Pitfalls and Artifacts Encountered in Clinical MR Imaging of the Spine

Author keywords

Images, artifact; Magnetic resonance (MR), artifact; Spine, MR, 30.1214

Indexed keywords

ARTIFACT; HUMAN; IMAGE PROCESSING; METHODOLOGY; NUCLEAR MAGNETIC RESONANCE IMAGING; PATHOLOGY; REVIEW; SPINE; SPINE DISEASE;

EID: 0032197240     PISSN: 02715333     EISSN: None     Source Type: Journal    
DOI: 10.1148/radiographics.18.6.9821197     Document Type: Article
Times cited : (86)

References (62)
  • 1
    • 0028472170 scopus 로고
    • Basic physics of MR imaging: An introduction
    • Hendrick RE. Basic physics of MR imaging: an introduction. RadioGraphics 1994; 14:829-846.
    • (1994) RadioGraphics , vol.14 , pp. 829-846
    • Hendrick, R.E.1
  • 2
    • 0028508850 scopus 로고
    • MR imaging instrumentation and image artifacts
    • Patton JA. MR imaging instrumentation and image artifacts. RadioGraphics 1994; 14:1083-1096.
    • (1994) RadioGraphics , vol.14 , pp. 1083-1096
    • Patton, J.A.1
  • 3
    • 0021681994 scopus 로고
    • Chemical shift misregistration effect in magnetic resonance imaging
    • Soila KP, Viamonte M, Starewicz PM. Chemical shift misregistration effect in magnetic resonance imaging. Radiology 1984; 153:819-820.
    • (1984) Radiology , vol.153 , pp. 819-820
    • Soila, K.P.1    Viamonte, M.2    Starewicz, P.M.3
  • 5
    • 0031044274 scopus 로고    scopus 로고
    • The influence of chemical shift artifact on magnetic resonance imaging of the ligamentum flavum at 0.5 Tesla
    • Whitehouse RW, Hutchinson CE, Laitt R, Jenkins JPR, Jackson A. The influence of chemical shift artifact on magnetic resonance imaging of the ligamentum flavum at 0.5 Tesla. Spine 1997; 22:200-202.
    • (1997) Spine , vol.22 , pp. 200-202
    • Whitehouse, R.W.1    Hutchinson, C.E.2    Laitt, R.3    Jenkins, J.P.R.4    Jackson, A.5
  • 6
    • 0025002359 scopus 로고
    • A comparison of default and reduced bandwidth MR imaging of the spine at 1.5 T
    • Ketonen L, Totterman S, Simon JH, et al. A comparison of default and reduced bandwidth MR imaging of the spine at 1.5 T. AJNR 1990; 11:9-15.
    • (1990) AJNR , vol.11 , pp. 9-15
    • Ketonen, L.1    Totterman, S.2    Simon, J.H.3
  • 7
    • 0023501075 scopus 로고
    • Potential false-negative MR images of the thoracic spine in disk disease with switching of phase- And frequency-encoding gradients
    • Enzmann DR, Griffin C, Rubin JB. Potential false-negative MR images of the thoracic spine in disk disease with switching of phase-and frequency-encoding gradients. Radiology 1987; 165:635-637.
    • (1987) Radiology , vol.165 , pp. 635-637
    • Enzmann, D.R.1    Griffin, C.2    Rubin, J.B.3
  • 8
    • 0021958023 scopus 로고
    • MR image artifacts from periodic motion
    • Wood ML, Henkelman RM. MR image artifacts from periodic motion. Med Phys 1985; 12:143-151.
    • (1985) Med Phys , vol.12 , pp. 143-151
    • Wood, M.L.1    Henkelman, R.M.2
  • 9
    • 0021235596 scopus 로고
    • The effect of motion on two-dimensional Fourier transformation magnetic resonance images
    • Schultz CL, Alfidi RJ, Nelson AD, Kopiwoda SY, Clampitt ME. The effect of motion on two-dimensional Fourier transformation magnetic resonance images. Radiology 1984; 152:117-121.
