Ventricular oversensing in 518 patients with implanted cardiac defibrillators: Incidence, complications, and solutions

Author keywords

Complication; ICD; Ventricular oversensing

Indexed keywords

ADULT; AGED; ARTICLE; ARTIFICIAL HEART PACEMAKER; CARDIOVASCULAR DISEASE; CONTROLLED STUDY; CONVULSIVE THERAPY; DEFIBRILLATOR; ELECTROMAGNETIC FIELD; FEMALE; FOLLOW UP; HEART VENTRICLE FIBRILLATION; HUMAN; INVASIVE PROCEDURE; MAJOR CLINICAL STUDY; MALE; PACEMAKER ELECTRODE; PRIORITY JOURNAL; CARDIOVERSION; COHORT ANALYSIS; ELECTROCARDIOGRAPHY; ELECTROCONVULSIVE THERAPY; HEART VENTRICLE; INCIDENCE; INSTRUMENTATION; METHODOLOGY; MIDDLE AGED; PATHOPHYSIOLOGY; RETROSPECTIVE STUDY;

EID: 40049087455     PISSN: 10995129     EISSN: 15322092     Source Type: Journal    
DOI: 10.1093/europace/eum195     Document Type: Article
Times cited : (47)

References (34)
  • 1
    • 0037149716 scopus 로고    scopus 로고
    • Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction
    • Moss AJ, Zareba W, Hall WJ, Klein H, Wilber DJ, Cannom DS et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med 2002;346:877-83.
    • (2002) N Engl J Med , vol.346 , pp. 877-883
    • Moss, A.J.1    Zareba, W.2    Hall, W.J.3    Klein, H.4    Wilber, D.J.5    Cannom, D.S.6
  • 2
    • 19944429413 scopus 로고    scopus 로고
    • Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure
    • Bardy GH, Lee KL, Mark DB, Poole JE, Packer DL, Boineau R et al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med 2005;352:225-37.
    • (2005) N Engl J Med , vol.352 , pp. 225-237
    • Bardy, G.H.1    Lee, K.L.2    Mark, D.B.3    Poole, J.E.4    Packer, D.L.5    Boineau, R.6
  • 3
    • 2442479695 scopus 로고    scopus 로고
    • Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure
    • Bristow MR, Saxon LA, Boehmer J, Krueger S, Kass DA, De MT et al. Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. N Engl J Med 2004;350:2140-50.
    • (2004) N Engl J Med , vol.350 , pp. 2140-2150
    • Bristow, M.R.1    Saxon, L.A.2    Boehmer, J.3    Krueger, S.4    Kass5    DA, D.M.6
  • 4
    • 9844224486 scopus 로고    scopus 로고
    • A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. The Antiarrhythmics versus Implantable Defibrillators (AVID) Investigators. N Engl J Med 1997;337:1576-83
    • A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. The Antiarrhythmics versus Implantable Defibrillators (AVID) Investigators. N Engl J Med 1997;337:1576-83.
  • 6
    • 0037109150 scopus 로고    scopus 로고
    • ACC/AHA/NASPE 2002 guideline update for implantation of cardiac pacemakers and antiarrhythmia devices: Summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/NASPE Committee to Update the 1998 Pacemaker Guidelines)
    • Gregoratos G, Abrams J, Epstein AE, Freedman RA, Hayes DL, Hlatky MA et al. ACC/AHA/NASPE 2002 guideline update for implantation of cardiac pacemakers and antiarrhythmia devices: summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/NASPE Committee to Update the 1998 Pacemaker Guidelines). Circulation 2002;106:2145-61.
    • (2002) Circulation , vol.106 , pp. 2145-2161
    • Gregoratos, G.1    Abrams, J.2    Epstein, A.E.3    Freedman, R.A.4    Hayes, D.L.5    Hlatky, M.A.6
  • 7
    • 0017706910 scopus 로고
    • A comparison of unipolar and bipolar electrograms for cardiac pacemaker sensing
    • DeCaprio V, Hurzeler P, Furman S. A comparison of unipolar and bipolar electrograms for cardiac pacemaker sensing. Circulation 1977;56:750-5.
    • (1977) Circulation , vol.56 , pp. 750-755
    • DeCaprio, V.1    Hurzeler, P.2    Furman, S.3
  • 8
    • 0020078923 scopus 로고
    • Unipolar sensing abnormalities: Incidence and clinical significance of skeletal muscle interference and undersensing in 228 patients
