Volumn 48, Issue 5, 2004, Pages 658-661

Pulmonary artery thermodilution cardiac output vs. transpulmonary thermodilution cardiac output in two patients with intrathoracic pathology

Author keywords

Cardiac output; Intrathoracic pathology; Transpulmonay cardiac output

Indexed keywords

ACCURACY; AGED; ALGORITHM; AORTA INJURY; ARTERY CATHETER; ARTICLE; BLEEDING; BOLUS TRANSPULMONARY THERMODILUTION TECHNIQUE; CALIBRATION; CARDIOVASCULAR DISEASE; CASE REPORT; CLINICAL FEATURE; CONTINUOUS PULMONARY ARTERY THERMODILUTION TECHNIQUE; CRITICAL ILLNESS; DEVICE; FEMALE; HEART LEFT RIGHT SHUNT; HEART OUTPUT; HEATING; HEMODYNAMICS; HUMAN; INTERMETHOD COMPARISON; PATIENT MONITORING; PRIORITY JOURNAL; PULMONARY ARTERY; PULMONARY VEIN MALFORMATION; SYMPTOM; THERMODILUTION; THORAX DISEASE; VALIDATION PROCESS;

EID: 2442477355     PISSN: 00015172     EISSN: None     Source Type: Journal    
DOI: 10.1111/j.1399-6576.2004.00375.x     Document Type: Article
Times cited : (16)