    • (1984) Radiology , vol.152 , pp. 117-121
    • Schultz, C.L.1    Alfidi, R.J.2    Nelson, A.D.3    Kopiwoda, S.Y.4    Clampitt, M.E.5
  • 10
    • 0024453934 scopus 로고
    • Importance of absence of CSF pulsation artifacts in the MR detection of significant myelographic block at 1.5 T
    • Quint DJ, Patel SC, Sanders WP, Hearshen DO, Boulos RS. Importance of absence of CSF pulsation artifacts in the MR detection of significant myelographic block at 1.5 T. AJNR 1989; 10:1089-1095.
    • (1989) AJNR , vol.10 , pp. 1089-1095
    • Quint, D.J.1    Patel, S.C.2    Sanders, W.P.3    Hearshen, D.O.4    Boulos, R.S.5
  • 11
    • 0347761011 scopus 로고
    • Utility of modifying phase/frequency gradient directions in clinical MRI
    • Kanal E, Prorok RJ, Brunberg JA, Davis PL. Utility of modifying phase/frequency gradient directions in clinical MRI (abstr). Magn Reson Imaging 1987; 5(suppl 1):75.
    • (1987) Magn Reson Imaging , vol.5 , Issue.1 SUPPL. , pp. 75
    • Kanal, E.1    Prorok, R.J.2    Brunberg, J.A.3    Davis, P.L.4
  • 12
    • 0029664763 scopus 로고    scopus 로고
    • Cerebrospinal fluid flow artifact: A possible pitfall on fast-spin-echo MR imaging of the spine simulating intradural pathology
    • Larsen DW, Teitelbaum GP, Norman D. Cerebrospinal fluid flow artifact: a possible pitfall on fast-spin-echo MR imaging of the spine simulating intradural pathology. Clin Imaging 1996; 20:140-142.
    • (1996) Clin Imaging , vol.20 , pp. 140-142
    • Larsen, D.W.1    Teitelbaum, G.P.2    Norman, D.3
  • 13
    • 0023832968 scopus 로고
    • FRODO pulse sequences: A new means of eliminating motion, flow, and wrap-around artifacts
    • Edelman RR, Atkinson DJ, Silver MS, Loaiza FL, Warren WS. FRODO pulse sequences: a new means of eliminating motion, flow, and wrap-around artifacts. Radiology 1988; 166:231-236.
    • (1988) Radiology , vol.166 , pp. 231-236
    • Edelman, R.R.1    Atkinson, D.J.2    Silver, M.S.3    Loaiza, F.L.4    Warren, W.S.5
  • 14
    • 0023217435 scopus 로고
    • CSF-gated MR imaging of the spine: Theory and clinical implementation
    • Rubin JB, Enzmann DR, Wright A. CSF-gated MR imaging of the spine: theory and clinical implementation. Radiology 1987; 163:784-792.
    • (1987) Radiology , vol.163 , pp. 784-792
    • Rubin, J.B.1    Enzmann, D.R.2    Wright, A.3
  • 15
    • 0023180462 scopus 로고
    • Motion artifact suppression technique (MAST) for MR imaging
    • Pattany PM, Phillips JJ, Chiu LC, et al. Motion artifact suppression technique (MAST) for MR imaging. J Comput Assist Tomogr 1987; 11:369-377.
    • (1987) J Comput Assist Tomogr , vol.11 , pp. 369-377
    • Pattany, P.M.1    Phillips, J.J.2    Chiu, L.C.3
  • 16
    • 0347130674 scopus 로고
    • Analysis of imaging axes significance in motion artifact suppression technique (MAST): Refocussing of moving spins
    • Duerk JL, Pattany PM. Analysis of imaging axes significance in motion artifact suppression technique (MAST): refocussing of moving spins (abstr). Magn Reson Imaging 1987; 5(suppl 1):108-109.
    • (1987) Magn Reson Imaging , vol.5 , Issue.1 SUPPL. , pp. 108-109
    • Duerk, J.L.1    Pattany, P.M.2
  • 17
    • 0023734532 scopus 로고
    • Correction of CSF motion artifact on MR images of the brain and spine by pulse sequence modification: Clinical evaluation
    • Szeverenyi NM, Kieffer SA, Cacayorin E. Correction of CSF motion artifact on MR images of the brain and spine by pulse sequence modification: clinical evaluation. AJNR 1988; 9:1069-1074.
    • (1988) AJNR , vol.9 , pp. 1069-1074
    • Szeverenyi, N.M.1    Kieffer, S.A.2    Cacayorin, E.3
  • 18
    • 0023943147 scopus 로고
    • The motion artifact suppression technique (MAST) in magnetic resonance imaging: Clinical results
    • Colletti PM, Raval JK, Benson RC, et al. The motion artifact suppression technique (MAST) in magnetic resonance imaging: clinical results. Magn Reson Imaging 1988; 6:293-299.