    • Secemsky SI, Hauser RG, Denes P, Edwards LM. Unipolar sensing abnormalities: incidence and clinical significance of skeletal muscle interference and undersensing in 228 patients. Pacing Clin Electrophysiol 1982;5:10-9.
    • (1982) Pacing Clin Electrophysiol , vol.5 , pp. 10-19
    • Secemsky, S.I.1    Hauser, R.G.2    Denes, P.3    Edwards, L.M.4
  • 9
    • 0035212343 scopus 로고    scopus 로고
    • Inappropriate arrhythmia detection in implantable defibrillator therapy due to oversensing of diaphragmatic myopotentials
    • Schulte B, Sperzel J, Carlsson J, Dursch M, Erdogan A, Pitschner HF et al. Inappropriate arrhythmia detection in implantable defibrillator therapy due to oversensing of diaphragmatic myopotentials. J Interv Card Electrophysiol 2001;5:487-93.
    • (2001) J Interv Card Electrophysiol , vol.5 , pp. 487-493
    • Schulte, B.1    Sperzel, J.2    Carlsson, J.3    Dursch, M.4    Erdogan, A.5    Pitschner, H.F.6
  • 10
    • 4644338784 scopus 로고    scopus 로고
    • Total failure to sense ventricular fibrillation with inappropriate defibrillator sensitivity adjustment
    • Lin D, Dixit S, Russo AM, Hsia HH. Total failure to sense ventricular fibrillation with inappropriate defibrillator sensitivity adjustment. Pacing Clin Electrophysiol 2004;27:1321-3.
    • (2004) Pacing Clin Electrophysiol , vol.27 , pp. 1321-1323
    • Lin, D.1    Dixit, S.2    Russo, A.M.3    Hsia, H.H.4
  • 11
    • 0037245991 scopus 로고    scopus 로고
    • Ventricular oversensing: A study of 101 patients implanted with dual chamber defibrillators and two different lead systems
    • Weretka S, Michaelsen J, Becker R, Karle CA, Voss F, Hilbel T et al. Ventricular oversensing: a study of 101 patients implanted with dual chamber defibrillators and two different lead systems. Pacing Clin Electrophysiol 2003;26:65-70.
    • (2003) Pacing Clin Electrophysiol , vol.26 , pp. 65-70
    • Weretka, S.1    Michaelsen, J.2    Becker, R.3    Karle, C.A.4    Voss, F.5    Hilbel, T.6
  • 13
    • 9344260258 scopus 로고    scopus 로고
    • T wave oversensing by a cardioverter defibrillator implanted in a patient with the Brugada syndrome
    • Porres JM, Brugada J, Marco P, Garcia F, Azcarate B. T wave oversensing by a cardioverter defibrillator implanted in a patient with the Brugada syndrome. Pacing Clin Electrophysiol 2004;27:1563-5.
    • (2004) Pacing Clin Electrophysiol , vol.27 , pp. 1563-1565
    • Porres, J.M.1    Brugada, J.2    Marco, P.3    Garcia, F.4    Azcarate, B.5
  • 14
    • 3242742134 scopus 로고    scopus 로고
    • High incidence of appropriate and inappropriate ICD therapies in children and adolescents with implantable cardioverter defibrillator
    • Korte T, Koditz H, Niehaus M, Paul T, Tebbenjohanns J. High incidence of appropriate and inappropriate ICD therapies in children and adolescents with implantable cardioverter defibrillator. Pacing Clin Electrophysiol 2004;27:924-32.
    • (2004) Pacing Clin Electrophysiol , vol.27 , pp. 924-932
    • Korte, T.1    Koditz, H.2    Niehaus, M.3    Paul, T.4    Tebbenjohanns, J.5
  • 16
    • 0034761426 scopus 로고    scopus 로고
    • Inappropriate discharges by fourth generation implantable cardioverter defibrillators in patients with ventricular arrhythmias
    • Washizuka T, Chinushi M, Tagawa M, Kasai H, Watanabe H, Hosaka Y et al. Inappropriate discharges by fourth generation implantable cardioverter defibrillators in patients with ventricular arrhythmias. Jpn Circ J 2001;65:927-30.
    • (2001) Jpn Circ J , vol.65 , pp. 927-930
    • Washizuka, T.1    Chinushi, M.2    Tagawa, M.3    Kasai, H.4    Watanabe, H.5    Hosaka, Y.6
  • 17
