Volumn 18, Issue 1, 2008, Pages 39-43

Is the PASCAL®-Tonometer suitable for measuring intraocular pressure in clinical routine? Long- and short-term reproducibility of dynamic contour tonometry

Author keywords

Dynamic contour tonometry; Glaucoma; Goldmann applanation tonometry; Reproducibility

Indexed keywords

ADULT; AGED; ARTICLE; CLINICAL ARTICLE; CLINICAL PRACTICE; CONTROLLED STUDY; CORNEA; CORRELATION ANALYSIS; EVALUATION; FEMALE; GLAUCOMA; HUMAN; INTRAOCULAR PRESSURE; MALE; PRESSURE MEASUREMENT; PRIORITY JOURNAL; REPRODUCIBILITY; STATISTICAL ANALYSIS; STATISTICAL SIGNIFICANCE; TONOMETER; TONOMETRY; ADOLESCENT; CLINICAL TRIAL; COMPARATIVE STUDY; CONTROLLED CLINICAL TRIAL; INSTRUMENTATION; METHODOLOGY; MIDDLE AGED; OCULOPLETHYSMOGRAPHY; PHYSIOLOGY; PROSPECTIVE STUDY; RANDOMIZED CONTROLLED TRIAL;

EID: 41149083154     PISSN: 11206721     EISSN: None     Source Type: Journal    
DOI: 10.1177/112067210801800107     Document Type: Article
Times cited : (18)

