F.W.4
  • 30
    • 0024341954 scopus 로고
    • A dangerous defect in a heat and moisture exchanger
    • Prados W. A dangerous defect in a heat and moisture exchanger. Anesthesiology 1989; 71: 804
    • (1989) Anesthesiology , vol.71 , pp. 804
    • Prados, W.1
  • 31
    • 0025344012 scopus 로고
    • Complications related to the use of a heat and moisture exchanger
    • Prasad KK, Chen L. Complications related to the use of a heat and moisture exchanger. Anesthesiology 1990; 72: 958
    • (1990) Anesthesiology , vol.72 , pp. 958
    • Prasad, K.K.1    Chen, L.2
  • 33
    • 0036190746 scopus 로고    scopus 로고
    • Rapid and complete occlusion of a heat and moisture exchange filter by pulmonary oedema
    • Williams DJ, Stacey MRW. Rapid and complete occlusion of a heat and moisture exchange filter by pulmonary oedema. Can J Anaesth 2002; 49: 126-31
    • (2002) Can. J. Anaesth. , vol.49 , pp. 126-131
    • Williams, D.J.1    Stacey, M.R.W.2
  • 35
    • 0021833766 scopus 로고
    • Role of tracheal and bronchial circulation in respiratory heat exchange
    • Baile EM, Dahlby RW, Wiggs BR, Pare PD. Role of tracheal and bronchial circulation in respiratory heat exchange. J Appl Physiol 1985; 58: 217-22
    • (1985) J. Appl. Physiol. , vol.58 , pp. 217-222
    • Baile, E.M.1    Dahlby, R.W.2    Wiggs, B.R.3    Pare, P.D.4
  • 36
    • 0015396511 scopus 로고
    • Effects of dry anesthetic gases on tracheobronchial ciliated epithelium
    • Chalon J, Loew DAY, Malebranche J. Effects of dry anesthetic gases on tracheobronchial ciliated epithelium. Anesthesiology 1972; 37: 338-43
    • (1972) Anesthesiology , vol.37 , pp. 338-343
    • Chalon, J.1    Loew, D.A.Y.2    Malebranche, J.3
  • 37
    • 0018353283 scopus 로고
    • Humidity and the anesthetized patient
    • Chalon J, Patel C, Ali M, et al. Humidity and the anesthetized patient. Anesthesiology 1979; 50: 195-8
    • (1979) Anesthesiology , vol.50 , pp. 195-198
    • Chalon, J.1    Patel, C.2    Ali, M.3
  • 38
    • 0017752489 scopus 로고
    • Optimum humidification of air administered to a tracheostomy in dogs. Scanning electron microscopy and surfactant studies
    • Tsuda T, Noguchi H, Takumi Y, Aochi O. Optimum humidification of air administered to a tracheostomy in dogs. Scanning electron microscopy and surfactant studies. Br J Anaesth 1977; 39: 965-77
    • (1977) Br. J. Anaesth. , vol.39 , pp. 965-977
    • Tsuda, T.1    Noguchi, H.2    Takumi, Y.3    Aochi, O.4
  • 39
    • 0016702558 scopus 로고
    • Effect of dry and humidified gases on the respiratory epithelium in rabbits
    • Marfatia S, Donahoe PK, Hendren WH. Effect of dry and humidified gases on the respiratory epithelium in rabbits. J Pediatr Surg 1975; 10: 583-92
    • (1975) J. Pediatr. Surg. , vol.10 , pp. 583-592
    • Marfatia, S.1    Donahoe, P.K.2    Hendren, W.H.3
  • 40
    • 0019517892 scopus 로고
    • Adult body temperature and heated humidification of anesthetic gases during general anesthesia
    • Stone DR, Downs JB, Paul WL, Perkins HM. Adult body temperature and heated humidification of anesthetic gases during general anesthesia. Anesth Analg 1981; 60: 736-41
    • (1981) Anesth. Analg. , vol.60 , pp. 736-741
    • Stone, D.R.1    Downs, J.B.2    Paul, W.L.3    Perkins, H.M.4
  • 41
    • 0015157999 scopus 로고
    • Humidification during positive-pressure ventilation of infants
    • Epstein RA. Humidification during positive-pressure ventilation of infants. Anesthesiology 1971; 35: 532-6
    • (1971) Anesthesiology , vol.35 , pp. 532-536
    • Epstein, R.A.1
  • 42
    • 0015543702 scopus 로고
    • Humidification and mucous flow in the intubated trachea
    • Forbes AR. Humidification and mucous flow in the intubated trachea. Br J Anaesth 1973; 45: 118-9
    • (1973) Br. J. Anaesth. , vol.45 , pp. 118-119
    • Forbes, A.R.1
  • 43
    • 0033030246 scopus 로고    scopus 로고
    • Do the components of heat and moisture exchanger filters affect their humidifying efficacy and the incidence of nosocomial pneumonia?