    • 0035113587 scopus 로고    scopus 로고
    • Avoiding inappropriate ventricular tachycardia detection due to T-wave oversensing in an implantable cardioverter defibrillator: A novel application of the electrogram width criterion
    • Duru F, Bauersfeld U, Candinas R. Avoiding inappropriate ventricular tachycardia detection due to T-wave oversensing in an implantable cardioverter defibrillator: a novel application of the electrogram width criterion. Europace 2001;3:80-4.
    • (2001) Europace , vol.3 , pp. 80-84
    • Duru, F.1    Bauersfeld, U.2    Candinas, R.3
  • 18
    • 0034835765 scopus 로고    scopus 로고
    • Incidence and predictors of pacemaker dysfunction with unipolar ventricular lead configuration. Can we identify patients who benefit from bipolar electrodes?
    • Wiegand UK, Bode F, Bonnemeier H, Tolg R, Peters W, Katus HA. Incidence and predictors of pacemaker dysfunction with unipolar ventricular lead configuration. Can we identify patients who benefit from bipolar electrodes? Pacing Clin Electrophysiol 2001;24:1383-8.
    • (2001) Pacing Clin Electrophysiol , vol.24 , pp. 1383-1388
    • Wiegand, U.K.1    Bode, F.2    Bonnemeier, H.3    Tolg, R.4    Peters, W.5    Katus, H.A.6
  • 19
    • 0028261802 scopus 로고
    • Inappropriate discharge by an implantable cardioverter defibrillator: Recognition of myopotential sensing using telemetered intracardiac electrograms
    • Sandler MJ, Kutalek SP. Inappropriate discharge by an implantable cardioverter defibrillator: recognition of myopotential sensing using telemetered intracardiac electrograms. Pacing Clin Electrophysiol 1994;17:665-71.
    • (1994) Pacing Clin Electrophysiol , vol.17 , pp. 665-671
    • Sandler, M.J.1    Kutalek, S.P.2
  • 20
    • 0031868579 scopus 로고    scopus 로고
    • Myopotential sensing by a dual chamber implantable cardioverter defibrillator: Two case reports
    • Deshmukh P, Anderson K. Myopotential sensing by a dual chamber implantable cardioverter defibrillator: two case reports. J Cardiovasc Electrophysiol 1998;9:767-72.
    • (1998) J Cardiovasc Electrophysiol , vol.9 , pp. 767-772
    • Deshmukh, P.1    Anderson, K.2
  • 21
    • 0034213167 scopus 로고    scopus 로고
    • Effect of sensing system on the incidence of myopotential oversensing during bradycardia pacing in implantable cardioverter-defibrillators
    • Mann DE, Otto L, Kelly PA, Reiter MJ. Effect of sensing system on the incidence of myopotential oversensing during bradycardia pacing in implantable cardioverter-defibrillators. Am J Cardiol 2000;85:1380-2.
    • (2000) Am J Cardiol , vol.85 , pp. 1380-1382
    • Mann, D.E.1    Otto, L.2    Kelly, P.A.3    Reiter, M.J.4
  • 24
    • 27844469791 scopus 로고    scopus 로고
    • Incidence of complications in patients with implantable cardioverter/defibrillator who receive additional transvenous pace/sense leads
    • Wollmann CG, Bocker D, Loher A, Kobe J, Scheld HH, Breithardt GE et al. Incidence of complications in patients with implantable cardioverter/defibrillator who receive additional transvenous pace/sense leads. Pacing Clin Electrophysiol 2005;28:795-800.
    • (2005) Pacing Clin Electrophysiol , vol.28 , pp. 795-800
    • Wollmann, C.G.1    Bocker, D.2    Loher, A.3    Kobe, J.4    Scheld, H.H.5    Breithardt, G.E.6
  • 25
    • 0037219077 scopus 로고    scopus 로고
    • Detection and management of an implantable cardioverter defibrillator lead failure: Incidence and clinical implications
    • Ellenbogen KA, Wood MA, Shepard RK, Clemo HF, Vaughn T, Holloman K et al. Detection and management of an implantable cardioverter defibrillator lead failure: incidence and clinical implications. J Am Coll Cardiol 2003;41:73-80.
