Volumn 12, Issue 3, 2008, Pages 399-405

An efficient motion-resistant method for wearable pulse oximeter

Author keywords

Blood oxygen saturation; Motion artifact; Photoplethysmography (PPG); Pulse oximetry; Telemedicine; Wearable medical device

Indexed keywords

ADAPTIVE FILTERS; ALGORITHMS; COMPUTATIONAL EFFICIENCY; HEMOGLOBIN OXYGEN SATURATION; LINEAR ALGEBRA; OXIMETERS; PLETHYSMOGRAPHY;

EID: 44449169431     PISSN: 10897771     EISSN: None     Source Type: Journal    
DOI: 10.1109/TITB.2007.902173     Document Type: Article
Times cited : (57)

References (22)
  • 1
    • 0028221837 scopus 로고
    • Use of two pulse oximeters to investigate a method of movement artifact rejection using photo-plethysmographic signals
    • A. R. Visram, R. D. M. Jones, M. G. Irwin, and J. Bacon-Shone, "Use of two pulse oximeters to investigate a method of movement artifact rejection using photo-plethysmographic signals," Brit. j. Anaesth. vol.72, pp. 388-392, 1994.
    • (1994) Brit. j. Anaesth , vol.72 , pp. 388-392
    • Visram, A.R.1    Jones, R.D.M.2    Irwin, M.G.3    Bacon-Shone, J.4
  • 2
    • 0027674239 scopus 로고
    • The pulse oximeter: Application and limitations: An analysis of 2000 incident reports
    • W. B. Runciman, R. K. Webb, L. Barker, and M. Curriie, "The pulse oximeter: Application and limitations: An analysis of 2000 incident reports," Anaaesth. Intensive Care, vol. 21, pp. 543-550, 1993.
    • (1993) Anaaesth. Intensive Care , vol.21 , pp. 543-550
    • Runciman, W.B.1    Webb, R.K.2    Barker, L.3    Curriie, M.4
  • 3
    • 44449138905 scopus 로고    scopus 로고
    • In vivo evaluation of Massimo SET and Nellcor N-395 oximetry accuracy compared to CO-oximetry in chronically desaturated children
    • F. A. Robertson, G. M. Hoffman, E. A. Stuth, S. Staudt, and R. Berens, "In vivo evaluation of Massimo SET and Nellcor N-395 oximetry accuracy compared to CO-oximetry in chronically desaturated children," Anesthesiology, vol. 99, p. A620, 2003.
    • (2003) Anesthesiology , vol.99
    • Robertson, F.A.1    Hoffman, G.M.2    Stuth, E.A.3    Staudt, S.4    Berens, R.5
  • 4
    • 0034968268 scopus 로고    scopus 로고
    • Artifact-resistant power-efficient design of finger-ring plethysmographic sensors
    • Jul
    • R. Sokwoo, B. H. Yang, and H. H. Asada, "Artifact-resistant power-efficient design of finger-ring plethysmographic sensors," IEEE Trans. Biomed. Eng., vol. 48, no. 7, pp. 795-805, Jul. 2001.
    • (2001) IEEE Trans. Biomed. Eng , vol.48 , Issue.7 , pp. 795-805
    • Sokwoo, R.1    Yang, B.H.2    Asada, H.H.3
  • 5
    • 18044373423 scopus 로고    scopus 로고
    • Reduction of motion artifact in pudse oximetry by smoothed pseudo Wigner-Ville distribution
    • Mar. 1
    • Y. S. Yan, C. C. Poon, and Y. T. Zhang, "Reduction of motion artifact in pudse oximetry by smoothed pseudo Wigner-Ville distribution," J. Neuroeng. Rehabil., vol. 2, no. 1, p. 3, Mar. 1, 2005.
    • (2005) J. Neuroeng. Rehabil , vol.2 , Issue.1 , pp. 3
    • Yan, Y.S.1    Poon, C.C.2    Zhang, Y.T.3
  • 6
    • 44449164400 scopus 로고    scopus 로고
    • Signal processing apparatus,
    • International Patent Application WO 96/12435, May 2
    • M. K. Diab, "Signal processing apparatus," International Patent Application WO 96/12435, May 2, 1996.
    • (1996)
    • Diab, M.K.1
  • 7
    • 44449095539 scopus 로고    scopus 로고
    • D. Parker, Optical monitor oximeter, etc, with motion artifact suppression, International Patent Application WO 94/03102, Feb. 17, 1994
    • D. Parker, "Optical monitor (oximeter., etc.) with motion artifact suppression," International Patent Application WO 94/03102, Feb. 17, 1994.
  • 8
    • 0000573306 scopus 로고    scopus 로고
    • Artifact reduction in photoplethysmography
    • M. J. Hayes and P. R. Smith, "Artifact reduction in photoplethysmography," Appl. Opt., vol. 37, pp. 7427-7446, 1998.
    • (1998) Appl. Opt , vol.37 , pp. 7427-7446
    • Hayes, M.J.1    Smith, P.R.2
  • 9
    • 0030017260 scopus 로고    scopus 로고
    • Clinical evaluation of a prototype motion artifact resistant pulse oximeter in the recovery room
    • C. Dumas, J. A. Wahr, and K. K. Tremper, "Clinical evaluation of a prototype motion artifact resistant pulse oximeter in the recovery room," Anesth. Analg., vol. 83, pp. 269-272, 1996.