    • (1988) Magn Reson Imaging , vol.6 , pp. 293-299
    • Colletti, P.M.1    Raval, J.K.2    Benson, R.C.3
  • 19
    • 0025142649 scopus 로고
    • Improvement of T2-weighted images of central nervous system by motion artifact suppression technique (MAST)
    • Iwasaki S, Nakagawa H, Uchida H, et al. Improvement of T2-weighted images of central nervous system by motion artifact suppression technique (MAST). Radiat Med 1990; 8:13-16.
    • (1990) Radiat Med , vol.8 , pp. 13-16
    • Iwasaki, S.1    Nakagawa, H.2    Uchida, H.3
  • 20
    • 0025944431 scopus 로고
    • Gradient moment nulling for steady-state free precession MR imaging of cerebrospinal fluid
    • Wood ML, Zur Y, Neuringer LJ. Gradient moment nulling for steady-state free precession MR imaging of cerebrospinal fluid. Med Phys 1991; 18:1038-1044.
    • (1991) Med Phys , vol.18 , pp. 1038-1044
    • Wood, M.L.1    Zur, Y.2    Neuringer, L.J.3
  • 21
    • 0023796922 scopus 로고
    • Magnetic susceptibility artifacts in gradient-recalled echo MR imaging
    • Czervionke LF, Daniels DL, Wehrli FW, et al. Magnetic susceptibility artifacts in gradient-recalled echo MR imaging. AJNR 1988; 9:1149-1155.
    • (1988) AJNR , vol.9 , pp. 1149-1155
    • Czervionke, L.F.1    Daniels, D.L.2    Wehrli, F.W.3
  • 22
    • 0026250576 scopus 로고
    • Quantitation of structural distortion of the cervical neural foramina in gradient-echo MR imaging
    • Tien RD, Buxton RB, Schwaighofer BW, Chu PK. Quantitation of structural distortion of the cervical neural foramina in gradient-echo MR imaging. JMRI 1991; 1:683-687.
    • (1991) JMRI , vol.1 , pp. 683-687
    • Tien, R.D.1    Buxton, R.B.2    Schwaighofer, B.W.3    Chu, P.K.4
  • 23
    • 0023905802 scopus 로고
    • Cervical spine: MR imaging with a partial flip angle, gradient-refocused pulse sequence
    • Enzmann DR, Rubin JB. Cervical spine: MR imaging with a partial flip angle, gradient-refocused pulse sequence. Radiology 1988; 166: 467-472.
    • (1988) Radiology , vol.166 , pp. 467-472
    • Enzmann, D.R.1    Rubin, J.B.2
  • 24
    • 0026016411 scopus 로고
    • Degenerative narrowing of the cervical spine neural foramina: Evaluation with high-resolution 3DFT gradient-echo MR imaging
    • Yousem DM, Atlas SW, Goldberg HI, Grossman RI. Degenerative narrowing of the cervical spine neural foramina: evaluation with high-resolution 3DFT gradient-echo MR imaging. AJNR 1991; 12:229-236.
    • (1991) AJNR , vol.12 , pp. 229-236
    • Yousem, D.M.1    Atlas, S.W.2    Goldberg, H.I.3    Grossman, R.I.4
  • 25
    • 0026023298 scopus 로고
    • Effects of magnetic susceptibility artifacts and motion in evaluating the cervical neural foramina on 3DFT gradient-echo MR imaging
    • Tsuruda JS, Remley K. Effects of magnetic susceptibility artifacts and motion in evaluating the cervical neural foramina on 3DFT gradient-echo MR imaging. AJNR 1991; 12:237-241.
    • (1991) AJNR , vol.12 , pp. 237-241
    • Tsuruda, J.S.1    Remley, K.2
  • 26
    • 0029915644 scopus 로고    scopus 로고
    • Cervical spondylosis: Three-dimensional gradient-echo MR with magnetization transfer
    • Melhem ER, Benson ML, Beauchamp NJ, Lee RR. Cervical spondylosis: three-dimensional gradient-echo MR with magnetization transfer. AJNR 1996; 17:705-711.