References (20)
  • 1
    • 0027938298 scopus 로고
    • Is continuous cardiac output measurement using thermodilution reliable in the critically ill patient?
    • Boldt J, Menges T, Wollbruck M, Hammermann H, Hempelmann G. Is continuous cardiac output measurement using thermodilution reliable in the critically ill patient? Crit Care Med 1994; 22: 1913-8.
    • (1994) Crit Care Med , vol.22 , pp. 1913-1918
    • Boldt, J.1    Menges, T.2    Wollbruck, M.3    Hammermann, H.4    Hempelmann, G.5
  • 2
    • 0029926874 scopus 로고    scopus 로고
    • Semi-continuous versus injectate cardiac output measurement in intensive care patients after cardiac surgery
    • Bottiger BW, Soder M, Rauch H et al. Semi-continuous versus injectate cardiac output measurement in intensive care patients after cardiac surgery. Intensive Care Med 1996; 22: 312-8.
    • (1996) Intensive Care Med , vol.22 , pp. 312-318
    • Bottiger, B.W.1    Soder, M.2    Rauch, H.3
  • 3
    • 0036198399 scopus 로고    scopus 로고
    • Continuous and intermittent cardiac output measurement: Pulmonary artery catheter versus aortic transpulmonary technique
    • Della RG, Costa MG, Pompei L, Coccia C, Pietropaoli P. Continuous and intermittent cardiac output measurement: pulmonary artery catheter versus aortic transpulmonary technique. Br J Anaesth 2002; 88: 350-6.
    • (2002) Br J Anaesth , vol.88 , pp. 350-356
    • Della, R.G.1    Costa, M.G.2    Pompei, L.3    Coccia, C.4    Pietropaoli, P.5
  • 4
    • 0026695419 scopus 로고
    • Continuous thermodilution cardiac output measurement in intensive care unit patients
    • Yelderman ML, Ramsay MA, Quinn MD et al. Continuous thermodilution cardiac output measurement in intensive care unit patients. J Cardiothorac Vasc Anesth 1992; 6: 270-4.
    • (1992) J Cardiothorac Vasc Anesth , vol.6 , pp. 270-274
    • Yelderman, M.L.1    Ramsay, M.A.2    Quinn, M.D.3
  • 5
    • 0027195184 scopus 로고
    • Computation of aortic flow from pressure in humans using a non-linear, three-element model
    • Wesseling KH, Jansen JR, Settels JJ, Schreuder JJ. Computation of aortic flow from pressure in humans using a non-linear, three-element model. J Appl Physiol 1993; 74: 2566-73.
    • (1993) J Appl Physiol , vol.74 , pp. 2566-2573
    • Wesseling, K.H.1    Jansen, J.R.2    Settels, J.J.3    Schreuder, J.J.4
  • 7
    • 0032734448 scopus 로고    scopus 로고
    • Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: Comparison with pulmonary arterial thermodilution
    • Goedje O, Hoeke K, Lichtwarck-Aschoff M et al. Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: comparison with pulmonary arterial thermodilution. Crit Care Med 1999; 27: 2407-12.
    • (1999) Crit Care Med , vol.27 , pp. 2407-2412
    • Goedje, O.1    Hoeke, K.2    Lichtwarck-Aschoff, M.3
  • 8
    • 0034869820 scopus 로고    scopus 로고
    • A comparison of cardiac output derived from the arterial pressure wave against thermodilution in cardiac surgery patients
    • Jansen JR, Schreuder JJ, Mulier JP et al. A comparison of cardiac output derived from the arterial pressure wave against thermodilution in cardiac surgery patients. Br J Anaesth 2001; 87: 212-22.
    • (2001) Br J Anaesth , vol.87 , pp. 212-222
    • Jansen, J.R.1    Schreuder, J.J.2    Mulier, J.P.3
  • 9
    • 0032847442 scopus 로고    scopus 로고
    • Comparison of pulmonary artery and arterial thermodilution cardiac output in critically ill patients
    • Sakka SG, Reinhart K, Meier-Hellmann A. Comparison of pulmonary artery and arterial thermodilution cardiac output in critically ill patients. Intensive Care Med 1999; 25: 843-6.
    • (1999) Intensive Care Med , vol.25 , pp. 843-846
    • Sakka, S.G.1    Reinhart, K.2    Meier-Hellmann, A.3
  • 10
    • 0033994066 scopus 로고    scopus 로고
    • Is the placement of a pulmonary artery catheter still justified solely for the measurement of cardiac output?
    • Sakka SG, Reinhart K, Wegscheider K, Meier-Hellmann A. Is the placement of a pulmonary artery catheter still justified solely for the measurement of cardiac output? J Cardiothorac Vasc Anesth 2000; 14: 119-24.
    • (2000) J Cardiothorac Vasc Anesth , vol.14 , pp. 119-124
    • Sakka, S.G.1    Reinhart, K.2    Wegscheider, K.3    Meier-Hellmann, A.4
  • 11
    • 0032802701 scopus 로고    scopus 로고
    • Thermal filament continuous thermodilution cardiac output delayed response limits its value during acute hemodynamic instability
    • Poli de Figueiredo LF, Malbouisson LM, Varicoda EY et al. Thermal filament continuous thermodilution cardiac output delayed response limits its value during acute hemodynamic instability. J Trauma 1999; 47: 288-93.
    • (1999) J Trauma , vol.47 , pp. 288-293
    • Poli De Figueiredo, L.F.1    Malbouisson, L.M.2    Varicoda, E.Y.3
  • 12
    • 0029992651 scopus 로고    scopus 로고
    • Delayed time response of the continuous cardiac output pulmonary artery catheter
    • Siegel LC, Hennessy MM, Pearl RG. Delayed time response of the continuous cardiac output pulmonary artery catheter. Anesth Analg 1996; 83: 1173-7.
    • (1996) Anesth Analg , vol.83 , pp. 1173-1177
    • Siegel, L.C.1    Hennessy, M.M.2    Pearl, R.G.3
  • 13
    • 0030880954 scopus 로고    scopus 로고
    • Continuous cardiac output monitoring during adult liver transplantation: Thermal filament technique versus bolus thermodilution
    • Greim CA, Roewer N, Thiel H, Laux G, Schulte am EJ. Continuous cardiac output monitoring during adult liver transplantation: thermal filament technique versus bolus thermodilution. Anesth Analg 1997; 85: 483-8.
    • (1997) Anesth Analg , vol.85 , pp. 483-488
    • Greim, C.A.1    Roewer, N.2    Thiel, H.3    Laux, G.4    Schulte Am, E.J.5
  • 14
    • 0036242474 scopus 로고    scopus 로고
    • Pulse contour analysis versus thermodilution in cardiac surgery patients
    • Rauch H, Muller M, Fleischer F et al. Pulse contour analysis versus thermodilution in cardiac surgery patients. Acta Anaesthesiol Scand 2002; 46: 424-9.
    • (2002) Acta Anaesthesiol Scand , vol.46 , pp. 424-429
    • Rauch, H.1    Muller, M.2    Fleischer, F.3
  • 15
    • 0028961244 scopus 로고
    • Continuous cardiac output measurements in the perioperative period
    • Jakobsen CJ, Melsen NC, Andresen EB. Continuous cardiac output measurements in the perioperative period. Acta Anaesthesiol Scand 1995; 39: 485-8.
    • (1995) Acta Anaesthesiol Scand , vol.39 , pp. 485-488
    • Jakobsen, C.J.1    Melsen, N.C.2    Andresen, E.B.3
  • 16
    • 0029168048 scopus 로고
    • Continuous versus intermittent cardiac output measurement in cardiac surgical patients undergoing hypothermic cardiopulmonary bypass
    • Bottiger BW, Rauch H, Bohrer H et al. Continuous versus intermittent cardiac output measurement in cardiac surgical patients undergoing hypothermic cardiopulmonary bypass. J Cardiothorac Vasc Anesth 1995; 9: 405-11.
    • (1995) J Cardiothorac Vasc Anesth , vol.9 , pp. 405-411
    • Bottiger, B.W.1    Rauch, H.2    Bohrer, H.3
  • 17
    • 0029016594 scopus 로고
    • Evaluation of a new continuous thermodilution cardiac output monitor in critically ill patients: A prospective criterion standard study
    • Haller M, Zollner C, Briegel J, Forst H. Evaluation of a new continuous thermodilution cardiac output monitor in critically ill patients: a prospective criterion standard study. Crit Care Med 1995; 23: 860-6.
    • (1995) Crit Care Med , vol.23 , pp. 860-866
    • Haller, M.1    Zollner, C.2    Briegel, J.3    Forst, H.4
  • 18
    • 0029153813 scopus 로고
    • The thermodilution method for the clinical assessment of cardiac output
    • Jansen JR. The thermodilution method for the clinical assessment of cardiac output. Intensive Care Med 1995; 21: 691-7.
    • (1995) Intensive Care Med , vol.21 , pp. 691-697
    • Jansen, J.R.1
  • 19
    • 0015542664 scopus 로고
    • Choice of detection site for the determination of cardiac output by thermal dilution: The injection-thermistor-catheter
    • Vliers AC, Oeseburg B, Visser KR, Zijlstra WG. Choice of detection site for the determination of cardiac output by thermal dilution: the injection-thermistor-catheter. Cardiovasc Res 1973; 7: 133-8.
    • (1973) Cardiovasc Res , vol.7 , pp. 133-138
    • Vliers, A.C.1    Oeseburg, B.2    Visser, K.R.3    Zijlstra, W.G.4
  • 20
    • 0015058059 scopus 로고
    • Limitations of thermal dilution curves for cardiac output determinations
    • Wessel HU, Paul MH, James GW, Grahn AR. Limitations of thermal dilution curves for cardiac output determinations. J Appl Physiol 1971; 30: 643-52.
    • (1971) J Appl Physiol , vol.30 , pp. 643-652
    • Wessel, H.U.1    Paul, M.H.2    James, G.W.3    Grahn, A.R.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.