    • (1996) Anesth. Analg , vol.83 , pp. 269-272
    • Dumas, C.1    Wahr, J.A.2    Tremper, K.K.3
  • 11
    • 4243980942 scopus 로고    scopus 로고
    • Evaluation of signal extraction technology (SET) in preventing false alarms when using pulse oximetry in the recovery room
    • P. R. Lichtenthal and L. D. Wade, "Evaluation of signal extraction technology (SET) in preventing false alarms when using pulse oximetry in the recovery room," Anesthesiology, vol. 86, pp. 278-279, 1996.
    • (1996) Anesthesiology , vol.86 , pp. 278-279
    • Lichtenthal, P.R.1    Wade, L.D.2
  • 12
    • 0007676990 scopus 로고
    • New pulse oximeter: Resistant to motion artifacts
    • N. S. Trivedi, N. K. Shah, and B. P. Jacobsen, "New pulse oximeter: Resistant to motion artifacts," Anesth. Analog., vol. 880, p. S510, 1995.
    • (1995) Anesth. Analog , vol.880
    • Trivedi, N.S.1    Shah, N.K.2    Jacobsen, B.P.3
  • 13
    • 0007676989 scopus 로고
    • Comparison of alarm conditions between new pulse oximeters during motion at normal and low oxygen saturations
    • N. Shah, S. Barker, and J. Hyatt, "Comparison of alarm conditions between new pulse oximeters during motion at normal and low oxygen saturations," Anesthesiology, vol. 83, p. A1067, 1995.
    • (1995) Anesthesiology , vol.83
    • Shah, N.1    Barker, S.2    Hyatt, J.3
  • 15
    • 0026335383 scopus 로고
    • Simple photon diffusion analysis of the effects of multiple scattering on pulse oximetry
    • Dec
    • J. M. Schmitt, "Simple photon diffusion analysis of the effects of multiple scattering on pulse oximetry," IEEE Trans. Biomed. Eng., vol. 38, no. 12, pp. 1194-1203, Dec. 1991.
    • (1991) IEEE Trans. Biomed. Eng , vol.38 , Issue.12 , pp. 1194-1203
    • Schmitt, J.M.1
  • 17
    • 0026186386 scopus 로고
    • Given rotation based least squares lattice and related algorithms
    • Jul
    • F. Ling, "Given rotation based least squares lattice and related algorithms," IEEE Trans. Signal Process., vol. 39, no. 7. pp. 1541-1551. Jul. 1991.
    • (1991) IEEE Trans. Signal Process , vol.39 , Issue.7 , pp. 1541-1551
    • Ling, F.1
  • 18
    • 0030032649 scopus 로고    scopus 로고
    • Signal processing methods for pulse oximetry
    • T. L. Rusch, R. Sankar, and J. E. Scarf, "Signal processing methods for pulse oximetry," Comput. Biol. Med., vol. 26, pp. 143-159, 1996.
    • (1996) Comput. Biol. Med , vol.26 , pp. 143-159
    • Rusch, T.L.1    Sankar, R.2    Scarf, J.E.3
  • 19
    • 0029703681 scopus 로고    scopus 로고
    • Neural networks and wavelet analysis in the computer interpretation of pulse oximetry data
    • presented at the, Kyoto, Japan
    • F. U. Dowla, P. G. Skowkowski, and R. R. Leach, "Neural networks and wavelet analysis in the computer interpretation of pulse oximetry data," presented at the IEEE Workshop Neural Netw. Signal Process., Kyoto, Japan, 1996.
    • (1996) IEEE Workshop Neural Netw. Signal Process
    • Dowla, F.U.1    Skowkowski, P.G.2    Leach, R.R.3
  • 20
    • 0036140133 scopus 로고    scopus 로고
    • A characterization of motion affecting pulse oximetry in 350 patients
    • R. M. Tobin, J. A. Pologe, and P. R. B. Batchelder, "A characterization of motion affecting pulse oximetry in 350 patients," Anesth. Analg. vol. 94, Suppl. 1, pp. S54-S61, 2002.
    • (2002) Anesth. Analg , vol.94 , Issue.SUPPL. 1
    • Tobin, R.M.1    Pologe, J.A.2    Batchelder, P.R.B.3
  • 21
    • 0036785830 scopus 로고    scopus 로고
    • Motion-resistant pulse oximetry: A comparison of new and old models
    • S. J. Barker, "Motion-resistant pulse oximetry: A comparison of new and old models," Anesth. Analg., vol. 95, pp. 967-972, 2002.
    • (2002) Anesth. Analg , vol.95 , pp. 967-972
    • Barker, S.J.1
  • 22
    • 0034435445 scopus 로고    scopus 로고
    • Determining the artifact sensitivity of recent pulse oximeter during laboratory benchmarking
    • S. W. Kastle and E. Konecny, "Determining the artifact sensitivity of recent pulse oximeter during laboratory benchmarking," J. Clin. Monit. Comput, vol. 16, pp. 509-522, 2000.
    • (2000) J. Clin. Monit. Comput , vol.16 , pp. 509-522
    • Kastle, S.W.1    Konecny, E.2


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.