    • (1996) AJNR , vol.17 , pp. 705-711
    • Melhem, E.R.1    Benson, M.L.2    Beauchamp, N.J.3    Lee, R.R.4
  • 27
    • 0026648242 scopus 로고
    • Fat-suppression failure artifacts simulating pathology on frequency-selective fat-suppression MR images of the head and neck
    • Anzai Y, Lufkin RB, Jabour BA, Hanafee WN. Fat-suppression failure artifacts simulating pathology on frequency-selective fat-suppression MR images of the head and neck. AJNR 1992; 13:879-884.
    • (1992) AJNR , vol.13 , pp. 879-884
    • Anzai, Y.1    Lufkin, R.B.2    Jabour, B.A.3    Hanafee, W.N.4
  • 28
    • 0029019622 scopus 로고
    • Low-cost device for avoiding bulk susceptibility artifacts in chemical-selective fat saturation MR of the head and neck
    • Cox IH, Dillon WP. Low-cost device for avoiding bulk susceptibility artifacts in chemical-selective fat saturation MR of the head and neck. AJNR 1995; 16:1367-1369.
    • (1995) AJNR , vol.16 , pp. 1367-1369
    • Cox, I.H.1    Dillon, W.P.2
  • 29
    • 0023756419 scopus 로고
    • Use of titanium wire in cervical spine fixation as a means to reduce MR artifacts
    • Mirvis SE, Geisler F, Joslyn JN, Zrebeet H. Use of titanium wire in cervical spine fixation as a means to reduce MR artifacts. AJNR 1988; 9: 1229-1231.
    • (1988) AJNR , vol.9 , pp. 1229-1231
    • Mirvis, S.E.1    Geisler, F.2    Joslyn, J.N.3    Zrebeet, H.4
  • 30
    • 0024325510 scopus 로고
    • Titanium wire internal fixation for stabilization of injury of the cervical spine: Clinical results and postoperative magnetic resonance imaging of the spinal cord
    • Geisler FH, Mirvis SE, Zrebeet H, Joslyn JN. Titanium wire internal fixation for stabilization of injury of the cervical spine: clinical results and postoperative magnetic resonance imaging of the spinal cord. Neurosurgery 1989; 25: 356-362.
    • (1989) Neurosurgery , vol.25 , pp. 356-362
    • Geisler, F.H.1    Mirvis, S.E.2    Zrebeet, H.3    Joslyn, J.N.4
  • 31
    • 0027517660 scopus 로고
    • A biomechanical evaluation of magnetic resonance imaging-compatible wire in cervical spine fixation
    • Scuderi GJ, Greenberg SS, Cohen DS, Latta LL, Eismont FJ. A biomechanical evaluation of magnetic resonance imaging-compatible wire in cervical spine fixation. Spine 1993; 18:1991-1994.
    • (1993) Spine , vol.18 , pp. 1991-1994
    • Scuderi, G.J.1    Greenberg, S.S.2    Cohen, D.S.3    Latta, L.L.4    Eismont, F.J.5
  • 32
    • 0024845945 scopus 로고
    • Anterior and posterior cervical spine fixation using titanium implants to facilitate magnetic resonance imaging evaluation
    • Savolaine ER, Ebraheim NA, Andreshak TG, Jackson WT. Anterior and posterior cervical spine fixation using titanium implants to facilitate magnetic resonance imaging evaluation. J Orthop Trauma 1989; 3:295-299.
    • (1989) J Orthop Trauma , vol.3 , pp. 295-299
    • Savolaine, E.R.1    Ebraheim, N.A.2    Andreshak, T.G.3    Jackson, W.T.4
  • 33
    • 0026062433 scopus 로고
    • Use of MR imaging-compatible Halifax interlaminar clamps for posterior cervical fusion
    • Aldrich EF, Crow WN, Weber PB, Spagnolia TN. Use of MR imaging-compatible Halifax interlaminar clamps for posterior cervical fusion. J Neurosurg 1991; 74:185-189.
    • (1991) J Neurosurg , vol.74 , pp. 185-189
    • Aldrich, E.F.1    Crow, W.N.2    Weber, P.B.3    Spagnolia, T.N.4
  • 34
    • 0027411830 scopus 로고
    • Magnetic resonance imaging evaluation of the spine with metal implants: General safely and superior imaging with titanium
    • Rupp R, Ebraheim NA, Savolaine ER, Jackson WT. Magnetic resonance imaging evaluation of the spine with metal implants: general safely and superior imaging with titanium. Spine 1993; 18:379-385.