    • Thomachot L, Vialet R, Arnaud S, Barberon B, Michel-Nguyen A, Martin C. Do the components of heat and moisture exchanger filters affect their humidifying efficacy and the incidence of nosocomial pneumonia? Crit Care Med 1999; 27: 923-8
    • (1999) Crit. Care Med. , vol.27 , pp. 923-928
    • Thomachot, L.1    Vialet, R.2    Arnaud, S.3    Barberon, B.4    Michel-Nguyen, A.5    Martin, C.6
  • 44
    • 0031733848 scopus 로고    scopus 로고
    • Comparing two heat and moisture exchangers, one hydrophobic and one hygroscopic, on humidifying efficacy and the rate of nosocomial pneumonia
    • Thomachot L, Viviand X, Arnaud S, Boisson C, Martin C. Comparing two heat and moisture exchangers, one hydrophobic and one hygroscopic, on humidifying efficacy and the rate of nosocomial pneumonia. Chest 1998; 114: 1383-9
    • (1998) Chest , vol.114 , pp. 1383-1389
    • Thomachot, L.1    Viviand, X.2    Arnaud, S.3    Boisson, C.4    Martin, C.5
  • 45
    • 0023128748 scopus 로고
    • Preservation of humidity and heat of respiratory gases during anaesthesia - A laboratory investigation
    • Bengtson JP, Sonander H, Stenqvist O. Preservation of humidity and heat of respiratory gases during anaesthesia - a laboratory investigation. Acta Anaesthesiol Scand 1987; 31: 127-31
    • (1987) Acta Anaesthesiol. Scand. , vol.31 , pp. 127-131
    • Bengtson, J.P.1    Sonander, H.2    Stenqvist, O.3
  • 46
    • 0037712126 scopus 로고    scopus 로고
    • Utilisation des filtres échangeurs de chaleur et d'humidité au cours de la ventilation méanique des patients de réanimation
    • Ricard JD, Markowicz P, Dreyfuss D. Utilisation des filtres échangeurs de chaleur et d'humidité au cours de la ventilation méanique des patients de réanimation. Réanimation 2001; 10: 44-52
    • (2001) Réanimation , vol.10 , pp. 44-52
    • Ricard, J.D.1    Markowicz, P.2    Dreyfuss, D.3
  • 47
    • 0030975466 scopus 로고    scopus 로고
    • Unfavourable mechanical effects of heat and moisture exchangers in ventilated patients
    • Iotti GA, Olivei M, Palo A, et al. Unfavourable mechanical effects of heat and moisture exchangers in ventilated patients. Intensive Care Med 1997; 23: 399-405
    • (1997) Intensive Care Med. , vol.23 , pp. 399-405
    • Iotti, G.A.1    Olivei, M.2    Palo, A.3
  • 48
    • 0030879788 scopus 로고    scopus 로고
    • A prospective, randomized comparison of an in-line heat moisture exchange filter and heated wire humidifiers: Rates of ventilator-associated early-onset (community-acquired) or late-onset (hospital-acquired) pneumonia and incidence of endotracheal tube occlusion
    • Kirton OC, Dehaven B, Morgan J, Morejon O, Civetta. J. A prospective, randomized comparison of an in-line heat moisture exchange filter and heated wire humidifiers: rates of ventilator-associated early-onset (community-acquired) or late-onset (hospital-acquired) pneumonia and incidence of endotracheal tube occlusion. Chest 1997; 112: 1055-9
    • (1997) Chest , vol.112 , pp. 1055-1059
    • Kirton, O.C.1    Dehaven, B.2    Morgan, J.3    Morejon, O.4    Civetta, J.5
  • 49
    • 0023111304 scopus 로고
    • Neonatal hygroscopic condenser humidifier
    • Gedeon A, Mebius C, Palmer K. Neonatal hygroscopic condenser humidifier. Crit Care Med 1987; 15: 51-4
    • (1987) Crit. Care Med. , vol.15 , pp. 51-54
    • Gedeon, A.1    Mebius, C.2    Palmer, K.3
  • 50
    • 0029848008 scopus 로고    scopus 로고
    • Comparison of the effects of heat and moisture exchangers and heated humidifiers on ventilation and gas exchange during weaning trials from mechanical ventilation
    • Le Bourdelles G, Mier L, Fiquet B, et al. Comparison of the effects of heat and moisture exchangers and heated humidifiers on ventilation and gas exchange during weaning trials from mechanical ventilation. Chest 1996; 110: 1294-8
    • (1996) Chest , vol.110 , pp. 1294-1298
    • Le Bourdelles, G.1    Mier, L.2    Fiquet, B.3
  • 51
    • 0029934198 scopus 로고    scopus 로고
    • Effects of heat and moisture exchangers on minute ventilation, ventilatory drive, and work of breathing during pressure-support ventilation in acute respiratory failure
    • Pelosi P, Solca M, Ravagnan I, Tubiolo D, Ferrario L, Gattinoni L. Effects of heat and moisture exchangers on minute ventilation, ventilatory drive, and work of breathing during pressure-support ventilation in acute respiratory failure. Crit Care Med 1996; 24: 1184-8
    • (1996) Crit. Care Med. , vol.24 , pp. 1184-1188
    • Pelosi, P.1    Solca, M.2    Ravagnan, I.3    Tubiolo, D.4    Ferrario, L.5    Gattinoni, L.6
  • 52
    • 0030896388 scopus 로고    scopus 로고
    • Safety of combined heat and moisture exchanger filters in long-term mechanical ventilation
    • Hurni J, Feihl F, Lazor R, Leuenberger P, Perret C. Safety of combined heat and moisture exchanger filters in long-term mechanical ventilation. Chest 1997; 111: 686-91
    • (1997) Chest , vol.111 , pp. 686-691
    • Hurni, J.1    Feihl, F.2    Lazor, R.3    Leuenberger, P.4    Perret, C.5
  • 53
    • 0025241214 scopus 로고
    • Effects of the heat-moisture exchangers on dynamic hyperinflation of mechanically ventilated COPD patients
    • Conti G, De Blasi RA, Rocco M, et al. Effects of the heat-moisture exchangers on dynamic hyperinflation of mechanically ventilated COPD patients. Intensive Care Med 1990; 16: 441-3
    • (1990) Intensive Care Med. , vol.16 , pp. 441-443
    • Conti, G.1    De Blasi, R.A.2    Rocco, M.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.