    • (2003) J Am Coll Cardiol , vol.41 , pp. 73-80
    • Ellenbogen, K.A.1    Wood, M.A.2    Shepard, R.K.3    Clemo, H.F.4    Vaughn, T.5    Holloman, K.6
  • 26
    • 0033762014 scopus 로고    scopus 로고
    • Intradevice interaction in a dual chamber implantable cardioverter defibrillator preventing ventricular tachyarrhythmia detection
    • Shivkumar K, Feliciano Z, Boyle NG, Wiener I. Intradevice interaction in a dual chamber implantable cardioverter defibrillator preventing ventricular tachyarrhythmia detection. J Cardiovasc Electrophysiol 2000;11:1285-8.
    • (2000) J Cardiovasc Electrophysiol , vol.11 , pp. 1285-1288
    • Shivkumar, K.1    Feliciano, Z.2    Boyle, N.G.3    Wiener, I.4
  • 27
    • 8044261416 scopus 로고    scopus 로고
    • Interactions between transvenous nonthoracotomy cardioverter defibrillator systems and permanent transvenous endocardial pacemakers
    • Geiger MJ, O'Neill P, Sharma A, Skadsen A, Zimerman L, Greenfield RA et al. Interactions between transvenous nonthoracotomy cardioverter defibrillator systems and permanent transvenous endocardial pacemakers. Pacing Clin Electrophysiol 1997;20:624-30.
    • (1997) Pacing Clin Electrophysiol , vol.20 , pp. 624-630
    • Geiger, M.J.1    O'Neill, P.2    Sharma, A.3    Skadsen, A.4    Zimerman, L.5    Greenfield, R.A.6
  • 28
    • 0037109150 scopus 로고    scopus 로고
    • ACC/AHA/NASPE 2002 guideline update for implantation of cardiac pacemakers and antiarrhythmia devices: Summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/NASPE Committee to Update the 1998 Pacemaker Guidelines)
    • Gregoratos G, Abrams J, Epstein AE, Freedman RA, Hayes DL, Hlatky MA et al. ACC/AHA/NASPE 2002 guideline update for implantation of cardiac pacemakers and antiarrhythmia devices: summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/NASPE Committee to Update the 1998 Pacemaker Guidelines). Circulation 2002;106:2145-61.
    • (2002) Circulation , vol.106 , pp. 2145-2161
    • Gregoratos, G.1    Abrams, J.2    Epstein, A.E.3    Freedman, R.A.4    Hayes, D.L.5    Hlatky, M.A.6
  • 32
    • 11244249523 scopus 로고    scopus 로고
    • The influence of non-ionizing electromagnetic fields on implantable cardiac medical devices
    • Souques M. The influence of non-ionizing electromagnetic fields on implantable cardiac medical devices. Presse Med 2004;33:1611-5.
    • (2004) Presse Med , vol.33 , pp. 1611-1615
    • Souques, M.1
  • 33
    • 0024990874 scopus 로고
    • Clinical interactions between pacemakers and automatic implantable cardioverter- defibrillators
    • Calkins H, Brinker J, Veltri EP, Guarnieri T, Levine JH. Clinical interactions between pacemakers and automatic implantable cardioverter- defibrillators. J Am Coll Cardiol 1990;16:666-73.
    • (1990) J Am Coll Cardiol , vol.16 , pp. 666-673
    • Calkins, H.1    Brinker, J.2    Veltri, E.P.3    Guarnieri, T.4    Levine, J.H.5
  • 34
    • 0037109150 scopus 로고    scopus 로고
    • ACC/AHA/NASPE 2002 guideline update for implantation of cardiac pacemakers and antiarrhythmia devices: Summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/NASPE Committee to Update the 1998 Pacemaker Guidelines)
    • Gregoratos G, Abrams J, Epstein AE, Freedman RA, Hayes DL, Hlatky MA et al. ACC/AHA/NASPE 2002 guideline update for implantation of cardiac pacemakers and antiarrhythmia devices: summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/NASPE Committee to Update the 1998 Pacemaker Guidelines). Circulation 2002;106:2145-61.
    • (2002) Circulation , vol.106 , pp. 2145-2161
    • Gregoratos, G.1    Abrams, J.2    Epstein, A.E.3    Freedman, R.A.4    Hayes, D.L.5    Hlatky, M.A.6


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.