    • (1993) Spine , vol.18 , pp. 379-385
    • Rupp, R.1    Ebraheim, N.A.2    Savolaine, E.R.3    Jackson, W.T.4
  • 35
    • 0028900149 scopus 로고
    • Magnetic resonance imaging of titanium anterior cervical spine plating systems
    • Tominaga T, Shimizu H, Koshu K, Kayama T, Yoshimoto T. Magnetic resonance imaging of titanium anterior cervical spine plating systems. Neurosurgery 1995; 36:951-955.
    • (1995) Neurosurgery , vol.36 , pp. 951-955
    • Tominaga, T.1    Shimizu, H.2    Koshu, K.3    Kayama, T.4    Yoshimoto, T.5
  • 36
    • 0030041439 scopus 로고    scopus 로고
    • Internal fixation of the spine using a braided titanium cable: Clinical results and postoperative magnetic resonance imaging
    • Doran SE, Papadopoulos SM, Miller LD. Internal fixation of the spine using a braided titanium cable: clinical results and postoperative magnetic resonance imaging. Neurosurgery 1996; 38:493-497.
    • (1996) Neurosurgery , vol.38 , pp. 493-497
    • Doran, S.E.1    Papadopoulos, S.M.2    Miller, L.D.3
  • 37
    • 0029979618 scopus 로고    scopus 로고
    • Postoperative magnetic resonance imaging with titanium implants of the thoracic and lumbar spine
    • Ortiz O, Pait TG, McAllister P, Sauter K. Postoperative magnetic resonance imaging with titanium implants of the thoracic and lumbar spine. Neurosurgery 1996; 38:741-745.
    • (1996) Neurosurgery , vol.38 , pp. 741-745
    • Ortiz, O.1    Pait, T.G.2    McAllister, P.3    Sauter, K.4
  • 39
    • 0025949259 scopus 로고
    • Magnetic resonance artifact in the postoperative cervical spine: A potential pitfall
    • Peterman SB, Hoffman JC, Malko JA. Magnetic resonance artifact in the postoperative cervical spine: a potential pitfall. Spine 1991; 16:721-725.
    • (1991) Spine , vol.16 , pp. 721-725
    • Peterman, S.B.1    Hoffman, J.C.2    Malko, J.A.3
  • 42
    • 0025283306 scopus 로고
    • Potential misinterpretation of cervical spondylosis with cord compression caused by metallic artifacts in magnetic resonance imaging of the postoperative spine
    • Levitt M, Benjamin V, Kricheff II. Potential misinterpretation of cervical spondylosis with cord compression caused by metallic artifacts in magnetic resonance imaging of the postoperative spine. Neurosurgery 1990; 27:126-130.
    • (1990) Neurosurgery , vol.27 , pp. 126-130
    • Levitt, M.1    Benjamin, V.2    Kricheff, I.I.3
  • 43
    • 0030975859 scopus 로고    scopus 로고
    • MRI artifacts following anterior cervical diskectomy
    • Salazar JL, Misra M, Bloom D, Dobben G. MRI artifacts following anterior cervical diskectomy. Surg Neurol 1997; 48:23-29.
    • (1997) Surg Neurol , vol.48 , pp. 23-29
    • Salazar, J.L.1    Misra, M.2    Bloom, D.3    Dobben, G.4
  • 46
  • 47
    • 0030012255 scopus 로고    scopus 로고
    • Optimizing imaging parameters for MR evaluation of the spine with titanium pedicle screws
    • Petersilge CA, Lewin JS, Duerk JL, Yoo JU, Ghaneyem AJ. Optimizing imaging parameters for MR evaluation of the spine with titanium pedicle screws. AJR 1996; 166:1213-1218.
    • (1996) AJR , vol.166 , pp. 1213-1218
    • Petersilge, C.A.1    Lewin, J.S.2    Duerk, J.L.3    Yoo, J.U.4    Ghaneyem, A.J.5
  • 48
    • 0028097780 scopus 로고
    • Metallic artifacts on MR images of the postoperative spine: Reduction with fast spin-echo techniques
    • Tartaglino LM, Flanders AE, Vinitski S, Friedman DP. Metallic artifacts on MR images of the postoperative spine: reduction with fast spin-echo techniques. Radiology 1994; 190:565-569.