References (17)
  • 1
    • 0027291601 scopus 로고
    • Sources of error with use of Goldmann-type tonometers
    • Whitacre MM, Stein R. Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol 1993; 38: 1-30.
    • (1993) Surv Ophthalmol , vol.38 , pp. 1-30
    • Whitacre, M.M.1    Stein, R.2
  • 2
    • 1842504236 scopus 로고    scopus 로고
    • Corneal thickness in glaucoma screening, diagnosis, and management
    • Brandt JD. Corneal thickness in glaucoma screening, diagnosis, and management. Curr Opin Ophthalmol 2004; 15:85-9.
    • (2004) Curr Opin Ophthalmol , vol.15 , pp. 85-89
    • Brandt, J.D.1
  • 3
    • 0031848594 scopus 로고    scopus 로고
    • Applanation tonometry and correction according to corneal thickness
    • Stodtmeister R. Applanation tonometry and correction according to corneal thickness. Acta Ophthalmol Scand 1998; 76: 319-24.
    • (1998) Acta Ophthalmol Scand , vol.76 , pp. 319-324
    • Stodtmeister, R.1
  • 5
    • 0031668042 scopus 로고    scopus 로고
    • Correlation of intraocular pressure and central corneal thickness in normal myopic eyes and after laser in situ keratomileusis
    • Emara B, Probst LE, Tingey DP, Kennedy DW, Willms LJ, Machat J. Correlation of intraocular pressure and central corneal thickness in normal myopic eyes and after laser in situ keratomileusis. J Cataract Refract Surg 1998; 24: 1320-5.
    • (1998) J Cataract Refract Surg , vol.24 , pp. 1320-1325
    • Emara, B.1    Probst, L.E.2    Tingey, D.P.3    Kennedy, D.W.4    Willms, L.J.5    Machat, J.6
  • 6
    • 0036272299 scopus 로고    scopus 로고
    • The Ocular Hypertension Treatment Study: Baseline factors that predict the onset of primary open-angle glaucoma
    • Gordon MO, Beiser JA, Brandt JD, et al. The Ocular Hypertension Treatment Study: baseline factors that predict the onset of primary open-angle glaucoma. Arch Ophthalmol 2002; 120: 714-20.
    • (2002) Arch Ophthalmol , vol.120 , pp. 714-720
    • Gordon, M.O.1    Beiser, J.A.2    Brandt, J.D.3
  • 8
    • 0029609715 scopus 로고
    • Ocular hypertension and central corneal thickness
    • Argus WA. Ocular hypertension and central corneal thickness. Ophthalmology 1995; 102: 1810-2.
    • (1995) Ophthalmology , vol.102 , pp. 1810-1812
    • Argus, W.A.1
  • 9
    • 13844318981 scopus 로고    scopus 로고
    • Assessment of the Pascal dynamic contour tonometer in monitoring intraocular pressure in unoperated eyes and eyes after LASIK
    • Siganos DS, Papastergiou GI, Moedas C. Assessment of the Pascal dynamic contour tonometer in monitoring intraocular pressure in unoperated eyes and eyes after LASIK. J Cataract Refract Surg 2004; 30: 746-51.
    • (2004) J Cataract Refract Surg , vol.30 , pp. 746-751
    • Siganos, D.S.1    Papastergiou, G.I.2    Moedas, C.3
  • 10
    • 3042667876 scopus 로고    scopus 로고
    • Dynamic contour tonometry for post-LASIK intraocular pressure measurements
    • Duba I, Wirthlin AC. Dynamic contour tonometry for post-LASIK intraocular pressure measurements. Klin Monatsbl Augenheilkd 2004; 221:347-50.
    • (2004) Klin Monatsbl Augenheilkd , vol.221 , pp. 347-350
    • Duba, I.1    Wirthlin, A.C.2
  • 12
    • 25644460363 scopus 로고    scopus 로고
    • Dynamic contour tonometry: Presentation of a new tonometer
    • Kanngiesser HE, Kniestedt C, Robert YC. Dynamic contour tonometry: presentation of a new tonometer. J Glaucoma 2005; 14: 344-50.
    • (2005) J Glaucoma , vol.14 , pp. 344-350
    • Kanngiesser, H.E.1    Kniestedt, C.2    Robert, Y.C.3
  • 13
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1: 307-10.
    • (1986) Lancet , vol.1 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 14
    • 24944580479 scopus 로고    scopus 로고
    • Dynamic contour tonometry versus Goldmann applanation tonometry: A comparative study
    • Pache M, Wilmsmeyer S, Lautebach S, Funk J. Dynamic contour tonometry versus Goldmann applanation tonometry: a comparative study. Graefes Arch Clin Exp Ophthalmol 2005; 243: 763-7.
    • (2005) Graefes Arch Clin Exp Ophthalmol , vol.243 , pp. 763-767
    • Pache, M.1    Wilmsmeyer, S.2    Lautebach, S.3    Funk, J.4
  • 15
    • 30344436327 scopus 로고    scopus 로고
    • Intraocular pressure measurement-comparison of dynamic contour tonometry and Goldmann applanation tonometry
    • Schneider E, Grehn F. Intraocular pressure measurement-comparison of dynamic contour tonometry and Goldmann applanation tonometry. J Glaucoma 2006; 15: 2-6.
    • (2006) J Glaucoma , vol.15 , pp. 2-6
    • Schneider, E.1    Grehn, F.2
  • 16
    • 33845260961 scopus 로고    scopus 로고
    • Reproducibility of dynamic contour tonometry. Comparison with TonoPenXL and Goldmann applanation tonometry: A clinical study on 323 normal eyes
    • Viestenz A, Langenbucher A. Reproducibility of dynamic contour tonometry. Comparison with TonoPenXL and Goldmann applanation tonometry: a clinical study on 323 normal eyes. Klin Monatsbl Augenheilkd 2006; 223: 813-9.
    • (2006) Klin Monatsbl Augenheilkd , vol.223 , pp. 813-819
    • Viestenz, A.1    Langenbucher, A.2
  • 17
    • 4344659799 scopus 로고    scopus 로고
    • Comparison of dynamic contour tonometry with goldmann applanation tonometry
    • Kaufmann C, Bachmann LM, Thiel MA. Comparison of dynamic contour tonometry with goldmann applanation tonometry. Invest Ophthalmol Vis Sci 2004; 45: 3118-21.
    • (2004) Invest Ophthalmol Vis Sci , vol.45 , pp. 3118-3121
    • Kaufmann, C.1    Bachmann, L.M.2    Thiel, M.A.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.