    • (1994) Radiology , vol.190 , pp. 565-569
    • Tartaglino, L.M.1    Flanders, A.E.2    Vinitski, S.3    Friedman, D.P.4
  • 50
    • 0030959636 scopus 로고    scopus 로고
    • Internally stabilized spine: Optimal choice of frequency-encoding gradient direction during MR imaging minimizes susceptibility artifact from titanium vertebral body screws
    • Frazzini VI, Kagetsu NJ, Johnson CE, Destian S. Internally stabilized spine: optimal choice of frequency-encoding gradient direction during MR imaging minimizes susceptibility artifact from titanium vertebral body screws. Radiology 1997; 204:268-272.
    • (1997) Radiology , vol.204 , pp. 268-272
    • Frazzini, V.I.1    Kagetsu, N.J.2    Johnson, C.E.3    Destian, S.4
  • 52
    • 0023917115 scopus 로고
    • The MR appearance of gray and white matter in the cervical spinal cord
    • Czervionke LF, Daniels DL, Ho PSP, et al. The MR appearance of gray and white matter in the cervical spinal cord. AJNR 1988; 9:557-562.
    • (1988) AJNR , vol.9 , pp. 557-562
    • Czervionke, L.F.1    Daniels, D.L.2    Ho, P.S.P.3
  • 54
    • 0025246764 scopus 로고
    • Pseudoatrophy of the cervical portion of the spinal cord on MR images: A manifestation of the truncation artifact?
    • Yousem DM, Janick PA, Atlas SW, et al. Pseudoatrophy of the cervical portion of the spinal cord on MR images: a manifestation of the truncation artifact? AJNR 1990; 11:373-377.
    • (1990) AJNR , vol.11 , pp. 373-377
    • Yousem, D.M.1    Janick, P.A.2    Atlas, S.W.3
  • 56
    • 0023803111 scopus 로고
    • Truncation artifact in MR images of the intervertebral disk
    • Breger RK, Czervionke LF, Kass EG, et al. Truncation artifact in MR images of the intervertebral disk. AJNR 1988; 9:825-828.
    • (1988) AJNR , vol.9 , pp. 825-828
    • Breger, R.K.1    Czervionke, L.F.2    Kass, E.G.3
  • 59
    • 0028034644 scopus 로고
    • Correction of nonuniformity in images of the spine and optic nerve from fixed receive-only surface coils at 1.5 T
    • Tofts PS, Barker GJ, Simmons A, et al. Correction of nonuniformity in images of the spine and optic nerve from fixed receive-only surface coils at 1.5 T. J Comput Assist Tomogr 1994; 18:997-1003.
    • (1994) J Comput Assist Tomogr , vol.18 , pp. 997-1003
    • Tofts, P.S.1    Barker, G.J.2    Simmons, A.3
  • 60
    • 0024358996 scopus 로고
    • Eddy-current-induced artifacts caused by an "MR-compatible" halo device
    • Malko JA, Hoffman JC, Jarrett PJ. Eddy-current-induced artifacts caused by an "MR-compatible" halo device. Radiology 1989; 173:563-564.
    • (1989) Radiology , vol.173 , pp. 563-564
    • Malko, J.A.1    Hoffman, J.C.2    Jarrett, P.J.3
  • 61
    • 0023893365 scopus 로고
    • Artifacts and diagnostic pitfalls on magnetic resonance imaging: A clinical review
    • Hahn FJ, Chu WK, Coleman PE, et al. Artifacts and diagnostic pitfalls on magnetic resonance imaging: a clinical review. Radiol Clin North Am 1988; 26:717-735.
    • (1988) Radiol Clin North Am , vol.26 , pp. 717-735
    • Hahn, F.J.1    Chu, W.K.2    Coleman, P.E.3
  • 62
    • 0026068768 scopus 로고
    • Magnetic resonance imaging of the normal craniovertebral junction
    • Ellis JH, Martel W, Lillie JH, Aisen AM. Magnetic resonance imaging of the normal craniovertebral junction. Spine 1991; 16:105-111.
    • (1991) Spine , vol.16 , pp. 105-111
    • Ellis, J.H.1    Martel, W.2    Lillie, J.H.3    Aisen, A.M.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.